ENTERAL ADMINISTRATION SET WITH SOLUTION CONTAINER, ENTERAL ADMINISTRATION SET, ENTERAL EXTENSION SET WITH MINITUBE CONN
Applicant
Acta Medical, LLC
Product Code
KNT · Gastroenterology, Urology
Decision Date
Mar 29, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To deliver liquid nutrition formula from storage container to an enteral access device (feeding tube)
Device Story
Acta Enteral Sets are medical-grade, non-DEHP PVC tubing systems designed for enteral nutrition delivery. Configurations include complete solution delivery systems with solution containers, enteral administration sets with spikes for external containers, and enteral extension sets. Devices feature various proximal and distal connectors, including Christmas tree connectors for gastric tube attachment. Used in clinical settings to facilitate the transfer of liquid nutrition from storage to a patient's feeding tube. Operation is manual, performed by healthcare providers. The device functions as a passive conduit for nutrition delivery; it does not incorporate active electronic components or software.
Clinical Evidence
Bench testing only. The device underwent biocompatibility testing per ISO 10993-1, including assessments for DEHP, extractables, and leachables. Performance testing was conducted per ISO 8536-4, specifically evaluating leak resistance and tensile strength.
Technological Characteristics
Constructed from non-DEHP PVC, silicone, and ABS. Features various connector types including Christmas tree connectors. Passive fluid delivery system. Tested for biocompatibility (ISO 10993-1) and performance (ISO 8536-4).
Indications for Use
Indicated for the delivery of liquid nutrition formula from a storage container to an enteral access device (feeding tube).
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Promedic Pediatric Enteral feeding tubes and Extension sets (K092628)
Dynarex Enteral Sets For Gravity and Pump Use (K082863)
Truecare Biomedix Intravascular Administration Set (K111351)
Submission Summary (Full Text)
{0}------------------------------------------------
K112863 Page 1 of 3
MAR 29 2012
# Acta Medical, LLC
Date Submitted: Submitted By:
September 28th, 2011 Acta Medical, LLC Inc 15 Minneakoning Road, Suite 203 Tel: 908-399-0279 Fax:908-788-9788
Submitter Contact:
Aaron Compton Acta Medical, LLC Contact email: acatutd@satx.rr.com Tel:214-533-1416 Fax:972-638-8549
Common Name of Device: Enteral Set
Predicate Device:
Promedic Pediatric Enteral feeding tubes and Extension sets (K092628) Dynarex Enteral Sets For Gravity and Pump Use(K082863) Truecare Biomedix Intravascular Administration Set (K111351)
Tubes, Gastrointestinal (and accessories)
Panel:
Product Code:
,
KNT
Device Classification: Class II
Classification Reference: 21CFR 876.5980
Appropriate Classification Panel: Gastroenterology/Urology
Manufacturing Location: Yangzhou Wei De Li Trade Co. Ltd.
> Li Xin Bridge Touqiao Township. Yangzhou City, Jiangsu Province. China Tel: +86-514-87897887 Fax: +86-514-87889967
{1}------------------------------------------------
K11 2063 Page 2/3
### Name & Model Numbers of Devices:
- 1. Enteral Administration Set with Solution Container, 1200ml capacity (VDENT001)
- 2. Enteral Administration Set (VDENTS001)
- 3. Enteral Extension Set with Minitube Connector(VDENT8-1222)
### Device Description:
Acta Enteral Sets are constructed from flexible, medical grade, Non-DEHP tubing and have various different connectors at each end of the tubing. One configuration may include an Enteral solution container constructed from same PVC material as the tubing with a Christmas tree connector to connect to a gastric tube (or an extension set) or another configuration may simply be an extension set which connects to patient gastric tube to the nutrition liquid delivery system. The following three configurations are most common, however, other configurations may be manufactured based on customer demand. (Further details, photos, drawings provided in Device Description, Section XI)
- 1. A complete solution delivery system: This consists of, a solution container, integrated delivery set and a Christmas tree connector to connect to the gastric tube or to an extension set
- 2. Enteral Administration Set: This consists of a delivery set with a spike to connect to a nutrition solution storage container and a Christmas tree connector to connect to gastric tube or an extension set
- 3. Enteral Extension Set: This consists of flexible PVC tubing set with proximal connector which can accept a standard Christmas tree connector and a distal connector designed to fit a custom gastric tube
#### Device Intended Use: Enteral Set
To deliver liquid nutrition formula from storage container to an enteral access device (feeding tube)
#### Substantial Equivalence:
Acta Medical, LLC Enteral Set is substantially equivalent to predicate devices as follows:
{2}------------------------------------------------
K112863 Page 3/3
A complete solution delivery system and Enteral Administration Set is substantially equivalent to predicate device Dynarex Enteral Set for Gravity and Pump Use (K082863) in its design, intended use, function, materials, performance
Enteral Extension Set is substantially equivalent to Promedic, Pediatric Enteral Feeding tubes and Extension sets (K092628) in its design, intended use, function, materials, performance
Enteral Set is substantially equivalent to Truecare Biomedix Intravascular administration set (K111351) in its materials and manufacturer. Enteral Sets are constructed from same materials (non-DEHP PVC, Silicone, ABS) and manufactured by same manufacturer (Yangzhou Wei De Li) at the same manufacturing location as K111351,
Acta Medical, LLC Enteral Sets construction materials have been tested for biocompatibility, DEHP, extractables and leachables, physico chemical tests for plastics. Furthermore, the manufactured sets have been tested per ISO 8536-4 for leak test, tensile strength. Acta Medical, LLC Enteral Sets meets all the requirements for biocompatibility (ISO 10993-1) and applicable performance criteria (ISO 8536-4)
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes on its wing, symbolizing health, services, and human needs. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Aaron Compton Acta Medical L.L.C. 929 Arbor Downs Drive PLANO TX 75023
MAR 2 9 2012
Re: K112863
Trade/Device Name: Enteral Set Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: March 22, 2012 Received: March 22, 2012
Dear Mr. Compton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerély wours.
Benjamin K. Eviko
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K112863
Device Name: Enteral Set
Indications For Use: To deliver liquid nutrition formula from storage container to an enteral access device (feeding tube)
Prescription Use _ x Use (Part 21 CFR 801 Subpart D)
・
AND/OR
Over-The-Counter
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Reproductive, Gastro-Renal, and
Urological Devices
| 510(k) Number | K112863 |
|---------------|---------|
| Page | 1 of 1 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.