K113355 · J. T. Posey Co. · OYS · May 2, 2012 · General Hospital
Device Facts
Record ID
K113355
Device Name
POSEY BED
Applicant
J. T. Posey Co.
Product Code
OYS · General Hospital
Decision Date
May 2, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6760
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Posey Bed 8040/8060 is an A-Frame canopy system. When attached to a compatible commercially available hospital bed (not included) the 8040/8060 is designed to help provide a safe, controlled environment for patients at extreme risk of injury from a fall or unassisted bed exit. The Posey Bed 8040/8060 is a less restrictive alternative to physical restraints such as belts, vests, or jackets for patients at least 46 inches tall, weighing between 46 and 300 pounds. The Posey Bed 8040/8060 is a restraint, and must be prescribed by a licensed physician that includes Rx use in the home environment.
Device Story
Enclosed A-Frame bed canopy system; attaches to compatible hospital beds to create safe, controlled environment; prevents falls and unassisted bed exits. Features zippered access panels and ports for medical lines/drainage. Acts as passive restraint; alternative to belts/vests/jackets. Used in clinical or home settings; operated by caregivers. Minimizes patient entrapment via specialized mattress compartment. Benefits patients by providing safety without traditional physical restraints.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993, human factors testing, and performance testing conducted to validate design, training regimen, and Instructions for Use.
Technological Characteristics
A-Frame canopy system; green nylon fabric, mesh, and zippers. Dimensions: 73 cubic-foot volume. Compatible with hospital beds (width ≤ 36", length 76-96"). Features zippered access panels and ports for IV/drainage. Non-powered, mechanical restraint system.
Indications for Use
Indicated for patients at least 46 inches tall, weighing 46-300 lbs, at extreme risk of injury from falls or unassisted bed exit. Must be prescribed by a licensed physician for use in clinical or home environments.
Regulatory Classification
Identification
A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.
{0}------------------------------------------------
510(k) Pre-Market Notification: Posey Bed 8040 and Posey Bed 8060
K 113355
i
MAY - 2 2012
## 510(k) Summary for the Posey Bed 8040 and Posey Bed 8060
SUBMITTER:
CONTACT PERSON:
# DATE PREPARED: DEVICE TRADE NAME: COMMON/USUAL NAME:
CLASSIFICATION NAME:
SUBSTANTIAL EQUIVALENCE:
## DEVICE DESCRIPTION:
J. T. Posey Company 5635 Peck Road Arcadia, California 91006 USA
Bonnie Bishop, RAC Vice President, Quality Assurance and Regulatory Affairs Phone: (626) 454- 2195 (direct) (626) 443-3143, extension 195 Fax: (626) 443-5012 E-mail: BBishop@posey.com
April 24, 2012
Posey Bed 8040 and Posey Bed 8060
Enclosed bed canopy system used as passive restraint
Protective Restraint (21 CFR §880.676020), Code OYS
The proposed device when used with a compatible commercially available hospital bed is substantially equivalent to the Posey Bed 8070 as described in K103817 previously cleared by the FDA via the 510(k) notification process. The proposed device is also substantially equivalent to the Soma Safe Enclosure, cleared under K963701.
The Posey Bed 8040 and Posey Bed 8060 (Posey Bed 8040/8060) is an enclosed bed canopy system with an adjustable enclosed mattress compartment. The Posey Bed 8040 features a 73 cubic-foot, A-Frame, green nylon canopy with one zippered access panel and one drainage port opening. The Posey Bed 8060 features a 73 cubic-foot, A-Frame green nylon canopy with four zippered access panels for easy patient access, and ten ports for intravenous lines, call bells and drainage bags. When attached to a compatible commercially available hospital bed (not included), the 8040/8060 is designed to help provide a safe, controlled
{1}------------------------------------------------
environment for patients at extreme risk of injury from a fall or unassisted bed exit.
