K113214 · Terumo Cardiovascular Systems Corp. · DXJ · Feb 7, 2012 · Cardiovascular
Device Facts
Record ID
K113214
Device Name
TLINK DATA MANAGEMENT SYSTEM (DMS)
Applicant
Terumo Cardiovascular Systems Corp.
Product Code
DXJ · Cardiovascular
Decision Date
Feb 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2450
Device Class
Class 2
Indications for Use
The TLink™ DMS is an electronic clinical record keeping and reporting system indicated for use in collecting, displaying, storing and managing data from external medical devices. The system facilitates the creation of electronic patient records and enables post-procedural case reviews. Data and records can be viewed on local workstations or transferred to a central computer or hospital network for storage and post-case analysis/reporting.
Device Story
TLink™ DMS is an electronic clinical record-keeping system for surgical/procedural environments. It interfaces with external medical devices (heart-lung machines, blood parameter monitors, anesthesia monitors) to collect physiological data automatically or via manual entry. Hardware includes touch-screen computers, barcode scanners, and mounting accessories. The system displays real-time data, graphs trends, and performs routine clinical calculations (e.g., Body Surface Area, Fluid Balance). Case templates are managed on a central computer and deployed to satellite workstations in procedure rooms. Data is stored centrally for post-case analysis and reporting (e.g., audit summaries, quality assurance). Clinicians use the system to manage patient records and review procedural outcomes. It benefits patients by centralizing clinical data, improving record accuracy, and facilitating post-procedure analysis.
Clinical Evidence
Bench testing only. System and software design verification and validation testing were conducted to ensure the device meets design inputs and user needs under simulated use conditions. No clinical data was required or provided.
Technological Characteristics
System comprises software, touch-screen computer, barcode scanner, and mounting hardware. Interfaces with external medical devices via network/serial connections. Performs data acquisition, storage, and calculation (e.g., BSA, fluid balance). Connectivity via hospital network or removable media. Software-based data management.
Indications for Use
Indicated for use in clinical settings to collect, display, store, and manage data from external medical devices (e.g., heart-lung machines, patient monitors) to create electronic patient records and facilitate post-procedural case reviews.
Regulatory Classification
Identification
A medical cathode-ray tube display is a device designed primarily to display selected biological signals. This device often incorporates special display features unique to a specific biological signal.
{0}------------------------------------------------
K113214
FEB - 7 2012
## Section 5: 510(k) Summary
This section includes a summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter Information | |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Terumo Cardiovascular Systems Corporation |
| Address | 6200 Jackson Road<br>Ann Arbor MI, 48103 |
| Phone number | Tel: (734) 663-4145 |
| Fax number | Fax: (734) 741-6069 |
| E-mail | Rebecca.andersen@terumomedical.com |
| Establishment<br>Registration Number | 1828100 |
| Name of contact person | Rebecca Andersen, PhD |
| Date prepared | October 28, 2011 |
| Device Information | |
| Trade or proprietary<br>name | TLink™ Data Management System (DMS) |
| Common or usual name | Clinical information management system |
| Classification name | Monitor, Physiological, Patient (Without Arrhythmia<br>Detection or Alarms) |
| Classification panel | 74 Cardiovascular |
| Regulation | 870.2450 |
| Product Code(s) | Medical Cathode-ray tube display<br>DXJ |
| Legally marketed<br>device(s) to which<br>equivalence is claimed | K012349, MetaVision Clinical Information System |
| Reason for 510(k) | Traditional 510(k) for new device |
| Device Information | |
| Device description | The TLink™ DMS consists of the TLink™ software and<br>hardware accessories including computers meeting<br>minimum requirements, data entry devices (barcode laser<br>scanner, touch screen stylus, keyboard), mounting trays and<br>brackets, and serial converters. The system can interface<br>with a variety of external medical devices including, but not<br>limited to, heart-lung machines, blood parameter monitoring<br>systems, centrifugal systems, blood gas devices, patient<br>monitors and anesthesia monitors. Case data can be<br>entered manually by the user or collected automatically from<br>independent medical devices. Screen layouts are<br>customizable to meet hospital and user requirements for<br>patient/case records. Physiological data can be graphed at<br>user defined time intervals for event recording. Certain<br>calculations routinely performed by the clinician during<br>surgery can be performed by the TLink™ DMS, e.g., Body<br>Surface Area (based on patient height and weight data) and<br>Fluid Balance (based on fluid input and output data). Case<br>templates and administrative information are developed on a<br>central computer and transferred to the satellite computer(s)<br>connected to the external medical device(s) in the procedure<br>rooms. Case records are then transferred back to the<br>central computer or hospital information system for central<br>storage and post-case analysis/reporting. All transfers<br>between satellite and central computers are via network<br>and/or removable media. A variety of post-procedure<br>reports are available including case report, clinical activity,<br>case checklist, quality assurance, audit summary report, and<br>audit detail report. |
| Indication for use | The TLink™ DMS is an electronic clinical record keeping<br>and reporting system indicated for use in collecting,<br>displaying, storing and managing data from external<br>medical devices. The system facilitates the creation of<br>electronic patient records and enables post-procedural<br>case reviews. Data and records can be viewed on local<br>workstations or transferred to a central computer or hospital<br>network for storage and post-case analysis/reporting. |
January 20, 2012
Image /page/0/Picture/6 description: The image shows the logo for Terumo. The logo consists of a circle with the letter T inside of it, followed by the word "TERUMO" in all capital letters. There is a small registered trademark symbol to the right of the word "TERUMO".
