K113128 · Microaire Surgical Instruments, LLC · QPB · Aug 17, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113128
Device Name
MICROAIRE SURGICAL INSTRUMENTS LLC
Applicant
Microaire Surgical Instruments, LLC
Product Code
QPB · General, Plastic Surgery
Decision Date
Aug 17, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MicroAire LipoTower System is indicated for use in the following surgical specialties when the fragmentation, emulsification, and aspiration of soft tissue is desired: -Neurosurgery - Gastrointestinal and Affiliated Organ Surgery - - Urological Surgery - Plastic and Reconstructive Surgery - -General Surgery - Orthopedic Surgery - Gynecological Surgery l - Thoracic Surgery - Laparoscopic Surgery The MicroAire LipoTower System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
Device Story
MicroAire LipoTower System is a modular, wheeled cart-based suction lipoplasty system. It integrates an adjustable rotary vane vacuum source (up to 29 inHg) and a peristaltic tumescent pump (50-600 ml/min). Operated by a physician in a surgical setting, the system uses an LCD touch screen for control and status monitoring, with foot controls for pump activation. It facilitates soft tissue fragmentation, emulsification, and aspiration. The system includes a bag holding pole, canister rack, and shelves; it requires a non-sterile biofilter on the vacuum inlet and non-sterile collection canisters (1200ml/2000ml). The device assists surgeons in body contouring and various surgical procedures by providing controlled suction and fluid delivery, potentially improving procedural efficiency and patient outcomes.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance standards (ISO 10079-1) was demonstrated.
Technological Characteristics
Modular wheeled cart; rotary vane vacuum pump; peristaltic tumescent pump; LCD touch screen interface; foot pedal control. Electrical: 115VAC, 50-60 Hz. Standards: IEC 60601-1 (Class I, Type B), IEC 60601-1-2, ISO 10079-1, ISO 13485, ISO 14971, IEC 62304. Connectivity: Standalone system. Sterilization: N/A (system components).
Indications for Use
Indicated for patients requiring soft tissue fragmentation, emulsification, and aspiration across neuro, GI, urological, plastic/reconstructive, general, orthopedic, gynecological, thoracic, and laparoscopic surgical specialties, and for subcutaneous fatty tissue aspiration for aesthetic body contouring.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Predicate Devices
Sound Surgical Technologies LLC - SoundVaser System (K022051)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Microaire Surgical Instruments, LLC Darren Reeves Consultant 7305 Hancock Village Dr Suite 109 Chesterfield, Virginia 23832
June 9, 2021
Re: K113128
Trade/Device Name: Microaire Surgical Instruments Llc Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Darren Reeves:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 17, 2012. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 1 7 2012
MicroAire Surgical Instruments, LLC % DP Distribution & Consulting, LLC Mr. Darren Reeves Consultant 15637 Fox Cove Circle Moseley, Virginia 23120
Re: K113128
Trade/Device Name: MicroAire LipoTower System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU Dated: July 5, 2012 Received: July 9, 2012
Dear Mr. Reeves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
## Page 2 - Mr. Darren Reeves
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
Mark A. Millman
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
11/3/28 510(K) Number known:
Device Name: MicroAire LipoTower System
#### Indications for Use:
The MicroAire LipoTower System is indicated for use in the following surgical specialties when the fragmentation, emulsification, and aspiration of soft tissue is desired:
- -Neurosurgery
- Gastrointestinal and Affiliated Organ Surgery -
- ー Urological Surgery
- Plastic and Reconstructive Surgery -
- -General Surgery
- Orthopedic Surgery -
- Gynecological Surgery l
- Thoracic Surgery -
- Laparoscopic Surgery
The MicroAire LipoTower System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
Prescription use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krave for MM
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113128
{4}------------------------------------------------
## 11 3 2 8
## Section 5: 510(k) Summary
- 1) Preparation Date: October 12, 2011
- 2) Submitted by:
MicroAire Surgical Instruments LLC 1641 Edlich Drive Charlottesville, Virginia 22911 Owner/Operator #: 9004658
Contact Person/Prepared by:
Darren Reeves Phone: (804) 307-7706 Fax: Email: dreeves@dpdconline.com
3) Device Identification:
Trade Name: MicroAire LipoTower System Common Name: Lipoplasty Suction System Classification: Lipoplasty Suction System (21 CFR 878.5040, Product Code MUU)
- 4) Predicate Device: Sound Surgical Technologies LLC - SoundVaser System (K022051)
- 5) Device Description: The MicroAire LipoTower System is a modular wheeled cart that contains an adjustable vacuum source and a tumescent pump (110/120 VAC, 60 Hz). The vacuum source is a rotary vane type pump that delivers 0-29 inHG vacuum and drives the suction of tissue. The tumescent pump is a peristaltic type that delivers 50-600 ml/min of fluid as required. The system includes a control module with an LCD touch screen for control and status of the system and foot controls for controlling the vacuum pump and tumescent pump. Accessories include a bag holding pole, canister rack and shelves with are included. The system requires a biofilter (non-sterile) on the vacuum inlet and collection canisters (non-sterile - 1200ml and 2000ml) which are available separately.