The Posey Bed 8040 and 8060 hospital bed compatibility guidelines are:
- . Length of bed frame must be ≤ 96" and does not retract < 76"
- . Width of bed frame must be ≤ 36"
- The height of the bed platform (deck of the . bed) from the floormust be able to adjust between:
- For Posey Bed 8040 -13 and 22 inches
- For Posey Bed 8060-16 and 25 inches
Depending on the features of the bed, operational limits must also be observed when a Posey Bed canopy is in use as described in the User Manual. For example, the head of bed angle can only be raised to ≤ 70°.
The Posey Bed 8040/8060 requires that the hospital bed side rails be kept in the "down" position. The canopy contains a specialized compartment for the mattress that helps prevent the mattress from moving within the canopy and the patient from crawling under the mattress. These features provide a system that minimizes the potential for patient entrapment.
Accessories that are specifically for use with the Posey Bed 8040/8060 include travel covers, universal bed straps, , and a pediatric canopy cover and skirt for a more child-friendly appearance. General bed accessories that can be used with the Posey Bed 8040/8060 or any other Posey Bed or hospital bed are filler cushions, a torso cushion, a support surface, an incontinence pad, and headboard/footboard pads ..
The Posey Bed 8040/8060 is an A-Frame bed canopy system. When attached to a compatible hospital bed (not included) the 8040/8060 is
#### INTENDED USE:
{2}------------------------------------------------
designed to help provide a safe, controlled environment for patients at extreme risk of injury from a fall or unassisted bed exit. The Posey Bed 8040/8060 is a less restrictive alternative to physical restraints such as belts, vests, or jackets for patients at least 46 inches tall, weighing between 46 and 300 pounds. The Posey Bed 8040/8060 is a restraint, and must be prescribed by a licensed physician that includes Rx use in the home environment.
### A comparison of device features demonstrates that the Posey Bed 8040/8060 when attached to a compatible hospital bed is substantially equivalent to the currently marketed Posey Bed 8070. The three Posey Bed devices (Models 8040, 8060 and 8070) utilize the same fabrics, mesh, and zippers and are framed, enclosed canopies. The Posey Bed 8040/8060 is also substantially equivalent to the Soma Safe Enclosure which was originally manufactured by Nova Technologies, Inc. (Bristol, Connecticut) and is now manufactured by Vivax Medical Corp (Torrington, Connecticut).
Results of biocompatibility (ISO 10993), human factors, and performance testing have established that the Posey Bed 8040/8060 is suitable for the intended use indicated and is substantially equivalent to the Posey Bed 8070 and the Soma Safe Enclosure. Human factors testing conducted by caregivers was used to validate the design of a training regimen and Instructions for Use for safe and effective device interactions.
#### TECHNICAL CHARACTERISTICS:
#### PERFORMANCE TESTING:
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Bonnie Bishop, RAC Vice President, Quality Assurance and Regulatory Affairs J.T. Posey Company 5635 Peck Road Arcadia, California 91006
MAY - 2 2012
Re: K113355
Trade/Device Name: Posey Bed 8040 and Posey Bed 8060 Regulation Number: 21 CFR 880.6760 Regulation Name: Protective Restraint Regulatory Class: I Product Code: OYS Dated: April 19, 2012 Received: April 23, 2012
Dear Ms. Bishop:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Bishop
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Ansin
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
## Indications for Use
K 11.2355
510(k) Number (If known):
Device Name: Posey Bed 8040 and Posey Bed 8060
Indications for Use:
The Posey Bed 8040/8060 is an A-Frame canopy system. When attached to a compatible commercially available hospital bed (not included) the 8040/8060 is designed to help provide a safe, controlled environment for patients at extreme risk of injury from a fall or unassisted bed exit. The Posey Bed 8040/8060 is a less restrictive alternative to physical restraints such as belts, vests, or jackets for patients at least 46 inches tall, weighing between 46 and 300 pounds. The Posey Bed 8040/8060 is a restraint, and must be prescribed by a licensed physician that includes Rx use in the home environment.
Prescription Use X (21 CFR §801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR §801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
evaluation (ODE)
5/1/2012
(Division'Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
CONFIDENTIAL
510(k) Number: .
510(k) Number: K113355
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.