K113357 · J. T. Posey Co. · OYS · Mar 21, 2012 · General Hospital
Device Facts
Record ID
K113357
Device Name
POSEY BED
Applicant
J. T. Posey Co.
Product Code
OYS · General Hospital
Decision Date
Mar 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6760
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Posey Bed 8070 is a hospital bed, canopy, and mattress system designed to help provide a safe, controlled environment for patients at extreme risk of injury from a fall or unassisted bed exit. The Posey Bed 8070 is a less restrictive alternative to physical restraints such as belts, vests, or jackets for patients at least 46 inches tall, weighing between 46 and 300 pounds. The Posey Bed 8070 is a restraint, and must be prescribed by a licensed physician that includes Rx use in the home environment.
Device Story
The Posey Bed 8070 is an enclosed bed canopy system integrated with an AC-powered adjustable hospital bed (Joerns Easy Care 2003DC or Invacare SC900DLX-POS). The system features a 70 cubic-foot green nylon canopy with zippered access panels and ports for IV lines, call bells, and drainage bags. A specialized mattress compartment prevents mattress movement and patient entrapment. The bed frame adjusts height (19-30 inches) for patient care or transfers. A perimeter guard provides temporary protection during caregiver access. Used in home or clinical settings, the device acts as a less restrictive alternative to traditional physical restraints (belts, vests, jackets). It provides a controlled environment to prevent falls and unassisted bed exits. Healthcare providers operate the system to manage patient safety; the physical enclosure serves as the primary intervention to mitigate injury risk.
Clinical Evidence
No clinical data. Evidence consists of biocompatibility testing (ISO 10993), performance testing, and human factors testing. Human factors testing validated the training regimen and Instructions for Use for safe/effective interaction in the home environment.
Technological Characteristics
Enclosed bed canopy system; 70 cubic-foot green nylon canopy; zippered access panels; integrated AC-powered adjustable hospital bed; specialized mattress compartment. Materials include fabrics, mesh, and zippers. Biocompatibility per ISO 10993. No software or electronic sensing/actuation beyond standard hospital bed motor controls.
Indications for Use
Indicated for patients at least 46 inches tall, weighing 46-300 lbs, at extreme risk of injury from falls or unassisted bed exit. Must be prescribed by a licensed physician for use in home or clinical environments.
Regulatory Classification
Identification
A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.
{0}------------------------------------------------
## KI13351
iv
510(k) Pre-Market Notification: Posey Bed 8070 (Rx Home Use/Additional Patient Environment)
MAR 2 1 2012
## 510(k) Summary for the Posey Bed 8070
| SUBMITTER: | J. T. Posey Company<br>5635 Peck Road ·<br>Arcadia, California 91006 USA |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Bonnie Bishop, RAC<br>Vice President, Quality Assurance and Regulatory<br>Affairs<br>Phone: (626) 454-2195 (direct)<br>(626) 443-3143, extension 195<br>Fax: (626) 443-5012<br>E-mail: BBishop@posey.com |
| DATE PREPARED: | November 9, 2011 |
| DEVICE TRADE NAME: | Posey Bed 8070 |
| COMMON/USUAL NAME: | Enclosed Bed System |
| CLASSIFICATION NAME: | Protective Restraint (21 CFR §880.6760), Code<br>FMQ |
## SUBSTANTIAL EQUIVALENCE:
DEVICE DESCRIPTION:
The proposed device, the Posey Bed 8070 with modified labeling to include home use, is the same unchanged device previously cleared by the FDA under K103817 via the 510(k) notification process, and is substantially equivalent for design and This 510(k) establishes that the Posey materials. Bed 8070 is also substantially equivalent to the Soma Safe Enclosure, as used with an AC-powered hospital bed, for the home use adjustable environment, (K963701).
