The Cardio7 electrocardiograph is one of the tools that clinicians can use to evaluate and diagnose patient cardiac function. The 12-lead ECG interpretive algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.
Device Story
Cardio7 is an EKG monitoring system used by trained operators in health facilities to acquire, analyze, store, and print 12-lead resting ECG waveforms. The device connects directly to patient leads via a patient cable; it processes signals using internal filters (AC, muscle, baseline drift, low pass) and a 12-lead interpretive algorithm to provide basic measurements (heart rate, PR interval, QRS duration, QT/QTc, P-R-T axis) and analysis of potential cardiac abnormalities. Data is displayed on a built-in color LCD, stored internally or on SD memory, and can be transferred to a PC via network or USB. The device assists clinicians in cardiac diagnosis; however, all computer-generated analysis must be confirmed by a physician using other relevant clinical information. Benefits include enhanced usability and data management for cardiac evaluation.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and compliance with recognized standards, specifically AAMI/ANSI EC11 and various IEC 60601 series standards.
Technological Characteristics
Electrocardiograph; 12-lead resting ECG; 500 samples/sec/channel; built-in color LCD; thermal printer; internal storage/SD slot; network/RS232/USB connectivity. Power: AC (95-240 VAC) or battery. Standards: EC11 (AAMI/ANSI), IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-2-25, ISO 10993-1:2003, ISO 14971:2007.
Indications for Use
Indicated for adult and pediatric patients to evaluate and diagnose cardiac function via 12-lead ECG acquisition and computer-generated analysis of potential cardiac abnormalities.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
K11330le
16 2012
## 1. 510(k) Summary - Basic Information
| 1.1 Submitter | |
|----------------|------------------------------------------------------------------------------------------------|
| Submitter: | Bionet Co., LTD |
| Address: | #1101, 11F E&C Dream Tower III, 197-33<br>Guro-dong, Guro-gu, Seoul, South Korea (ZIP 152-050) |
| Contact: | Marc Goodman |
| Company: | Noblitt & Rueland<br>5405 Alton Parkway, Suite A530<br>Irvine, CA 92604 |
| Voice: | (714) 258-4646 |
| FAX: | (714) 258-3990 |
| Cell: | (949) 872-1011 |
| Email: | marcgood@ca.rr.com |
| Date Prepared: | October 7, 2011 |
### 1.2 Device Name
| Device Name: | Cardio7 |
|----------------------|-----------------------------------------|
| Common Name: | Electrocardiograph |
| Classification Name: | Electrocardiograph (870.2340, Class II) |
### 1.3 Identification of Legally Marketed Device
Substantial equivalence is claimed to the Bionet CardioXP (K102767).
### 1.4 Device Description
Cardio7 is an EKG monitoring system that provides standard EKG and (optional) pulmonary diagnostic functions. It offers enhanced usability, data analysis and data management. It consists of the several components depicted in Figure 1 that perform EKG and pulmonary monitoring, data analysis and data management functions.
## Figure 1: Cardio7 Block Diagram Depiction
Image /page/0/Figure/10 description: The image shows a diagram of the Cardio7 System Unit and its connections to other devices. The Cardio7 System Unit is connected to a Remote PC via a network connector, a Local PC via RS232, and an SD Memory via a built-in SD Slot. The Cardio7 System Unit is also connected to Patient Leads via a Patient Cable.
{1}------------------------------------------------
# 1.5 Intended Use
Cardio7 is intended for use as a diagnostic tool by trained operators in health facilities. It provides the following functions:
- Acquire ECG waveform data from up to twelve (12) leads through surface electrodes . adhered to the patient's body.
- . Input patient data.
- . View, store and print captured data.
- . Analyze captured data; display and print analysis results.
- Retain captured data and analysis results for up to 120 patients. .
- . Transfer retained data to a PC (via server IP) or directly to insertable USB Memory.
# 1.6 Comparison to Cleared Device
Table 1 compares the relevant features of Cardio7 to the Bionet CardioXP (K102767).
