EASYPROPHY SERIES AIR POLISH DEVICES, MODELS EASYPROPHY 200 AND EASYPROPHY 400
K112869 · Beyes Dental Canada, Inc. · EFB · Dec 16, 2011 · Dental
Device Facts
Record ID
K112869
Device Name
EASYPROPHY SERIES AIR POLISH DEVICES, MODELS EASYPROPHY 200 AND EASYPROPHY 400
Applicant
Beyes Dental Canada, Inc.
Product Code
EFB · Dental
Decision Date
Dec 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The EasyProphy Series Air Polish Devices are intended for use in dental applications to remove stains and plaque deposits from the teeth by shooting a mixture of sodium bicarbonate powers, air, and water onto tooth surfaces.
Device Story
Hand-held dental air-polishing device; connects to standard dental unit turbine tube for air/water supply. Features capped powder chamber and nozzle. Operation: device mixes air, water, and sodium bicarbonate powder into a spray; clinician directs spray onto tooth surfaces to remove stains and plaque. Used in dental clinics by dental professionals. Output is a mechanical cleaning spray; aids in dental hygiene and prophylaxis.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Hand-held air-powered dental handpiece. Materials: Type 304 Stainless Steel for patient-contacting and waterline components. Energy source: pneumatic (air) and hydraulic (water) from dental unit. Connectivity: standard turbine tube connection. No software or electronic components.
Indications for Use
Indicated for dental patients requiring removal of stains and plaque deposits from tooth surfaces via air-polishing.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a series of small, dark, irregular shapes scattered across a white background. The shapes vary in size and are distributed unevenly, with some clustered together and others more isolated. The overall impression is one of randomness and contrast between the dark shapes and the bright background.
K112869
## 510(k) Summarv
#### Summary Prepared Date: 12/14/2011
## Submission Sponsor:
Beyes Dental Canada Inc. 595 Middlefield Road Unit 9, Toronto, M1V 3S2 Canada
Mr. Ted Thompson Toll Free: 1-877-703-3562 info@beves.ca
## Submission Correspondent:
Mr. Anthony Hopkins Regulatory Affairs Specialist
Mr. Leon Lu Director of Quality and Requlatory Affairs
#### MEDevice Services, LLC
3500 South Dupont Highway Dover, Delaware 19901 USA Phone: 1-877-202-1588 Fax: 1-888-202-8884 Email: medeviceservices@gmail.com www.medeviceservices.com
Trade/Device Name: EasyProphy Series Air Polish Devices, Models EasyProphy 200 and EasyProphy 400
Common or Usual Name: handpiece, air-powered, dental Device Class: I Classification Name: handpiece, air-powered, dental Requlation Number: 21 CFR 872.4200 Product Code: EFB Review Panel: Dental
{1}------------------------------------------------
#### Predicate Device:
- K032395, Nakanishi, Incorporated . Prophy-Mate
- K092289, Ems Electro Medical Systems Sa ● Ems Air-Flow Handy Perio
## Device Description:
This device is a hand-held device that contains air and water lines, a capped chamber for the cleaning powder, and a nozzle. This device connects to a standard turbine tube that supplies air and water. When the Air Polisher is connected and activated, a stream comprised of powder, air, and water sprav is generated. This spray can be directed on the tooth surface to clean and polish the tooth.
## Intended Use:
The EasyProphy Series Air Polish Devices are intended for use in dental applications to remove stains and plaque deposits from the teeth by shooting a mixture of sodium bicarbonate powers, air, and water onto tooth surfaces.
## Comparison to Predicate Devices:
The EasyProphy Series Air Polish Devices and the predicate device are used for cleaning and polishing of teeth. Both the proposed device and the predicate device connect to a standard turbine connection on a dental operative unit. Both deliver a mixture of water, air, and dental powder to a treatment site.
The EasyProphy Series Air Polish Devices is essentially the same as or similar to the predicate device in terms of the intended use, design and construction. performance characteristics, and materials. The patient contact materials used are well-known biocompatible, so no new issues of biocompatibility are raised with regard to this device.
## Discussion of Non-Clinical Tests Performed:
Testing conducted demonstrates that the easyProphy Series Air Polish Devices fulfills the prospectively defined performance specifications.
{2}------------------------------------------------
K112869
Therefore, we conclude that the EasyProphy Series Air Polish Devices is both safe and effective for its intended use.
## Biocompatibility:
The patient-contacting portions and waterline composition of the EasyProphy Series Air Polish Devices, Models EasyProphy 200 and EasyProphy 400 are made of type 304 Stainless Steel. Biocompatibility testing on the type 304 Stainless Steel was not performed as it is commonly used for implant device and other medical devices, and has a history of biocompatibility.
#### Discussion of Clinical Tests Performed:
None
#### Conclusion:
The EasyProphy Series Air Polish Devices as safe and effective as the predicate devices. The proposed device has the same intended uses and indications. similar technological characteristics, and principles of operation as its predicate device. The minor differences between the proposed device and its predicate devices raise no new issues of safety or effectiveness. Thus, the EasyProphy Series Air Polish Devices, models EasyProphy 200 and EasyProphy 400 are substantially equivalent to its predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake entwined around it, and three horizontal bars above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES -
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Beyes Dental Canada Incorporated C/O Mr. Anthony Hopkins Regulatory Affairs Specialist MEDevice Services, LLC 3500 South Dupont Highway Dover. Delaware 19901
DEC 1 6 2011
Re: K112869
Trade/Device Name: EasyProphy Series Air Polish Devices, Models EasyProphy 200 and EasyProphy 400 Regulation Number: 21 CFR 872.4200
Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: September 25, 2011 Received: September 30, 2011
Dear Mr. Hopkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Hopkins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety /ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
K112869
# Indications for Use
#### 510(k) Number (if known):
Device Name:
EasyProphy Series Air Polish Devices, Models EasyProphy 200 and EasyProphy 400
#### Indications for Use:
The EasyProphy Series Air Polish Devices are intended for use in dental applications to remove stains and plaque deposits from the teeth by shooting a mixture of sodium bicarbonate powers, air, and water onto tooth surfaces.
Prescription Use × Over-The-Counter Use J AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
C
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112869
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.