The Rumble Tuff Electric Breast Pump is an electrically powered single-user device to express and collect milk from the breasts of lactating women. The device is not intended for hospital use.
Device Story
The Rumble Tuff Electric Breast Pump is a powered, single-user device for expressing and collecting breast milk. It utilizes a diaphragm-type vacuum pump driven by a microcontroller-controlled DC electric motor. The system includes a vacuum relief valve operated by a solenoid to regulate suction. Users control cycle speed and vacuum levels via a roller-type adjuster. Models feature either LED indicators or an LCD screen for displaying pumping information. Power options include AA batteries, a 7.4V Li-ion battery, or an AC adapter. The device is intended for home use by the lactating woman. By providing adjustable suction and cycle speeds, the device facilitates milk expression, potentially benefiting the user by supporting breastfeeding goals. Patient-contacting components are made of polypropylene and silicone.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included vacuum level and cycle speed verification. Biocompatibility testing conducted per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (skin sensitization and irritation). Safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Materials: Polypropylene and silicone. Principle: Diaphragm-type vacuum pump with solenoid-driven vacuum relief valve. Energy: 4 AA batteries, 7.4V Li-ion battery, or 12V AC/DC adapter. Connectivity: None. Software: Microcontroller-based control of motor and vacuum. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for lactating women requiring an electrically powered, single-user device to express and collect breast milk. Not for hospital use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Lansinoh Affinity Double Electric Breast Pump (DEBP) (K092783)
Submission Summary (Full Text)
{0}------------------------------------------------
# Acute Ideas Co., Ltd.
3F, No.11, Lane 35, Jihu Road, Neihu Dist., Taipei, Taiwan
3F, No.11, Lane 35, Jihu Road, Neihu Dist., Taipei, Taiwan
Sr, No.11, Lane 35, Jihu Road, Neihu Dist., Taipei, Taiwan
TEL: 886-2-8751 4868 FAX: 886-2-8751 5868
# NOV 2 0 2012
.
# 510(K) Summary
ر م
:
| Applicant: | Acute Ideas Co., Ltd. |
|----------------------|----------------------------------------------------------------------|
| | 3F, No. 11, Lane 35, Jihu Road, |
| | Neihu Dist, Taipei, Taiwan, 114, R.O.C. |
| | Tel: 886-2-87514868 |
| | Fax: 886-2-87515868 |
| | www.acuteideas.com |
| | Attn: Jessica Wang |
| Contact: | Matthew Kho |
| | A Cute Baby, Inc. |
| | 865 N 1430 W |
| | Orem UT 84057 USA |
| | Tel: 801.609.8168 |
| | Fax: 801.796.2688 |
| | |
| Date of Submission: | 11/01/2012 |
| Proprietary Name: | Rumble Tuff Electric Breast Pump |
| | (Models: PA200S, PA201S, PA201D, PA203S and PA203D) |
| Common Name: | Electric Breast Pump |
| Classification name: | Pump, Breast, Powered |
| Regulatory Class: | Class II |
| Product Codes: | HGX |
| Indication for Use: | The Rumble Tuff Electric Breast Pump is an electrically powered |
| | single-user device used to express and collect milk from the breasts |
| | of lactating women. The device is not intended for hospital use. |
| | |
| Predicate Device(s): | Swing Breastpump (K053052) By Medela, Inc. and Lansinoh |
| | Affinity Double Electric Breast Pump (DEBP) (K092783) By |
| | ENKO Ltd. |
| | 1 of 5 |
{1}------------------------------------------------
Acute Ideas Co., Ltd 3F, No.11, Lane 35, Jihu Road, Neihu Dist., Taipei, Taiwan TEL: 886-2-8751 4868 FAX: 886-2-8751 5868
