← Product Code [HRX](/productcode/HRX) · K113362

# SPINEVU ENDOSCOPIC SPINE SYSTEM (SESS) (K113362)

_Spine View, Inc. · HRX · Feb 10, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K113362

## Device Facts

- **Applicant:** Spine View, Inc.
- **Product Code:** [HRX](/productcode/HRX.md)
- **Decision Date:** Feb 10, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Surgical accessories are indicated for facilitating endoscopic access and visualization in the surgical area of the cervical, thoracic, or lumbar spine for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy.

## Device Story

SpineVu Endoscopic Spine System (SESS) is a collection of single-use, sterile arthroscopic surgical accessories. System components include enVue Cannula (with articulating jaws for tissue retraction), enVue Sheath, 16G Introducer Cannula with Stylet, Guidewire, Dilator, Beveled Cannula, Ball-Tipped Probe, Infusion Cannula, and Suction Cannula. The system facilitates access and visualization at the surgical site in the spine. The physician uses the introducer and guidewire to establish access, followed by the dilator and beveled cannula to create a pathway. The enVue Cannula, which accommodates a flexible endoscope and surgical instruments, is then inserted. The articulating jaws on the enVue Cannula allow the surgeon to mechanically retract tissue away from the cannula tip to improve visualization. The system is used in a clinical/surgical setting by trained spinal surgeons. The output is direct visualization of the surgical site via an endoscope, enabling the surgeon to perform procedures like discectomy. The system benefits patients by providing a minimally invasive approach to spinal surgery.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included visual and dimensional verification, jaw cycle and activation integrity, device-to-device compatibility, tensile strength, flow rate, luer attachment, leakage, biocompatibility, design validation, packaging, and sterility testing.

## Technological Characteristics

Single-use, sterile (irradiated) arthroscopic accessories. Materials include stainless steel, PEBAX, polyurethane, PVC, polycarbonate, nylon-12, and medical-grade adhesives (Loctite 4018, 4013, 3311). Features include articulating jaws for mechanical tissue retraction, working channels for endoscopes (up to 2mm) and instruments, and luer-lock connections for fluid management. No electronic or software components.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Spine View SpineVu Endoscopic Spine System ([K081051](/device/K081051.md))
- Spine View enSpire Discectomy System ([K110992](/device/K110992.md))
- Endius Atavi System ([K061345](/device/K061345.md))
- Richard Wolf Minimally Invasive Spinal Surgery Set ([K994363](/device/K994363.md))
- Karl Storz Percutaneous Foraminoscopy Set ([K001918](/device/K001918.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K11 3362 & lot 13

# 510(k) Summary

# FEB 1 0 2012 .

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

## A. Name and Address of Applicant Spine View, Inc. 48810 Kato Road, Suite 100E Fremont CA 94538 Phone: (510) 743-5090 Fax: (510) 623-1093

- B. Contact Person Sevrina Ciucci, RAC Regulatory Affairs Consultant
Phone: (408) 316-4837

### C. Date Prepared

November 11, 2011

- Device Name D.

| Trade Name:          | SpineVu Endoscopic Spine System (SESS) |
|----------------------|----------------------------------------|
| Common Name:         | Arthroscope                            |
| Classification Name: | Arthroscope                            |

### E. Device Classification

| Classification: | 21 CFR §888.1100 |
|-----------------|------------------|
| Product Code:   | HRX              |
| Device Class:   | Class II         |

### Predicate Device F.

The modified SpineVu Endoscopic Spine System (SESS) is substantially equivalent to the Spine View SpineVu Endoscopic Spine System (SESS) (K081051), Spine View enSpire Discectomy System (K110992), Endius Atavi System (K061345), Richard Wolf Minimally Invasive Spinal Surgery Set (K994363), and Karl Storz Percutaneous Foraminoscopy Set (K001918).

### Device Description G.

The SpineVu Endoscopic Spine System (SESS) is a collection of arthroscopic surgical accessories provided sterile (irradiated) and intended for single-use only. As a group,

{1}------------------------------------------------

K113362 p. 2 of 13

the accessories are provided to facilitate delivery of an endoscope and other instruments to the targeted treatment site. The SpineVu Endoscopic Spine System (SESS) consists of the following devices.

