AUTOSURE VOICE II PLUS, AUTOSURE PLUS BLOOD GLUCOSE TEST STRIPS,CONTREX PLUS III
Applicant
Apex BioTechnology Corp.
Product Code
NBW · Clinical Chemistry
Decision Date
Jan 17, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Indications for Use
The AutoSure Voice II Plus Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. The meter includes voice functionality to assist visually impaired users. It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient and should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. The AutoSure Plus Blood Glucose Test Strips are to be used with the AutoSure Voice II Plus Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from a fingertip, palm, or forearm. They are not indicated for the diagnosis or screening of diabetes or for neonatal use. The purpose of the control solution test is to validate the performance of the blood glucose monitoring system using a testing solution with a known range of glucose. A control test that falls within the acceptable range indicates the user's technique is appropriate and the test strip and meter are functioning properly.
Device Story
System comprises blood glucose meter, test strips, and control solutions. Meter accepts capillary whole blood samples via test strips; performs quantitative glucose measurement. Includes voice functionality for visually impaired users. Used at home by patients for self-monitoring of diabetes. Healthcare providers use output to assess glycemic control. Control solutions validate system performance and user technique. Benefits include accessible, accurate, and convenient glucose monitoring for diabetes management.
Clinical Evidence
Bench testing only. Accuracy evaluated against YSI reference method using 145 capillary samples. Results showed 100% of samples within ±15 mg/dL (for <75 mg/dL) and 100% within ±20% (for ≥75 mg/dL). Lay-user study confirmed performance across fingertip, palm, and forearm sites. Robustness testing validated 5-year simulated use for cleaning/disinfection. Altitude, temperature, humidity, and hematocrit (30-55%) effects were within acceptable limits.
Indicated for lay use by people with diabetes to monitor blood glucose levels using fresh capillary whole blood from fingertips, forearm, or palm. Not for diagnosis/screening of diabetes or neonatal use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
AutoSure Voice II Blood Glucose Monitoring System (k102037)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
K113098
B. Purpose for Submission:
New device
C. Measurand:
Capillary whole blood glucose
D. Type of Test:
Quantitative, amperometric assay, glucose oxidase
E. Applicant:
Apex Biotechnology Corporation
F. Proprietary and Established Names:
AutoSure Voice II Plus Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345, Glucose test system
21 CFR 862.1660, Quality Control Material (assayed and unassayed)
2. Classification:
Class II, Class I (Reserved)
3. Product code:
NBW, System, Test, Blood Glucose, Over The Counter
CGA, Glucose Oxidase, Glucose
JJX, Quality Control Material
4. Panel:
{1}
Clinical Chemistry (75)
## H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indication(s) for use:
**AutoSure Voice II Plus Blood Glucose Monitoring System**
The AutoSure Voice II Plus Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. The meter includes voice functionality to assist visually impaired users. It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient and should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.
**AutoSure Plus Blood Glucose Test Strips**
The AutoSure Plus Blood Glucose Test Strips are to be used with the AutoSure Voice II Plus Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from a fingertip, palm, or forearm. They are not indicated for the diagnosis or screening of diabetes or for neonatal use.
**Contrex Plus II Glucose Control Solutions**
The purpose of the control solution test is to validate the performance of the blood glucose monitoring system using a testing solution with a known range of glucose. A control test that falls within the acceptable range indicates the user's technique is appropriate and the test strip and meter are functioning properly.
3. Special conditions for use statement(s):
For Over the Counter use
Not intended for use on neonates
Not for the diagnosis of or screening for diabetes mellitus
Not to be used for patients who are dehydrated, hypotensive, in shock, critically ill or in a hyperosmolar state
Allows testing on the fingertip, palm, or forearm.
Alternative site testing can be used only during steady-state blood glucose conditions
Alternative site testing (AST) should not be used to calibrate continuous glucose monitors (CGMs) nor for use in insulin dose calculations.
4. Special instrument requirements:
The AutoSure Voice II Plus Blood Glucose Meter
## I. Device Description:
The AutoSure Voice II Plus Blood Glucose Monitoring System consists of the AutoSure
{2}
Voice II Plus Blood Glucose Meter, AutoSure Plus Blood Glucose Test strips, and three levels of control solution (Low, Level 1, and Level 2). The control solutions for use with the system (Contrex Plus III Control Solutions) have been previously cleared (k102816). The AutoSure Voice II Plus Blood Glucose Monitoring System is used for testing of blood glucose by self-testers at home.
