The Acme Monaco Slidewire® is intended for use in diagnostic or interventional procedures to assist in directing a catheter to the desired anatomical locations in the coronary or peripheral vasculature.
Device Story
Acme Monaco Slidewire is a single-use, sterile guidewire used by clinicians to navigate catheters through coronary or peripheral vasculature. Device consists of a Nickel Titanium (Nitinol) core wire covered by a polyurethane jacket blended with Tungsten for radiopacity; exterior features a hydrophilic coating to reduce friction during insertion and manipulation. Clinicians operate the device manually under medical imaging guidance to reach target anatomical sites. The device provides kink resistance and maintains original shape in tortuous anatomy. Benefits include enhanced maneuverability and visibility during diagnostic or interventional procedures.
Clinical Evidence
Bench testing only. Performance testing conducted per FDA Coronary and Cerebrovascular Guidewire Guidance (1995) and ISO 11070:1998. Tests included tensile strength, torque strength, torqueability, tip flexibility, coating adherence, catheter compatibility, surface lubricity, dimensional verification, and corrosion resistance. Biocompatibility testing performed per ISO 10993-1:2003, including cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity, and hemocompatibility (hemolysis, thromboresistance, clotting time, complement activation).
Technological Characteristics
Core: Nickel Titanium (Nitinol). Jacket: Polyurethane resin blended with Tungsten. Coating: Hydrophilic. Form factor: Various lengths/diameters/distal configurations. Sterilization: Sterile, single-use. Standards: ISO 10993 (biocompatibility), ISO 11070 (catheter introducers).
Indications for Use
Indicated for patients undergoing diagnostic or interventional procedures requiring catheter placement in coronary or peripheral vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Terumo Radifocus® Guidewire M Hydrophilic Coated Guidewires (K863138, K063372)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for ACME MONACO. The logo is black and white and features the word "ACME" stacked on top of the word "MONACO". The words are contained within a stylized box-like shape. There is a registered trademark symbol in the upper right corner of the logo.
K113162
p. 1 of 3
## 510(k) SUMMARY
This 510(k) summary is provided per the requirements of Section 807.92(c).
| Submitted by: | Acme Monaco Corporation<br>75 Winchell Road<br>New Britain, CT 06052<br>Phone: 860-224-1349<br>Fax: 860-827-9982 | JUL 20 2012 |
|----------------------|------------------------------------------------------------------------------------------------------------------|-------------|
| Contact Person: | Benjamin Sweeney.<br>Manager of Device Compliance<br>Phone: 860-224-1349<br>bsweeney@acmemonaco.com | |
| Trade Name: | Acme Monaco Slidewire® | |
| Common Name: | Guidewire | |
| Classification Name: | Device – Wire, Guide, Catheter | |
| Regulation: | Catheter guide wire Per 21 CFR Part 870.1330 | |
| Product Code: | DQX | |
| Classification: | Class II | |
| Date prepared: | June 22, 2012 | |
## Legally marketed device to which equivalence is claimed:
Terumo Radifocus® Guidewire M Hydrophilic Coated Guidewires, Premarket Notification K863138, and K063372.
#### 5.1 Description of the Device:
The Acme Monaco Hydrophilic Guidewire, hereafter referred to as Acme Monaco Slidewire®, is composed of a Nickel Titanium (aka NiTi or Nitinol) core wire, jacketed with a durable polymer that is receptive to a coating. The polymer resin is blended with Tungsten, a radiopaque metal and affixed to the core wire via adhesive. The jacketed core wire is then coated with a hydrophilic coating. The Acme Monaco Slidewire of can be constructed to various lengths,
{1}------------------------------------------------
diameters and distal configurations, is a single-use device and provided sterile for commercial distribution.
### 5.2 Scientific concepts that form the basis for the device:
The Acme Monaco Slidewire® has a core wire constructed of Nickel Titanium (aka NiTi or Nitinol). The core wire is coated with a polymer jacket. A radiopaque metal is blended into the polymer resin in order to enhance visibility of the guidewire on medical imaging devices. The polymer jacket is coated with a hydrophilic coating designed to provide a low coefficient of friction when wet, thus enhancing the movement and maneuverability of the device.
