K113280 · Tursiop Technologies, LLC · MOS · Nov 23, 2011 · Radiology
Device Facts
Record ID
K113280
Device Name
CAROTID COIL GE 1.5T
Applicant
Tursiop Technologies, LLC
Product Code
MOS · Radiology
Decision Date
Nov 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Carotid Coil GE1.5T is a receive-only 2-channel diagnostic MR imaging coil designed for use at a static magnetic field strength of 1.5T with the GE Twinspeed HDx 1.5T MRI scanner/system. It is used for obtaining diagnostic images of the carotid arteries in the neck and lower cranium.
Device Story
Receive-only 2-channel RF coil; designed for 1.5T MRI systems (GE Twinspeed HDx). Input: hydrogen nuclei excitation signals from carotid arteries; Output: high-resolution MR images. Used in clinical imaging settings by trained MRI technologists/radiologists. Coil consists of two independent rectangular loops (7 cm x 4.5 cm) with active/PIN diode-enabled blocking circuitry to prevent coupling during transmission. Provides optimized signal-to-noise ratio (SNR) for a 7 cm field of view. Output images assist clinicians in visualizing carotid artery anatomy for diagnostic purposes.
Clinical Evidence
Bench testing only. Side-by-side performance comparison with the GE 3-inch General Purpose Coil demonstrated equivalent SNR and image quality. Compliance with IEC 60601-1, IEC 60601-2-33, UL 94, ISO 10993-1, and NEMA MS-6/MS-9 standards verified safety and effectiveness.
Technological Characteristics
Receive-only 2-channel phased array coil; 1.5T static magnetic field compatibility; polycarbonate housing; active/PIN diode-enabled blocking circuitry; 7 cm x 4.5 cm element dimensions; 7 cm field of view. Standards: IEC 60601-1, IEC 60601-2-33, UL 94 (flammability), ISO 10993-1 (biocompatibility), NEMA MS-6, NEMA MS-9.
Indications for Use
Indicated for patients requiring diagnostic MR imaging of the carotid arteries in the neck and lower cranium using a GE Twinspeed HDx 1.5T MRI system.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
NOV 23 2011
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## A. GENERAL INFORMATION
| Classification Name | Magnetic Resonance Diagnostic Device (21 CFR 892.1000) | |
|-------------------------------|---------------------------------------------------------------------------------------------------------|--|
| Device Trade Name: | Carotid Coil GE1.5T | |
| Applicant's Name and Address: | Tursiop Technologies, LLC<br>11000 Cedar Ave. Suite 280<br>Cleveland OH 44106 | |
| Submitter: | Raju Viswanathan<br>Chief Technology Officer<br>Tursiop Technologies, LLC<br>11000 Cedar Ave. Suite 280 | |
| | Telephone: (216) 658-4521<br>Email: raju@tursiop.com | |
#### B. INTENDED USE
The Carotid Coil GE1.5T is a receive-only 2-channel diagnostic MR imaging coil designed for use at a static magnetic field strength of 1.5T with the GE Twinspeed HDx 1.5T MRI scanner/system. It is used for obtaining diagnostic images of the carotid arteries in the neck and lower cranium.
## C. DEVICE DESCRIPTION
The Carotid Coil GE1.5T is a receive-only 2-channel coil for operation at a magnetic field strength of 1.5T with the GE Twinspeed HDx system. The coil is a two-element coil consisting of two independent rectangular loops, each of approximate size 7 cm x 4.5 cm. Coupling of either coil element to the transmitted field is prevented through respective active blocking circuits.
The Carotid Coil GE1.5T is specially designed for MR imaging of the carotid arteries in the region of the neck and lower cranium. The coil provides optimum signal to noise ratio and coverage in the area of the carotid arteries, allowing high-resolution imaging, while the sensitive region of the coil covers an approximately 7 cm Field of View.
## D. PERFORMANCE TESTING & STANDARDS
- IEC 60601-1 (1988): Medical electrical equipment Part 1: General requirements for safety, . including Amendment 1 (1991) and Amendment 2 (1995).
- . IEC 60601-2-33 (2006): Medical electrical equipment-Part 2-33: Particular requirements for the safety of magnetic resonance equipment for medical diagnosis
- . UL 94; Tests for Flammability of Plastic Materials for parts in Devices and Appliance
- . ISO 10993-1 (2009): Biological evaluation of medical devices-Evaluation and testing within a risk management process
- . NEMA: MS-6 (2008) Characterization of Special Purpose Coils for Diagnostic Magnetic Resonance Images
{1}------------------------------------------------
- . NEMA: MS-9 (2008) Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images
- . NEMA Standardized 510(k) Pre-Market Notification Submission Template for RF Coil Accessories Intended for Use with Magnetic Resonance Diagnostic Imaging Devices
## E. PREDICATE DEVICES
The Carotid Coil GE1.5T is substantially equivalent to the predicate devices. The table below identifies similarities between the devices. Although the GE 3-inch General Purpose Coil is in commercial distribution it is not included in the FDA 510(k) database and/or has been purged from the FDA PMA Tursiop Technologies LLC has conducted side by side testing between the Carotid Coil database. GE1.5T and the GE 3-inch General Purpose Coil to verify substantial equivalence.
