K113337 · MeVis Medical Solutions AG · LNH · Dec 30, 2011 · Radiology
Device Facts
Record ID
K113337
Device Name
VISIA DYNAMIC REVIEW
Applicant
MeVis Medical Solutions AG
Product Code
LNH · Radiology
Decision Date
Dec 30, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic resonance imaging (MRI) studies. Visia™ Dynamic Review supports evaluation of dynamic MR data. Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient motion and visualizes different enhancement characteristics (parametric image maps). Furthermore, it performs other user-defined post-processing functions such as image subtractions; multi-planar reformats and maximum intensity projections. The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source image. Visia™ Dynamic Review can also be used to provide measurements for diameters, areas and volumes. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues. Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation. When interpreted by a skilled physician, Visia™ Dynamic Review provides information that may be useful in diagnosis. Patient management decisions should not be made based solely on the results of Visia™ Dynamic Review analysis.
Device Story
Visia™ Dynamic Review is a software package for viewing/analyzing dynamic MRI studies. Inputs: serial MR image data. Processing: automatic registration of serial patient motion; generation of parametric image maps; image subtractions; multi-planar reformats; maximum intensity projections; segmentation of tissues for uptake characteristic evaluation; diameter, area, and volume measurements. Output: processed images/parametric maps displayed on workstation. Usage: clinical environment; operated by radiologists/clinicians/physicians. Workflow: integrates into network via standard PC as standalone or server-client. Clinician uses output to assist in diagnosis; patient management decisions require physician interpretation and must not rely solely on software results.
Clinical Evidence
Bench testing only. Validation testing evaluated all input/output functions and operational modes against predetermined acceptance criteria. Results demonstrated that the software meets all design requirements and risk management controls.
Technological Characteristics
Software-only device; no patient contact. Operates on standard off-the-shelf PC. Supports standalone or server-client network configurations. Functions include automated motion registration, parametric mapping, image subtraction, multi-planar reformatting, and maximum intensity projection. Does not control life-sustaining equipment. No specific algorithm architecture or material standards cited.
Indications for Use
Indicated for use by clinicians to view and analyze dynamic magnetic resonance (MR) imaging studies. Supports evaluation of segmented tissue uptake characteristics, diameter, area, and volume measurements. Not for primary diagnostic interpretation of non-MR modalities.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MeVis Medical Solutions. The logo is in black and white, with the word "MeVis" in large, bold letters. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in smaller letters. The logo is simple and professional, and it is likely used to represent the company's brand.
DEC 3 0 2011
K113337
p'1/3
# 510(k) Summary
| Submitted By: | MeVis Medical Solutions AG<br>Universitaetsallee 29<br>28359 Bremen, Germany | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | |
| Establishment<br>Name: | MeVis Medical Solutions AG | |
| Establishment<br>Registration Number: | Applied for and awaiting assignment by FDA | |
| Contact Person: | Thomas E. Tynes<br>President & CEO<br>MeVis Medical Solutions, Inc.<br>N27 W24075 Paul Court - Suite 100<br>Pewaukee, WI 53072<br>USA<br>Phone: +1-262-691-9530<br>Fax: +1-262-691-9531 | |
| Date Prepared: | 11/10/2011 | |
| Trade Name: | Visia™ Dynamic Review | |
| Common Name: | Medical Image Processing Software | |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) | |
| Classification<br>Regulation Number: | 892.1000 | |
| Class: | II | |
| Panel: | Radiology | |
| Product Code: | LNH | |
{1}------------------------------------------------
#### Device Description
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic resonance imaging (MRI) studies. Visia™ Dynamic Review supports evaluation of dynamic MR data.
Visia™ Dynamic Review integrates within typical clinical workflow patterns through receiving and transferring medical images over a computer network. The software can be loaded on a standard offthe-shelf personal computer (PC) and can operate as a stand-alone workstation or in a distributed server-client configuration across a computer network.
Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient motion and visualizes different enhancement characteristics (parametric image maps). Furthermore, it performs other user-defined post-processing functions such as image subtractions; multi-planar reformats and maximum intensity projections.
The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source images can also be displayed based on physician preferences using configurable viewing options or hanging protocols.
Visia™ Dynamic Review provides the clinician with a broad set of viewing and analysis tools to annotate, measure, and output selected image views or user defined reports. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues.
Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation.
#### Intended Use
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic resonance imaging (MRI) studies. Visia™ Dynamic Review supports evaluation of dynamic MR data.
Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient motion and visualizes different enhancement characteristics (parametric imaqe maps). Furthermore, it performs other user-defined post-processing functions such as image subtractions: multi-planar reformats and maximum intensity projections. The resulting information can be displayed in a variety of formats, including a parametric image overlaid onto the source image. Visia™ Dynamic Review can also be used to provide measurements for diameters, areas and volumes. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues.
Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation.
When interpreted by a skilled physician, Visia™ Dynamic Review provides information that may be useful in diagnosis. Patient management decisions should not be made based solely on the results of Visia™ Dynamic Review analysis.
> Visia™ Dynamic Review 510(k) 510(k) Summary -- 11/10/2011 Page 2 of 3
{2}------------------------------------------------
# Predicate Device Information
| Product | Predicate Device Name | Predicate 510(k)<br>Submission Reference |
|-----------------------|-----------------------|------------------------------------------|
| Visia™ Dynamic Review | Syngo BreVis | K090038 |
# Safety and Effectiveness
The Visia™ Dynamic Review labeling contains instructions for use and necessary cautions, warnings and notes to provide for safe and effective use of the device. Risk Management is ensured via MeVis Medical Solution AG's Risk Management procedure, which is used to identify potential hazards. These potential hazards are controlled via software development and verification & validation testing.
# Nonclinical Testing and Performance Information
The complete system configuration has been assessed and tested at the manufacturer's facility and has passed all in-house testing criteria. The Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the software in each operational mode and followed the process documented in the Validation Test Plan.
Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification activities and that the results demonstrated that the predetermined acceptance criteria were met.
#### Technological Characteristics
Visia™ Dynamic Review is a software device that does not contact the patient, nor does it control any life sustaining devices. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians.
A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.
#### Conclusion
The 510(k) Pre-Market Notification for Visia™ Dynamic Review contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device. MeVis Medical Solutions has determined that its device, Visia™ Dynamic Review, is substantially equivalent to the identified predicate device listed above. A comparison with the legally marketed predicate device indicates that it is substantially equivalent to this device, and that it does not raise any new safety or efficacy concerns. Nonclinical tests demonstrate that the device is safe, effective, and is substantially equivalent to the predicate device.
> Visia™ Dynamic Review 510(k) 510(k) Summary - 11/10/2011 Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. To the left of the eagle is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
MeVis Medical Solutions AG % Mr. Thomas E. Tynes President, CEO MeVis Medical Solutions, Inc. N27 W24075 Paul Court, Suite 100 PEWAUKEE WI 53072
DEC 3 0 2011
Re: K113337
Trade/Device Name: Visia™ Dynamic Review Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: November 10, 2011 Received: November 14, 2011
Dear Mr. Tynes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MeVis Medical Solutions. The logo is in black and white, with the word "MeVis" in large, bold letters. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in smaller letters. The logo is simple and professional, and it is likely used to represent the company's brand.
# Indications for Use Statement
510(k) Number (if known): K113337
Device Name: Visia™ Dynamic Review
Indications for Use:
Visia™ Dynamic Review is a software package intended for use in viewing and analyzing magnetic Visia ~ Dynamic Review is a sommer Review supports evaluation of dynamic MR data.
Visia™ Dynamic Review automatically registers serial patient motion to minimize the impact of patient visia Dynamic Koviow automatics (parametric image maps). Furthermore, it motion and viousless andrer on one processing functions such as image subtractions; multi-planar reformats performs other user delinou por processions information can be displayed in a variety of formats, and maximally projections in the source image. Visia™ Dynamic Review can also be used including a paraments for diameters, areas and volumes. Furthermore, Visia™ Dynamic Review can evaluate the uptake characteristics of segmented tissues.
Visia™ Dynamic Review also displays images from a number of other imaging modalities; however, these images must not be used for primary diagnostic interpretation.
When interpreted by a skilled physician, Visia™ Dynamic Review provides information that may be useful while merplosed by a chilled by a child not be made based solely on the results of Visia™ Dynamic Review analysis.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
ن پن
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S Postel
Office of In Vitro Diagnostic Device
510(k) K.11337
Visia™ Dynamic Review 510(k) Indications for Use Statement - 11/10/2011 Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.