The AutoSure Voice II Plus Glucose Monitoring System is based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 6 seconds. The control solutions are used to test the performance of the device.
# J. Substantial Equivalence Information:
1. Predicate device name(s): AutoSure Voice II Blood Glucose Monitoring System
2. Predicate 510(k) number(s): k102037
3. Comparison with predicate:
| Item | Candidate Devices AutoSure Voice II Plus Blood Glucose Monitoring System | Predicate (k102037) AutoSure Voice II Blood Glucose Monitoring System |
| --- | --- | --- |
| Indications for use | The AutoSure Voice II Plus Blood Glucose Monitoring Systems are intended for the quantitative measurement of glucose in fresh capillary whole blood taken from fingertips, palm, or forearm. Testing is done outside the body (In Vitro Diagnostic use). It is indicated for lay use by people with diabetes as an aid to monitoring levels in Diabetes Mellitus and should only be used on a single patient. | Same |
| Test Principle | Electrochemical biosensor with carbon electrodes that measures current produced by a chemical reaction | Same |
| Enzyme | Glucose oxidase | Same |
| Sample Type | Fresh capillary whole blood | Same |
{3}
| Item | Candidate Devices
AutoSure Voice II Plus Blood Glucose Monitoring System | Predicate
(k102037)
AutoSure Voice II
Blood Glucose
Monitoring System |
| --- | --- | --- |
| Sample Site | Fingertip, the palm, the forearm | Same |
| Memory feature | 300 tests | Same |
| Day average | 7-, 14-, 30- day average glucose result | Same |
| Speaking function | Yes | Same |
| Measuring time | 6 sec | Same |
| Measurement range | 20-600 mg/dL | Same |
| Sample Volume | 0.8 μL | 1 μL |
| Meter dimensions (mm) | 93(L)x58(W)x21(H) mm | Same |
| Weight (g) | 79 g | Same |
| Test strip | AutoSure Plus Test Strip | AutoSure Test Strip |
| Autocoding | Yes | Same |
K. Standard/Guidance Document Referenced (if applicable):
- ISO15197:2003- In vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
- CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach
L. Test Principle:
The test is based on electrochemical biosensor technology and the principle of capillary action. The electrical current generated by the reaction of glucose with the reagent of the strip is measured by the meter and is displayed as the corresponding blood glucose level. The strength of the current produced by the reaction depends on the amount of glucose in the blood sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Within-run and between-run precision studies were performed. For the within-run precision studies, 10 replicates of each of 6 spiked venous whole blood glucose levels, were analyzed using 10 meters, and 3 lots of test strips. These tests were performed every day for 20 days. Results are summarized below.
{4}
| | Test Strip Lot 1 | | |
| --- | --- | --- | --- |
| Samples | Mean | SD | % CV |
| Interval 1
40-60 mg/dL | 50 | 2.7 | 5.4 |
| Interval 2
51-110 mg/dL | 85 | 3.6 | 4.2 |
| Interval 3
100-140 mg/dL | 122 | 3.2 | 2.7 |
| Interval 4
180-250 mg/dL | 202 | 6.3 | 3.1 |
| Interval 5
300-380 mg/dL | 326 | 8.6 | 2.6 |
| Interval 6
400-500 mg/dL | 440 | 5.7 | 1.3 |
| | Test Strip Lot 2 | | |
| --- | --- | --- | --- |
| Samples | Mean | SD | % CV |
| Interval 1
40-60 mg/dL | 51 | 3.3 | 6.5 |
| Interval 2
51-110 mg/dL | 84 | 2.9 | 3.4 |
| Interval 3
100-140 mg/dL | 123 | 1.9 | 1.6 |
| Interval 4
180-250 mg/dL | 202 | 6.9 | 3.4 |
| Interval 5
300-380 mg/dL | 327 | 11.5 | 3.4 |
| Interval 6
400-500 mg/dL | 439 | 16.2 | 3.6 |
| | Test Strip Lot 3 | | |
| --- | --- | --- | --- |
| Samples | Mean | SD | % CV |
| Interval 1
40-60 mg/dL | 51 | 2.3 | 4.5 |
| Interval 2
51-110 mg/dL | 86 | 3.1 | 3.6 |
| Interval 3
100-140 mg/dL | 122 | 3.8 | 3.1 |
| Interval 4
180-250 mg/dL | 203 | 5.7 | 2.8 |
| Interval 5
300-380 mg/dL | 327 | 8.0 | 2.4 |
{5}
A between-run precision study consisting of 10 replicates per day, for 20 days using each of 6 spiked venous whole blood glucose levels and 2 levels of control solution. Each sample was tested using 10 meters and 3 lots of test strips. Blood samples were prepared fresh daily over the 20 days. Results are summarized below.
