K112875 · Blue Sky Bio, LLC · DZE · Jan 27, 2012 · Dental
Device Facts
Record ID
K112875
Device Name
BLUE SKY BIO DENTAL TAPER HEX IMPLANT SYSTEMS
Applicant
Blue Sky Bio, LLC
Product Code
DZE · Dental
Decision Date
Jan 27, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Blue Sky Bio 3mm Double Hex implants are primarily intended for restoration of mandibular incisors and maxillary lateral incisors Blue Sky Bio 3mm Double Hex Dental Implant System is intended for implantation into the fully and partially edentulous ridge for the support of a dental prosthesis, and for single tooth replacement in the anterior area of the jaws Implants can be placed in immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. Multi-unit abutments for fixed restorations are to be used only for multiple splinted restorations. In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases. Blue Sky Bio 3mm Double Hex Dental Implant System is compatible with Astra Osseospeed Narrow Dental Implant System.
Device Story
Root form dental implant system; includes implants, abutments, and surgical/restorative instruments. Used by clinicians to restore chewing function in edentulous or partially edentulous patients. Implants placed into jaw bone; support cement-retained, screw-retained, or overdenture prostheses. Features internal double hex with taper for anti-rotation. Compatible with Astra Osseospeed Narrow system. Provides restorative options for single or multiple tooth applications. Immediate function possible with sufficient primary stability and appropriate occlusal loading.
Clinical Evidence
Bench testing only. Fatigue testing conducted per ISO 14801; compatibility testing performed to assess fit with predicate systems. No clinical data provided.
Technological Characteristics
Materials: CP Titanium Grade 4, Ti-6Al-4V. Design: 2-stage root form, 3mm diameter, internal double hex with taper. Surface: Blasted with resorbable medium or Aluminum Oxide and acid etched. Sterilization: Gamma. Connectivity: N/A (mechanical device).
Indications for Use
Indicated for fully or partially edentulous patients requiring dental prosthesis support or single tooth replacement in the anterior jaw (specifically mandibular incisors and maxillary lateral incisors). Contraindicated for immediate loading in fully edentulous patients unless 4 or more implants are used.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Blue Sky Bio, LLC
JAN 2 7 2012
888 E BELVIDERE SUITE 212 GRAYSLAKE, IL 60030 Tel: 718-376 0422 Fax: 888-234 3685 Email: azickmann@blueskybio.com
### 510(K) Summary
General Information
| Classification Name: | Endosseous Implant | |
|----------------------|----------------------------------------------|--|
| Common Name: | Prosthetic Dental Implant System | |
| Trade Name: | Blue Sky Bio Dental Taper Hex Implant System | |
| Submitter's Name: | Blue Sky Bio, LLC | |
| Address: | 888 E Belvidere Rd., Suite 212 | |
| | Grayslake, IL 60030 | |
| Telephone: | 847-548 8499 | |
| Fax: | 847-548 8491 | |
| Contact: | Michele Vovolka | |
| Date of Summary | June 2011 | |
#### Device Description
The modification of the Blue Sky Bio Dental Implant System consists of root form dental implants of various lengths and associated abutment systems, which provide the clinician with cement retained , screw retained and overdenture-type restorative options. Modifications to the existing system do not introduce new issues of safety or efficacy. The implants and components are supplied sterile or not sterile and are labeled accordingly.
#### Intended Use .
- Blue Sky Bio 3mm Double Hex implants are primarily intended for restoration of mandibular . incisors and maxillary lateral incisors
- Blue Sky Bio 3mm Double Hex Dental Implant System is intended for implantation into the . fully and partially edentulous ridge for the support of a dental prosthesis, and for single tooth replacement in the anterior area of the jaws
- Implants can be placed in immediate function on single-tooth and/or multiple tooth applica-. tions when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function.
- Multi-unit abutments for fixed restorations are to be used only for multiple splinted restora-. tions.
- In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded . cases.
- Blue Sky Bio 3mm Double Hex Dental Implant System is compatible with Astra Osseospeed . Narrow Dental Implant System.
Blue Sky Bio, LLC 510(k)
ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴ
Page 9
August 2011
{1}------------------------------------------------
... ..........................................................................................................................................................................
.
