CLO-SURPLUS P.A.D.

K112961 · Scion Cardio-Vascular, Inc. · QSY · Oct 31, 2011 · SU

Device Facts

Record IDK112961
Device NameCLO-SURPLUS P.A.D.
ApplicantScion Cardio-Vascular, Inc.
Product CodeQSY · SU
Decision DateOct 31, 2011
DecisionSESE
Submission TypeSpecial
Device ClassClass U
AttributesTherapeutic

Indications for Use

The Scion Cardio-Vascular Clo-SurPUS P.A.D. is intended for the local management of bleeding wounds and to provide a barrier to bacterial penetration of the dressing in all patients and for the promotion of rapid control (hemostasis) of bleeding in patients following hemodialysis and for those on anticoagulation therapy. The dressing is indicated for the following wounds: lacerations, nose bleeds, surgical debridement sites, skin surface puncture sites, vascular sites, and sites involving percutaneous catheters, tubes and pins.

Device Story

Clo-SurPLUS P.A.D. is a sterile, soft, nonwoven topical hemostasis pad; composed of poly-D-glucosamine and poly-N-acetylglucosamine derived from chitosan. Device functions as a physical barrier to bacteria and promotes rapid hemostasis via natural biological properties of chitosan. Used in clinical settings for management of bleeding wounds; optional slit design facilitates placement around percutaneous catheters, tubes, and pins. Provides antibacterial activity for up to 144 hours. Benefits include optimal wound healing environment, biocompatibility, and biodegradability. No active electronic components or software.

Clinical Evidence

No clinical data provided. Evidence relies on established biocompatibility, biodegradability, and hemostatic properties of the chitosan-derived material, supported by historical scientific literature and in-vitro antibacterial testing demonstrating efficacy against various pathogens (e.g., E. coli, S. aureus) for up to 144 hours.

Technological Characteristics

Soft, nonwoven pad composed of poly-D-glucosamine and poly-N-acetylglucosamine (chitosan-derived). Sterile, single-use, E-beam radiation sterilized (10^-6 SAL). Provides antibacterial barrier and hemostatic activity. No energy source, software, or electronic components.

Indications for Use

Indicated for all patients requiring local management of bleeding wounds and bacterial barrier protection. Specifically indicated for hemostasis in patients post-hemodialysis and patients on anticoagulation therapy. Applicable to lacerations, abrasions, nose bleeds, surgical debridement sites, skin surface puncture sites, vascular procedure sites, and sites involving percutaneous catheters, tubes, and pins.