- enVue Cannula a device which provides access into the body for performing . procedures in and around the spine. The enVue Cannula is provided with two different tip configurations (Standard Jaw and Long Jaw).
- . enVue Sheath - a device that may be used to facilitate the delivery of a flexible endoscope (up to 2 mm diameter) through the working lumen of the enVue Cannula.
- . 16G Introducer Cannula with Stylet - facilitates initial access to the treatment site. The Introducer Cannula is used in conjunction with the Introducer Stylet.
- . Guidewire - (16" long with a diameter of 0.058") may be used to facilitate exchange of the Introducer Cannula for the Dilator.
- . Dilator - is designed to fit over the Guidewire and inside the Beveled Cannula. The Dilator is used to bluntly dissect soft tissue in order to provide a pathway for the Beveled Cannula.
- . Beveled Cannula - is designed to fit over the Dilator. The Beveled Cannula comes in two lengths (5.4" and 6.1"). It provides an access path to the targeted treatment site for other instruments, such as the enVue Cannula.
- . Ball-Tipped Probe - is used to probe or palpate tissue or bone.
- Infusion Cannula is used to introduce fluids to the treatment site. .
- . Suction Cannula - is used to aspirate materials from the treatment site.

### H. Intended Use

Surgical accessories are indicated for facilitating endoscopic access and visualization in the surgical area of the cervical, thoracic, or lumbar spine for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy.

### 1. Technological Comparison

The SpineVu Endoscopic Spine System (SESS) has similar features as compared to the predicate devices as shown in the tables below:

{2}------------------------------------------------

3 of 13 હ K113362

Spine View, Inc.

Special 510(K) Special 510(K) Notification
pineVu Endoscopic Spine System (SESS)

