← Product Code [CAH](/productcode/CAH) · K112958

# HUMID-VENT HEPA (K112958)

_Teleflex Medical · CAH · Sep 21, 2012 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K112958

## Device Facts

- **Applicant:** Teleflex Medical
- **Product Code:** [CAH](/productcode/CAH.md)
- **Decision Date:** Sep 21, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5260
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

The Humid-Vent HEPA Filters are used for humidification and bacterial/viral filtration during anesthesia and ventilator care.

## Device Story

Humid-Vent HEPA is a combined heat and moisture exchanger (HME) and mechanical bacterial/viral filter. Device incorporates a hydrophobic glass fibre filter and HME media made of corrugated hygroscopic paper within a housing. Features ISO standard 15M/22F conical connectors at both machine and patient ends. Used in clinical settings during anesthesia and ventilator care to provide humidification and filtration of breathing gases. Operates passively via patient respiration; no external power or active processing required. Benefits patient by maintaining airway moisture and reducing risk of cross-contamination during mechanical ventilation.

## Clinical Evidence

Bench testing only. The device underwent a full battery of performance testing against pre-determined acceptance criteria. Results verified that the device meets all specifications and is substantially equivalent to predicate devices.

## Technological Characteristics

Combined HME and mechanical bacterial/viral filter. Materials: hydrophobic glass fibre filter, corrugated hygroscopic paper HME media. Connectors: ISO standard 15M/22F conical. Passive operation; no energy source. Standalone device.

## Regulatory Identification

A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.

## Predicate Devices

- Gibeck Iso-Gard Angled and Straight Filters
- Gibeck Humid-Vent 2 HME
- Hygroster Filter / HME combination bacterial / viral filter (HEPA)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP

## 112 112 958 510(K) SUMMARY

#### 1. SUBMITTER:

TELEFLEX MEDICAL INMED MFG.SDN.BHD, RUSCH P.O. Box 28, Kamunting Industrial Estate, Kamunting Perak, Malaysia 34600 Telephone# 605-891-5111 Establishment Registration Number: 8040412

Official contact: Telephone: Date Prepared:

Elizabeth Paul, Director, International/Domestic RA 919-433-8076 September 23, 2011

#### 2. DEVICE:

Tradename: Humid-Vent® HEPA Classification Name: Filter, Bacterial, Breathing-Circuit Classification: Class II Common Name: Breathing circuit bacterial filter Classification Panel: Anesthesiology Product Code: CAH Regulation Number: 868.5260

#### 3. PREDICATE DEVICES:

- . Teleflex Medical's Gibeck Iso-Gard Angled and Straight Filters,
- . Teleflex Medical's Gibeck Humid-Vent 2 HME, and
- Covidien's Hygroster Filter / HME combination bacterial / viral filter (HEPA) . and heat / moisture exchanger (HME)

#### 4. DEVICE DESCRIPTION:

The Humid-Vent® HEPA is a combined Heat and Moisture Exchanger (HME) and a mechanical Bacterial/Viral Filter. The device is used for humidification and bacterial/viral filtration during anesthesia and ventilator care. The device consists of a housing which incorporates a hydrophobic glass fibre filter and an HME media made of corrugated hygroscopic paper. It contains ISO Standard 15M / 22F conical connectors at the machine end as well as the patient end.

{1}------------------------------------------------

#### 5. INDICATIONS FOR USE:

The Humid-Vent HEPA Filters are used for humidification and bacterial/viral filtration during anesthesia and ventilator care.

### COMPARISON OF CHARACTERISTICS: ર્દ્ર.

Comparisons of the proposed and predicate devices show that the technological characteristics such as materials, performance characteristics and packaging are identical or substantially equivalent to the currently marketed predicate devices.

#### 7. PERFORMANCE DATA:

The Humid-Vent® HEPA was subjected to a full battery of performance testing including pre-determined acceptance criteria. As expected, the device met all acceptance criteria. The performance data verified that the Humid-Vent® HEPA is substantially equivalent to the currently marketed predicate devices, adequately meets its intended use, and is acceptable for commercial distribution.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is simple and recognizable, representing the department's role in promoting the health and well-being of the nation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Teleflex Medical Ms. Elizabeth Paul Director, International/Domestic Regulatory Affairs 2917 Weck Drive Research Triangle Park, North Carolina 27709

SEP 21 2012

Re: K112958

Trade/Device Name: Humid-Vent® HEPA Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: II Product Code: CAH Dated: September 18, 2012 Received: September 19, 2012

Dear Ms. Paul:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

# Page 2- Ms. Paul

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Ph for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K112958

Device Name: Gibeck Humid-Vent® HEPA

Indications for Use:

The Humid-Vent HEPA Filters are used for humidification and bacterial/viral filtration during anesthesia and ventilator care.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Y Schulto

1

Page of

(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices

**510(k) Number:**

---

**Source:** [https://fda-staging.innolitics.com/device/K112958](https://fda-staging.innolitics.com/device/K112958)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com