#### 6) Indications for Use:
The MicroAire LipoTower System is indicated for use in the following surgical specialties when the fragmentation, emulsification, and aspiration of soft tissue is desired:
- l Neurosurgery
- -Gastrointestinal and Affiliated Organ Surgery
- Urological Surgery l
- Plastic and Reconstructive Surgery -
- General Surgery |
- Orthopedic Surgery 1
- -Gynecological Surgery
- Thoracic Surgery -
- Laparoscopic Surgery -
The MicroAire LipoTower System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
AUG 1 7 2012
{5}------------------------------------------------
## 7) Comparison to Predicate:
',
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
- .
| General Technical | Predicate - Sound Surgical Technologies, LLC | Proposed device | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|------------------|
| Characteristics | (K022051) | | |
| Indications for use | The MicroAire LipoTower System is indicated for use in<br>the following surgical specialties when the<br>fragmentation, emulsification, and aspiration of soft<br>tissue is desired:<br>- Neurosurgery<br>- Gastrointestinal and Affiliated Organ Surgery<br>- Urological Surgery<br>- Plastic and Reconstructive Surgery<br>- General Surgery<br>- Orthopedic Surgery<br>- Gynecological Surgery<br>- Thoracic Surgery<br>- Laparoscopic Surgery<br>The System is indicated for use when the<br>fragmentation, emulsification, and aspiration of soft<br>tissue is desired or subcutaneous fatty tissues for<br>aesthetic body contouring is desired | Same | |
| | Size (cm) (H x W x D) | 89 x 46 x 43 | 106 x 56 x 69 |
| | Input Power | 115 & 230 VAC. 50-60 Hz | 115VAC, 50-60 Hz |
| | Handpiece | | N/A |
| | Diameter (cm) | 2.5 | N/A |
| | Length (cm) | 17.5 | N/A |
| | Transducer | PZT crystal | N/A |
| | Vibration Frequency | 36-55 kHz | N/A |
| | Probes | | N/A |
| | Material | Titanium alloy | N/A |
| | Diameter (mm) | 2.9-3.7 | N/A |
| | Length (cm) | 7.7-27.0 | N/A |
| | Irrigation | | |
| | Type | Adjustable Flow | Same |
| | Pump Type | Peristaltic | Same |
| | Flow Rate | | 50-600ml/min |
| Suction | | | |
| Vacuum | 24 in Hg | Adjustable up to 29 in Hg | |
| Pump Type | | Rotary Vane | |
| Sterilization | Steam | N/A | |
| Sterilization Validation | | N/A | |
| Residuals | | N/A | |
| Accessories | | N/A | |
| AC Power cord | 3 prong, Hosp Grade | Same | |
| Handpiece cable | 3 m, silicone | N/A | |
| Footswitch control | Yes | Yes | |
| Collection Canister | 1250ml | 1200ml, 2000ml | |
| Biofilter | N/A | Yes | |
| Prescription | Yes | Yes | |
| IEC Classification | Class I, Type BF | Class I, Type B | |
| | Predicate Used | MicroAire | |
| IEC EN 60601-1- Medical Electrical Equipment, Part 1: General Requirements for Safety - 2nd Edition | | | |
| Radiated emissions | | | |
| CISPR 11<br>for IEC 60601-1-2 clause 36.201.1 | Compliant | Compliant | |
| Conducted emissions | | | |
| CISPR 11<br>for IEC 60601-1-2 clause 36.201.1 | Compliant | Compliant | |
| Harmonics | | | |
