K113311 · Biocer Entwicklungs GmbH · FTL · Jul 17, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113311
Device Name
TIO2MESH TM
Applicant
Biocer Entwicklungs GmbH
Product Code
FTL · General, Plastic Surgery
Decision Date
Jul 17, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TiO2Mesh™ is a surgical mesh implant specifically indicated for repair of tissue defects of the abdominal wall, where a non-reabsorbable support material is required. Relevant applications include the repair of inguinal and incisional hernia in all common surgical methods. TiO2Mesh™ can be used for both laparoscopic and other open procedures.
Device Story
TiO2Mesh™ is a surgical mesh implant composed of monofilament polypropylene thread with a titanium dioxide surface coating. It features a large-pore structure and blue orientation stripes to assist in surgical placement. The mesh is highly flexible and elastic, designed to accommodate body dynamics. It is available in various shapes and sizes for use by surgeons in both laparoscopic and open abdominal wall repair procedures. The device provides non-reabsorbable structural support for tissue defects, such as inguinal and incisional hernias. It is intended for prescription use only.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including cytotoxicity, irritation, sensitization, acute systemic toxicity, implantation studies (7 and 90 days), extractable organic substance analysis (GC/MS), and bacterial endotoxin testing.
Technological Characteristics
Material: Polypropylene monofilament with titanium dioxide coating. Structure: Large-pore (2.8 mm), flexible, elastic fabric with blue orientation stripes. Grammage: 43 g/m². Sterilization: Ethylene oxide (ETO). Connectivity: None (implantable).
Indications for Use
Indicated for patients requiring non-reabsorbable support for the repair of abdominal wall tissue defects, including inguinal and incisional hernias, via laparoscopic or open surgical procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
K113311
# JUL 1 7 2012
Biocer Entwicklungs GmbH Traditional 510(k) Premarket Submission TiO2Mesh'"
# Section 5 – 510(k) Summary for TiO2Mesh™
#### 1. Submission Sponsor
Biocer Entwicklungs GmbH Address: Ludwig-Thomas Strasse 36c, Bayreuth COUNTRY: Germany Phone: 49 0 921-78-77-70-0 Fax: 49 0 921-78-77-70-79 Contact: Martina Feldmann Ph.D., Product Manager.
## 2. Submission Correspondent
Emergo Group 611 West 5th Street, Third Floor Austin, TX 78701 Cell Phone: Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Cheryl Fisher, Sr. QA/RA Consultant Email: project.management@emergogroup.com
#### Date Prepared ന്
17 July 2012
#### 4. Device Name
Trade/Proprietary Name: TiO2Mesh™ Common/Usual Name: Surgical Mesh Classification Name: Surgical Mesh Polymeric Classification Regulation: 878.3300 Surgical Mesh Classification Panel: 878 General and Plastic Surgery Product Code: OXJ, Mesh, Surgical, non-absorbable, large Abdominal Wall defects FTL, Mesh, Surgical, polymeric
Device Class: 2
FDA Establishment Registration #: Not available yet.
#### 5. Predicate Devices
GfE Medizintechnik GmbH ; TiMeshTC Bard; SoftMesh
{1}------------------------------------------------
# ા 33ના
Biocer Entwicklungs GmbH Traditional 510(k) Premarket Submission TiO2Mesh™
#### 6. Device Description
TiO2Mesh™ is made from a monofilament polypropylene thread and has a large-pore structure with blue orientation stripes. The surgical mesh implant has a titanium dioxide coated surface. The fabric can be stretched in both directions and is highly flexible to react to body dynamics in terms of elasticity. TiO2Mesh™ is available in different shapes and sizes (product flyer TiO2Mesh™ ).
## 7. Intended Use
TiO₂Mesh™ is a surgical mesh implant specifically indicated for repair of tissue defects of the abdominal wall, where a non-reabsorbable support material is required. Relevant applications include the repair of inguinal and incisional hernia in all common surgical methods. TiO2Mesh™ can be used for both laparoscopic and other open procedures.
