IPS is prescription use software used by dentists and dental lab technicians for the visualization and image segmentation of DICOM data from medical scanners such as CT. The software aids the user in the creation of 3D models of oral maxillofacial region and in planning dental surgical treatments and placement of dental implants.
Device Story
IPS is a dental 3D planning software; inputs DICOM-compliant radiological data (CT/Cone beam CT); transforms data into 3D models of oral maxillofacial region; provides visualization, segmentation, and measurement tools (distance, angle, bone density); used by dentists and dental lab technicians in clinical settings; output compatible with CAD and rapid prototyping machines; aids in surgical treatment planning and implant placement; facilitates clinical decision-making through anatomical visualization and measurement.
Clinical Evidence
No clinical data. Bench testing only; validation performed to demonstrate DICOM conformance, including importing, sorting, and grouping of DICOM data.
Technological Characteristics
Pure software device; PC-based; DICOM standard compliant; compatible with Windows XP, Vista, 7, and Intel-based Mac (booted in Windows); outputs compatible with CAD and rapid prototyping systems.
Indications for Use
Indicated for dentists and dental lab technicians for visualization and segmentation of DICOM data (e.g., CT) to create 3D models of the oral maxillofacial region, plan dental surgical treatments, and plan dental implant placement.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
510k Summary
DEC - 6 2011
Device Owner Address:
Telephone: Fax:
Official Correspondent
Telephone: Email:
Symbvo Dental, LLC, 15310 Amberly Drive suite#250 Tampa FL 33647 (877) 479-6296 (866) 697-8559
Amberry Drive suite#250
FL 33647
479-6296
697-8559
Parul Chansoria 87 Hulme Ct Apt. 216 Stanford CA 94305 877-479-6296 parul@symbyo.com
## Device Name and Classification
| Trade Name/Product Name: | IPS | Areas of<br>Comparison | Simplant system<br>K033849 | Sicat implant<br>K090119 | IPS |
|-------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Dental 3D Planning Software | Description | Pure software device | Pure software device | Pure software device |
| Classification Name: | System, Image Processing, Radiological; 21CFR892.2050 | Input source | CT scanners, MRI<br>machines | CT scanners, DVT<br>scanners | Cone beam CT<br>machines, CT<br>scanners |
| Product Code: | LLZ | Data conformance | DICOM | DICOM | DICOM |
| Regulation Number: | Picture Archiving and Communication System, 21CFR892.2050 | Regulation<br>description | Picture archiving and<br>communication system | Picture archiving and<br>communication system | Picture archiving and<br>communication system |
| Panel: | Radiology | Intended use | Visualization and<br>planning simulation that<br>aids the user in the<br>placement of dental<br>implants and planning<br>for surgical treatments | Visualization and planning<br>simulation that aids the<br>user in the placement of<br>dental implants and<br>planning for surgical<br>treatments | Visualization and<br>planning simulation<br>that aids the user in the<br>placement of dental<br>implants and planning<br>for surgical treatments |
| Class | 2 | Intended user | Well trained dentist \<br>dental technicians | Well trained dentist | Dentists, dental<br>technician |
| Substantial Equivalence claimed to: | MATERIALISE N.V. Simplant system (K033849) | Main tools | Visualization, implant<br>placement,<br>measurement of<br>distance, angle<br>measurement, density<br>determination,<br>segmentation tool | Visualization, implant<br>placement, measurement<br>of distance, angle<br>measurement, density<br>determination | Visualization, implant<br>and nerve placement,<br>measurement of<br>distance, angle<br>measurement, density<br>determination,<br>segmentation tool |
| | siCAT GmbH & Co. KG Sicat implant (K090119) | Host platform | PC | PC | PC |
| Operating system | Not specified | Not specified | Windows XP, vista, 7 | | |
| Output<br>compatibility | CAD, rapid prototyping<br>machines | CAD, rapid prototyping<br>machines | CAD, rapid<br>prototyping machines | | |
| Operating system<br>compatibility | - Windows XP or vista<br>- Intel based MAC<br>(computer should be<br>booted from windows<br>not MAC OS) | - Windows XP<br>- Windows Vista<br>- WINDOWS 7 | - Windows XP<br>- Windows Vista<br>- WINDOWS 7 | | |
### Device Description
IPS is a Dental 3D Planning Software that aids dentists and dental lab technicians in the creation of 3D models of oral maxillofacial region and in planning dental surgical treatments and placement of dental implants. The data used for planning the treatment course is obtained from radiological devices (e.g., CT) complying with DICOM format. IPS provides dentists with various visualization and measurement tools which facilitate in generating dental panoramic view, segmenting of the dental anatomical structures and measuring bone density and distance between the dental implants.
#### Indications for Use
IPS is prescription use software used by dentists and dental lab technicians for the visualization and image segmentation of DICOM data from medical scanners such as CT. The software aids the user in the creation of 3D models of oral maxillofacial region and in planning dental surgical treatments and placement of dental implants.
## Substantial Equivalence
Symbyo Dental LLC. wishes to use the following devices as a predicates: MATERIALISE N.V. Simplant system (K033849) siCAT GmbH & Co. KG Sicat implant (K090119)
The claim of substantial equivalence to the predicate devices is based on both intended use and technology used.
{1}------------------------------------------------
.
Predicate Comparison Table
{2}------------------------------------------------
## Based Upon Intended Used:
IPS and its predicate devices have intended use; which is Visualization and planning simulation that aids the user in the placement of dental implants and planning for dental surgical treatment.
#### Based Upon Technology Used:
IPS, like the predicate devices, is a pure software device with DICOM data conformance.
The technological characteristics of IPS vs. the predicate devices are summarized below.
Our software conforms to DICOM standard and it has been verified and validated for it.
Validation and verification was performed to demonstrate DICOM conformance
### Following test cases were used to validate DICOM conformance:
- Importing DICOM Data.
- Sorting DICOM data.
- Grouping multiple DICOM case
These test cases have been adopted to perform a validation process according to the IPS DICOM Conformance statement. These test cases are performed on the patient's DICOM images to ensure that the IPS application's modules are fully compliant with the DICOM conformance.
#### Conclusion
The results of all testing demonstrate that IPS does not raise any new issues of safety, effectiveness or performance when compared to the existing predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room – WO66-G609 Silver Spring, MD 20993-0002
Symbyo Dental, LLC. % Mr. Mark Job -- Official Correspondent 1394 25th Street NW Buffalo MN 94305
Re: K113404 Trade/Device Name: Dental 3D Planning Software IPS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 17, 2011 Received: November 18, 2011
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
DEC - 6 2011
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Device Name: Dental 3D Planning Software IPS
#### Indications for Use:
IPS is prescription use software used by dentists and dental lab technicians for the visualization and image segmentation of DICOM data from medical scanners such as CT. The software aids the user in the creation of 3D models of oral maxillofacial region and in planning dental surgical treatments and placement of dental implants.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
.l.O.D.R.
ostic Device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.