EVOLUTION MP ADAPTIVE CS INSERT

K113325 · Wrightmedicaltechnologyinc · JWH · Dec 9, 2011 · Orthopedic

Device Facts

Record IDK113325
Device NameEVOLUTION MP ADAPTIVE CS INSERT
ApplicantWrightmedicaltechnologyinc
Product CodeJWH · Orthopedic
Decision DateDec 9, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EVOLUTION® MP Adaptive CS Insert is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: 1. Noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2. Inflammatory degenerative joint disease including rheumatoid arthritis; 3. Correction of functional deformity; 4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques. The EVOLUTION® Total Knee System is for cemented use only.

Device Story

The EVOLUTION® MP Adaptive CS Insert is a tibial insert component designed to enable compatibility between the EVOLUTION® Total Knee System femoral component and the ADVANCE® Knee System tibial base. It is manufactured from UHMWPE and is available in 11 sizes with thicknesses ranging from 10mm to 20mm. The device is intended for use by orthopedic surgeons in a clinical/surgical setting during knee arthroplasty procedures. It functions as a bearing surface within the knee joint prosthesis. The device is implanted to restore joint function and alleviate pain associated with degenerative joint disease or trauma. Clinical benefit is derived from its role in a semi-constrained cemented prosthesis system, allowing for revision or primary arthroplasty in patients with specific joint pathologies.

Clinical Evidence

Bench testing only. The device was evaluated through mechanical testing and engineering analyses, including static stability, contact area, and range of motion testing. No clinical data was provided.

Technological Characteristics

Tibial insert manufactured from UHMWPE. Available in 11 sizes (left/right) with 10-20mm thickness. Designed for cemented use as part of a semi-constrained patellofemorotibial/femorotibial prosthesis. Mechanical testing performed for stability and range of motion.

Indications for Use

Indicated for skeletally mature patients undergoing knee arthroplasty for noninflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, revision procedures, or unmanageable fractures. Cemented use only.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KI13325 ### 510(K) Summary of Safety and Effectiveness In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the use of the EVOLUTION® MP Adaptive CS Insert. | Submitted By: | Wright Medical Technology, Inc.<br>5677 Airline Rd, Arlington TN, 38002<br>(800) 238-7188 | |--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date: | November 9, 2011 | | Contact Person: | Danielle Mueller<br>Regulatory Affairs Specialist II | | Proprietary Name: | EVOLUTION® MP Adaptive CS Insert | | Common Name: | Tibial Insert | | Classification Name and Reference: | 21 CFR 888.3560 Knee joint Patellofemorotibial<br>Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis Class I<br><br>21 CFR 888.3530 Knee joint Femorotibial<br>Metal/Polymer Semi-Constrained Cemented Prosthesis Class II | | Subject Product Code and Panel Code: | Orthopedics/87/ JWH, HRY | | Predicate Devices: | EVOLUTION® MP Total Knee System<br>ADVANCE® Total Knee System<br>AXIOM® Total Knee System<br>510(k)s: K894334, K972626, K972770, K033890<br>K093552, K102380 | #### DEVICE INFORMATION #### A. Intended Use The EVOLUTION® MP Adaptive CS Insert is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: - 1. Noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; - 2. Inflammatory degenerative joint disease including rheumatoid arthritis; - Correction of functional deformity; 3. - Revision procedures where other treatments or devices have failed; and treatment of fractures 4. that are unmanageable using other techniques. The EVOLUTION® Total Knee System is for cemented use only. \$\rho\$ 1.\$f2\$ {1}------------------------------------------------ KN-3325 #### B. Device Description The purpose of this submission is to introduce a new EVOLUTION® Adaptive CS Tibial Insert that allows use of a femoral component from the EVOLUTION® Total Knee System with a tibial base from the ADVANCE® Knee System. The design features of the EVOLUTION® Adaptive CS Insert are summarized below: - . Tibial inserts manufactured from UHMWPE - . Available in 11 sizes, left and right - . Tibial insert thickness: 10 - 20mm The EVOLUTION® Adaptive CS Insert was evaluated via mechanical testing and engineering analyses; including static stability, contact area, and range of motion testing. A review of these results indicates that the subject device is equivalent to predicate devices and is capable of withstanding expected in vivo loading without failure. #### C. Substantial Equivalence Information The indications for use of the EVOLUTION® Adaptive CS Insert are identical to the previously cleared predicate devices. The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the EVOLUTION® Adaptive CS Insert are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification. p. 2 of 2 {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Wright Medical Technologies, Inc. % Ms. Danielle Mueller Regulatory Affairs Specialist II 5677 Airline Road Arlington Tenessee 38002 DEC - 9 2011 ## Re: K113325 Trade/Device Name: EVOLUTION® MP Adaptive CS Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, HRY Dated: November 9, 2011 Received: November 10, 2011 Dear Ms. Mueller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ ## Page 2 - Ms. Danielle Mueller forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, ely yours, for N. Malkerson Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ KI13325 # Indications for Use 510(k) Number (if known): Device Name: EVOLUTION® MP Adaptive CS Insert Indications For Use: The EVOLUTION® MP Adaptive CS Insert is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: - Noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or 1. avascular necrosis; - 2. Inflammatory degenerative joint disease including rheumatoid arthritis; - 3. Correction of functional deformity; - 4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques. The EVOLUTION® Total Knee System is for cemented use only. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K113325 Page 1 of _ 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...