The Rumble Tuff Electric Breast Pump is a powered Breast Device Description: Pump. With different models, the Pumping can be performed on one breast or on both breasts at the same time; the Rumble Tuff Electric Breast Pump can be powered by 4 AA batteries, one 7.4V Li batteries or an AC Adapter provided with the pump. The pumping system consists of a diaphragm-type vacuum pump which is driven by a microcontroller controlled DC electric motor. Rumble Tuff Electric Breast Pump has five models; PA200S, PA201S, PA201D, PA203S, and PA203D. The suffix "S" stands for the model with single collection kit, and "D" stands for the model with dual collection kits. The user interface of PA200S uses 4 LEDs for indicating pumping cycle and uses a roller type adjuster for pumping level adjustment and indicating pumping level. Models PA201S, PA201D, PA203S, and PA203D all use the same user interface, a LCD screen for showing the pumping information; it also uses a roller type adjuster for pumping level adjustment. For all models, the user is able to control the cycle speed and vacuum level. The summary of the 5 models are as follows:
| Model | PA200S | PA201S | PA203S |
|----------------------|----------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------|
| Pumping Suction | 125 - 250 mmHg | 115 - 250 mmHg | 115 - 250 mmHg |
| Suction Settings | 7 Levels | a) Stimulation (8 Levels)<br>b) Expression (8 Levels) | a) Stimulation (8 Levels)<br>b) Expression (8 Levels) |
| Power Supply | a) 4 AA Alkaline Batteries<br>b) 12V AC/DC Adapter | a) 7.4V Li-ion Battery<br>b) 12V AC/DC Adapter | a) 4 AA Alkaline Batteries<br>b) 12V AC/DC Adapter |
| Pumping Option | Single | Single | Single |
| Back Flow Protection | Yes | Yes | Yes |
| Interface (display) | LED indicators | LCD Screen | LCD Screen |
| Pumping Cycles | up to 1.67/sec | Up to 1.83/sec | Up to 1.11/sec |
#### Single Pumping Models
{2}------------------------------------------------
## Acute Ideas Co., Ltd
3F, No.11, Lane 35, Jihu Road, Neihu Dist., Taipei, Taiwan TEL: 886-2-8751 4868 FAX: 886-2-8751 58
| Model | PA201D | PA203D |
|----------------------|---------------------------|----------------------------|
| Pumping Suction | 85 – 250 mmHg | 85 – 250 mmHg |
| Suction Settings | a) Stimulation (8 levels) | a) Stimulation (8 levels) |
| | b) Expression (8 levels) | b) Expression (8 levels) |
| Power Supply | a) 7.4V Li-Ion Battery | a) 4 AA Alkaline Batteries |
| | b) 12V AC/DC Adapter | b) 12V AC/DC Adapter |
| Pumping Option | Single or Double | Single or Double |
| Back Flow Protection | Yes | Yes |
| Interface (display) | LCD Screen | LCD Screen |
| Pumping Cycles | Up to 1.23/sec | Up to 1.23/sec |
#### Double Pumping Models
The patient-contacting portions of the Rumble Tuff Electric Breast Materials Used: Pump are composed of polypropylene and silicone. The patientcontacting portions of the predicate devices also included polypropylene. The polypropylene and silicone in the Rumble Tuff were evaluated for biocompatibility per ISO 10993 (see testing description in the "Testing" section) and deemed biocompatible. Testing: The Rumble Tuff Electric Breast Pump is tested to meet: . IEC 60601-1:2005, "Medical Electrical Equipment, General Requirements for Safety" ● IEC 60601-1-2:2007, "Medical electrical equipment. General requirements for basic safety and essential performance. Collateral standard. Electromagnetic compatibility. " ● ISO 10993-5: 2009, "In Vitro Cytotoxicity Test of Expression collection Kit"
- ISO 10993-10: 2010, "Skin Sensitization Test of Expression collection Kit (polar and non-polar)"
- ISO 10993-10: 2010, "Skin irritation Test of Expression collection Kit {polar and non-polar}"
- Additionally, the tests of vacuum levels and cycle speeds conducted by Acute Ideas are provided to demonstrate the performance of Rumble Tuff Electric Breast Pump.
{3}------------------------------------------------
#
TEL: 886-2-8751 4868 FAX: 886-2-8751 5868
## Comparison to Predicate Devices:
The following table summarizes the basic similarities and differences between Rumble Tuff Electric Breast Pump and predicate devices.
