As provided in Sec. 870.1340 the PAJUNK®s Vascular Sono cannula/needle is an introducer used to facilitate placing an intravascular or intravenous device for diagnostically or interventional use through the skin into a vein or artery.
Device Story
Vascular Sono is a single-use, sterile, non-pyrogenic introducer cannula/needle designed for percutaneous access to veins or arteries. The device features enhanced ultrasound visibility (Cornerstone-technique) to assist clinicians in needle placement under ultrasound guidance. It consists of a stainless steel cannula, polycarbonate hub, and epoxy resin glue; an optional polymeric NanoLine coating is available. The device is used by clinicians in clinical settings to facilitate the insertion of intravascular devices. It does not include safety mechanisms or tear-away features found in some predicate devices. Performance is verified through bench testing, including bending stability, hub-to-needle bonding, breakage resistance, and insertion force. The device provides improved visualization during ultrasound-guided procedures, potentially increasing accuracy and reducing complications during vascular access.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including packaging integrity (DIN EN 868-5, ASTM E 515-05, ASTM F 1929-98), sterilization validation (EtO residuals per ISO 10993-7), and mechanical performance testing (bending rigidity, hub-to-needle bonding, breakage resistance, and insertion force). Biocompatibility is established via compliance with ISO 10993-1 and equivalence to previously cleared Pajunk devices (K040965, K000722, K053283).
Technological Characteristics
Materials: stainless steel tubing, polycarbonate hub, epoxy resin, optional polymeric NanoLine coating. Sensing: enhanced ultrasound visibility (Cornerstone-technique). Energy: none (manual). Connectivity: none. Sterilization: Ethylene Oxide (EtO). Standards: ISO 10993-1 (biocompatibility), ISO 10993-7 (EtO residuals), DIN EN 868-5 (packaging).
Indications for Use
Indicated for use as a catheter introducer to facilitate the placement of intravascular or intravenous devices for diagnostic or interventional procedures through the skin into a vein or artery.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
# 510(k) Summary
# K113196
## Date of Preparation: October 27th 2011
### Submitter Information/ production site:
Paiunk GmbH Karl-Hall-Strasse 01 78187 Geisingen Germany Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 Establishment Registration Number: 9611612
#### USA Contact:
PAJUNK MEDICAL SYSTEMS 6611 Bay Circle, Suite 100 Norcross, GA 30071 Phone: (770) 493 - 6832 ext.111 (678) 514 - 3388 Fax: (770) 330 - 2724 Cell: richard.fischer@pajunk-usa.com
### Contact:
Christian G. H. Quass Director Regulatory Affairs, Safety Official Patricia Weisbrod, Regulatory Affairs Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 E-Mail: christian.guass@pajunk.com E-Mail: patricia.weisbrod@pajunk.com
### Contact
Richard Fischer MD President Fon: +01(0)770-493-6832 Ext 111 Fax: 678 5143388 E-Mail: Richard.fischer@pajunk-usa.com
## Contract Sterilizer:
External service provider located in Europe
#### Device Information:
| Device Name: | Vascular Sono vascular introducer cannula/needle |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Trade Names: | Vascular Sono |
| Common Name: | Introducer |
| Classification Name: | Catheter introducer |
| Classification Reference: | 21 CFR § 870.1340 April 1, 2011 |
| | PART 870 -- CARDIOVASCULAR DEVICES |
| | Subpart B--Cardiovascular Diagnostic Devices |
| | Sec. 870.1340 Catheter introducer |
| | (a) Identification. A catheter introducer is a sheath used to facilitate<br>placing a catheter through the skin into a vein or artery. |
| Establishment Registration<br>Number: | 9611612 |
| Regulatory Class: | II |
| Product Code: | DYB, OFC, OFD, OEX |
| Panel: | Cardiovascular |
| Predicate Devices: | 1. K020834 BD Introsyte Precision Introducer, BD Medical Systems<br>2. K030135 Safety Introducer Needle, BBraun Medical, Allentown |
{1}------------------------------------------------
#### Indications for use
As provided in Sec. 870.1340 the PAJUNK®s Vascular Sono cannula/needle is an introducer used to facilitate placing an intravenous device for diagnostical or interventional use through the skin into a vein or artery.
### Device Description:
The Vascular Sono intravenous introducer cannula/needle is a single use sterile and nonpvrogenic medical device used to gain entry or puncture arteries and veins.
#### Predicate Devices:
Predicate devices with identical or at least partial indications of use are:
- 1. K020834 BD Introsvte Precision Introducer, BD Medical Systems
- 2. K030135 Safety Introducer Needle, BBraun Medical, Allentown
Predicate Devices are marketed with additional equipment like safety mechanism and tear-away cannula. Basically, the introducer cannula of the predicate devices and the subject device is made from stainless steel.
The detailed discussion of substantial equivalence can be found in Section 12 of this submission.