| Manufacturer     | Device Name                           | 510(k) #                                                       | Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                           | Product Code                                                                                                                      | Classification Section | Spine View, Inc.<br>SpineVu<br>Endoscopic Spine<br>System                                                                                                                         | Spine View, Inc.<br>SpineVu<br>Endoscopic Spine<br>System                                                                                                                                                                      | Manufacturer     | Device Name                                                                            | 510(k) #        | Design                                              | Endoscope<br>Compatibility       | Supplied<br>Sterile/Sterilization Method | Single Use<br>Only |              | Company, K#                                                                                                                                                                                                                                                                                                                                                                   | Device Name                                                                                                                                                                    | Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate                                                                                                                              | Subject                                                                                                                                                                                                                             | Company, K#                                                                                                                                                                                                                                                                 | Predicate                                                                                                                                                                                                                                           | Predicate                                                                                                                                                                                                   | Subject                                                                                                                                                                                                       | Company, K#                                                                                                                                                | Predicate         | Subject                                              |                                                     |
|------------------|---------------------------------------|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|----------------------------------------------------------------------------------------|-----------------|-----------------------------------------------------|----------------------------------|------------------------------------------|--------------------|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|------------------------------------------------------|-----------------------------------------------------|
| Zimmer (Endius)  | Atavi System                          | K061345                                                        | Posterior or anterior access and visualization in the surgical area of the cervical, thoracic, or lumbar spine allowing the surgeon to perform any type of surgical spinal procedures such as discectomy, nucleotomy, spinal fusion, spinal decompression, and insertion of spinal implants. Other examples of generic surgical use of the Endius Atavi System would be for use in the knee, ankle, shoulder, hand, wrist, and temporomandibular joint (TMJ). | HRX                                                                                                                               | Class II               | K081051                                                                                                                                                                           | K113362                                                                                                                                                                                                                        | Zimmer (Endius)  | Atavi System                                                                           | K061345         | Dilatation<br>Access<br>Retraction<br>Visualization | OD up to<br>approximately 0.944" | No                                       | No                 |              | Endius, K061345                                                                                                                                                                                                                                                                                                                                                               | Atavi System                                                                                                                                                                   | Posterior or anterior access and visualization in the surgical area of the cervical, thoracic, or lumbar spine allowing the surgeon to perform any type of surgical spinal procedures such as discectomy, nucleotomy, spinal fusion, spinal decompression, and insertion of spinal implants. Other examples of generic surgical use of the Endius Atavi System would be for use in the knee, ankle, shoulder, hand, wrist, and temporomandibular joint (TMJ). |                                                                                                                                        |                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                             | Endius, K061345                                                                                                                                                                                                                                     | Spine View, SpineVu<br>Endoscopic Spine System,<br>K081051                                                                                                                                                  | Spine View, SpineVu Endoscopic Spine System<br>K113362                                                                                                                                                        |                                                                                                                                                            | Endius, K061345   | Spine View, SpineVu Endoscopic Spine System, K081051 | Spine View, SpineVu Endoscopic Spine System K113362 |
| Spine View, Inc. | enSpire Discectomy System             | K110992                                                        | Cutting, grinding and aspirating intervertebral disc material during discectomy procedures in the cervical, thoracic and lumbar spine.                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                              | Same                   | Same                                                                                                                                                                              | Same                                                                                                                                                                                                                           | Spine View, Inc. | enSpire Discectomy<br>System                                                           | K110992         | Dilatation<br>Access<br>Excision                    | OD up to 0.058"                  | Yes/Gamma radiation                      | Yes                |              |                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Spine View, SpineVu Endoscopic Spine System, K081051                                                                                   | Spine View, SpineVu Endoscopic Spine System K113362                                                                                                                                                                                 | Device Name                                                                                                                                                                                                                                                                 | Atavi System                                                                                                                                                                                                                                        | SpineVu Balloon Cannula                                                                                                                                                                                     | enVue Cannula                                                                                                                                                                                                 | enVue Sheath                                                                                                                                               | Device Name       | Atavi System                                         | enVue Cannula                                       |
| Richard Wolf     | Minimally Invasive Spinal Surgery Set | K994363                                                        | Examination, diagnosis, and/or therapy by personnel trained and qualified in connection with endoscopically used accessories in various medical disciplines, such as orthopedic and spinal surgery.                                                                                                                                                                                                                                                           | Same                                                                                                                              | Same                   | Same                                                                                                                                                                              | Same                                                                                                                                                                                                                           | Richard Wolf     | Minimally Invasive<br>Spinal Surgery Set                                               | K994363         | Dissection<br>Manipulation<br>Excision              | OD up to 0.271"                  | No                                       | No                 |              |                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                               | SpineVu Balloon Cannula                                                                                                                | enVue Cannula                                                                                                                                                                                                                       | Product Code                                                                                                                                                                                                                                                                | HRX                                                                                                                                                                                                                                                 | HRX                                                                                                                                                                                                         | Same                                                                                                                                                                                                          | Same                                                                                                                                                       | Target Anatomy    | Intervertebral procedures                            | Same                                                |