### 5.3 Intended Use of the Device:
The Acme Monaco Slidewire® is intended for use in diagnostic or interventional procedures to assist in directing a catheter to the desired anatomical locations in the coronary or peripheral vasculature.
### 5.4 Comparison of technological characteristics to legally marked device:
The Acme Monaco Slidewire® is being compared for substantial equivalence against the Terumo Radifocus® Guidewire M, Premarket Notification K863138 and K063372.
The Acme Monaco Slidewire® and the predicate device both contain a Nickel Titanium core. Both products incorporate a polyurethane jacket which covers the core wire. Both the predicate device and the Acme Monaco Slidewire® contain Tungsten that is blended into the polyurethane resin in order to improve visibility on medical images. Both devices employ a hydrophilic coating over the polyurethane jacket in order to provide reduced friction between the medical devices and tissue. Both the predicate device and the Acme Monaco device retain their lubricity over multiple uses during a single procedure. Both devices provide for kink resistance and retain original conformations, even when subjected to tortuous anatomical environments.
The differences between the Acme Monaco Slidewire® and the predicate device lie in the geometric configurations of the distal end of the device, and the method of polymer jacket adherence to the core wire. These differences do not affect the intended use of the Acme Monaco Slidewire . The Acme Monaco device has been completely tested for biocompatibility in accordance with FAD General Program Memorandum #G95-1, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing", May 1, 1995.
### 5.5 Performance Testing
Performance testing has been conducted in accordance with FDA guidance document Coronary and Cerebrovascular Guidewire Guidance, January 1995, ISO 11070:1998(E), Sterile single-use intravascular catheter introducers and ISO 10993-1:2003(E), Biological evaluation of medical devices - Part 1: Evaluation and testing. The results of the following performance tests have demonstrated substantial equivalence to the predicate device.
{2}------------------------------------------------
-
- Tensile Strength ●
- Torque Strength
- Torqueability .
- . Tip Flexibility
- . Coating Adherence/Integrity
- Catheter Compatibility .
- Surface Lubricity ●
- Dimensional Verification .
- Resistance to corrosion .
#### Biocompatibility Testing 5.6
The following biocompatibility testing, performed in accordance with ISO-10993, has been performed on the Acme Monaco Slidewire® :
- . Cytotoxicity (ISO 10993-5)
- Kligman Maximization/2 Extracts/35 animals/Concurrent (+) controls (ISO 10993-10) .
- Intracutaneous Injection/2 extracts (ISO 10993-10)
- Acute Systematic Injection/2 extract (ISO 10993-11)
- Rabbit Pyrogen-Material Mediated (ISO 10993-11) .
- Hemolysis Complete (Direct & Indirect) -ASTM (ISO 10993-4)
- . In Vitro Hemocompatibility (Direct) (ISO 10993-4)
- . Dog Thromboresistance/2 animals (ISO 10993-4)
- Lee & White Clotting Time/Human Blood (Direct) (ISO-10993-4) .
- Complement Activation Assay (C3a), Direct Contact (ISO 10993-4) ●
- . Complement Activation Assay (SC5-b-9), Direct Contact (ISO 10993-4)
#### 5.7 CONCLUSIONS
Based on the evaluation of performance and biocompatibility testing, the Acme Monaco Slidewire® is considered to be appropriate for its intended use and performs in an equivalent manner as the Terumo Radifocus Guidewire M. The materials of construction and intended use are virtually identical.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 20 2012
Acme Monaco Corp. Mr. Benjamin Sweeney Manager, Device Compliance 75 Winchell Road New Britain, CT 06052
Re: K113162
Trade/Device Name: Acme Monaco Slidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guidewire Regulatory Class: Class II Product Code: DQX Dated: July 17, 2012 Received: July 18, 2012
Dear Mr. Sweeney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Mr. Benjamin Sweeney
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
My Hillehen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number: K113162
Acme Monaco Slidewire™ (Trademark Pending) Device Name:
Indications for Use: The Acme Monaco Slidewire™ (Trademark Pending) is intended for use in diagnostic or interventional procedures to assist in directing a catheter to the desired anatomical locations in the coronary or peripheral vasculature.
## AND/OR Over-The-Counter Use _ Prescription Use _ X
# (PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
.g. Hillhen
Division Sign-Off) (Division Sign-Sign-Sign-Scular Devices
510(k) Number K113162
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.