{1}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
### Section 5: 510(k) Summary
{2}------------------------------------------------
#### Section 5: 510(k) Summary
## Substantial Equivalence - Summary of the technological characteristics of the TLink™ DMS compared to the predicate device - MetaVision
| Characteristic | TLink™ DMS | MetaVision - K012349 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The TLink™ DMS is an electronic<br>clinical record keeping and<br>reporting system indicated for use<br>in collecting, displaying, storing<br>and managing data from external<br>medical devices. The system<br>facilitates the creation of electronic<br>patient records and enables post-<br>procedural case reviews. Data<br>and records can be viewed on<br>local workstations or transferred to<br>a central computer or hospital<br>network for storage and post-case<br>analysis/reporting. | For use in data collection,<br>display, management, and<br>storage in the intensive care unit.<br>The system is used in<br>conjunction with independent<br>patient bedside devices and<br>systems, connected via a<br>network. The way the system is<br>used for generating patient<br>records, computation of drug and<br>fluid dosage and research tasks<br>is determined by the health care<br>providers, in terms of their<br>environment and requirements.<br>The MetaVision application is<br>resident on a workstation that<br>provides for data input and<br>patient data display - to health<br>care professionals. Typically, a<br>MetaVision system comprises<br>several workstations connected<br>via a network system to one or<br>more servers. Data is stored and<br>managed by servers. The<br>MetaVision system network can<br>communicate with a number of<br>remotely located patient care<br>units. |
| System<br>Components | • System software<br>• Touch screen computer (local<br>to monitoring devices)<br>• Barcode laser scanner and<br>holder<br>• Touch screen stylus | • System software<br>• Local workstations connected<br>via hospital network<br>• Bar code scanner for<br>scanning drug vial bar codes |
| Characteristic | TLink™ DMS | MetaVision-K012349 |
| Functionality | • Hardware for mounting touch screen computer (trays, brackets)<br>• Capture device data during procedure and display data in electronic record format customized by the user. Case data can be displayed graphically over time. Clinical events can be marked for subsequent analysis<br>• Generate reports including diagnoses, clinical data, procedures and outcomes. Reports used for post-procedure analysis and quality assurance | • Import data from hospital information systems for pre-op evaluations and patient preparation, such as patient medications, lab reports and imaging studies<br>• Capture device data during procedure and display data in electronic record format customized by the user. Case data can be displayed graphically over time. Clinical events can be marked for subsequent analysis<br>• Generate reports including diagnoses, clinical data, procedures and outcomes. Reports used for post-procedure analysis, quality assurance and billing purposes |
January 20, 2012
TERUMO
{3}------------------------------------------------
## Section 5: 510(k) Summary
1
{4}------------------------------------------------
#### Section 5: 510(k) Summary
## Summary of Performance Tests Conducted for Determination of Substantial Equivalence
| Characteristic | Report | Results Summary |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| System & Software<br>Design Verification<br>Testing | Provides documented evidence that the<br>design outputs for TLink™ DMS V. 2.0<br>continue to meet the existing design inputs of<br>prior software release versions and meet the<br>new design inputs as well. This protocol<br>covers all of the software-related system<br>requirements and includes the additional<br>hardware components. | Pass - Test results<br>demonstrate that the<br>design outputs meet<br>the design input<br>requirements (pre-<br>defined acceptance<br>criteria). |
| System & Software<br>Design Validation<br>Testing | Validates that TLink™ DMS V. 2.0 meets the<br>user needs and intended use under<br>simulated use conditions by addressing the<br>following three areas:<br>Packaging, Labeling and<br>Miscellaneous Installation and Set-up Simulated Use | Pass - Test results<br>demonstrate that<br>intended use and<br>user needs are<br>fulfilled. |
# Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The TLink™ DMS is substantially equivalent to the MetaVision Clinical Data Management System because it has the same intended use and substantially equivalent performance specifications.
Image /page/4/Picture/8 description: The image shows the logo for Terumo Corporation. The logo consists of a circle with the letter 'T' inside, followed by the word 'TERUMO' in capital letters. There is a small circle above and to the right of the 'O' in 'TERUMO'. The logo is simple and recognizable.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB - 7 -2012
Terumo Cardiovascular Systems c/o Dr. Rebecca Andersen 6200 Jackson Rd Ann Arbor, MI 48103
Re: K113214
Trade/Device Name: TLink Data Management System Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-ray tube display Regulatory Class: Class II Product Code: DXJ Dated: January 17, 2012 Received: January 19, 2012
Dear Dr. Andersen:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becare ined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use sured in the encrease) to tog-sey - the Medical Device Amendments, or to conninered pror conner of the in accordance with the provisions of the Federal Food, Drug, de vices that have been require approval of a premarket approval application (PMA). and Cosmetic 71ct () the ) that the device, subject to the general controls provisions of the Act. The I ou may, merciors, manovate a ect include requirements for annual registration, listing of general connects provisions practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability addiciation: Flease noter
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is blacklined (boo acon ) = ) = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = finay be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may found in the Oode of I caeral cogniteming your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Dr. Rebecca Andersen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be advised that I Drimination that your device complies with other requirements of the Act that 1 DA has intact a and regulations administered by other Federal agencies. You must or any I city all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CI K 1 at 607); adeling (21 CFR 803); good manufacturing practice requirements as set de vice-related adverse events) (21 CF 1000), government 820); and if applicable, the electronic form in the quants of ovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rr you active you fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Terumo Cardiovascular Systems Corporation
TLink™ DMS 510(k)
#### Section 4: Indications for Use
510(k) Number: K113214
#### TLink™ Data Management System (DMS) Device Name:
Indications for Use:
The TLink™ DMS is an electronic clinical record keeping and reporting system indicated for use in collecting, displaying, storing and managing data from external medical devices. The system facilitates the creation of electronic patient records and enables post-procedural case reviews. Data and records can be viewed on local workstations or transferred to a central computer or hospital network for storage and post-case analysis/reporting.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
510(k) Number K113214
Section 4 - Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.