| Karl Storz       | Percutaneous Foraminoscopy Set        | K001918                                                        | Visualize and treat vertebral disc herniations in the lumbar region of the spine using a posterolateral approach and fluoroscopic control.                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                              | Same                   | 15G Needle<br>SpineVu Guide Wire<br>SpineVu Dilator<br>SpineVu Introducer Sheath<br>SpineVu Infusion Cannula<br>SpineVueBalloon Cannula<br>SpineVue Debrider<br>SpineVu CoagProbe | 1. 16G Introducer Cannula w/ Stylet<br>2. Guidewire<br>3. Dilator<br>4. Beveled Cannula<br>5. enVue Cannula (Standard Jaw or Long Jaw)<br>6. enVue Sheath<br>7. Ball Tipped Probe<br>8. Suction Cannula<br>9. Infusion Cannula | Karl Storz       | Percutaneous<br>Foraminoscopy Set                                                      | K001918         | Dilatation<br>Access<br>Visualization               | OD up to 0.236"                  | No                                       | No                 |              |                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Cutting, grinding and aspirating intervertebral disc material during discectomy procedures in the cervical, thoracic and lumbar spine. | Facilitating endoscopic access and visualization in the surgical area of the cervical, thoracic, or lumbar spine for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy. Same                         | Design                                                                                                                                                                                                                                                                      | Articulating portal with retracting<br>jaws mechanically actuated to<br>retract tissue away from end of<br>cannula                                                                                                                                  | Toroidal balloon can be inflated<br>to mechanically retract tissue<br>away from the end of the<br>cannula                                                                                                   | Articulating Top Jaw can be<br>mechanically actuated to retract<br>tissue away from end of cannula<br>Distal tip: Articulating jaw (Nylon)<br>for tissue retraction (Standard and<br>Long jaw configurations) | When combined with the enVue<br>Cannula, can accommodate the<br>following:<br>a) Flexible Endoscope<br>b) Endoscopic Instruments<br>c) Irrigation/Drainage | Supplied Sterile? | Yes                                                  | Same                                                |
| Spine View, Inc. | SpineVu Endoscopic Spine System       | K081051                                                        | Endoscopic access and visualization in the surgical area of the cervical, thoracic, or lumbar spine and are accessorized with surgical and coagulation tools for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy.                                                                                                                                                                                                            | Same                                                                                                                              | Same                   | Spine View, Inc.                                                                                                                                                                  | SpineVu<br>Endoscopic Spine<br>System                                                                                                                                                                                          | K081051          | Dilatation<br>Access<br>Retraction<br>Excision<br>Aspiration/Infusion<br>Visualization | OD up to 0.032" | Yes/E-beam                                          | Yes                              |                                          |                    |              |                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                | enVue Sheath                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                        | Articulating single lumen portal<br>with retracting jaws used to dilate<br>and maintain access which can<br>accommodate the following:<br>a) Flexible Endoscope<br>b) Endoscopic Instruments<br>c) Irrigation/Drainage              | Single, multi-lumen device which<br>can accommodate the following:<br>a) Flexible Endoscope<br>b) Endoscopic Instruments<br>c) Irrigation/Drainage                                                                                                                          |                                                                                                                                                                                                                                                     |                                                                                                                                                                                                             | Single Use?                                                                                                                                                                                                   | Yes                                                                                                                                                        | Same              |                                                      |                                                     |
| Spine View, Inc. | SpineVu Endoscopic Spine System       | K113362                                                        | Surgical accessories are indicated for facilitating endoscopic access and visualization in the surgical area of the cervical, thoracic, or lumbar spine for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy.                                                                                                                                                                                                                 | Same                                                                                                                              | Same                   | Spine View, Inc.                                                                                                                                                                  | SpineVu<br>Endoscopic Spine<br>System                                                                                                                                                                                          | K113362          | Dilatation<br>Access<br>Retraction<br>Aspiration/Infusion                              | OD up to 0.079" | Yes/E-beam                                          | Yes                              |                                          |                    | Intended Use | 1. Access portal with working channel for instruments - facilitating endoscopic access and visualization in the surgical area of the spine for use during interventional spinal procedures<br>2. Distal feature for retracting tissue away from end of cannula -articulating portal with retracting jaws that mechanically actuate to retract tissue away from end of cannula | 1. Access cannula with working channels for endoscope, endoscopic instruments, and irrigation and drainage<br>2. Distal feature for retracting tissue away from end of cannula | 1. Access cannula with working channel for endoscope, endoscopic instruments, and irrigation and drainage<br>2. Distal feature for retracting tissue away from end of cannula<br>1. Intended for use with the enVue Cannula to provide the access ports for endoscopic instruments and endoscope, and irrigation                                                                                                                                              | Dimensions                                                                                                                             | 1. OD: Unknown<br>2. Working Length: Unknown<br>3. Working channel for<br>endoscopic instruments and<br>endoscope: 0.945" x 1.18"<br>4. Length, Tissue Retraction<br>Feature: 1.181"<br>5. OD, Tissue Retraction<br>Feature: 2.362" | 1. OD: 0.140"<br>2. Working Length: ~7.5"<br>3. Length, Tissue Retraction<br>Feature: up to 0.315"<br>4. OD, Tissue Retraction<br>Feature: 0.248"<br>5. Multiple Working channels<br>for endoscopic instruments:<br>~0.049"<br>6. Working channel for<br>endoscope: ~0.039" | 1. OD: 0.282"<br>2. Working Length: 7.5"<br>3. Length, Tissue Retraction<br>Feature 0.394"<br>4. OD, Tissue Retraction<br>Feature: 0.354"<br>5. Single Working Channel for<br>enVue Sheath (for<br>endoscopic instruments and<br>endoscope): 0.240" | 1. OD: 0.236" - Designed to be<br>delivered through enVue<br>Cannula<br>2. Working Length: 8.8"<br>3. Working Channel for<br>endoscopic instruments:<br>0.122"<br>4. Working channel for<br>endoscope: 2 mm |                                                                                                                                                                                                               | SpineVu Balloon Cannula                                                                                                                                    | enVue Sheath      |                                                      |                                                     |
| Materials        | Unknown                               | Stainless Steel, PEBAX,<br>Polyurethane, PVC,<br>Polycarbonate | Nylon-12, Stainless Steel, Solder<br>(Sn/Ag), Polycarbonate, Medical<br>Grade Adhesive (Loctite 4018,<br>Loctite 4013, Loctite 3311)                                                                                                                                                                                                                                                                                                                          | Stainless Steel, Solder (Sn/Ag),<br>Polycarbonate, Tygon, Medical<br>Grade Adhesive (Loctite 4018,<br>Loctite 4013, Loctite 3311) |                        | Same                                                                                                                                                                              | Same                                                                                                                                                                                                                           |                  |                                                                                        |                 |                                                     |                                  |                                          |                    |              |                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                        |                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                     |                                                                                                                                                                                                             |                                                                                                                                                                                                               |                                                                                                                                                            |                   |                                                      |                                                     |
|                  | Same                                  | Same                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                   |                        |                                                                                                                                                                                   |                                                                                                                                                                                                                                |                  |                                                                                        |                 |                                                     |                                  |                                          |                    |              |                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                        |                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                     |                                                                                                                                                                                                             |                                                                                                                                                                                                               |                                                                                                                                                            |                   |                                                      |                                                     |