The Posey Bed 8070 is an enclosed bed canopy system which includes and attaches to a fully automatic. AC-powered adjustable hospital bed (either the Joerns Healthcare, Inc., Stevens Point, Wisconsin, Easy Care® 2003DC, Model B684DC or the Invacare Corporation, Elyria, Ohio, Model SC900DLX-POS Low Electric Bed), which has a six-inch enclosed mattress compartment to help reduce the risk of patient entrapment. The device also features a 70 cubic-foot rectangular green nylon canopy with four zippered access panels for easy access to patients and four ports for intravenous lines, call bells and drainage bag The bed has a control that allows the openings.
{1}------------------------------------------------
v
frame to be raised to 30 inches for patient care or lowered to 19 inches for ease of wheelchair transfers. A perimeter guard is available as a temporary guard for use when patient care is being given.
The Posey Bed 8070 does not have side rails, headboards or footboards. Contained within the canopy is a specialized compartment for the mattress to prevent the mattress from moving within the canopy and the patient from crawling under the These features provide a system that mattress. minimizes the potential for patient entrapment.
Other safety features include a perimeter guard (reference: Soft Rails) that provides an additional measure to protect the patient from falling when the canopy is opened and a health care provider is attending to the patient.
Optional accessories include filler cushions, torso cushions, bed support surface, incontinence pads, and travel covers.
The Posey Bed 8070 is a hospital bed, canopy, and mattress system designed to help provide a safe, controlled environment for patients at extreme risk of injury from a fall or unassisted bed exit. The Posey Bed 8070 is a less restrictive alternative to physical restraints such as belts, vests, or jackets for patients at least 46 inches tall, weighing between 46 and 300 pounds. The Posey Bed 8070 is a restraint, and must be prescribed by a licensed physician that includes Rx use in the home environment.
The Posey Bed 8070 is the same product as that previously cleared under K103817, however, it is now being proposed for expanded use in the home environment. A comparison of device features in K103817 demonstrated that the Posey Bed 8070 is substantially equivalent to the currently marketed Soma Safe Enclosure, as used with an AC-powered adjustable hospital bed. Both devices utilize equivalent fabrics, mesh and zippers and are framed, enclosed canopies.
#### INTENDED USE:
#### TECHNICAL CHARACTERISTICS:
{2}------------------------------------------------
vi
# PERFORMANCE TESTING:
Results of biocompatibility (ISO 10993), human factors, and performance testing have established that the Posey Bed 8070 for use in a home care environment is suitable for the intended use indicated and is substantially equivalent to the previously cleared Posey Bed 8070 and the Soma Safe Enclosure.
The human factors testing conducted by caregivers was used to validate the design of a training regimen and Instructions for Use to include safe and effective device interactions in a home environment.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Bonnie Bishop, RAC Vice President, Quality Assurance and Regulatory Affairs J.T. Posey Company 5635 Peck Road Arcadia, California 91006
MAR 2 1 2012
Re: K113357
Trade/Device Name: Posey Bed 8070 (for home use) Regulation Number: 21 CFR 880.6760 Regulation Name: Protective Restraint Regulatory Class: I Product Code: OYS Dated: February 22, 2012 Received: February 23, 2012
Dear Ms. Bishop:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Ms. Bishop
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
5 10(k) Pre-Market Notification: Posey Bed 8070 (Rx Home Use/Additional Patient Environment)
K113357
## Indications for Use
510(k) Number (If known):
Device Name: Posey Bed 8070 (for home use)
Indications for Use:
The Posey Bed 8070 is a hospital bed, canopy, and mattress system designed to help provide a safe, controlled environment for patients at extreme risk of injury from a fall or unassisted bed exit. The Posey Bed 8070 is a less restrictive alternative to physical restraints such as belts, vests, or jackets for patients at least 46 inches tall, weighing between 46 and 300 pounds. The Posey Bed 8070 is a restraint, and must be prescribed by a licensed physician that includes Rx use in the home environment.
Prescription Use X (21 CFR 8801 Subpart D)
.4
AND/OR
Over-the-Counter Use (21 CFR §801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
> Concurrence of CDRH, Office of Device Evaluation (ODE) S
> > /Division Sign-Off) Division of Anesthesiology, General Hospital ifection Control, Dental Devices
510(k) Number: K113357
iii
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.