#### Sterilization
The contract sterilizer and the sterilizing process are identical to the process and sterilizer used for all PAJUNK® - manufactured and purchased devices which are already cleared for market or exempt. The Cornerstone-technique does neither influence sterilization process nor shelf life properties.
Cleaning and Sterilization method, which ensures an SAL of 10th as well as compliance with limits for chemical burden, bioburden (i.e. LAL) and EtO-residuals as well as shelf life have been validated and are safe and effective. Efficacy of sterile product's lifecycle has been proven for a period of 10 years now. Shelf life is set to 5 years.
### Biocompatibility:
All cannulas comply with ISO 10993-1, 20d and 3rd edition.
The stainless steel tubing of the Sono-needles/cannulas is identical to stainless steel tubing of the NanoLine-needles/cannulas as they were cleared for market in K040965, K000722 and K053283 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).
The polycarbonate hub of the Sono-needles/cannulas is identical to the polycarbonate hub of the needles/cannulas as they were cleared for market in K040965, K000722 and K053283 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).
The epoxy resin glue of the Sono-needles/cannulas is identical to the epoxy resin glue of the needles/cannulas as they were cleared for market in K040965, K000722 and K053283 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).
The optional polymeric NanoLine coating of the Sono-needles/cannulas is identical to the polymeric NanoLine coating of the NanoLine-needles/cannulas as they were cleared for market in K053283 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).
{2}------------------------------------------------
### Technology Characteristics:
The Vascular Sono is provided with enhanced ultrasound visibility. So is the predicate manufactured by BBraun. Additional Safety features like BBraun and BD offer are not part of the equipment. BD in Addition does provide a tear away cannula.
The components are listed in a table in section 11 of this submission. Shelf life and impact of sterilization and storage on the devices has been proven and found to be safe and effective.
#### Performance Testing
The needles/ cannulas have been subjected to standard testing applicable for all cannulas. Standard testing consists of bending stability and breaking resistance testing as well as of hubto-needle-stability testing. Due to technological equivalence the subject device is tested the same way as each cannula is tested at PAJUNK® GmbH Medizintechnologie. There are no special testing requirements defined, neither in incoming and in-process inspection routines nor in final testing.
| Test Description | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Packaging: DIN EN 868-5:2009-09 Packaging for terminally sterilized medical devices -<br>Part 5: Sealable pouches and reels of porous materials and plastic film construction -<br>Requirements and test methods | |
| Packaging: "Bubble Emission test" acc. To ASTM E 515-05 4a | |
| Packaging: Rhodamin-B-Test, DYE-Test acc. To DIN 868-1 and ASTM F 1929-98 | |
| Packaging: Peel-Test acc. To 868-10:2009-09 Annex C | |
| Microbiology/Particles: LAL-Testing (Pyrogene) | |
| Microbiology/Particles: Bioburden | |
| Microbiology/Particles: Chemical Burden | |
| Microbiology/Particles: Ethylene-Oxide Residuals acc. 10993-7 | |
| Microbiology/Particles: Test for sterility (per batch) | |
| Performance: Bending Rigidity/ Bending Stability | |
| Performance: Hub to needle bonding | |
| Performance: Breakage resistance | |
| Performance: Insertion force at point of entry | |
#### Summary of testing conducted on the Vascular SONO
### Conclusion:
The comparison between the predicate devices and the subject device in section 12 of this submission as well as the validated sterilization process and the results of the bench testing and bench marking demonstrates that the proposed devices are substantially equivalent to the predicate devices and identical in technical description to devices already cleared for market and therefore demonstrated to be safe and effective. Based on the clinical evaluation, the biocompatibility testing and the bench testing conducted, safety and effectiveness as well as efficacy of the Cornerstone/ Sono-technique is proven for each type of cannula.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
PAJUNK GmbH C/O Mr. Christian G.H. Quass Director Regulatory Affairs, Safety Official Karl-Hall-Strasse 01 78187 Geisingen Germany
FEB - 6 2012
Re: K113196/S001
Trade/Device Name: Vascular Sono vascular introducer cannula Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: January 9, 2012 Received: January 12, 2012
Dear Mr. Quass:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{4}------------------------------------------------
Page 2 - Mr. Christian G.H. Quass
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
M. A. Hilleben
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for use
510(k) Number:
KII 3196
Device Name: Indications for Use:
Vascular Sono vascular introducer cannula
As provided in Sec. 870.1340 the PAJUNK®s Vascular Sono cannula/needle is an introducer used to facilitate placing an intravascular or intravenous device for diagnostically or interventional use through the skin into a vein or artery.
| Prescription Use<br>(Per 21 CFR 801.109) | <b>X</b> |
|------------------------------------------|------------------------------------------------|
| AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number_1(318) كلك الكارك الكارك الكارك الكارك الكارك الكارك الكارك ا
**Page 1 of 1**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.