{3}------------------------------------------------

Special 510(k) Notification
pineVu Endoscopic Spine System (SESS)

{4}------------------------------------------------

pineVu Endoscopic Special 510(K) Notification
Concessoric Spine System (SESS)

7.50613

|くいろろん2

{5}------------------------------------------------

Special 510(K) Special 510(K) Notification
SpineVu Endoscopic Spine System (SESS)

KI13362 p. 6 of 13

# enVue Cannula and enVue Sheath

{6}------------------------------------------------

Special 510(k) Special 510(k) Notification
Controlu Endoscopic Spine System (SESSS)

K113362 p. 7 . F 13

{7}------------------------------------------------

Spine View, Inc.

# Special 510(K) Notification
pineVu Endoscopic Spine System (SESS)

# nula and Sty

|                        | 6G Introducer Cannula and Stylet                                                                                                                   |                                                                                                                                                                                                                          | Guidewire                                                                                                                                          |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
|                        | Predicate                                                                                                                                          | Subject                                                                                                                                                                                                                  | Predicate                                                                                                                                          |
| Company, K#            | Spine View, enSpire<br>Discectomy System, K110992                                                                                                  | Spine View, SpineVu<br>Endoscopic Spine System                                                                                                                                                                           | SpineVu Endoscopic Spine<br>System, K081051                                                                                                        |
| Device Name            | Introducer Cannula with<br>Stylet                                                                                                                  | 16G Introducer Cannula and<br>Stylet                                                                                                                                                                                     | SpineVu Guide Wire                                                                                                                                 |
| Indications for<br>Use | Cutting, grinding and aspirating<br>intervertebral disc material<br>during discectomy procedures in<br>the cervical, thoracic and lumbar<br>spine. | Facilitating endoscopic access<br>and visualization in the surgical<br>area of the cervical, thoracic, or<br>lumbar spine for interventional<br>spinal procedures such as<br>discectomy, nucleotomy and<br>foraminotomy. | Cutting, grinding and aspirating<br>intervertebral disc material<br>during discectomy procedures in<br>the cervical, thoracic and lumbar<br>spine. |
| Intended Use           | Penetrate skin and provide a<br>conduit to targeted treatment<br>site.                                                                             | Same                                                                                                                                                                                                                     | Maintaining access to the target<br>site for exchange of the other<br>accessories                                                                  |
| Product Code           | HRX                                                                                                                                                | Same                                                                                                                                                                                                                     | HRX                                                                                                                                                |