| | Test Strip Lot 1 | | |
| --- | --- | --- | --- |
| Samples | Mean | SD | % CV |
| Interval 1 40-60 mg/dL | 50 | 1.2 | 2.4 |
| Interval 2 51-110 mg/dL | 85 | 0.7 | 1.9 |
| Interval 3 100-140 mg/dL | 122 | 1.3 | 1.0 |
| Interval 4 180-250 mg/dL | 202 | 2.8 | 1.4 |
| Interval 5 300-380 mg/dL | 326 | 3.9 | 1.2 |
| Interval 6 400-500 mg/dL | 440 | 3.6 | 0.8 |
| Control Level 1 | 110 | 1.2 | 1.9 |
| Control Level 2 | 208 | 1.6 | 0.8 |
| | Test Strip Lot 2 | | |
| --- | --- | --- | --- |
| Samples | Mean | SD | % CV |
| Interval 1 40-60 mg/dL | 51 | 1.2 | 2.4 |
| Interval 2 51-110 mg/dL | 84 | 0.9 | 1.1 |
| Interval 3 100-140 mg/dL | 123 | 1.7 | 1.4 |
| Interval 4 180-250 mg/dL | 202 | 2.6 | 1.3 |
| Interval 5 300-380 mg/dL | 327 | 4.2 | 1.3 |
| Interval 6 400-500 mg/dL | 439 | 3.1 | 0.7 |
| Control Level 1 | 110 | 1.3 | 1.9 |
| Control Level 2 | 209 | 1.5 | 0.7 |
{6}
| | Test Strip Lot 3 | | |
| --- | --- | --- | --- |
| Samples | Mean | SD | % CV |
| Interval 1
40-60 mg/dL | 51 | 1.5 | 2.9 |
| Interval 2
51-110 mg/dL | 86 | 3.1 | 3.6 |
| Interval 3
100-140 mg/dL | 122 | 1.9 | 1.6 |
| Interval 4
180-250 mg/dL | 203 | 3.6 | 1.8 |
| Interval 5
300-380 mg/dL | 327 | 5.3 | 1.6 |
| Interval 6
400-500 mg/dL | 442 | 4.1 | 0.9 |
| Control Level 1 | 111 | 1.3 | 1.9 |
| Control Level 2 | 208 | 1.3 | 0.6 |
b. Linearity/assay reportable range:
Linearity was evaluated using 3 lots of test strips, 10 meters, and 8 venous whole blood samples with glucose levels ranging from 13-651 mg/dL, obtained by spiking pooled venous blood with a glucose solution. Each glucose level was analyzed 40 times over 3 test strip lots. Linear regression analysis for each test strip lot compared to the YSI resulted in:
$$
y = 1.018x - 0.3134; R^2 = 0.9983 \text{ for Test Strip Lot 1}
$$
$$
y = 1.031x - 3.5555; R^2 = 0.9986 \text{ for Test Strip Lot 2}
$$
$$
y = 1.021x - 2.1524; R^2 = 0.9979 \text{ for Test Strip Lot 3}
$$
The claimed range of measurement for this device is 20 to 600 mg/dL. Data from bench studies and software verification studies were provided to demonstrate that if a sample is less than 20 mg/dL, the result is flagged by the meter as LO. If a sample result exceeds 600 mg/dL, the result is flagged by the meter as HI.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Three levels of control material, Contrex Plus Control Solution Low (40-70mg/dL), Contrex Plus Control Solution Level 1 (87-131 mg/dL) and Contrex Plus Control Solution Level 2 (186-280 mg/dL), are available for use with this test system. These control solutions have been previously cleared (k102816). See k102816 for control solutions traceability, shelf life, storage, and stability information. The Contrex Plus control solutions were qualified for use with the AutoSure Voice II Plus system. The
{7}
control solutions are not provided with the meter.
