The EmbryoScope (Version D) provides an environment with controlled temperature, CO2 (and other gases) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development. EmbryoViewer is an optional accessory software package for use in displaying, comparing, storing, and transferring EmbryoScope (Version D) generated images. The software includes a user annotation function for capturing information on embryo development parameters, treatment data, and outcome data. The EmbryoViewer software does not control any hardware components in the EmbryoScope (Version D) device.
Device Story
EmbryoScope (Version D) is an embryo incubator with integrated inverted microscope for time-lapse imaging; logs incubation conditions and captures images of embryos during 5-day development. EmbryoViewer is an optional accessory software for displaying, comparing, storing, and transferring these time-lapse images. Software includes manual annotation tools for embryologists to record development parameters, treatment data, and outcomes. Device used in clinical embryology labs; operated by trained personnel. Software does not control hardware; acts as a viewing and data management platform. Output allows clinicians to inspect high-resolution images and track embryo development, facilitating clinical decision-making regarding embryo selection for transfer.
Clinical Evidence
Bench testing only. Software validation testing conducted to verify data transfer and annotation accuracy. Electrical safety, electromagnetic compatibility, and environmental/operating performance testing performed per IEC 60601-1.
Technological Characteristics
Incubator with integrated inverted microscope for time-lapse imaging. Software: EmbryoViewer accessory for image display and manual annotation. Compliance: IEC 60601-1 (electrical/EMC), EN 62304 (software lifecycle).
Indications for Use
Indicated for embryos in an assisted reproduction environment requiring controlled incubation (temperature, CO2/gases) and time-lapse imaging for up to 120 hours post-fertilization to day 5 of development.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
DEC 2 1 2011
Image /page/0/Picture/1 description: The image shows the logo for UNISENSE FERTILITECH. The logo features a stylized graphic of cells or molecules connected by a line, positioned above the company name. The text "UNISENSE" is in a bold, sans-serif font, with a circled "R" indicating a registered trademark. Below "UNISENSE" is the word "FERTILITECH", also in a bold, sans-serif font.
## K113075
## 510(k) Summary - EmbryoScope (Version D) and EmbryoViewer software
## Administrative information:
| Name:<br>Address: | Unisense FertiliTech A/S<br>Tueager 1<br>DK-8200 Aarhus N<br>Denmark |
|---------------------------------------------|----------------------------------------------------------------------|
| Contact person: | Mette Munch, QA Consultant |
| Telephone Number:<br>Fax Number:<br>E-mail: | +45 89449500<br>+45 89449549<br>mm@unisense.com |
| Date of summary: | 19-Dec-2011 |
Propose of submission: The propose of this submission is to add the optional accessory EmbryoViewer software.
Device name :
| Trade name | Common name | Class | CFR Reference | Procode |
|--------------------------------------------------------------|--------------------------------|-------|---------------|---------|
| EmbryoScope<br>(Version D) and<br>EmbryoViewerTM<br>software | Embryo<br>monitoring<br>system | II | 884.6120 | 85 MQG |
## Indications for Use:
The EmbryoScope (Version D) provides an environment with controlled temperature, CO2 (and other gases) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development.
EmbryoViewer is an optional accessory software package for use in displaying, comparing, storing, and transferring EmbryoScope (Version D) generated images. The software includes a user annotation function for capturing information on embryo development parameters, treatment data, and outcome data. The EmbryoViewer software does not control any hardware components in the EmbryoScope (Version D) device.
Phone: +45 89 44 95 00 Fax: +45 89 44 95 49 sales@fertilitech.com www.fertilitech.com
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for a company called UNISENSE FERTILITECH. The logo features the company name in bold, uppercase letters, with the word "UNISENSE" appearing above the word "FERTILITECH". To the top of the company name is a graphic of three circles, with a line going through one of the circles. The "R" symbol is to the right of the word UNISENSE.
## Device Description:
The EmbryoScope 1M - (Version D) is an embryo incubator, which performs timelapse microscopy at multiple planes and logging of incubation conditions on individual embryos during their development. The EmbryoScope (Version D) included in this submission is identical to that cleared in previous 510(k) submissions (K092183 and K111715).
The EmbryoViewer software is an accessory to the EmbryoScope (Version D). The Embryo Viewer™ software does not perform any diagnostics, but only shows data from the EmbryoScope™ (Version D) and data filled in by the user. The EmbryoViewer™ software assists the user by allowing for inspection of high resolution time-lapse images of embryo development, detailed annotation tools and inspection of EmbryoScope™ (version D) running conditions.