| Predicate                                           | Subject                                                               |           |                                                |                                      |                                                                                                                                     |  |                                                                            |                  |                                                                             |                           |                      |             |
|-----------------------------------------------------|-----------------------------------------------------------------------|-----------|------------------------------------------------|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--|----------------------------------------------------------------------------|------------------|-----------------------------------------------------------------------------|---------------------------|----------------------|-------------|
| u Endoscopic Spine<br>System, K081051               |                                                                       | Subject   | Spine View, SpineVu<br>Endoscopic Spine System | Guidewire                            | Solid Stainless Steel wire with blunt tip                                                                                           |  | Length: 16"                                                                | Diameter: 0.058" | Same                                                                        | Same                      | Same                 | Same        |
| neVu Guide Wire                                     | Guidewire                                                             | Predicate | SpineVu Endoscopic Spine<br>System, K081051    | SpineVu Guide Wire                   | Sold Stainless Steel wire with<br>sharp tip                                                                                         |  | Length: 16.5"                                                              | Diameter: 0.046" | Stainless Steel                                                             | Intervertebral procedures | Yes                  | Yes         |
| Grinding and aspirating<br>bral disc material       | Facilitating endoscopic access<br>and visualization in the surgical   | Subject   |                                                | 16G Introducer Cannula and<br>Stylet | Same                                                                                                                                |  | Same                                                                       |                  | Same                                                                        | Same                      | Same                 | Same        |
| iscectomy procedures in<br>cal, thoracic and lumbar | area of the cervical, thoracic, or<br>lumbar spine for interventional | Predicate |                                                |                                      | 16G Introducer: Hollow Stainless Steel tube with a proximal luer<br>Stylet: Solid Stainless Steel instrument with a tri-beveled tip |  | Cannula Length: 7"<br>Stylet Length 7.7"<br>OD: 16G (0.068")<br>ID: 0.060" |                  | Stainless Steel, Polycarbonate,<br>Medical Grade Adhesive<br>(Loctite 3311) | Intervertebral procedures | Yes                  | Yes         |
|                                                     | spinal procedures such as                                             |           | Company, K#                                    | Device Name                          | Design                                                                                                                              |  | Dimensions                                                                 |                  | Materials                                                                   | Target<br>Anatomy         | Supplied<br>Sterile? | Single Use? |
|                                                     | discectomy, nucleotomy and<br>foraminotomy.                           |           |                                                |                                      |                                                                                                                                     |  |                                                                            |                  |                                                                             |                           |                      |             |
| ng access to the target                             | Same                                                                  |           |                                                |                                      |                                                                                                                                     |  |                                                                            |                  |                                                                             |                           |                      |             |
| xchange of the other<br>es                          | Same                                                                  |           |                                                |                                      |                                                                                                                                     |  |                                                                            |                  |                                                                             |                           |                      |             |

7

{8}------------------------------------------------

Special 510(k) Special 510(k) Notification
SpineVu Endoscopic Spine System (SESS)

9 of 13

ﺓ

11 3362

{9}------------------------------------------------

K113362 p. 10 of 13

A

Spine View, Inc.

Special 510(k) Special 510(k) Notification
pineVu Endoscopic Spine System (SESSS)