The AutoSure Plus test strips are identical to the test strips cleared under k102816, and differ only in trade name. Stability testing protocols and acceptance criteria for the test strips were reviewed in k102816 and were found to be acceptable. The manufacturer claims a shelf life stability of 17 months and an open-vial stability of 3 months at the recommended storage temperatures of $5^{\circ}\mathrm{C}$ to $30^{\circ}\mathrm{C}$ for the AutoSure Plus test strips.
# d. Detection limit:
The reportable range is 20 to $600\mathrm{mg / dL}$ based on linearity/reportable range studies above (section M.1.b.).
# e. Analytical specificity:
The sponsor tested substances for interference using 1 lot of test strips (10 test strips per interferent level), 10 meters, and 3 levels of glucose (achieved by adjusting human venous blood glucose levels to the ranges of 70-90, 110-130, and $300 - 330\mathrm{mg / dL}$ ). Samples were then spiked with the following interfering substances. Each sample was analyzed 10 times. The labeling states that elevated concentration of L-DOPA ( $>1.5$ mg/dL), ibuprofen ( $>40$ mg/dL), tolazamide ( $>60$ mg/dL), ascorbic acid ( $>5.0$ mg/dL), fructose ( $>15$ mg/dL), uric acid ( $>15$ mg/dL), cholesterol ( $>400$ mg/dL), bilirubin conjugated ( $>25$ mg/dL), bilirubin unconjugated ( $>15$ mg/dL), triglyceride ( $>2000$ mg/dL), and methyl-dopa ( $>7.5$ mg/dL) may affect test results. The following table lists the concentrations of each substance at which no significant interference ( $\leq 10\%$ ) was detected:
| Interfering Substance | Therapeutic/ Physiological Levels (mg/dL) | Test Levels (mg/dL) |
| --- | --- | --- |
| Acetaminophen | 1 – 3 | 10 and 20 |
| Tolbutamide | 5.4 – 10.8 | 10 and 64 |
| Dopamine | 0.03 | 0.05 and 0.1 |
| Salicylic acid | 10 – 30 | 40 and 65 |
| Methyl-Dopa | 1 – 7.5 | 7.5 and 15.0 |
| Tetracycline | 0.2 – 0.5 | 0.5 and 1.5 |
| Ephedrine | 0.014 | 0.02 and 0.05 |
| Mannitol | 10 | 15 and 30 |
| Mannose | 1.2 | 2.0 and 4.0 |
| Sorbitol | 0.05 | 0.1 and 0.2 |
| Maltose | -- | 20 and 50 |
| EDTA | -- | 1x and 5x |
| Lactose | -- | 10 and 25 |
| Heparin | -- | 1x and 5x |
| Maltotriose | -- | 120 and 240 |
| Maltotetraose | -- | 60 and 120 |
{8}
9
| Xylitol | 0.02 | 0.05 and 0.1 |
| --- | --- | --- |
| Xylose | -- | 10 and 25 |
| Hemoglobin | 100 - 200 | 100 and 200 |
| Creatinine | 0.6 – 1.3 | 2.0 and 5.0 |
| Galactose | < 5 | 10 and 15 |
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Reference Method Comparison:
The sponsor performed a system accuracy evaluation comparing the AutoSure Voice II Plus to YSI. Healthcare professionals tested 145 capillary finger samples in 6 concentration categories (50-80, 81-120, 121-200, 201-300, 301-400, and >400 mg/dL) per strip, using 8 meters and 3 lots of test strips using the AutoSure Voice II Plus meter and the YSI (the reference method). Seven additional glycolyzed samples were also tested to obtain concentrations < 50 mg/dL. Results are summarized below.
