← Product Code [CJE](/productcode/CJE) · K113269

# ELITTECH CLINICAL SYSTEMS ALP IFCC SL (K113269)

_Elitechgroup · CJE · Dec 29, 2011 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K113269

## Device Facts

- **Applicant:** Elitechgroup
- **Product Code:** [CJE](/productcode/CJE.md)
- **Decision Date:** Dec 29, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1050
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

ELITech Clinical Systems ALP IFCC SL is intended for the quantitative in vitro diagnostic determination of alkaline phosphatase in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings. Alkaline phosphatase or its isoenzymes measurements are used in the diagnosis and treatment of liver, bone, parathyroid and intestinal diseases.

## Device Story

The ELITech Clinical Systems ALP IFCC SL is an enzymatic colorimetric assay kit used on Selectra ProM analyzers. It utilizes two reagents: R1 (AMP buffer, Mg2+, Zn2+) and R2 (p-Nitrophenylphosphate). The device measures alkaline phosphatase activity in human serum or lithium heparinized plasma. The principle of operation involves the hydrolysis of p-NPP by alkaline phosphatase into inorganic phosphate and p-nitrophenol, which produces a yellow color measured spectrophotometrically at 405 nm. The rate of p-nitrophenol formation is directly proportional to alkaline phosphatase activity. The device is intended for use in clinical laboratories by trained personnel. Results are used by clinicians to aid in the diagnosis and management of liver, bone, parathyroid, and intestinal disorders. The assay provides quantitative results within a measuring range of 20–1023 U/L.

## Clinical Evidence

Bench testing only. Performance validated per CLSI guidelines: EP05-A2 (precision), EP06-A (linearity), EP07-A2 (interference), EP17-A (LoD/LoQ), and EP09-A2 (method/matrix comparison). Precision: Total %CV 2.6–4.4%. Linearity: 20–1023 U/L (R²=0.9996). Method comparison (n=100) vs. predicate: slope 1.025, r=0.998. Matrix comparison (n=40) serum vs. lithium heparin plasma: slope 1.10, r=1.000. No significant interference observed for triglycerides, bilirubin, hemoglobin, ascorbic acid, acetylsalicylic acid, or acetaminophen.

## Technological Characteristics

Kinetic colorimetric assay; measures p-nitrophenol production at 405 nm. Reagents: R1 (AMP buffer pH 10.45, Mg2+, Zn2+), R2 (p-NPP, sodium azide). Liquid, ready-to-use format. Designed for use on Selectra ProM analyzers. Storage: 2-8°C, protected from light. Software-controlled automated analysis on Selectra platform.

## Regulatory Identification

An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

## Predicate Devices

- Roche Diagnostics ALP2S (Alkaline phosphatase acc. to IFCC Gen.2) ([K033185](/device/K033185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k113269

B. Purpose for Submission:
New Device

C. Measurand:
Alkaline phosphatase

D. Type of Test:
Quantitative, enzymatic colorimetric

E. Applicant:
ELITechGroup

F. Proprietary and Established Names:
ELITech Clinical Systems ALP IFCC SL

G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1050; alkaline phosphatase/isoenzymes test system
2. Classification:
Class II
3. Product code:
CJE; Nitrophenylphosphate, alkaline phosphatase or isoenzymes test system
4. Panel:
Clinical Chemistry (75)

H. Intended Use:
1. Intended use:
See Indications for Use
2. Indications for use:
ELITech Clinical Systems ALP IFCC SL is intended for the quantitative in vitro diagnostic determination of alkaline phosphatase in human serum and plasma on ELITech Clinical Systems Selectra analyzers.

It is not intended for use in Point of Care settings.

Alkaline phosphatase or its isoenzymes measurements are used in the diagnosis and treatment of liver, bone, parathyroid and intestinal diseases.