### Comparison to Predicate Device:
Table 1: Comparison of EmbryoScope (Version D) and EmbryoViewer™ software to EmbryoScope (Version D) -- K092183 and K111715.
| | EmbryoScope (Version D) and<br>EmbryoViewer software | EmbryoScope (Version D) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | The EmbryoScope (Version<br>D) provides an environment<br>with controlled<br>temperature, CO2 (and<br>other gases) for the<br>development of embryos.<br>This model has an<br>integrated inverted<br>microscope and imaging<br>system for embryo viewing.<br>Device use is limited to five<br>days (120 hr) covering the<br>time from post-fertilization<br>to day 5 of development.<br><br>EmbryoViewer is an<br>optional accessory software<br>package for use in<br>displaying, comparing,<br>storing, and transferring<br>EmbryoScope (Version D)<br>generated images. The<br>software includes a user<br>annotation function for<br>capturing information on<br>embryo development<br>parameters, treatment data,<br>and outcome data. The<br>EmbryoViewer software | The EmbryoScope (Version<br>D) provides an<br>environment with<br>controlled temperature,<br>CO2 (and other gases) for<br>the development of<br>embryos. This model has<br>an integrated inverted<br>microscope and imaging<br>system for embryo viewing.<br>Device use is limited to<br>five days (120 hr) covering<br>the time from post-<br>fertilization to day 5 of<br>development. |
Phone: +45 89 44 95 00 Fax: +45 89 44 95 49 sales@fertilitech.com www.fertilitech.com
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for UNISENSE FERTILITECH. The logo consists of the words "UNISENSE" and "FERTILITECH" stacked on top of each other. Above the words is a graphic of three circles connected together with a line going through them. The logo is in black and white.
| Discussion of<br>NonClinical<br>Tests performed<br>for the<br>determination of<br>Substantial<br>Equivalence | does not control any<br>hardware components in<br>the EmbryoScope (Version<br>D) device.<br>Underwent and passed<br>electrical safety<br>electromagnetic<br>compatibility,<br>environmental and<br>operating performance<br>testing. Classification<br>according to IEC 60601-1.<br>The EmbryoViewer<br>software is fulfilling the<br>requirements of the EN<br>62304 standard according<br>to software testing. | Underwent and passed<br>electrical safety<br>electromagnetic<br>compatibility,<br>environmental and<br>operating performance<br>testing. Classification<br>according to IEC 60601-1. |
|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Summary of comparison to predicate device:
The current submission includes addition of the optional EmbryoViewer to be used with the predicate device The EmbryoScope (Version D). This extension results in a new Indication for Use reflection the addition of annotation tool.
The EmbryoViewer software does not introduce changes to the EmbryoScope (Version D) thus the components in this device are identical in the predicate device.
As the EmbryoViewer software does not control any functions on the EmbryoScope (Version D) device it can not affect the operation of the EmbryoScope (Version D) device.
The purpose of the EmbryoViewer software is to view images generated by the EmbryoScope (version D), and software validation testing has been conducted to verify data transfer.
The annotation functions of the device do not raise any concerns because the information inputted by the physician is the same that would typically be collected by an embryologist and maintained on paper patient records. Also, software testing has been conducted to verify correct annotation for a specific patient/embryo.
In the event of a software malfunction, there will be no affect on the EmbryoScope device.
Conclusion: The addition of the optional EmbryoViewer software to the EmbryoScope (Version D) does not raise new types of safety and effectiveness questions.
Phone: +45 89 44 95 00 Fax: +45 89 44 95 49 sales@fertilitech.com www.fertilitech.com
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter of the circle. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
DEC 2 1 2011
Mette Munch, Ph.D. QA Consultant Unisense FertiliTech A/S Tueager 1 DK-8200 AARHUS N DENMARK
Re: K113075
Trade/Device Name: EmbryoScope (Version D) and EmbryoViewer Software Regulation Number: 21 CFR\$ 884.6120 Regulation Name: Assisted reproduction accessories Regulatory Class: II Product Code: MQG Dated: October 14, 2011 Received: October 24, 2011
Dear Dr. Munch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Hubert Lemmer MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: EmbryoScope (Version D) and EmbryoViewer software
Indications For Use:
The EmbryoScope (Version D) provides an environment with controlled temperature, CO2 (and other gases) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development.
EmbryoViewer is an optional accessory software package for use in displaying, comparing, storing, and transferring EmbryoScope (Version D) generated images. The software includes a user annotation function for capturing information on embryo development parameters, treatment data, and outcome data. The EmbryoViewer software does not control any hardware components in the EmbryoScope (Version D) device.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Tom Whing
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.