K113206 · Draeger Medical Systems, Inc. · LBI · Feb 17, 2012 · General Hospital
Device Facts
Record ID
K113206
Device Name
NANOBLUTM 500
Applicant
Draeger Medical Systems, Inc.
Product Code
LBI · General Hospital
Decision Date
Feb 17, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Dräger NanoBlu 500 LED Phototherapy Light is intended to treat neonatal hyperbilirubinemia by providing phototherapeutic light to the body of the patient. It is intended for use on the recommendation and under the supervision of healthcare professionals. Additionally, this product can be used with an underbaby phototherapy light to increase patient coverage.
Device Story
The NanoBlu 500 is an LED-based phototherapy device used to treat neonatal hyperbilirubinemia. It consists of an LED module (5 LEDs), a cooling fan, a microprocessor controller, and an alphanumeric display/keyboard interface. The device is available in hood-mount (for incubator placement) or trolley-mount configurations. Healthcare professionals operate the device to deliver phototherapeutic light (400-550 nm) to the patient's body to degrade bilirubin. The microprocessor allows clinicians to program light functions via the keyboard. The device may be used in conjunction with under-baby phototherapy lights to enhance coverage. It is intended for clinical environments such as NICUs or nurseries.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including spectral characterization of the LED source, optical radiation hazard analysis, and electrical safety/EMC testing per IEC 60601-1, UL 60601-1, IEC 60601-1-2, and IEC 60601-2-50.
Technological Characteristics
LED-based phototherapy unit; 5 LEDs; 400-550 nm wavelength; minimum intensity 40 μW/cm2·nm. Microprocessor-controlled with alphanumeric display and keyboard. Available as hood-mount or trolley-mount. Electrical requirements: 90-240 VAC. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-50, UL 60601-1, CSA/CAN C22.2 601.1.
Indications for Use
Indicated for the treatment of neonatal hyperbilirubinemia in neonates. Intended for use under the supervision of healthcare professionals.
Regulatory Classification
Identification
A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.
{0}------------------------------------------------
510(k) Summarv NanoBluTM 500
**Dräger** K1132.06
FEB 1 7 2012
## 510(k) Summary
acc. to 807.92
Draeger Medical Systems, Inc Submitter's Name and Address: 3135 Quarry Road Telford, PA 18969 USA
Contact Person:
Bryan Overton Phone: 215.660.2135 Fax: 215.721.5424
Date submission was prepared:
2011-10-31
Device Name:
Proprietary Name: Common Name: Classification Name: Regulation Number: Product Code: Class:
NanoBlu™ 500 Phototherapy Light, Bili Light Neonatal Phototherapy Unit 21 CRF 880.5700 LBI 2
Legally Marketed Device Identification: Natus® Blue Light Phototherapy, K022196
#### Device Description:
The Dräger NanoBlu 500 is an LED phototherapy light. It consists of an LED module with 5 LEDs and a fan, a microprocessor controller, and an alphanumeric display and keyboard. The display and keyboard are used to program the functions of the light.
The Dräger NanoBlu 500 is available in two versions, the hood mount and the trolley mount. The hood mount version has suction cups on the bottom that are used when placed on the top of an incubator.
The trolley mount version is mounted on a rolling trolley to allow use above an incubator or radiant warmer. The trolley consists of a base supported by three 2 in (5.08 cm) castors. A column, with an adjustable and articulating arm, connects them. Both of these features allow the light to be positioned for optimum light over the patient.
There are no single-use components or accessories for the NanoBlu™ 500.
#### Intended Use:
The Dräger NanoBlu 500 LED Phototherapy Light is intended to treat neonatal hyperbilirubinemia by providing phototherapeutic light to the body of the patient. It is intended for use on the recommendation and under the supervision of
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The letters are black and have a slightly pixelated appearance. The two dots above the "a" are also black and bold.
healthcare professionals. Additionally, this product can be used with an underbaby phototherapy light to increase patient coverage.
