EPICARE YAG

K113115 · Light Age, Inc. · GEX · Apr 3, 2012 · General, Plastic Surgery

Device Facts

Record IDK113115
Device NameEPICARE YAG
ApplicantLight Age, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateApr 3, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Light Age Inc. EpiCare-YAG™ Laser System is intended for use in general and plastic surgery, dermatology, and podiatry for the treatment of vascular lesions, benign pigmented lesions, for the removal of dark tattoo inks and reduction of hypertrophic and keloid scars and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hair regrowth after a treatment regime on all skin types (Fitzpatrick I-VI including tanned skin.) Specifically the 1064 nm wavelength is indicated for: Removal of unwanted hair, for stable long term or permanent hair reduction on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins, and spider veins and poikiloderma of civatte. Coagulation and hemostasis of soft tissue. Benign cutaneous lesions such as warts, scars, striae, and psoriasis. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. Pigmented lesion size reduction in patients with lesions who would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of pseudofolliculitis barbae (PFB). Treatment of wrinkles including but not limited to periocular wrinkles and perioral wrinkles.

Device Story

Class IV Nd:YAG laser system; emits 1064 nm pulsed energy via flexible optical fiber and handpiece; includes laser source, microprocessor control unit, laptop user interface, and foot pedal. Operated by clinicians in general/plastic surgery, dermatology, and podiatry settings. Device delivers energy to target skin areas for photocoagulation, hemostasis, and tissue treatment. Microprocessor manages laser parameters; clinician monitors via laptop interface. Benefits include stable long-term hair reduction, lesion removal, and scar management. Derived from EpiCare-DUO™ by removing Alexandrite 755nm modules; maintains identical 1064nm operation.

Clinical Evidence

No clinical data presented. Safety and effectiveness supported by nearly one decade of use of the EpiCare™ product family with over 1 million treatments performed.

Technological Characteristics

Nd:YAG laser source; 1064 nm wavelength; microprocessor-controlled; flexible optical fiber delivery; laptop user interface; foot pedal activation. Complies with 21 CFR 1040.10, 1040.11, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-1-22.