| Dilator             |                                                                                                                                        |                                                                                                                                                                                                        | Beveled Can                                                                             |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
|                     | Predicate                                                                                                                              | Subject                                                                                                                                                                                                | Pr-                                                                                     |
| Company, K#         | SpineVu Endoscopic Spine System, K081051                                                                                               | Spine View, SpineVu Endoscopic Spine System                                                                                                                                                            | SpineVu E                                                                               |
| Device Name         | SpineVu Dilator                                                                                                                        | Dilator                                                                                                                                                                                                | Syste                                                                                   |
| Indications for Use | Cutting, grinding and aspirating intervertebral disc material during discectomy procedures in the cervical, thoracic and lumbar spine. | Facilitating endoscopic access and visualization in the surgical area of the cervical, thoracic, or lumbar spine for interventional spinal procedures such as discectomy, nucleotomy and foraminotomy. | SpineVu Int Cutting, grinding intervertebral during discectomy the cervical, the spine. |
| Intended Use        | Provide a larger pathway through tissue for other instruments                                                                          | Same                                                                                                                                                                                                   | Provides a co instruments to                                                            |
| Product Code        | HRX                                                                                                                                    | Same                                                                                                                                                                                                   | HRX                                                                                     |
| Design              | Stainless Steel tube with tapered tip                                                                                                  | Same                                                                                                                                                                                                   | Stainless Steel tip                                                                     |
| Dimensions          | Length: 9"<br>ID: 0.050"<br>OD: 0.142"                                                                                                 | Length: 9"<br>ID: 0.063" (at tip)<br>OD: 0.285"                                                                                                                                                        | Length: 7"<br>ID: 0.144"<br>OD: 0.150"                                                  |
| Materials           | Stainless Steel                                                                                                                        | Same                                                                                                                                                                                                   | Stainless Steel Polycarbonate                                                           |
| Target Anatomy      | Intervertebral procedures                                                                                                              | Same                                                                                                                                                                                                   | Intervertebral                                                                          |
| Supplied Sterile?   | Yes                                                                                                                                    | Same                                                                                                                                                                                                   | Yes                                                                                     |
| Single Use?         | Yes                                                                                                                                    | Same                                                                                                                                                                                                   | Yes                                                                                     |

| Beveled Cannula | Predicate                                                                                                                                          | Subject                                                                                                                                                                                                                  |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                 | SpineVu Endoscopic Spine System, K081051                                                                                                           | Spine View, SpineVu<br>Endoscopic Spine System                                                                                                                                                                           |
|                 | SpineVu Introducer Sheath                                                                                                                          | Beveled Cannula                                                                                                                                                                                                          |
|                 | Cutting, grinding and aspirating<br>intervertebral disc material<br>during discectomy procedures in<br>the cervical, thoracic and lumbar<br>spine. | Facilitating endoscopic access<br>and visualization in the surgical<br>area of the cervical, thoracic, or<br>lumbar spine for interventional<br>spinal procedures such as<br>discectomy, nucleotomy and<br>foraminotomy. |
|                 | Provides a conduit for other<br>instruments to treatment site.                                                                                     | Same                                                                                                                                                                                                                     |
|                 | HRX                                                                                                                                                | Same                                                                                                                                                                                                                     |
|                 | Stainless Steel tube with straight<br>tip                                                                                                          | Stainless Steel tube with<br>beveled tip                                                                                                                                                                                 |
|                 | Length: 7"<br>ID: 0.144"<br>OD: 0.150"                                                                                                             | Length, Short: 5.4"<br>Length, Long: 6.1"<br>ID: 0.291"<br>OD: 0.312"                                                                                                                                                    |
|                 | Stainless Steel, ABS,<br>Polycarbonate                                                                                                             | Stainless Steel                                                                                                                                                                                                          |
|                 | Intervertebral procedures                                                                                                                          | Same                                                                                                                                                                                                                     |
|                 | Yes                                                                                                                                                | Same                                                                                                                                                                                                                     |
|                 | Yes                                                                                                                                                | Same                                                                                                                                                                                                                     |

{10}------------------------------------------------

KII 3362 p 11 of 13

Spine View, Inc.

