K113338 · Oricare, Inc. · BTI · Aug 10, 2012 · Anesthesiology
Device Facts
Record ID
K113338
Device Name
MEDICAL AIR COMPRESSOR
Applicant
Oricare, Inc.
Product Code
BTI · Anesthesiology
Decision Date
Aug 10, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.6250
Device Class
Class 2
Indications for Use
The Oricare C4500 Medical Air Compressor is indicated to supply dry filtered and compressed air to a Medical Ventilator that operates within the C4500 Compressor manufacturer specifications.
Device Story
The Oricare C4500 is a portable medical air compressor designed for use in medical care facilities. It operates from an AC voltage source to provide a continuous supply of clean, oil-free, pressurized air to medical ventilators. The device utilizes an oil-free wobble-type rocker piston to compress ambient air, which is then cooled, dried, and filtered (5-micron) before delivery. It features automatic condensed water separation and pressure regulation. The system includes acoustic and optical alarms for high internal temperature (>80°C), power loss, and low output pressure (<25.4 psig). Visual indicators confirm AC power, ready status, and air source. By providing a reliable, independent gas supply, the device ensures the continuous operation of connected ventilators, supporting patient respiratory therapy.
Clinical Evidence
Bench testing only. Performance was verified through comparison of output flow, pressure, filtration, and electrical/mechanical safety standards (IEC 60601-1:1988, +A1:1991, +A2:1995) against the predicate device.
Indicated for supplying dry, filtered, compressed air to medical ventilators operating within manufacturer specifications in medical care facilities.
Regulatory Classification
Identification
A portable air compressor is a device intended to provide compressed air for medical purposes, e.g., to drive ventilators and other respiratory devices.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.
{0}------------------------------------------------
8
3
# 3
=
k
# 10(K) Summar
David Jamison, Executive Vice Presider
Oricare, Inc.
Thone: 215-538-2470 KOM DK50D Medical Compresso 500 Medical Air Compress Oricare, Inc.
1900 AM Drive, Suite 10
Quakertown, PA. 18951 mpressor, Air, Portab edical Air Compress ectober 19, 201 1CFR868.6256 060781 вті ame: Common / Usua evice Trade Name assification Nam egulation Numbe edicate 510(k) { Predicate Device ontact Persor Date Prepared Product Code Submitter
# Substantial Equivalence Summary
stantial Equivalence Summary
C4500 Medical Compressor is substantially equivalent in intended use, physical characteristics, performance, and
acteristics to the EKOM DK50K, c
{1}------------------------------------------------
**Intended Use**
Intended Use:
This air compressor can be used as a gas supply for a critical care ventilat
cations for Use:
Oricare C4500 Medical Air Compressor is indicated to supply dry filtered and compressed air to a Medical Ventilator
France within the C4500 Compresses monufa perates within the C4500 Compressor manufacturer specification
Device Description:
The Oricare C4500 Medical Air Compressor operates from an AC Voltage source and produces air from the normal environment
supply compressed air for nedical
# Device Testing
eners Four easing has been on case in accordance with vality ratous inh valier recognized
tandard, include tester 1999, 42:1995 and EC 6001-2-2.2007, Additionally, Air Esti
{2}------------------------------------------------
| # | Item of comparison | EKOM DK50 D Medical Compressor<br>(K060781) | Oricare C4500 Air Compressor | Discussion of the<br>differences |
|----|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Intended use of the<br>device | The DK50 D Medical Compressor is<br>indicated for supplying compressed<br>air for medical ventilators. | This air compressor can be used as a gas<br>supply for a critical care ventilator. | No differences |
| 2 | Indications for use | The EKOM DK50 D, DK50 DM is<br>indicated for use as medical air<br>compressor to supply a source of<br>clean, oil-free pressurized air for use<br>with medical ventilators. | The Oricare C4500 Medical Air<br>Compressor is indicated to supply dry<br>filtered and compressed air to a Medical<br>Ventilator that operates within the C4500<br>Compressor manufacturer specifications. | No differences |
| 3 | Environment of Use | Medical Care Facilities | Medical Care Facilities | No Differences |
| 4 | Principle of<br>operation | Compression, cooling, drying and<br>filtering of oil-free air | Compression, cooling, drying and filtering<br>of oil-free air | No Differences |
| 5 | Output flow/pressure | 40 l/min at 51 psig | 40 l/min at 50 psig | No Differences |
| 6 | Peak Flow | 200 pm for 2 sec | Peak Flow 180 Lpm for 0.6 sec | No Differences |
| 7 | Power | 120V/60 Hz | Power 115V/60 Hz | AC Mains line voltage<br>varies from 103VAC to<br>127VAC. Both systems<br>are within the range. |
| 8 | Power consumption | 5.6 A | Nominal current (amps) <600VA | Oricare device consumes<br>slightly less power<br>meaning it is slightly<br>more efficient. |
| # | Item of comparison | EKOM DK50 D Medical Compressor<br>(K060781) | Oricare C4500 Air Compressor | Discussion of the<br>differences |
| 10 | Air filtration | 5 micron | 5 micron | No Difference |
| 11 | Pressure dew point | @ 40 Lpm 20°C - 5°C below ambient<br>temperature | @ 40 Lpm 3°C below ambient temperature | No real discernible<br>difference. |