#### Predicate Devices:
510(k) Number K022196
Device Name Natus Blue Light Phototherapy Unit
#### Substantial Equivalence:
The NanoBlu™ 500 and the Natus® Blue Light Phototherapy Unit have the same intended use (treatment of hyperbilirubinemia), use the same operating principle (delivery of light to degrade bilirubin), and are similar in their hardware configuration. See the Comparison Table below for details.
| Feature | NanoBlu 500 | Natus® Blue Light<br>(K022196) |
|------------------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Intended Use | For the treatment of neonatal<br>hyperbilirubinemia | For the treatment of neo-<br>natal hyperbilirubinemia |
| Target population | Neonates | Neonates |
| Physical Design | | |
| Type | Freestanding device | Freestanding device |
| Mounting hardware | Roll Stand, 3 legs w/casters,<br>3<br>locking | Roll Stand, 5 legs<br>w/casters, 2<br>locking |
| Light attachment | Lights mounted in enclosure | Lights mounted in enclo-<br>sure |
| Performance | | |
| Light source | Light Emitting Diodes (LED) | Light Emitting Diodes<br>(LED) |
| Wavelength | 400-550.nm | 400-550 nm |
| Intensity | Minimum 40 μW/cm2·nm | 35 ± 3.5 pW/cm2/nm @<br>High setting |
| Electrical Require-<br>ments | | |
| Operating Voltage | 90 VAC to 240 VAC | 85 VAC to 264 VAC |
| Standards and Safety | | |
| Electrical Safety | IEC 60601 -1 and 1-1 -2<br>UL 60601-1<br>CSA/CAN C22.2 601.1<br>IEC 60601-2-50 | EN 60601 -1 and 1-1 -2<br>UL 2601-1<br>CSA/CAN C22.2 601.1<br>EN 60601 -2-50 |
#### Summary Comparison Table NanoBlu 500 vs. Natus® Blue Light
#### · Summary of Nonclinical Testing
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The "a" in the word has an umlaut, which is a diacritical mark consisting of two dots placed over the letter. The text is black and the background is white.
This submission includes the results of testing prototype devices to specifications, spectral characterization of the LED light source, and an analysis of the potential optical radiation hazard of the blue LED light source.
Testing was also performed to assess compliance to the following standards:
- IEC60601-1, Medical Electrical Equipment, Part 1: General Requirements . for Safety
- UL60601-1, Medical Electrical Equipment, Part 1: General Requirements . for Safety
- IEC60601-1-2, Medical Electrical Equipment, Part 1-2: General Require-. ments for Safety Collateral Standard: Electromagnetic Compatibility
- IEC60601-2-50, Medical Electrical Equipment, Part 2: Particular require-. ments for the Safety of Infant Phototherapy Equipment
#### Clinical Tests - Not Applicable
The results were as expected and no new issues of safety or effectiveness were raised as a result of the nonclinical testing.
#### Conclusions
Based on the data and information presented in this submission, the NanoBlu™ 500 Phototherapy Light is substantially equivalent to the currently legally marketed Natus® Blue Light Phototherapy Unit manufactured and distributed by Natus Medical, Inc.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, topped by a single wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Bryan Overton Director, Quality Assurance & Compliance Draeger Medical Systems, Inc. 3135 Quarry Road Telford, Pennsylvania 18969
·FEB 1 7 2012
Re: K113206
Trade/Device Name: NanoBlu™ 500 Regulation Number: 21 CFR 880.5700 Regulation Name: Neonatal Phototherapy Unit Regulatory Class: II Product Code: LBI Dated: January 26, 2012 Received: January 27, 2012
Dear Mr. Overton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of Amendinents, or to ac rices that in tio Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 wire), it may of basyon . Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Overton
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the For of any voth all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing (reporting of increats as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to Compilance: "Theo, provins 201CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to ad verse evelife ander and mo Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may obtain other geturers, International and Consumer Assistance at its toll-free Diviser (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
el
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
# Indications for Use Form
Indications for Use
510(k) Number (if known): ____
Device Name: NanoBlu™ 500_
Indications for Use:
The Drager NanoBlu 500 LED Phototherapy Light is intended to treat neonatal hyperbilirubinemia by providing phototherapeutic light to the body of the patient. It is intended for use on the recommendation and under the supervision of healthcare professionals.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
RSC Chape 2/15/12
(Division Sign-Off)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.