#### For single pumping models:
| | New Device | New Device | New Device | Predicate Device |
|-----------------------------------------------|------------------------------------------------------------|----------------------------------------------------------------------|------------------------------------------------------------|--------------------------------------------------------|
| Device Name | Rumble Tuff Electric<br>Breast Pump -<br>PA200S | Rumble Tuff Electric<br>Breast Pump -<br>PA2015 | Rumble Tuff Electric<br>Breast Pump -<br>PA203S | Medela Swing |
| 510(k) # | K113315 | K113315 | K113315 | K053052 |
| Product Code | HGX | HGX | HGX | HGX |
| Classification | 2 | 2 | 2 | 2 |
| Intended use | to express milk | to express milk | to express milk | to express milk |
| Power Source | a) 4 AA alkaline<br>batteries<br>b) 12V AC/DC<br>Adapter | a) 7.4V Li-Ion<br>Rechargeable<br>Battery<br>b) 12V AC/DC<br>Adapter | a) 4 AA Alkaline<br>Batteries<br>b) 12V AC/DC<br>Adapter | a) 4 AA alkaline<br>batteries<br>b) 4.8V DC<br>Adapter |
| Pump type | Diaphragm type | Diaphragm type | Diaphragm type | Diaphragm type |
| Pumping<br>Suction | 125 – 250 mmHg | 115 - 250 mmHg | 115 – 250 mmHg | 0 - 250 mmHg |
| Adjustable<br>suction level | Yes | Yes | Yes | Yes |
| Pumping Option | Single | Single | Single | Single |
| Back Flow<br>Protection | Yes | Yes | Yes | No |
| Let Down<br>Function | Yes | Yes | Yes | Yes |
| Cycling /<br>Suction Control<br>Mechanism | Microcontroller | Microcontroller | Microcontroller | Microcontroller |
| Design of<br>vacuum relief | Using a vacuum<br>relief valve driven<br>by a solenoid | Using a vacuum<br>relief valve driven<br>by a solenoid | Using a vacuum<br>relief valve driven<br>by a solenoid | Using a vacuum<br>relief valve driven<br>by a solenoid |
| Material (that<br>may contact<br>user's body) | PP for Breastshield;<br>Silicone for<br>Breastshield cover | PP for Breastshield;<br>Silicone for<br>Breastshield cover | PP for Breastshield;<br>Silicone for<br>Breastshield cover | PP for Breast Shield |
{4}------------------------------------------------
# Acute Ideas Co., Ltd.
3F, No.11, Lane 35, Jihu Road, Neihu Dist., Taipei, Taiwan
| | New Device | New Device | Predicate Device |
|-----------------------------------------------|----------------------------------------------------------------|------------------------------------------------------------|-----------------------------------------------------|
| Device Name | Rumble Tuff Electric Breast<br>Pump - PA201D | Rumble Tuff Electric Breast<br>Pump - PA203D | Lansinoh DEBP |
| 510(k) # | K113315 | K113315 | K092783 |
| Product Code | HGX | HGX | HGX |
| Classification | 2 | 2 | 2 |
| Intended use | to express milk | to express milk | to express milk |
| Power Source | a) 7.4V Li-Ion<br>Rechargeable Battery<br>b) 12V AC/DC Adapter | a) 4 AA Alkaline Batteries<br>b) 12V AC/DC Adapter | a) 6 AA alkaline batteries<br>b) 9V DC Adapter |
| Pump type | Diaphragm type | Diaphragm type | Diaphragm type |
| Pumping Suction | 85 - 250 mmHg | 85 - 250 mmHg | 50 - 250 mmHg |
| Adjustable<br>suction level | Yes | Yes | Yes |
| Pumping Option | Single or Double | Single or Double | Single or Double |
| Back Flow<br>Protection | Yes | Yes | Yes |
| Let Down<br>Function | Yes | Yes | Yes |
| Cycling/Suction<br>Control<br>Mechanism | Microcontroller | Microcontroller | Microcontroller |
| Design of vacuum<br>relief | Using a vacuum relief valve<br>driven by a solenoid | Using a vacuum relief valve<br>driven by a solenoid | Using a vacuum relief valve<br>driven by a solenoid |
| Material (that<br>may contact user's<br>body) | PP for Breastshield;<br>Silicone for Breastshield<br>cover | PP for Breastshield;<br>Silicone for Breastshield<br>cover | PP + TPR for Breast Shield |
#### For double pumping models:
#### Conclusion:
The Rumble Tuff Electric Breast Pumps are substantially equivalent to the predicate devices. Based upon the test data submitted, the devices provide sufficient vacuum pressure to safely and effectively express and collect milk from lactating women.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird-like figure.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
#### November 20, 2012
Acute Ideas Co., Ltd. % Mr. Matthew Kho Director A Cute Baby, Inc. 865 N 1430 W OREM UT 84057
Re: K113315
Trade/Device Name: Rumble Tuff® Electric Breast Pump (Models: PA200S, PA201S, PA201D, PA203S and PA203D) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered breast pump Regulatory Class: II Product Code: HGX Dated: November 6, 2012 Received: November 6, 2012
Dear Mr. Kho:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and I va must compy "Harth labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### 510(k) Number: K113315
Device Name: Rumble Tuff® Electric Breast Pump (Models: PA200S, PA201S, PA201D, PA203S and PA203D)
Indication for Use:
The Rumble Tuff Electric Breast Pump is an electrically powered single-user device to express and collect milk from the breasts of lactating women. The device is not intended for hospital use.
Prescription Use ___
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
(Please DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fisher -S 2012.11.20 15:05:38 -05'00'
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K113315 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.