| Para<br>mete<br>r | Carotid Coil GE1.5T | TT1 General Purpose Coil<br>(K083434) | Predicate Coil: GE 3-inch General<br>Purpose Coil |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| d/Indications for<br>Use<br>Int. | The Carotid Coil GE1.5T is a<br>receive-only 2-channel diagnostic<br>MR imaging coil designed for use<br>at a static magnetic field strength of<br>1.5T with the GE Twinspeed HDx<br>1.5T MRI scanner/system. It is used<br>for obtaining diagnostic images of<br>the carotid arteries in the region of<br>the neck and lower cranium. | The TT1 General Purpose Coil is a<br>receive-only diagnostic MR imaging<br>coil designed for use at a static<br>magnetic field strength of 1.5T with<br>the GE Signa® 1.5T system. It is<br>used for obtaining diagnostic images<br>of a variety of small-to-medium<br>sized anatomical regions, such as<br>jaw, spine, neck, shoulder, thigh,<br>foot, ankle and joints. | The GE 3-inch General Purpose Coil<br>is a receive-only RF coil designed for<br>1.5T MR imaging of various<br>anatomical regions, such as jaw, spine,<br>neck, shoulder, thigh, foot, ankle and<br>joints. |
| Dimensions | Dimensions of each channel<br>element:<br>Coil length - 7 cm<br>Coil width - 4.5 cm<br>Similar in weight | Coil length - 7.5 cm<br>Coil width - 7.5 cm<br>Similar in weight | Coil length - 7.5 cm<br>Coil width – 7.5 cm<br>Similar in weight |
| Coil Architecture | Two channel receive-only for 1.5T<br>operation<br>Housing Type - Polycarbonate | Single channel receive-only for 1.5T<br>operation<br>Housing Type - ABS plastic | Single channel receive-only for 1.5T<br>operation<br>Housing Type - Fiberglass |
| | Primary Decoupling - Active/PIN<br>diode-enabled blocking circuitry for<br>each channel | Primary Decoupling - Active/PIN<br>diode-enabled blocking circuitry | Primary Decoupling - Active/PIN<br>diode-enabled blocking circuitry |
| Performance<br>& Safety | IEC 60601-1, IEC60601-2-33<br>Flammability UL 94, ISO 10993-<br>1,NEMA MS6, NEMA MS9 | IEC 60601-1, IEC60601-2-33<br>Flammability UL 94, ISO 10993-1,<br>NEMA MS6 | IEC 60601-1, IEC60601-2-33<br>Flammability UL 94, ISO 10993-1,<br>NEMA MS6 |
{2}------------------------------------------------
| Principle of<br>Operation | Hydrogen nuclei excitation for<br>imaging of the scanned organ | Hydrogen nuclei excitation for<br>imaging of the scanned organ | Hydrogen nuclei excitation for<br>imaging of the scanned organ |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Technological<br>Characteristics | Optimal Signal to Noise Ratio<br>(SNR) and coverage<br>allowing high-resolution imaging,<br>while the sensitive region of the<br>coil covers an approximately 7 cm<br>Field of View. | Optimal Signal to Noise Ratio<br>(SNR) and coverage<br>allowing high-resolution imaging,<br>while the sensitive region of the coil<br>covers an approximately 10 cm Field<br>of View. | Optimal Signal to Noise Ratio<br>(SNR) and coverage and high<br>resolution imaging |
## F. SAFETY AND EFFECTIVENESS
Tursiop Technologies, LLC complies with FDA's GMPs and is ISO 13485-certified for the design and manufacture of magnetic resonance imaging coils. The Carotid Coil GE1.5T complies with voluntary standards for safety/effectiveness (IEC 60601, UL 94, ISO 10993-1) all of which mandate that relevant components are tested to minimize hazards (electrical, mechanical, flammability, biocompatibility). The patient contacting surface is polycarbonate painted with biocompatible paint. ISO-mandated risk management practices have been utilized to assess potential risks throughout the device life cycle to mitigate unacceptable levels of risk by conducting risk assessment (FMEA) procedures.
Tursiop Technologies, LLC has conducted testing to establish the safety and effectiveness concerning the Carotid Coil GE1.5T. Side by side comparison testing was conducted with one of the predicate devices, GE 3-inch General Purpose Coil which demonstrated substantially equivalent performance. The Carotid Coll GE1.5T does not introduce any new potential safety risks and operates in a manner similar to the predicate devices.
## G. SUBSTANTIAL EQUIVALENCE STATEMENT
The Carotid Coil GE1.5T is substantially equivalent to the GE 3-inch General Purpose Coil and the TT1 General Purpose Coil (K083434).
This opinion is based on the fact that comparing the Carotid Coil GE1.5T technological characteristics, coil architecture and operating principles with that of the predicate devices reveals that the devices comply with the same or equivalent standards and have the same or equivalent intended uses.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Tursiop Technologies LLC. Mr. Mark Job Responsible Third Party Official % Regulatory Technology Services, LLC. 1394 25th Street NW BUFFALO MN 55313
NOV 2 3 2011
Re: K113280
Trade/Device Name: Carotid Coil GE 1.5T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: November 4, 2011 Received: November 7, 2011
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K113280
Device Name: Carotid Coil GE1.5T
Indications for Use:
The Carotid Coil GE1.5T is a receive-only 2-channel diagnostic MR imaging coil designed for use at a static magnetic field strength of 1.5T with the GE Twinspeed HDx 1.5T MRI scanner/system. It is used for obtaining diagnostic images of the carotid arteries in the region of the neck and lower cranium.
Prescription Use x x AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Markel D'Oyle
sion Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of 1
510(k) 113280
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.