K113053 · Respironics, Inc. · MNS · Feb 1, 2012 · Anesthesiology
Device Facts
Record ID
K113053
Device Name
BIPAP A 30 VENTILATORY SUPPORT SYSTEM
Applicant
Respironics, Inc.
Product Code
MNS · Anesthesiology
Decision Date
Feb 1, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The BiPAP A30 Ventilator is intended to provide non-invasive ventilatory support to treat adult and pediatric patients weighing over 10kg (22lbs) with Obstructive Sleep Apnea (OSA) and Respiratory Insufficiency. It is intended to be used in both the home and clinical settings, such as hospitals, sleep laboratories, sub-acute care institutions.
Device Story
Microprocessor-controlled, blower-based positive pressure ventilatory system with integrated heated humidifier; provides non-invasive ventilatory support. Operates in home and clinical settings (hospitals, sleep labs, sub-acute care). Inputs: patient breathing parameters via patient circuit (tubing, exhalation device, mask). Therapy modes: CPAP, Spontaneous, Spontaneous/Timed, Pressure Control; includes Bi-Flex or AVAPS therapy features. Graphical user interface displays settings and data; system includes fixed and user-adjustable alarms to alert users to treatment changes. Healthcare professionals configure therapy; device assists patient breathing. Benefits: provides ventilatory support for patients with OSA and respiratory insufficiency.
Clinical Evidence
Bench testing only. System verification and side-by-side testing against predicate devices performed using common protocols. Testing confirmed compliance with ISO 10651-6, IEC 60601-1, IEC 60601-1-2, and EN ISO 8185. All tests met predetermined acceptance criteria.
Technological Characteristics
Microprocessor-controlled blower-based positive pressure system. Includes integrated heated humidifier. Connectivity via patient circuit (tubing, exhalation device, mask). Complies with IEC 60601-1, IEC 60601-1-2, ISO 10651-6, and EN ISO 8185. Software-controlled therapy modes (CPAP, Spontaneous, S/T, Pressure Control) with Bi-Flex/AVAPS features.
Indications for Use
Indicated for adult and pediatric patients >10kg (22lbs) with Obstructive Sleep Apnea (OSA) and Respiratory Insufficiency requiring non-invasive ventilatory support.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
BiPAP C Series Ventilatory Support System (K092818)
{0}------------------------------------------------
FEB - 1 2012
Addition Information Request - K113053
Philips Respironics BiPAP A30 Ventilatory Support System
Section 05: 510(k) Summary
1/3053
# Administrative Information and Device Identification
| Name and address of the<br>manufacturer and sponsor of the<br>510(k) submission: | Manufacturer: |
|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Respironics, Inc.<br>312 Alvin Drive<br>New Kensington, PA 15068 |
| | Sponsor: |
| | Respironics<br>1740 Golden Mile Highway<br>Monroeville, PA 15146<br>Office: 724-387-7562<br>Fax: 724-387-7490 |
| FDA registration number of the<br>manufacturer of the new device: | 2518422 |
| Official contact person for all<br>correspondence: | Elaine Larkin<br>Regulatory Affairs Engineer<br>Respironics<br>1740 Golden Mile Highway<br>Monroeville, PA 15146<br>Office: 724-387-5350<br>Fax: 724-387-7490<br>Email: elaine.larkin@philips.com |
| Submission Date | October 12, 2011 |
| Classification Reference | 21 CFR 868.5895<br>a) Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.<br>(b) Classification. Class II (performance standards). |
| Panel Code: | MNS - ventilator, continuous, non-life<br>supporting |
| Classification Panel: | Anesthesiology |
{1}------------------------------------------------
.
,
.
.
.
Philips Respironics BiPAP A30 Ventilatory
Support System
.
.
。
| Common/Usual Name | Ventilatory Support System |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary name of new device: | • Respironics BiPAP A30 Ventilatory Support<br>System |
| Predicate Device Name(s) and<br>510(k) numbers: | • BiPAP C Series Ventilatory Support System<br>(K092818)<br>• V60 Ventilator<br>(K082660)<br>• ResMed Stellar 150/100<br>(K103167) |
| Reason for submission: | Device modifications and additional accessories |
:
·
{2}------------------------------------------------
Philips Respironics BiPAP A30 Ventilatory Support system
# Intended Use
The BiPAP A30 Ventilator is intended to provide non-invasive ventilatory support to treat adult and pediatric patients weighing over 10kg (22lbs) with Obstructive Sleep Apnea (OSA) and Respiratory Insufficiency. It is intended to be used in both the home and clinical settings, such as hospitals, sleep laboratories, sub-acute care institutions.