{1}

3. Special conditions for use statements:
- For prescription use only
- For in vitro diagnostic use only
- For use in clinical laboratories only

4. Special instrument requirements:
ELITech Clinical Systems Selectra ProM Analyzer

I. Device Description:

ALP IFCC SL is available as a kit only. It consists of 2 reagents, R1 and R2. Reagent 1 contains 2-Amino-2-methyl-1-propanol (AMP) buffer (pH 10.45), Magnesium ions (2.4 mM), and Zinc ions (1.2 mM). Reagent 2 contains p-Nitrophenylphosphate (p-NPP; 80 mM) and sodium azide (&lt;0.1%).

J. Substantial Equivalence Information:

1. Predicate device name(s):
Roche Diagnostics ALP2S (Alkaline phosphatase acc. to IFCC Gen. 2)

2. Predicate k number(s):
k033185

{2}

3. Comparison with predicate:

|   | ELITech Clinical Systems Device (ALP IFCC SL) (k113269) | Predicate device (Roche Diagnostics ALP2S) (k033185)  |
| --- | --- | --- |
|  Intended use | Same | In vitro test for the quantitative determination of alkaline phosphatase in human serum and plasma.  |
|  Indication for Use | Same | Alkaline phosphatase or its isoenzymes measurements are used in the diagnosis and treatment of liver, bone, parathyroid and intestinal diseases.  |
|  Appearance of reagents | Same | Liquid form, ready to use  |
|  Sample type | Same | Serum Lithium heparinized plasma  |
|  Assay Principle | Same | Colorimetric  |
|  Reagent storage | Same | Store at 2-8 °C the reagent is stable until the expiry date stated on reagent.  |
|  Instrument | Selectra ProM Analyzer | Cobas c111  |
|  Measuring range | 20 – 1023 U/L | 3 – 1200 U/L  |
|  Calibration Frequency | 1 day | 5 days  |
|  On board stability | 1 day | 10 days  |
|  Calibrator | Recommended calibration material (not included): ELITech Clinical Systems ELICAL 2 | Recommended calibration material (not included): Roche Calibrator f.a.s.  |
|  Controls | Recommended quality control material (not included): ELITech Clinical Systems ELITROL I (Normal control) ELITech Clinical Systems ELITROL II (Pathologic control) | Recommended quality control material (not included): Roche Precinorm U Roche Precipath U  |

{3}

K. Standard/Guidance Document Referenced (if applicable):

- CLSI EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Second Edition
- CLSI EP06-A: Evaluation of the Linearity of the Measurement of Quantitative Procedures: a Statistical Approach
- CLSI EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline—Second Edition
- CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantification; Approved Guideline
- CLSI EP09-A2: Method Comparison and Bias estimation Using Patient Samples; Approved Guideline—Second Edition

L. Test Principle:

In the presence of Mg²⁺, Zn²⁺ and AMP as a phosphate acceptor, alkaline phosphatase hydrolyzes p-nitrophenyl phosphate substrate into inorganic phosphate and p-nitrophenol (yellow compound with an absorbance peak at 405 nm). The rate of the formation of p-nitrophenol is directly proportional to the alkaline phosphatase activity as follows:

$$p\text{-NPP} + \mathrm{H}_2\mathrm{O} \longrightarrow \text{Inorganic phosphate} + p\text{-Nitrophenol}$$ (yellow; can be measured using a 405 nm spectrophotometer)

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

This study followed the CLSI EP5-A2 “Evaluation of Precision Performance of Clinical Chemistry Devices” guideline. Within-run and Total Precision studies were performed for alkaline phosphatase. Studies were performed by conducting 2 runs per day, two measurements per run, for 3 levels of serum-based samples and 2 levels of controls, on 2 Selectra ProM analyzers for 20 operating days. Data for level 1 is from a diluted human serum pool sample, level 2 is from a natural serum pool sample, and level 3 is from a spiked human serum pool sample. Two lots of reagent were used for the study. The results are summarized in the table below:

{4}

|  ALP Levels | Level 1 | Level 2 | Level 3 | Elitrol I | Elitrol II  |
| --- | --- | --- | --- | --- | --- |
|  No. of Samples | 80 | 80 | 80 | 80 | 80  |
|  No. of Replicates (per run) | 2 | 2 | 2 | 2 | 2  |
|  Mean (U/L) | 57 | 144 | 262 | 82 | 224  |
|  Total SD | 2.5 | 5.4 | 7.7 | 2.2 | 6.4  |
|  Total %CV | 4.4 | 3.8 | 2.9 | 2.6 | 2.8  |
|  Within-Run SD | 0.8 | 1.3 | 1.5 | 0.6 | 1.9  |
|  Within-Run %CV | 1.3 | 0.9 | 0.6 | 0.7 | 0.8  |

# b. Linearity/assay reportable range:

A linearity study was conducted based on the CLSI EP06-A guideline by comparing observed versus expected values for 11 equally-spaced serum samples. Two pools of patient serum samples were prepared to obtain one high and one low concentration. A high concentration of sample was obtained by spiking the serum pool (1023 U/L) and a low concentration of sample was prepared by diluting the sample pool with buffered saline solution (20 U/L). Using the high and low pooled samples, eleven levels of inter-mixtures with equidistant activities were prepared, with all samples assayed in triplicate on one Selectra ProM instrument. Data was analyzed using  $1^{\text{st}}$ ,  $2^{\text{nd}}$ , and  $3^{\text{rd}}$  order least square regressions. A  $1^{\text{st}}$  order linear regression was generated as follows:

|  T h | Slope | Intercept | R2 | r | Standard Error (U/L) | Concentration Ranges tested  |
| --- | --- | --- | --- | --- | --- | --- |
|  eALP | 0.9681 | 3.1406 | 0.9996 | 0.9998 | 7 | 20-1023 U/L  |

The results of the study support the sponsor's claim that the ALP IFCC SL assay is linear across the measuring range of 20 to  $1023\mathrm{U / L}$ .

# Dilution Study:

The sponsor performed a 1:10 manual dilution study with 10 spiked samples using saline as the diluent. Ten spiked serum samples with ALP analyte concentrations between 991 and  $9874\mathrm{U / L}$  were diluted 1:10. The  $\%$  recoveries between the expected and observed values are within  $10\%$ . Therefore, the sponsor claimed that the ALP concentration greater than the upper claimed measuring range (1023 U/L) can be diluted manually 1:10 with saline.

{5}

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability: The sponsor claims that the Elitech ALP IFCC SL assay is traceable to the IFCC formulation (Tietz, 1983), by manual measurement.

Stability of the reagent: The shelf-life of ALP IFCC SL reagents is 18 months (tested using 2 lots) when stored at 2 to 8°C. The shelf-life stability is based on their real-time stability study, which is still on-going. The on-board stability and calibration frequency is 1 day. Shelf-life and on-board stability study protocols have been provided and found to be adequate.

d. Detection limit:

The Limits of Blank (LoB), Detection (LoD) and Quantitation (LoQ) studies were conducted following the CLSI EP17-A guideline for ALP IFCC SL. Two Selectra Pro M analyzers and two lots of a validated reagent were used. The LoB values were determined by analyzing a blank sample 60 times using 2 reagent lots over 2 analyzers over multiple days. The LoD and LoQ studies each used 4 diluted sample pools prepared from 4 different patient sample pools of known activities in order to obtain an activity close to four times the calculated value of the LoB or the expected LoQ, respectively. LoD and LoQ values were determined by measuring each of the 4 diluted sample pools 15 times (measuring low activity of analyte for LoD and samples diluted in saline for LoQ) in one run across two analyzers using two reagent lots on multiple days (2-5 days). Results are summarized below:

|   | LoB | LoD | LoQ  |
| --- | --- | --- | --- |
|  Detection Limits | 4 U/L | 6 U/L | 20 U/L  |

e. Analytical specificity:

