The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
Device Story
AutoPulse Resuscitation System Model 100 is an automated, battery-powered external cardiac compressor. Device consists of a reusable platform, single-use chest compression assembly (LifeBand), rechargeable battery, and multi-chemistry charger. Platform houses mechanical drive, control system, and electronics to generate chest compression force per AHA protocols. Used in clinical settings to provide blood flow during cardiac arrest. Operator (clinician) places device on patient; system performs automated compressions. Output is mechanical chest compression. Benefits include consistent, automated CPR delivery. Subject 510(k) introduces Li-Ion battery and multi-chemistry charger; no changes to platform, LifeBand, or core operating principles.
Clinical Evidence
Bench testing only. Verified system-level compatibility of new Li-Ion battery and multi-chemistry charger with existing AutoPulse platform. Testing confirmed identical performance to NiMH battery-powered system. Charger functionality verified for charging, battery maintenance, and end-of-life identification. Compliance with international standards (IEC 60601-1, IEC 60601-1-2, IEC 60068, IEC 61000, CISPR11, IEC 62133) confirmed via environmental, electrical safety, and electromagnetic compatibility testing.
Indicated for adult patients in clinical death (lack of spontaneous breathing and pulse) as an adjunct to manual CPR.
Regulatory Classification
Identification
An external cardiac compressor is an externally applied prescription device that is electrically, pneumatically, or manually powered and is used to compress the chest periodically in the region of the heart to provide blood flow during cardiac arrest. External cardiac compressor devices are used as an adjunct to manual cardiopulmonary resuscitation (CPR) when effective manual CPR is not possible (e.g., during patient transport or extended CPR when fatigue may prohibit the delivery of effective/consistent compressions to the victim, or when insufficient EMS personnel are available to provide effective CPR).
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Nonclinical performance testing under simulated physiological conditions must demonstrate the reliability of the delivery of specific compression depth and rate over the intended duration of use.
(2) Labeling must include the following:
(i) The clinical training necessary for the safe use of this device;
(ii) Adjunctive use only indication prominently displayed on labels physically placed on the device and in any device manuals or other labeling;
(iii) Information on the patient population for which the device has been demonstrated to be effective (including patient size and/or age limitations,
*e.g.,* adult, pediatric and/or infant); and(iv) Information on the time necessary to deploy the device as demonstrated in the performance testing.
(3) For devices that incorporate electrical components, appropriate analysis and testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(4) Human factors testing and analysis must validate that the device design and labeling are sufficient for effective use by the intended user, including an evaluation for the time necessary to deploy the device.
(5) For devices containing software, software verification, validation, and hazard analysis must be performed.
(6) Components of the device that come into human contact must be demonstrated to be biocompatible.
Predicate Devices
AutoPulse Resuscitation System Model 100 (K072527)
AutoPulse Resuscitation System Model 100 (K063602)
AutoPulse Resuscitation System Model 100 (K040453)
AutoPulse Resuscitation System Model 100 (K032852)
AutoPulse Resuscitation System Model 100 (K022345)
AutoPulse Resuscitation System Model 100 (K011046)
Submission Summary (Full Text)
{0}------------------------------------------------
# 3 510(k) Summary
510(K) SUMMARY FOR AutoPulse® Resuscitation System Model 100
Submitter's Name, Address, Telephone Number, and Contact Person
ZOLL Circulation, Inc. 650 Almanor Avenue Sunnyvale, CA 94085 Establishment Registration Number: 3003793491
Contact: Sam Nanavati . Phone: 408-541-2140 Fax: 408-541-1030 Email: snanavati@zollcirculation.com
## Name of Device
AutoPulse® Resuscitation System Model 100
## Common or Usual Name
Automatic Mechanical Chest Compressor
## Classification Name
21 CFR 870.5200 External cardiac compressor
## Device Class
Class III
## Predicate Devices
| Device Name | 510(k) # | Decision<br>Date | Decision |
|---------------------------------------------|----------|------------------|-------------------------------|
| AutoPulse Resuscitation<br>System Model 100 | K072527 | 03/19/2008 | Substantially equivalent (SE) |
| | K063602 | 12/21/2006 | Substantially equivalent (SE) |
| | K040453 | 03/11/2004 | Substantially equivalent (SE) |
| | K032852 | 11/12/2003 | Substantially equivalent (SE) |
| | K022345 | 08/15/2002 | Substantially equivalent (SE) |
| | K011046 | 10/25/2001 | Substantially equivalent (SE) |
{1}------------------------------------------------
#### Indications for Use
The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
The Indications for Use are unchanged.