| 12 | Outlet connection | DISS | 1- DISS | No Difference |
| 13 | Sound level | < 51 dB(A) | < 52 dB(A) | No real discernible<br>difference. The audible tone<br>is for alarm annunciation<br>purposes. An increase in the<br>dB(A) would provide an<br>advantage to user. |
| 14 | Mode of operation | Continuous - SI | Continuous - SI | No Difference |
| 15 | Separation of<br>condensed water | Automatic | Automatic | No Difference |
| 16 | Operating pressure of<br>safety valve | 116 psig | 65 psig | Device regulates pressure at<br>40 psig. |
| 17 | Adjustment of pressure | Pressure regulator | Pressure regulator | No Difference |
| 18 | Alarm for cooling<br>failure / high | Acoustic and optical if increase in<br>internal temperature > 80°C | Acoustic and optical if increase in internal<br>temperature > 80°C | No Difference |
| 19 | Automatic turn-on<br>pressure | When central distribution pressure <<br>40.6 psig | When central distribution pressure < 40.6<br>psig | No Difference |
| 20 | Output pressure | Pressure gauge | Pressure gauge | No Difference |
| 21 | Alarm - loss of power | No, but instructions require connection<br>to equipment with this alarm | Yes | Advantage for Oricare<br>product Safety |
| # | Item of comparison | EKOM DK50 D Medical Compressor<br>(K060781) | Oricare C4500 Air Compressor | Discussion of the<br>differences |
| 22 | Additional Visual<br>Indicators | Not included | "AC Power" light<br>"Ready" light - outlet pressure ready<br>"Air Source - Wall" light | Safety Advantage for<br>determining AC Mains<br>Power is connected |
| 23 | Indication of drying | Pressure gauge | None | design includes chiller and<br>external water trap to<br>remove excess moisture. |
| 24 | Alarm for low pressure | None instructions require connection<br>to equipment with this alarm | Yes, triggers when output pressure <25.4 psig | Oricare device does not rely<br>on 3rd party device to be<br>connected. |
| 25 | Material in gas pathway | Aluminum, brass, nickel plated brass,<br>polyurethane and silicon tubing, copper<br>tubing, Plastic (PA, PET, POM, PBT, PC,<br>Acetyl, Polyester), polyurethane foam,<br>NBR rubber, Stainless steel, die cast<br>zinc | Aluminum, brass, nickel plated brass,<br>polyurethane and silicon tubing, copper<br>tubing, Plastic (PA, PET, POM, PBT, PC,<br>Acetyl, Polyester), polyurethane foam, NBR<br>rubber, Stainless steel, die cast zinc | No Difference |
| 26 | Air tank capacity | 5L | 2L | Excess capacity not needed<br>to supply sufficient peak<br>flow. Predicate device is<br>larger because extra<br>capacity requires additional<br>weight and physical size. |
| 27 | Compressor Type | Oil free piston and ring- positive | Oil free wobble type rocker piston- positive | No Difference |
| 28 | Type of lubrication | Oil-less | Oil-less | No Difference |
| # | Item of comparison | EKOM DK50 D Medical Compressor<br>(K060781) | Oricare C4500 Air Compressor | Discussion of the<br>differences |
| 29 | Ambient Environment | 41 to 104 degrees F; up to 95% RH | 50 to 104 degrees F; 15 – 90% RH | Oricare device range is<br>typical of many ventilators. |
| 30 | Dimensions | 19(L)x20(W)x33(H) in | 16.22 (L)x17.4(W)x15.75(H) in | Oricare device is smaller |
| 31 | Weight | 101 lbs | 66 lbs | Oricare device is lighter |
| 32 | Electrical Safety | EN 60601-1 | IEC 60601-1:1988, +A1:1991, +A2:1995 | No difference, both comply |
| 33 | Mechanical Safety | EN 60601-1 | IEC 60601-1:1988, +A1:1991, +A2:1995 | No difference, both comply |
.
omparison to Predicate Device
Page 3 of 6
{3}------------------------------------------------
.
.
:
·
・
Page 4 of 6
・
.
5
6
7
8
·
.
:
:
.
{4}------------------------------------------------
·
.
----
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
·
Page 5 of 6
:
ge 50+
.
.
{5}------------------------------------------------
.
・
.
.
:
・
.
:
.
:
. . . .
.
:
:
:
.
100 - 100 -
:
:
:
.
. . . . . . .
:
:
{6}------------------------------------------------
Image /page/6/Picture/11 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
-AUG-1-0-2012-
ORICARE, Incorporated Mr. Fred Cowdery Manager Regulatory Affairs and Quality Assurance 1900 AM Drive, Suite 100 Ouakertown. Pennsylvania 18951
Re: K113338
Trade/Device Name: C4500 Medical Air Compressor Regulation Number: 21 CFR 868.6250 Regulation Name: Portable Air Compressor Regulatory Class: II Product Code: BTI Dated: July 31, 2012 Received: August 02, 2012
Dear Mr. Cowdery:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
## Page 2- Mr. Cowdery
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies-You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Cintan Om
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
# 510(k) Number: -- K113338 US -------------------------------------------------------------------------------------------------------------------------------------------------
Device Name: C4500 Medical Air Compressor
Indications for Use:
The Oricare C4500 Medical Air Compressor is indicated to supply dry filtered and compressed air to a Medical Ventilator that operates within the C4500 Compressor manufacturer specifications.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
for l. Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices
510(k) Number: K113328
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.