# Table 1: Comparison of Cardio7 to Predicate Device (CardioXP)
| CardioXP | Cardio7 | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Operating<br>Principle | electrocardiographs | same |
| Target<br>Population | Adult and pediatric patients | same |
| ECG<br>Acquisition | Simultaneous 12 leads resting ECG connects to Patient<br>Signal Module. CardioXP System Unit receives data from<br>Patient Signal Module using either (1) direct cable<br>connection or (2) Bluetooth communication. | Simultaneous 12 leads<br>resting ECG connect<br>directly to Cardio7 System<br>Unit. |
| Basic<br>measurement | Heart rate, PR int, QRS dur, QT/OTc, P-R-T axis | same |
| Sampling<br>Rate | 500 samples/sec/channel | same |
| Filters | • AC (50/60 Hz, -20dB or better),<br>• Muscle (25~35Hz, -3dB or better),<br>• Base line drift (0.1Hz, -3dB or better),<br>• Low pass filter(off, 40Hz, 100Hz, 150Hz) | same |
| Display Type | built in color LCD | same |
| Display<br>Functions | View and adjust ECG waveforms. | same |
| Monitor<br>display | ID, HR, Sensitivity, Speed, Date, Power status | HR, ID, date, AC or Battery<br>state, sensitivity, speed,<br>number of saved data,<br>printing form, rhythm lead |
| Patient Data | ID, name, age, birth, gender, height, weight, smoke,<br>department, room number, race | ID, name, age, sex, height,<br>weight, smoke, race |
| User Input | touch screen and rotary push-knob | same |
| Signal quality<br>control | Disconnected lead detection.<br>Pacemaker pulse detection | same |
| Data Storage | • Internal storage for up to 120 ECG<br>• Additional storage available using built in SD RAM slot | same |
| Weight | Approx. 4 kg (8.8 lb) | Approx. 3.5 kg (7.7 lb) |
| Dimensions | 300 x 299 x 123mm<br>(11.8 x 11.8 x 4.8 in) | 296 x 305 x 92 mm<br>(11.7 x 12 x 3.6 in) |
| Power Source | AC power supply (95 ~ 240 VAC, 50/60Hz, 60 VA max)<br>or battery power | same |
{2}------------------------------------------------
| | CardioXP | Cardio7 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Battery | 3 hours of normal use or print 300 ECG pages. Battery recharge to full capacity in 8 hours. | 1 hours of normal use or print 100 ECG pages. Battery recharge to full capacity in 4 hours. |
| Sterility | Wipe exterior of patient cable and electrocardiograph with<br>damp cloth using mild detergent diluted in water.<br>Disinfect patient cable using damp cloth of chemical<br>disinfectants containing one of the following:<br>ethanol (70% - 80%)<br>propanol (70% - 80%)<br>aldehydes (2% - 4%) | same |
| Safety | Recognized and applicable standards:<br>EC11 (AAMI/ANSI) IEC 60601-1-1 IEC 60601-1-2 IEC 60601-1-4 IEC 60601-1-6 IEC 60601-1-8 IEC 60601-2-25 ISO 10993-1:2003 ISO 14971:2007 ETSI EN 301 489 ETSI EN 300 328 EN 60950-1 | Recognized and applicable<br>standards:<br>EC11 (AAMI/ANSI) IEC 60601-1-1 IEC 60601-1-2 IEC 60601-1-4 IEC 60601-1-6 IEC 60601-1-8 IEC 60601-2-25 ISO 10993-1:2003 ISO 14971:2007 |
| Printer | Thermal printer (internal) | same |
| Network<br>Connection | Hardwired Ethernet | same |
| Indication for<br>Use | The CardioXP electrocardiograph is one of the tools that<br>clinicians can use to evaluate and diagnose patient<br>cardiac function. The 12-lead ECG interpretive algorithm<br>provides a computer-generated analysis of potential<br>patient cardiac abnormalities, which must be confirmed<br>by a physician with other relevant clinical information. | same |
#### 1.6.1 Differences between Cardio7 and CP200
The meaningful differences between Cardio7 and CardioXP are the method of ECG data acquisition and the availability of an external monitor.
- . Whereas CardioXP acquires ECG data from its Patient Signal Module, Cardio7 acquires ECG data directly from the Patient Cable. (CardioXP's Patient Signal Module is capable of transmitting ECG data to its system unit from a distance via wireless Bluetooth communication).
- CardioXP provides for an external monitor and Cardio7 does not. .
# 2. Performance Information
Because Cardio7 complies with the EC11 standard, assessment of performance data is not necessary to determine of Substantial Equivalence for this type of device.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 16 2012
Bionet Co., Ltd. c/o Marc Goodman Regulatory Consultant 5405 Alton Parkway 5A, #530 Irvine, CA 92604-3718
Re: K113306
Trade/Device Name: Cardio7 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: July 5, 2012 Received: July 9, 2012
Dear Mr. Goodman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Marc Goodman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm.
Sincerely yours,
Bram Zuckerman, MD
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement
510 (k) Number (if known)
Device Name Cardio7
Indications for Use
The Cardio7 electrocardiograph is one of the tools that clinicians can use to evaluate and diagnose patient cardiac function. The 12-lead ECG interpretive algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K113306
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.