Interference studies were performed according to the CLSI EP7-A2 guideline for alkaline phosphatase. Seven to nine increasing concentrations of potential interferents, tested in triplicate, were spiked into low (150 U/L), medium (250 U/L), and high (900 U/L) levels of pooled patient serum samples. The sponsor's definition of no significant interference is ≤10% difference between the tested and the control samples. There was no significant interference between alkaline phosphatase and interferents in the concentration ranges indicated in the table below:

{6}

|  Interferents | Highest level tested with no interference  |
| --- | --- |
|  Triglycerides | 3141 mg/dL  |
|  Unconjugated bilirubin | 30.0 mg/dL, (513 μmol/L)  |
|  Conjugated bilirubin | 29.5 mg/dL, (504 μmol/L)  |
|  Hemoglobin | 500 mg/dL  |
|  Ascorbic acid | 20.0 mg/dL  |
|  Acetylsalicylic acid | 200 mg/dL  |
|  Acetaminophen | 30 mg/dL  |

The sponsor notes in the label that other compounds may interfere based on the following literature references:

-Young, D. S., Effects of preanalytical variables on clinical laboratory tests, 2nd Ed., AACC Press, (1997).
-Young, D. S., Effects of drugs on clinical laboratory tests, 4th Ed., AACC Press, (1995).
-Berth, M. &amp; Delanghe, J. Protein precipitation as a possible important pitfall in the clinical chemistry analysis of blood samples containing monoclonal immunoglobulins: 2 case reports and a review of literature, Acta Clin Belg., (2004), 59, 263.

f. Assay cut-off: Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

A method comparison study was performed using the ALP IFCC SL reagent on a Selectra ProM analyzer and Roche Diagnostics Alkaline phosphatase acc. to IFCC Gen.2 reagent on a cobas c111 analyzer according to the CLSI EP09-A2 guideline. A total of 100 serum patient samples were used, including 5 diluted and 5 spiked samples. The sample range tested on the candidate device was 20 to 1023 U/L. The result of the linear regression is shown below:

|  Slope (95% Confidence Interval) | Intercept (95% Confidence Interval) | r | R²  |
| --- | --- | --- | --- |
|  1.025 (1.017 to 1.034) | 0.909 (-4 to 3) | 0.998 | 0.997  |

{7}

b. Matrix comparison:

A matrix comparison study was performed with 40 paired serum and plasma (in lithium heparin) samples using one Selectra ProM analyzer, according to the CLSI EP09-A2 guideline. Two samples were diluted and 3 samples were spiked. The sample range tested was 21 to 878 U/L. The result of the linear regression is shown below:

|  Slope (95% Confidence Interval) | Intercept (95% Confidence Interval) | r | R²  |
| --- | --- | --- | --- |
|  1.10 (1.096 to 1.111) | -3 (-5 to -1) | 1.000 | 0.999  |

Based on the data, the sponsor claims that lithium heparin is an acceptable anticoagulant for the collection of plasma samples.

3. Clinical studies:

a. Clinical Sensitivity: Not applicable.
b. Clinical specificity: Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.

4. Clinical cut-off: Not applicable.

5. Expected values/Reference range:

Expected values are cited from “Panteghini, M., Bais, R., Enzyme, Tietz Fundamentals of Clinical Chemistry, 6th Ed., Burtis, C.A., Ashwood, E.R., Bruns, D.E., (Saunders), (2008), 317.”

|   | Men: | Women:  |
| --- | --- | --- |
|  20-50 years old: | 53-128 U/L | 42-98 U/L  |
|  ≥ 60 years old: | 56-119 U/L | 53-141 U/L  |

{8}

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

9

---

**Source:** [https://fda-staging.innolitics.com/device/K113269](https://fda-staging.innolitics.com/device/K113269)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
