POWDER FREE LATEX PATIENT EXAM GLOVE WITH PROTEIN LABELING CLAIM (50 MICROGRAMS PER DM SQUARED OR LESS OF WATER SOLUBLE

K112988 · Northstar Healthcare Holdings Limited · LYY · Nov 9, 2011 · General Hospital

Device Facts

Record IDK112988
Device NamePOWDER FREE LATEX PATIENT EXAM GLOVE WITH PROTEIN LABELING CLAIM (50 MICROGRAMS PER DM SQUARED OR LESS OF WATER SOLUBLE
ApplicantNorthstar Healthcare Holdings Limited
Product CodeLYY · General Hospital
Decision DateNov 9, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1
Attributes3rd-Party Reviewed

Indications for Use

The latex examination glove is a disposable device intended for medical purposes to be worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for residual protein levels in accordance with ASTM D5712 Standard and contain 50µg/dm2 or less of water soluble protein.

Device Story

Disposable latex patient examination glove; smooth or textured surface; natural color (off-white). Worn by healthcare examiners on hands or fingers to provide a barrier against contamination between patient and examiner. Manufactured to contain 50 µg/dm² or less of water-soluble protein. Used in clinical or medical settings. Provides protection for both patient and examiner during examinations.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM D3578 (dimensions, physical properties), ASTM D5151 (pinholes), ASTM D6124 (powder residual), ASTM D5712 (protein levels), and ISO 10993-10 (biocompatibility/irritation/sensitization).

Technological Characteristics

Latex material; smooth or textured surface. Complies with ASTM D3578, ASTM D6124, ASTM D5151, and ASTM D5712. Biocompatibility per ISO 10993-10. Protein content ≤ 50 µg/dm². Powder-free.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KII2988 Traditional 510(k) Premarket Notification Powder Free Latex Patient Exam Glove, Textured & Smooth 510(k) Summary Page 1 NOV - 9 2011 ## 510(k) SUMMARY | DATE: | June 30, 2011 | |----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | OWNER: | Northstar Healthcare Holdings<br>Hamilton House<br>10 Queen Street<br>Hamilton, Bermuda HM11 | | OFFICIAL<br>CORRESPONDENT: | Michael Riordan<br>Northstar Healthcare Holdings Limited<br>3300 Cork Airport Business Park, Kinsale Road<br>Cork, Ireland<br>Telephone: 353 21 4548255<br>Fax: 353 21 4548294 | | DEVICE NAME: | Trade Name: Powder Free Latex Patient Exam Glove, Smooth<br>and Textured Natural Color (Off White) with Protein<br>Labeling Claim (50 $\mu$ g/dm² or less of water soluble protein | | | Common Name: Patient Examination Gloves | | | Classification: Patient Examination Gloves | | | Class: Class I | | | Product Code: LYY | ## PREDICATE DEVICE(S): | Predicate<br>510(k) | Device Name | Indication | Clearance<br>Date | Company | |---------------------|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|----------------------------------------------------------------------| | K932521 | Powder Free Latex<br>Examination Gloves | The patient examination glove is a<br>disposable device intended for medical<br>purposes that is worn on the examiner's hand<br>to prevent contamination between patient and<br>examiner. | 13 Dec 1994 | Top Glove<br>SDN.<br>BHD.<br>Selangor<br>Darul<br>Ehsan,<br>Malaysia | | DEVICE<br>DESCRIPTION: | Powder Free Latex Patient Exam Glove, Smooth<br>and Textured Natural Color (Off White) with Protein<br>Labeling Claim (50 µg/dm² or less of water soluble protein) | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------| |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------| {1}------------------------------------------------ K112988 Traditional 510(k) Premarket Notification Powder Free Latex Patient Exam Glove, Textured & Smooth 510(k) Summary Page 2 #### STATEMENT OF INTENDED USE: The latex examination glove is a disposable device intended for medical purposes to be worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for residual protein levels in accordance with ASTM D5712 Standard and contain 50µg/dm2 or less of water soluble protein. #### TECHNOLOGICAL CHARACTERISTICS: The Powder Free Latex Patient Exam, Textured and Smooth with Protein Labeling Claim is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with latex using similar manufacturing processes. ## ASSESSMENT OF NONCLINICAL DATA: | Characteristic | Standard | Device Performance | |-------------------------|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------| | Dimension | ASTM Standard D3578 | Meets | | Physical Properties | ASTM Standard D3578 | Meets | | Freedom from Pinholes | 21 CFR 800.20; ASTM D5151 | Meets | | Powder Residual | ASTM Standard D6124 | Meets<br>Results generated<br>values below 2mg of<br>residual powder | | Biocompatibility | Primary Skin Irritation in rabbits<br>(ISO 10993-10)<br>Dermal Sensitization in the guinea<br>pig (ISO 10993-10) | Gloves are non-<br>irritating<br>Gloves do not display<br>any potential for<br>sensitization | | Residual Protein Levels | ASTM Standard D5712 | Meets | #### CONCLUSIONS: The Powder Free Latex Patient Exam Gloves meet the requirements of established standards ASTM D3578-05, ASTM D6124-06, ASTM D5151-06, ASTM D5712-05e and ISO 10993-10. Based on the comparison of intended use, design, materials and performance, the Powder Free Latex Patient Exam Gloves, Textured and Smooth, With Protein Labeling Claim are substantially equivalent to the predicate device. Northstar Healthcare Holdings {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Northstar Healthcare Holding Limited C/O Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062 NOV - 9 2011 Re: K112988 Trade/Device Name: Powder Free Latex Patient Exam Glove, Smooth and Textured. Natural Color (Off White) with Protein Labeling Claim (50ug/dm2 or less of water soluble protein) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: October 26, 2011 Received: October 27, 2011 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ## Page 2 – Mr. Devine Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Wh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ## Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K112988 Device Name: Powder Free Latex Patient Exam Glove, Smooth and Textured Natural Color (Off White) with Protein Labeling Claim (50 µg/dm² or less of water soluble protein) Indications for Use: The patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Elith F. Clamens Wills Page 1 of (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K112988
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...