| IEC 61000-3-2 for IEC 60601-1-2 clause 36.201.3.1 | N/A | Not applicable (to devices whose<br>voltage < 220V). | |
| Flicker | | | |
| IEC 61000-3-3 for IEC 60601-1-2 clause 36.201.3.2 | N/A | Not applicable (to devices whose<br>voltage < 220V or line-frequency of<br>60Hz | |
| ESD immunity | | | |
| IEC 61000-4-2 for IEC 60601-1-2 clause 36.202.2 | | | |
| UL 801-2 | Compliant | Compliant to level of 2kV | |
| Air ESD | Unknown | Compliant | |
| Radiated immunity | | | |
| IEC 61000-4-3 for IEC 60601-1-2 clause 36.202.3 | | | |
| UL 801-3 | Compliant | Compliant | |
| EFT immunity | | | |
| IEC 61000-4-4 for IEC 60601-1-2 clause 36.202.4 | | | |
| UL 801-4 | Compliant | Compliant to level of 1.65kV | |
| Surge immunity | | | |
| IEC 61000-4-5 for IEC 60601-1-2 clause 36.202.5 | | | |
| UL 801-5 | Compliant | Compliant | |
| Conducted immunity | | | |
| IEC 61000-4-6 for IEC 60601-1-2 clause 36.202.6 | Unknown | Compliant | |
| Voltage dips and short-interruptions immunity | | Compliant.Note: Using 1kVA<br>exclusion of 36.202.7 a)1) for 60%<br>voltage dip | |
| IEC 61000-4-11 for IEC 60601-1-2 clause 36.202.7 | Unknown | | |
| Power-frequency magnetic-field immunity | | | |
| IEC 61000-4-8 for IEC 60601-1-2 clause 36.202.8.1 | Unknown | Compliant | |
| CAN/CSA C22.2 No. 601.1 - Medical Electrical Equipment, Part 1: General Requirements for<br>Safety - 1st Edition | Compliant | Compliant | |
| UL 60601-1- Medical Electrical Equipment, Part 1: General Requirements for Safety - 1st<br>Edition | Compliant | Compliant | |
| ASME Y 14.5M-1994 - Mathematical Definition of Dimensioning and Tolerancing | Unknown | Compliant | |
| ISO 10079-1 - Medical suction equipment -- Part 1: Electrically powered suction equipment --<br>Safety requirements | Unknown | Compliant | |
| ISO 15223-1- Medical devices -- Symbols to be used with medical device labels, labelling and<br>information to be supplied -- Part 1: General requirements | Unknown | Compliant | |
| ISO 13485 - Medical devices -- Quality management systems -- Requirements for regulatory<br>purposes | Unknown | Compliant | |
| ISO 14971 - Medical devices -- Application of risk management to medical devices | Unknown | Compliant | |
| IEC 62304- Medical device software - Software life cycle processes | Unknown | Compliant | |
UII 3128
{6}------------------------------------------------
# UU3I28
1 - 3 of You
:
:
{7}------------------------------------------------
UU 3128
Yofy
## 8) Conclusion:
The information in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to devices already in commercial distribution. Equivalence is demonstrated through intended use, materials, design and tested methods.
## 9) Similarities/Differences of the proposed device when compared to the predicate:
The data within this submission demonstrates that there are no significant differences between the application device and the predicate, indicating that the application device is safe, effective and substantially equivalent for marketing in the U.S.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.