Regulatory Classification

Identification

To temporarily control bleeding and cover external wounds.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized eagle. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Scion Cardio-Vascular, Inc. c/o C2C Development, LLC Craig Pagan 1050 W NASA Boulevard, Suite 136 Melbourne, Florida 32901 July 28, 2023 Re: K112961 Trade/Device Name: Clo-SurPLUS P.A.D. Regulatory Class: Unclassified Product Code: QSY, LYA Dear Craig Pagan: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 31, 2011 and correction letter dated May 1, 2012. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product codes QSY and LYA. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov. Sincerely. Julie A. Morabito -S Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure in profile, with three overlapping faces. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular pattern around the figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Scion Cardio-Vascular, Inc. % C2C Development, LLC Mr. Craig Pagan 1050 W NASA Boulevard, Suite 136 Melbourne, Florida 32901 Re: K112961 Trade/Device Name: Clo-Sur-LUS P.A.D. Regulatory Class: Unclassified Product Code: FRO Dated: September 29, 2011 Received: October 4, 2011 MAY 1 2012 Dear Mr. Pagan: This letter corrects our substantially equivalent letter of October 31, 2011. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {2}------------------------------------------------ Page 2 - Mr. Craig Pagan CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mada M Millan Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## SECTION 5 - INDICATIONS FOR USE STATEMENT 112961 510(k) Number: Clo-Surplus P.A.D. Device Name: ## INDICATIONS: The Scion Cardio-Vascular Clo-SurPUS P.A.D. is intended for the local management of bleeding wounds and to provide a barrier to bacterial penetration of the dressing in all patients and for the promotion of rapid control (hemostasis) of bleeding in patients following hemodialysis and for those on anticoagulation therapy. The dressing is indicated for the following wounds: lacerations, nose bleeds, surgical debridement sites, skin surface puncture sites, vascular sites, and sites involving percutaneous catheters, tubes and pins. Prescription Use _ > (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Danill Ksomefu MXM (Division Sign Off) (Division Sign-Division of Surgical, Orthopedic. and Restorative Devices 510(k) Number K112961 Page 1 of 1 Scion Cardio-Vascular, Inc. Special 510(k) Submission - Modified Clo-Sur PLUS PAD Page 18 of 47 {4}------------------------------------------------ K112961 page 1 of 2 OCT 3 1 2011 ## SECTION 6 - 510(k) SUMMARY Submitted by: Scion Cardio-Vascular, Inc. 14256 S.W. 119th Avenue Miami, FL 33186 Phone: (305) 259-8880 Fax: (305) 259-8878 Contact Person: Lou Rose 14256 S.W. 119th Avenue Miami, FL 33186 Phone: (305) 259-8880 Ext 103 Fax: (305) 259-8878 Date Prepared: September 14, 2011 Clo-Surpl.us PAD™ Proprietary Name: Common Name: Classification: Classification Name: Predicate Device: Device Description: Intended Use: Topical Hemostasis Pad Unclassified Topical Wound Dressing Pad Scion Cardio-Vascular, Inc, K092552, CLO-SURPLUS P.A.D., The Scion Cardio-Vascular Clo-SurPLUS PADTM is a soft, nonwoven topical pad that provides an optimal wound healing environment, combining an effective antibacterial barrier activity with exudates management. An optional slit Clo-SurfLUS PAD™ allows for easier placement of the dressing around pins and tubes. Clo-Sur PLUS PADTM has demonstrated in-vitro antibacterial activity for up to 144 hours (6 days) in certain strains shown to be detrimental to wound healing such as: Escherichia Coli, Staphylococcus Aureus, Streptococcus pyogenes, Pseudomonas aeruginosa, Bacillus cereus, Enterococcus faecium, Candida Albicans and Asperigillus brasiliensis. Clo-SurPLUS PAD™ is a sterile topical hemostasis pad, packed in a foil pouch and sterilized by E-beam radiation to a 10°6 SAL. The Scion Cardio-Vascular Clo-Sur-Lus P.A.D. is intended for the local management of bleeding wounds and to provide a barrier to bacterial penetration of the dressing in all patients and for the promotion of rapid control (hemostasis) of bleeding {5}------------------------------------------------ K112961 page 2 of 2 in patients following hemodialysis and for those on anticoagulation therapy. The dressing is indicated for the following wounds: lacerations, abrasions, nose bleeds, surgical debridement sites, skin surface puncture sites, vascular procedure sites, and sites involving percutaneous catheters, tubes and pins. Technological Characteristics The Scion Cardio-Vascular Clo-SurPLUS PADTM is a soft, nonwoven pad made of a proprietary formulation of poly-Dglucosamine and poly-N-acetylglucosamine derived from chitosan. The natural biological properties of this material gives the Clo-Sur-LUS PADTM an advantage as an effective bacterial barrier while providing for an optimal wound healing environment. Several biomedical applications of poly-D-glucosamine and poly-N-acetyglucosamine have been reported. The studies represent research on the safety and use of these materials, which have been published over a period of decades by scientists from around the world. The scientific literature satisfies the requirement that a general recognition of safety requires common knowledge about the substance throughout the scientific community. This formulation has many useful and advantageous properties in their application as a wound dressing, namely biocompatibility, biodegradability, hemostatic activity, anti-infectional activity. The technological characteristics of the modified Clo-SurPLus PAD™ are the same as the predicate device. The Scion Cardio-Vascular modified Clo-SuPLUS PAD™ works in the same manner as the approved predicate device. Scion Cardio-Vascular, Inc. Special 510(k) Submission - Modified Clo-Sur PLUS PAD
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...