#### Device Description
The AutoPulse Resuscitation System Model 100 is an automated, portable, batterv powered device that compresses the chest of an adult human as an adjunct to manual CPR. A brief description of the device is provided below.
The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. The AutoPulse Platform contains the mechanical drive mechanism, control system, software and electronics necessary to generate and control the force required to perform mechanical chest compressions. User controls and indicators are contained in the User Control Panel. The AutoPulse Platform and LifeBand are unchanged. A new battery and battery charger are the subject devices of this 510(k).
#### Principles of Operation
The principle of operation remains unchanged. The AutoPulse Resuscitation System Model 100 is an electrically powered external cardiac compressor that compresses the chest in the region of the heart per AHA recommended protocol to provide blood flow during cardiac arrest. This principle of operation of the modified AutoPulse System that is the subject device of this 510(k) is identical to the principal of operation of the currently marketed AutoPulse System.
#### Summary of Testing
The changes included in this 510(k) application include the introduction of new Li-lon battery and a multi-chemistry charger, capable of charging the current NiMH battery and the new Li-lon battery. The charger has the built-in functional testing features to automate battery testing and battery maintenance tasks. There were no changes made to the AutoPulse Model 100 hardware, software or the operating specifications. There were no changes to the LifeBand.
The changes were extensively tested to verify the safety and efficacy of the entire system. First, the system level compatibility was verified to ensure that the AutoPulse powered with the new Li-lon battery meets the operating parameters in a safe and intended manner. The data showed that the new battery chemistry did not alter functioning of the AutoPulse in any way and that it operates identically to the NiMH Battery powered AutoPulse.
Testing also included verification of the compatibility between the new multi-chemistry charger and the current NiMH battery. The bench testing showed that the new charger was capable of safely and consistently charging the NiMH battery, testing the battery and to
{2}------------------------------------------------
correctly identify when the end of life has been reached. The new charger also successfully demonstrated its ability to charge, test, maintain and identify the end of life conditions per the Li-lon battery specifications.
Extensive bench testing was conducted to verify ability of the Li-Ion battery to meet the performance safety specifications. The testing included charging/discharging characteristics of the battery, ability to power the AutoPulse for the specified runtime throughout the specified life of the battery, environmental testing, electrical safety and electromagnetic compatibility. A partial list of international standards applied in the testing includes but is not limited to: applicable parts of IEC 60601-1, IEC 60601-1-2, IEC 60068 series, IEC 61000, CISPR11, IEC 62133 etc.
### Summary of the Basis for Finding of Substantial Equivalence
The proposed AutoPulse Resuscitation System Model 100 is identical to the previously cleared device. The proposed system:
- . has the same intended use;
- . uses the same operating principles;
- incorporates the same device design; .
- . has the same performance specifications;
- incorporates a new battery and battery charger which have . the same operating principles as those of the predicate device;
- . incorporates the same hardware and software functions.
## Rational for Equivalence
There are two (2) minor technological changes to the currently marketed AutoPulse System that are proposed in this 510(k):
- i) a new rechargeable battery pack using Lithium lon (Li-lon) chemistry battery cells; and
- a new battery charger (the AutoPulse Multi-Chemistry Charger) capable of ii) charging both the currently marketed nickel metal hydride (NiMH) battery and the new Li-lon Battery. The addition of the new battery to the AutoPulse System and charger will not alter the function or performance of the AutoPulse System.
There are no changes proposed to the Indications for Use, the AutoPulse Platform, and the single use, non-sterile LifeBand. The instructions for use are updated with directions for using the new battery and battery charger. Bench Testing results were sufficient to assure that the design changes raised no new issues of Safety and Effectiveness. The subject AutoPulse Resuscitation System Model 100 is thus substantially equivalent to its predicate.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized emblem featuring a symbol that resembles an abstract caduceus or a stylized representation of human figures.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR 1 5 2012
Zoll Circulation c/o Mr. Sam Nanavati Vice President, Quality & Regulatory Affairs 650 Almanor Avenue Sunnyvale, CA 94085
Re: K112998
AutoPulse Resuscitation System Model 100 Regulation Number: 21 CFR 870.5200 Regulation Name: External Cardiac Compressor Regulatory Class: Class III Product Code: DRM Dated: February 27, 2012 Received: February 28, 2012
Dear Mr. Nanavati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{4}------------------------------------------------
Page 2 - Mr. Sam Nanavati
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K112998
Device Name: AutoPulse Resuscitation System Model 100
Indications For Use:
The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Cardiovas 510(k) Number
Page 1 of *_*_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.