# Technological Characteristic Comparison
| Feature | Subject Device<br>Blue Sky Bio 3mm Dental<br>Implant System (Double Hex<br>3mm) | Predicate Device<br>Astra Tech 3mm Dental<br>Implant System K080396. |
|----------------------------------------------|---------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Material | CP Titanium Grade 4,<br>Ti-6Al-4V | CP Titanium<br>Grade 4 |
| 1 Stage/ 2 Stage | 2 Stage | 2 Stage |
| Surface | Blasted with resorbable<br>medium, or Aluminum Oxide<br>and acid etched | Blasted with titanium oxide and acid<br>etched |
| Body Diameter<br>(mm) | 3 mm | 3mm |
| Platform<br>Diameter<br>(mm) | 3mm | 3mm |
| Lengths (mm) | 10, 11, 13, 15, 17 mm | 11, 13, 15, 17 mm |
| External Screw<br>Threads | Yes | Yes |
| Anti-rotational<br>Feature | Internal Double Hex with taper, | Internal Double Hex with taper |
| Gamma Sterilized | Yes | Yes |
| Solid abutment | Yes | Yes |
| Two-Piece<br>Screwed<br>Abutment | Yes | Yes |
| Overdenture<br>abutment | Yes | Yes |
| Instruments<br>(surgical and<br>restorative) | Yes | Yes |
| Angled Abutment | Yes Multiunit | Yes TiDesign |
| Single Unit | Yes | Yes |
| Multiple Unit | Yes | Yes |
Blue Sky Bio, LLC 510(k)
August 2011
: |
{2}------------------------------------------------
## Safety and Efficacy
The material, technology and facilities used to produce the modified Blue Sky Bio Dental Implant Systems are the same. Therefore it is substantially equivalent to other commercially available Dental Implant Systems including predicate devices Osseospeed Narrow Dental Implant System Astratech (K080396) and to Blue Sky Bio Dental Implant System (K073713)
The technical comparison charts list the primary technical aspects and specifications that are pertinent to Dental Implant Systems. The 3mm Blue Sky Bio dental implant system is as safe and effective as the predicate devices.
#### Non-Pyrogenic Statement
No claims are made that the products are non-pyrogenic.
#### Performance Tests
Compatibility tests with other systems according to Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document; Root-form Endosseous Dental Implants and Endosseous Dental Abutments: These tests were performed to assess compatibility with predicate devices. The tests showed that the new devices are compatible with predicate devices and the fit is adequate.
Fatigue testing for angled abutments and narrow diameter implants: This test has been conducted according to ISO 14801 for predicate devices. The new devices have the same wall thickness and equal or smaller angulation than the predicate devices and are therefore equivalent or stronger than the predicate devices.
#### Conclusion
The Blue Sky Bio Dental Implant system, subject to this submission and the predicate devices are believed to be substantially equivalent. The device constitutes a safe, reliable and effective medical device. meeting all declared requirements of its intended use and the device does not introduce new risks and does not present any adverse health effects or safety risks to patients when used as intended.
Blue Sky Bio, LLC 510(k)
August 2011
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 2 7 2012
Ms. Michele Vovolka Correspondent Blue Sky Bio, LLC 888 E Belvidere Road, Suite 212 Grayslake, Illinois 60030
Re: K112875
Trade/Device Name: Blue Sky Bio Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: January 8, 2012 Received: January 19, 2012
Dear Ms. Vovolka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Vovolka
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
G. ler. Anthony Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Blue Sky Bio, LLC
888 E BELVIDERE SUITE 212 GRAYSLAKE, IL 60030 Tel: 718-376 0422 Fax: 888-234 3685 Email: azickmann@blueskybio.com
#### Indications for Use Statement
Page _1_ of _1_
510(k) Number (if Known): K112875
Device Name: Blue Sky Bio Dental Implant System
Indications for Use:
- Blue Sky Bio 3mm Double Hex implants are primarily intended for restoration of mandibular . incisors and maxillary lateral incisors
- more Sky Bio 3mm Double Hex Dental Implant System is intended for implantation into the . fully and partially edentulous ridge for the support of a dental prosthesis, and for single tooth replacement in the anterior area of the jaws
- Implants can be placed in immediate function on single-tooth and/or multiple tooth applica-. tions when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function.
- enowing raileriem.
The abutments for fixed restorations are to be used only for multiple splinted restora-. tions.
- In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded ◆ cases.
- Blue Sky Bio 3mm Double Hex Dental Implants are compatible with Astra Osseospeed . Narrow abutments.
- Rurow abutinents!
Blue Sky Bio 3mm Double Hex Abutments are compatible with Astra Osseospeed Narrow . Implants.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801:109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Division of Anesthesiologie Befleral Hospital
Page 14
Proprietary & Confidentianfection Control, Dental Devices
510(k) Number:
(Division Sigh-Off)
Blue Sky Bio, LLC 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.