Special 510(k) Special 510(k) Notification
Calletter

| Ball Tip Probe         |                                                                                                                                                                                                                       |                                                                                                                                                                                                                          | Suction Cann                                                                          |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
|                        | Predicate                                                                                                                                                                                                             | Subject                                                                                                                                                                                                                  |                                                                                       |
| Company, K#            |                                                                                                                                                                                                                       |                                                                                                                                                                                                                          | Pre                                                                                   |
|                        | Richard Wolf, Minimally<br>Invasive Spinal Surgery Set,<br>K994363                                                                                                                                                    | Spine View, SpineVu<br>Endoscopic Spine System<br>Ball Tip Probe                                                                                                                                                         | Percutaneous<br>Set,                                                                  |
| Device Name            | Probe                                                                                                                                                                                                                 |                                                                                                                                                                                                                          | Suction                                                                               |
| Indications for<br>Use | Examination, diagnosis, and/or<br>therapy by personnel trained<br>and qualified in connection with<br>endoscopically used<br>accessories in various medical<br>disciplines, such as orthopedic<br>and spinal surgery. | Facilitating endoscopic access<br>and visualization in the surgical<br>area of the cervical, thoracic, or<br>lumbar spine for interventional<br>spinal procedures such as<br>discectomy, nucleotomy and<br>foraminotomy. | Visualize and<br>disc herniation<br>region of the<br>posterolateral<br>fluoroscopic c |
| Intended Use           | Bluntly probe tissue or bone                                                                                                                                                                                          | Same                                                                                                                                                                                                                     | Aspirate fluid to<br>treatment site                                                   |
| Product Code           | HRX                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                     | HRX                                                                                   |
| Design                 | Retractable solid ball-tip wire<br>with flexible blunted tip                                                                                                                                                          | Solid ball-tip wire shaft with ball<br>with blunt tip                                                                                                                                                                    | Metal hypotube<br>female luer hub<br>distal tip                                       |
| Dimensions             | Length: 13.8"<br>Diameter: 0.098"                                                                                                                                                                                     | Length: 12.4"<br>Diameter, Shaft: 0.047"<br>Diameter, Tip: 0.094"                                                                                                                                                        | Length: 11.8"<br>ID: 0.027"<br>OD: 0.157"                                             |
| Materials              | Unknown                                                                                                                                                                                                               | Stainless Steel, Brazing Paste<br>(Ag/Cu/ Zn/Sn)                                                                                                                                                                         | Stainless Steel                                                                       |
| Target<br>Anatomy      | Intervertebral procedures                                                                                                                                                                                             | Same                                                                                                                                                                                                                     | Intervertebral                                                                        |
| Supplied<br>Sterile?   | No                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                     | No                                                                                    |
| Single Use?            | No                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                     | No                                                                                    |

| Suction Cannula | Predicate                                                                                                                                               | Subject                                                                                                                                                                                                                  |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                 | Karl Storz<br>Percutaneous Foraminoscopy<br>Set, K001918                                                                                                | Spine View, SpineVu<br>Endoscopic Spine System                                                                                                                                                                           |
|                 | Suction Cannula                                                                                                                                         | Suction Cannula                                                                                                                                                                                                          |
|                 | Visualize and treat verterbral<br>disc herniations in the lumbar<br>region of the spine using a<br>posterolateral approach and<br>fluoroscopic control. | Facilitating endoscopic access<br>and visualization in the surgical<br>area of the cervical, thoracic, or<br>lumbar spine for interventional<br>spinal procedures such as<br>discectomy, nucleotomy and<br>foraminotomy. |
|                 | Aspirate fluid from the targeted<br>treatment site                                                                                                      | Aspirate fluid from the targeted<br>treatment site                                                                                                                                                                       |
|                 | HRX                                                                                                                                                     | Same                                                                                                                                                                                                                     |
|                 | Metal hypotube with proximal<br>female luer hub and beveled<br>distal tip                                                                               | Metal hypotube with proximal<br>barb fitting<br>Blunt distal tip<br>Designed to fit through working<br>lumen of enVue Cannula/enVue<br>Sheath                                                                            |
|                 | Length: 11.8"<br>ID: 0.027"<br>OD: 0.157"                                                                                                               | Length: 12.2"<br>ID: 0.106"<br>OD: 0.120"                                                                                                                                                                                |
|                 | Stainless Steel                                                                                                                                         | Stainless Steel, Acrylic, Medical<br>Grade Adhesive (Loctite 3311)                                                                                                                                                       |
|                 | Intervertebral procedures                                                                                                                               | Same                                                                                                                                                                                                                     |
|                 | No                                                                                                                                                      | Same                                                                                                                                                                                                                     |
|                 | No                                                                                                                                                      | Same…

---

**Source:** [https://fda-staging.innolitics.com/device/K113362](https://fda-staging.innolitics.com/device/K113362)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
