LEGION NARROW OXINIUM CRUCIATE RETAINING AND POSTERIOR STABILIZED FENORAL COMPONENTS
K112941 · Smith & Nephew, Inc. · JWH · Dec 20, 2011 · Orthopedic
Device Facts
Record ID
K112941
Device Name
LEGION NARROW OXINIUM CRUCIATE RETAINING AND POSTERIOR STABILIZED FENORAL COMPONENTS
Applicant
Smith & Nephew, Inc.
Product Code
JWH · Orthopedic
Decision Date
Dec 20, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact. Smith & Nephew, Inc. Legion Narrow Oxinium CR and PS Femoral Components are indicated for single use only and for use only with bone cement.
Device Story
Legion Narrow Oxinium CR and PS Femoral Components are orthopedic implants for total knee replacement. Manufactured from oxidized zirconium alloy (Oxinium). Designed for primary or revision surgery to restore joint function in patients with arthritis or failed prior knee procedures. Implants are used by orthopedic surgeons in a clinical/hospital setting. The device replaces the femoral surface of the knee joint, articulating with a polyethylene insert and metal tibial base. The surgeon selects the appropriate size and configuration (CR or PS) based on patient anatomy and ligamentous stability. The device is fixed using bone cement. Benefits include restoration of knee joint mechanics and reduction of pain associated with degenerative or inflammatory joint disease.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including tibiofemoral constraint testing (A/P draw, M/L shear, and rotary laxity), tibiofemoral contact area testing, and patellofemoral subluxation resistance testing compared to predicate devices.
Technological Characteristics
Implantable femoral components manufactured from oxidized zirconium alloy (Oxinium). Designed as semi-constrained, cemented, patellofemoral-tibial prostheses. Available in multiple sizes and configurations (CR and PS). Mechanical design features include specific geometry for narrow anatomy. No software or electronic components.
Indications for Use
Indicated for patients requiring primary or revision total knee arthroplasty due to rheumatoid, post-traumatic, or degenerative arthritis, or failed previous procedures (osteotomy, unicompartmental, or total knee replacement). Posterior stabilized systems indicated when anterior/posterior cruciate ligaments are incompetent and collateral ligaments are intact. For single use only; must be used with bone cement.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness Legion Narrow Oxinium CR and PS Femoral Components and Device Specific Instruments
| Submitted By: | Smith & Nephew, Inc.<br>Orthopaedics<br>1450 Brooks Road<br>Memphis, TN 38116 |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Date: | October 3, 2011 |
| Contact Person: | David Henley, Regulatory Affairs Project Manager<br>Tel: (901) 399-6487 Fax: (901) 566-7079 |
| Proprietary Name: | Legion Narrow Oxinium CR and PS Femoral<br>Components and Instruments |
| Common Name: | Total Knee Prosthesis |
| Classification Name and Reference: | 21 CFR 888.3560, Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis |
| Device Class | Class II |
| Device Product Code and Panel Code: | JWH / Orthopedics / 87 |
### Device Description:
The implant devices described in this premarket notification are comprised of Legion Narrow Oxinium Cruciate Retaining (CR) and Posterior Stabilized (PS) Femoral Components. All described implant components are manufactured from oxidized zirconium alloy material (trade name: Oxinium). The subject devices are available in the following configurations and sizes:
| Device Type | Hand | Available Sizes |
|---------------------------------------------|-------|-----------------|
| Legion Narrow Oxinium CR Femoral Components | Left | 3, 4, 5 and 6 |
| Legion Narrow Oxinium CR Femoral Components | Right | 3, 4, 5 and 6 |
| Legion Narrow Oxinium PS Femoral Components | Left | 3, 4, 5 and 6 |
| Legion Narrow Oxinium PS Femoral Components | Right | 3, 4, 5 and 6 |
Device specific instruments are also described in this premarket notification in select sections and exhibits.
#### Intended Use:
Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
Smith & Nephew, Inc. Legion Narrow Oxinium CR and PS Femoral Components are indicated for single use only and for use only with bone cement.
{1}------------------------------------------------
#### Technological Characteristics:
Components comprising Legion Narrow Oxinium CR and PS Femoral Components are verv similar to the legally marketed devices listed below. When compared to the predicates, the proposed devices share very similar or identical Indications for Use, intended use, are manufactured from similar or identical materials, and incorporate very similar technological design characteristics.
#### Substantial Equivalence Information:
When compared to the implant and device specific instrument predicate devices listed below. substantial equivalence is based on similarities with regard to overall indications for use, material composition, and technological design characteristics.
- · Genesis II Total Knee System K951987
- · Genesis II and Profix Zirconium Femoral Components K962557
- · Legion Primary Knee System K093746
### Preclinical Testing:
To further support a determination of substantial equivalence, various types of pre-clinical testing were conducted on the subject, implantable devices in comparison against one or more of the previously cleared predicate devices described above. The specific types of pre-clinical testing included:
- · Tibiofemoral Constraint Testing of femorals against interfacing polyethylene articular insert components seated in appropriate metal tibial bases in A/P Draw, M/L Shear and rotary laxity (R/L)
- · Tibiofemoral Contact Area Testing of femorals against interfacing polyethylene articular insert components seated in appropriate metal tibial bases
- Patellofemoral Subluxation Testing of the femoral to determine the amount of resistance to lateral subluxation of the patellofemoral interface
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 2 0 2011
Smith & Nephew, Inc. Orthopaedics Division % Mr. David Henley Regulatory Affairs Project Manager 7135 Goodlet Farms Parkway Cordova, Tennessee 38016
Re: K112941
Trade/Device Name: Legion Narrow Oxinium Cruciate Retaining (CR) and Posterior Stabilized (PS) Femoral Components and Device Specific Instruments Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: October 3, 2011 Received: October 4, 2011
Dear Mr. Henley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. David Henley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Erine Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Premarket Notification
# Indications for Use Statement
510(k) Number (if known): _ K || 29 H |
Device Name: Legion Narrow Oxinium C/R and P/S Femoral Components
Indications for Use:
Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
Smith & Nephew, Inc. Legion Narrow Oxinium C/R and P/S Femoral Components are indicated for use only with cement and are single use only devices.
x Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (Part 21 CFR 807 Subpart C)
(Please do not WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Mark Melkerson
(Division Sign-Off) (Division Sign-On), Orthopedic, Division of Restorative Devices
K112941
510(k) Number .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.