For glucose concentrations < 75 mg/dL
| within ± 5 mg/dL | within ± 10 mg/dL | within ± 15 mg/dL |
| --- | --- | --- |
| 13/28 (46%) | 25/28 (89%) | 28/28 (100.0%) |
For glucose concentrations ≥ 75 mg/dL
| within ± 5 % | Within ± 10 % | within ± 15 % | within ± 20 % |
| --- | --- | --- | --- |
| 77/124 | 111/124 | 123/124 | 124/124 |
| (62%) | (90%) | (99%) | (100.0%) |
Linear Regression Analysis:
| Comparison | N | Slope and y-intercept | R² |
| --- | --- | --- | --- |
| AutoSure Voice II Plus vs. YSI | 152 | Y = 0.9878x + 1.3379 | 0.9899 |
b. Matrix comparison:
Not applicable. Capillary whole blood is the only indicated matrix.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
{9}
c. Other clinical supportive data (when a. and b. are not applicable):
# Fingerstick and Alternative Site Testing Comparison Studies:
The sponsor performed a lay-user study where accuracy of the device was tested using 145 fingerstick samples obtained by the lay-user and 145 samples per each alternative site (the palm and the forearm) obtained by the lay-user. Participants, who were able to read the User's Manual in English, were instructed to read the manual and perform testing on the finger and then the alternative sites. A technician collected capillary blood for measurements on YSI. Results were obtained for each of the alternative sites. Samples in the study contained glucose concentrations that ranged from 62 to $492\mathrm{mg / dL}$ . Results are summarized below.
For glucose concentrations $< 75\mathrm{mg / dL}$
| Sites | within ± 5 mg/dL | within ± 10 mg/dL | within ± 15 mg/dL |
| --- | --- | --- | --- |
| Finger | 10/21 (48%) | 18/21 (86%) | 20/21 (95%) |
| Palm | 14/21 (67%) | 19/21 (90%) | 21/21 (100.0%) |
| Forearm | 11/21 (52%) | 20/21 (95%) | 21/21 (100.0%) |
For glucose concentrations $\geq 75\mathrm{mg / dL}$
| Site | Within ± 5 % | within ± 10 % | within ± 15 % | within ± 20 % |
| --- | --- | --- | --- | --- |
| Finger | 60/124 (48%) | 106/124 (85%) | 121/124 (98%) | 124/124 (100.0%) |
| Palm | 61/124 (49%) | 103/124 (83%) | 120/124 (97%) | 124/124 (100.0%) |
| Forearm | 63/124 (51%) | 103/124 (83%) | 118/124 (95%) | 124/124 (100.0%) |
Linear Regression Analysis:
| Comparison | N | Slope and y-intercept | R2 |
| --- | --- | --- | --- |
| Finger vs. YSI | 145 | Y = 0.9930x + 2.1102 | 0.9822 |
| Palm vs. YSI | 145 | Y = 0.9944x + 0.3915 | 0.9802 |
| Forearm vs. YSI | 145 | Y = 0.9910x - 0.6339 | 0.9830 |
# 4. Clinical cut-off:
Not applicable.
# 5. Expected values/Reference range:
| Time of day | Range, Non-diabetes |
| --- | --- |
| Before meals | Less than 100 mg/dL |
| After meals | Less than 140 mg/dL |
{10}
The sponsor references: American Diabetes Association. Standards of Medical Care in Diabetes, Diabetes Care. 2010;33:S11-S61.
## N. Instrument Name:
The AutoSure Voice II Plus Blood Glucose Meter
## O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?:
Yes ______ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?:
Yes ______ or No ☐.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ or No ______
The applicant has provided documentation that indicates the device was designed and developed under good software life-cycle processes.
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood from the finger, palm, and forearm. Since the whole blood sample is applied directly to the test strip there are no special handling or storage issues.
5. Calibration:
The meter is a non-coding meter, therefore no coding is required by the user.