## 8. Technological Characteristics and Substantial Equivalence
The following table compares the TiO2MeshTM to the TiMesh-TC with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | Biocer Entwicklungs<br>GmbH | GfE Medizintechnik<br>GmbH | Bard |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | TiO2 Mesh | TiMesh-TC | Soft Mesh |
| 510(k) Number | Not Available yet | KO31225 | KO52155 |
| Product Code | FTL | FTL | FTL |
| Regulation<br>Number | 878.3300 | 878.3300 | 878.3300 |
| Regulation<br>Name | Surgical mesh. | Surgical mesh.<br>The TiMESH also | Surgical mesh.<br>Bard® Large Pore |
| Indications for<br>use: | TiO2Mesh™ is a<br>surgical mesh<br>implant specifically<br>indicated for repair<br>of tissue defects<br>of the abdominal<br>wall, where a non-<br>reabsorbable<br>support is<br>required. Relevant<br>applications<br>include the repair | known as TiMESH-TC,<br>along with Ethicon<br>Inc. Prolene Soft<br>Mesh, U.S.S.C.'s Non<br>Absorbable<br>Polypropylene<br>Surgical Mesh are<br>intended to be used<br>for the<br>reinforcement of<br>tissue during surgical<br>repair. Thus, the<br>TiMESH also known<br>as TiMESH-TC mesh<br>and all the predicates | Soft Mesh is<br>indicated for use in<br>the reconstruction<br>of soft tissue<br>deficiencies, such<br>as for the repair of<br>hernias and chest<br>wall defects. Bard<br>Large<br>Pore Soft Mesh<br>Pre-Shaped is<br>indicated for the<br>repair of inguinal<br>hernia defects |
| TiO2Mesh™<br>Manufacturer | Biocer Entwicklungs<br>GmbH | GfE Medizintechnik<br>GmbH | Bard |
| Trade Name | TiO2 Mesh | TiMesh-TC | Soft Mesh |
| | of inguinal and<br>incisional hernia in<br>all common<br>surgical methods.<br>TiO2Mesh™ can<br>be used for both<br>laparoscopic and<br>other open<br>procedures. | have the same<br>intended use. The<br>TiMESH is specifically<br>indicated for<br>laparoscopic and<br>open surgery for the<br>repair of direct or<br>indirect inguinal,<br>femoral, umbilical,<br>and incisional<br>hernias; rectal,<br>vaginal, and apical<br>prolapses; and<br>parietal<br>reinforcement of<br>tissues and<br>abdominal wall<br>repair. The TiMESH<br>has the same<br>indications for use as<br>a combination of the<br>predicated devices<br>TIMESH is a<br>perscriptive device<br>and should only be<br>used by a licensed<br>physician. | |
| Material | PP + Titanium | PP + Titanium | PP |
| Grammage | 43 | 65 | 44 |
| Pore size {mm} | 2.8 | 1 | 2.5 |
| Biocompatible | Yes | Yes | Yes |
| Physical<br>Properties | Comparable<br>FDA Guidance<br>used | Comparable<br>FDA Guidance used | Comparable<br>(unknown whether<br>FDA Guidance was<br>used) |
#### Comparison Table
{2}------------------------------------------------
K113311
Biocer Entwicklungs GmbH Traditional 510(k) Premarket Submission TiO2Mesh™
## 9. Non-Clinical Testing
- a. In vitro Cytotoxicity Assay: Cell Growth Analysis via BCA-Staining with an Extract of TiO2Mesh™
- ف Irritation Test (Intracutaneous Reactivity) with TiO2Mesh™
- Test for Sensitisation (Local Lymph Node Assya LLNA) TiO2Mesh™ C.
·
{3}------------------------------------------------
Biocer Entwicklungs GmbH Traditional 510(k) Premarket Submission TiO2Mesh™
- Acute Systemic Toxicity in the Mouse with 4 Extracts of TiO2Mesh™ di
- Test for the Local Effects after Implantation (7 days period) with TiO2Mesh™ e.
- f. Test for the Local Effects after Implantation (90 days period) with TiO2Mesh™
- g. Investigation of Extractable Organic Substances (GC/MS Fingerprint) after Liquid
- Extraction of a Surgical Mesh Implant h.
- Test for Bacterial Endotoxins using Limulus-Amoebocyte-Lysate (LAL-Test) Kinetic i.
- j. Turbidimetric Assay (KTA): Test for inhibition or enhancement with TiO2Mesh™
- k. Transport simulation and packaging of TiO2Mesh™ - Validation Report
- 1. ETO Sterilization TiO2Mesh™ - Validation report
#### 10. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The substantial equivalence of the device is supported by the non-clinical testing. The verification and validation testing of the device was found to be acceptable and supports the claims of substantial equivalence and product safety and effectiveness assuming substantially equivalent technical characteristics are the basis for the safety and effectiveness of the device.
#### 11. Conclusion
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
The TiO2Mesh™, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 1 7 2012
Biocentw % Emergo Group Ms. Cheryl Fisher Senior QA/RA, Consultant 611 West 5th Street, 3rd Floor Austin, Texas 78701
Re: K113311
Trade/Device Name: Ti02Mesh™ Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL, OXJ Dated: June 01, 2011 Received: June 04, 2011
## Dear Ms. Fisher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
# Page 2 - Ms. Cheryl Fisher
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
Mark N. Melkers
Division of Surgical, Orthopedic
and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Director
Enclosure
{6}------------------------------------------------
Biocer Entwicklungs GmbH Traditional 510(k) Premarket Submission TiO2Mesh™
# Section 4 - Indications for Use Statement
## 510(k) Number (if known): K113311
Device Name: TiO2Mesh™
#### Indications for Use:
TiO Mesh™ is a surgical mesh implant specifically indicated for repair of tissue defects of the abdominal wall, where a non-reabsorbable support material is required. Relevant applications include the repair of inguinal and incisional hernia in all common surgical methods. TiO9Mesh™ can be used for both laparoscopic and other open procedures.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use ____________ (21 CFR 801 Subpart C)
510(k) Num
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Knaebel MXM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.