## Device Description
The Respironics BiPAP A30 Ventilatory Support System is a microprocessor controlled blower based positive pressure ventilatory system with integrated heated humiditier. The device platform being used as the key topic for this submission was previously cleared in K092818. The same ventilation modalities and therapy features, previously cleared in K071509 is also included in the BiPAP A30 Ventilatory Support System, which is the topic of this submission. These modes and therapy features include: CPAP, Spontaneous, Spontaneous/Timed, Pressure Control modes with Bi-Flex or AVAPS therapy features available if enabled by the health care professional.
A Graphical user interface displays device data and device settings.
The BiPAP A30 Ventilatory Support System is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.
Like its predicates, the BiPAP A30 Ventilatory Support System is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen tubing, an exhalation device, and a patient interface device. ·
#### Performance Data
Design and Verification activities were performed on the BiPAP A30 as a result of the risk analysis and product design requirements. All tests confirmed the product met the predetermined acceptance criteria. Performance testing comprises pressure performance, trigger and cycling, as well as volume assured pressure support ventilation. In addition to system verification testing was performed using common protocols for BiPAP A30 and the predicate device. The side-by-side testing demonstrated that the BiPAP A30 is Substantially Equivalent to the predicate devices.
าใ
{3}------------------------------------------------
This device has been tested to appropriate ISO and IEC standards and other applicable requirements
passing all test protocols. The BIPAP A30 was designed and tested according to:
- IEC 60601-1:1988, Medical electrical equipment Part 1: General requirements for basic safety . and essential performance and its Amendments Al1:1991 and A2:1 995
- . IEC 60601-1-2:2007, Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- . EN ISO 8185 - Humidifiers for Medical Use - General Requirements for Humidification Systems
- ISO 10651-6:2004, Lung ventilators for medical use Particular requirements for basic safety and . essential performance. Part 6: Home care ventilatory support devices.
The new device complies with the applicable requirements referenced in the FDA guidance documents:
- FDA Draft Reviewer Guidance for Ventilators (July 1995) .
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical .
- Devices (May 11, 2005) .
### Non-Clinical Testing
This device has been tested to appropriate collateral and particular ISO, ASTM, and IEC standards and other applicable requirements passing all test protocols. The Respironics BiPAP A30 Ventilatory Support Systems was designed and tested according to guidance outlined in:
- FDA Draft Reviewer Guidance for Premarket Notification Submissions Anesthesiology and 1. Respiratory Devices Branch; Division of Cardiovascular, Respiratory, and Neurological Devices (November 1993);
- 2. FDA Draft Reviewer Guidance for Ventilators July 1995; and
- FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical 3. Devices" (May 11, 2005).
## Substantial Equivalence
The modified device has the following similarities to the previously cleared predicate devices:
- o Same intended use.
- D Same operating principle.
- 0 Same technology.
- Same manufacturing process. 0
Design verification tests were performed on the Respironics BiPAP A30 Ventilatory Support as a result of the risk analysis and product requirements. All tests were verified to meet the required acceptance criteria. Respironics has determined that the modifications have no impact on the safety
{4}------------------------------------------------
and effectiveness of the device. In summary, the device described in this submission is substantially equivalent to the predicate devices.
The modified device complies with the applicable standards referenced in the Guidance for FDA Reviewers and Industry "Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices," May 2006.
## Conclusion:
Bench testing and comparative analysis has confirmed that the BiPAP A30 Ventilatory Support System performs equivalently to the cited predicate devices. The Respironics BiPAP A30 Ventilatory Support System is substantially equivalent to the predicate devices listed above and the device, as changed, does not raise any new issues of safety and effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the wings and body of the bird.
Food and Drug Administration. 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Respironics, Incorporated C/O Ms. Elaine Larkin Engineer, Regulatory Affairs Sleep & Home Respiratory Group 1740 Golden Mile Highway Monroeville, Pennsylvania 15146
FEB - 1 2012
Re: K113053
Trade/Device Name: BiPAP A30 Ventilatory Support System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MNS Dated: January 23, 2012 Received: January 25, 2012
Dear Ms. Larkin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Ms. Larkin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Rh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Section 4.0 Indications for Use
#### Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: BiPAP A30 Ventilatory Support System
The BiPAP A30 ventilator is intended to provide non-invasive ventilatory support to treat adult and pediatric patients weighing over 10 kg (22lbs) with Obstructive Sleep Apnea (OSA and Respiratory Insufficiency. It is intended to be used in both the home and clinical settings, such as hospitals, sleep laboratories, and sub-acute care institutions.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schult
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.