6. Quality Control:
The sponsor manufactures three levels of glucose control solution, Low, Level 1, and Level 2 to be used with the AutoSure Voice II Plus Blood Glucose Monitoring System. These control solutions must be purchased separately and are not provided with the
11
{11}
device kits. Instructions for how to purchase the control solution are provided in the user manuals. To perform a control test the user is instructed to press the down button while the blood drop symbol is flashing. The "ctl" symbol will then appear on the display. An acceptable range for each control level is printed on the test strip vial label. If the control values fall outside these ranges, the user is referred to the user manual and customer support for problems and more information
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. The device is intended for single-patient use. Disinfection studies were performed on the AutoSure Voice II Plus meter and lancet device by outside commercial laboratory testing services to determine the disinfection efficacy of the meter and lancing device to the recommended cleaning and disinfection protocol, and its effectiveness in preventing the spread of bloodborne pathogens, particularly hepatitis B virus (HBV). Dispatch Hospital Cleaner Disinfectant Towels with Bleach disposable wipes (EPA Reg. No: 56392-8) were validated, demonstrating complete inactivation of live virus for use with the meter and lancing device. The sponsor also conducted robustness studies and demonstrated that there was no change in performance or in the external materials of the meter and lancing device after 1,825 cleaning and disinfection cycles to simulate 5 years of use by lay-users. Each robustness cycle tested consisted of one pre-clean wipe and one disinfecting wipe.
2. The effect of different hematocrit levels was evaluated with 3 lots of test strips using 10 test strips at each of 5 concentration ranges of glucose (30-50, 51-70, 111-150, 250-350, 500-600 mg/dL). The glucose samples were prepared from venous blood samples at 6 hematocrit levels at approximately 30, 33, 35, 43, 50 and 55%. Each of the results were compared to the value obtained from the same plasma glucose concentration obtained by YSI. The data vs. YSI was reviewed and was found to be acceptable to support the claimed hct range of 30-55%.
3. The effect of altitude was evaluated at five whole blood samples with glucose concentrations ranging from 48 to 470 mg/dL, testing at sea level and 3150 meters (10,334 feet above sea level). Each glucose concentration was measured 20 times at each altitude. The bias was calculated relative to YSI at sea level and at 3150 meters. The results demonstrate that the system meets the acceptance criteria for testing at altitudes up to 3,150 meters (10,334 feet above sea level).
4. The sponsor performed temperature and humidity studies at the combined extremes of 10°C/RH: 18%, 40°C/RH: 20%, 10°C/RH: 88%, and 40°C/RH: 89%, with venous blood samples (50-400 mg/dL) that demonstrated that the AutoSure Voice II Plus meter can be used at temperatures of 50 to 104°F (10 to 40°C) and 20% to 90% relative humidity.
5. Insufficient sample studies were performed at volumes of 0.2 to 1.5μL on ten meters and 3 test strip lots. Three glucose concentrations were tested ranging from approximately 50-350 mg/dL, as determined by the YSI. Appropriate sample volume was determined
12
{12}
if the meter testing could start properly and if the meter results matched the YSI results. A blood volume ≥ 0.8 μL met the criteria.
6. The sponsor provided a readability study and obtained Flesch-Kincaid Grade Level Scores of 8.0, 7.6, and 7.8 for the AutoSure Plus test strip insert, the AutoSure Voice II Plus User's Manual, and Contrex Plus III package insert, respectively.
7. The sponsor stated that they conformed to the following guidelines and provided the appropriate documentation to demonstrate compliance:
- IEC/EN 60601-1-2: Medical electrical equipment, Part 2. Electromagnetic compatibility, 2007.
- EN 61000-3-2: Medical electrical equipment Part 2. Electromagnetic compatibility, 2006.
- EN 61000-3-3: Medical electrical equipment Part 3. Electromagnetic compatibility, 2008.
- The AutoSure Voice II Plus meter is identical to the cleared meter of k102037 and differs only in trade name. EMC testing was evaluated and certified by QuieTek of Taiwan (R.O.C.) and a letter of attestation was issued to Apex Biotechnology Corp. See k102037.
8. The AutoSure Voice II Plus meter is identical to the cleared meter of k102037 and differs only in trade name. The voice functionality feature of the meter of k102037 was tested and appropriate software verification and validation information for this feature was provided. See k102037. Voice functionality testing results is applicable to this current submission for the AutoSure Voice II Plus meter.
9. A study with visually impaired users was previously conducted to support the claim for use with visually impaired users for the meter of k102037 (See k073137). The AutoSure Voice II Plus meter is identical to the cleared meter of k102037 and differs only in trade name. Testing results from visually impaired users is applicable to this current submission for the AutoSure Voice II Plus meter to support a claim for use with visually impaired users.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.