Indications for Use

Indicated for patients of all skin types (Fitzpatrick I-VI, including tanned) requiring treatment for vascular/pigmented lesions, unwanted hair, dark tattoo inks, hypertrophic/keloid scars, soft tissue coagulation, benign cutaneous lesions, pseudofolliculitis barbae, and wrinkles.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) APR - 3 2012 # 510(k) Summary EpiCare-YAG™ Laser System October 18, 2011 .K113/15 page w of (4) Submittal Information Post Approval Contact: Dr. Donald F. Heller Chief Executive Officer Elizabeth Reddington Director of Regulatory Affairs Light Age, Inc 500 Apgar Drive Somerset, NJ 08873 Tel: 732-563-0600 Fax: 732-563-1571 Device Name and Classification 510(k) Number: Proprietary Name: Common Name: Classification Name: Classification Panel: C.F.R Section Device Class: Product Code: Light Age EpiCare-Y AGTM Laser System Nd:YAG Laser System Class IV Laser Surgical Instrument General and Plastic Surgery Devices 878.4810 II GEX #### Predicate Devices Light Age, Inc. EpiCare-DUO™ Laser System (K091625) Candela, Gentle Y AG® (K033172) Cynosure, Apogee Elite™ (K034030) #### Device Description The Light Age Inc. EpiCare-YAG™ Laser System is a Class IV laser surgical instrument for use in general medical / cosmetic procedures, general and plastic surgery and dermatology. Using neodymium YAG crystal rods, pulsed energy is emitted at 1064 nanometers in the near infrared portion of the spectrum under the guidance of a visible aiming beam. The device consists of the following components and accessories: - 1. Laser source and onboard microprocessor based control unit - 2. Laptop user interface - 3. Flexible optical fiber and handpiece delivers energy from the laser to the target area via optical fiber with handpiece, which produces a circular beam on the skin. - 4. Foot pedal switch activates delivery when lasers are enabled. {1}------------------------------------------------ ## Intended Use: The Light Age EpiCare-YAG™ Laser System delivers variable pulse laser light in the near infrared portion of the spectrum during procedures. The Light Age Inc. EpiCare-YAG™ Laser System is intended for use in general and plastic surgery, dermatology, and podiatry for the treatment of vascular lesions, benign pigmented lesions, for the removal of dark tattoo inks and reduction of hypertrophic and keloid scars and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hair regrowth after a treatment regime on all skin types (Fitzpatrick I-VI including tanned skin.) No new indications were sought in this 510(k) and no clinical data is presented. The EpiCare-YAG™ has a subset of the capabilities of our EpiCare-DUO™ predicate device. The EpiCare-DUO™ has a modular design format. To create the EpiCare-Y AG™ Light Age, Inc removed the Alexandrite 755nm modules from our previously approved EpiCare-DUO™ Laser System. From both a design and clinical perspective, the predicate and candidate laser devices are of the same technology and have the same intended use. Based on an analysis of the overall performance characteristics for the devices, Light Age, Inc. believes that no significant differences exist. Specifically the 1064 nm wavelength is indicated for: - . Removal of unwanted hair, for stable long term or permanent hair reduction on all skin types Fitzpatrick I-VI including tanned skin. - . Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins, and spider veins and poikiloderma of civatte. Coagulation and hemostasis of soft tissue. - . Benign cutaneous lesions such as warts, scars, striae, and psoriasis - . Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. - Pigmented lesion size reduction in patients with lesions who would potentially . benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. - . Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar - Treatment of pseudofolliculitis barbae (PFB). . - Treatment of wrinkles including but not limited to periocular wrinkles and . perioral wrinkles. {2}------------------------------------------------ ## Performance Standards: - The EpiCare-YAG™ Laser System complies with applicable performance . standards for light emitting products as outlined in 21 CFR 1040.10 and 21 CFR 1040.11. - . The device also conforms to the voluntary electrical equipment standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-1-22. ### Substantial Equivalence The reason for this 510(K) is a design change. Light Age, Inc has removed some modules (Alexandrite) from our EpiCare-DUO™ Laser System which has been cleared and in use since 2009. The candidate device operates in 1064 nm wavelength mode and is identical in design and operation to the 1064nm wavelenth mode of the previously EpiCare-DUO™ Laser System. ### Safety and Effectiveness The Light Age EpiCare-YAG™ Laser System should not raise any concerns regarding its overall safety and effectiveness. In the nearly one decade of use with over 1 million treatments performed, the EpiCare™ product family has prove to be clinically safe with no reports of significant patient or operator injury. The Light Age EpiCare-YAG™ is designed in accordance with both mandatory and voluntary standards ensuring it is both safe and effective for cosmetic / medical procedures indicated above. No new clinical indications are to be provided by the introduction of this device as compared to the predicate devices, identified above, which have previously demonstrated clinical effectiveness. ### Indications for Use: The Light Age Inc. EpiCare-YAG™ Laser System is intended for use in general and plastic surgery, dermatology, and podiatry for the treatment of vascular lesions, benign pigmented lesions, for the removal of dark tattoo inks and reduction of hypertrophic and keloid scars and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hair regrowth after a treatment regime on all skin types (Fitzpatrick I-VI including tanned skin.) No new indications were sought in this 510(k) and no clinical data is presented. The EpiCare-Y AGTM has a subset of the capabilities of our EpiCare-DUO™ predicate device. Both systems are modular in design. Light Age, Inc. is removing a capability (Alexandrite 755nm) from our previously approved EpiCare-DUO™ Laser System. From both a design and clinical perspective, the predicate and candidate laser devices are of the same technology and have the same intended use. Based on an analysis of the {3}------------------------------------------------ overall performance characteristics for the devices, Light Age, Inc. believes that no significant differences exist. Specifically the 1064 nm wavelength is indicated for: - t Removal of unwanted hair, for stable long term or permanent hair reduction on all skin types Fitzpatrick I-VI including tanned skin. - . Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins, and spider veins and poikiloderma of civatte. Coagulation and hemostasis of soft tissue. - . Benign cutaneous lesions such as warts, scars, striae, and psoriasis - . Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. - . Pigmented lesion size reduction in patients with lesions who would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. - . Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar - . Treatment of pseudofolliculitis barbae (PFB). - . Treatment of wrinkles including but not limited to periocular wrinkles and perioral wrinkles. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 APR - 3 2012 Light Age, Inc. % Ms. Elizabeth Reddington Director of Regulatory Affairs 500 Apgar Drive Somerset, New Jersey 08873 Re: K113115 Trade/Device Name: Light Age, Inc. EpiCare YAG™ Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: March 30, 2012 Received: April 2, 2012 Dear Ms. Reddington: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 – Ms. Elizabeth Reddington Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # INDICATIONS FOR USE STATEMENT KII3//5 #### Device Name: Light Age, Inc. EpiCare-YAGTM Laser System The Light Age Inc. EpiCare-YAG™ Laser System is intended for use in general and plastic surgery, dermatology, and podiatry for the treatment of vascular lesions, benign pigmented lesions, for the removal of dark tattoo inks and reduction of hypertrophic and keloid scars and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hair regrowth after a treatment regime on all skin types (Fitzpatrick I-VI including tanned skin.) Specifically the 1064 nm wavelength is indicated for: - Removal of unwanted hair, for stable long term or permanent hair reduction on all . skin types Fitzpatrick I-VI including tanned skin. - Photocoagulation and hemostasis of pigmented and vascular lesions such as but . not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins, and spider veins and poikiloderma of civatte. Coagulation and hemostasis of soft tissue. - Benign cutaneous lesions such as warts, scars, striae, and psoriasis - Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar � lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. - Pigmented lesion size reduction in patients with lesions who would potentially . benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. - Reduction of red pigmentation in hypertrophic and keloid scars where vascularity . is an integral part of the scar - Treatment of pseudofolliculitis barbae (PFB). . - Treatment of wrinkles including but not limited to periocular wrinkles and . perioral wrinkles. OR (Please Do Not Write Below This Line - Continue on Another Page if Needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X (Per 21 CFR 801 .109) Over the Counter Use (Optional Format 1-2-06) Melanil P. Rele for Paul Ozdoba (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number k113115 Page 12
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...