K112839 · Varian Medical Systems · IYE · Nov 14, 2011 · Radiology
Device Facts
Record ID
K112839
Device Name
VARIAN HIGH ENERGY LINEAR ACCELERATOR
Applicant
Varian Medical Systems
Product Code
IYE · Radiology
Decision Date
Nov 14, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Varian High Energy Linear Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Device Story
Varian High Energy Linear Accelerator delivers high-energy radiation for stereotactic radiosurgery and precision radiotherapy. Device utilizes photon treatment modes, including Flattening Filter Free (FFF) high-intensity mode. System includes modified couch motion rules for targeted anatomy below couch surface and enhanced safety interlocks. Operated by clinical staff in radiation oncology settings. System transforms programmed dose parameters into controlled radiation delivery. Healthcare providers use output to treat lesions and tumors; device enables precise dose delivery to target volumes. Benefits include expanded treatment capabilities for complex anatomy and increased dose delivery rates for specific photon energies.
Clinical Evidence
Bench testing only. Verification and validation testing performed to ensure compliance with functional requirements, use cases, and hazard mitigations. Testing confirmed adherence to international standards including IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and IEC 60601-2-1 (medical accelerator safety). No clinical data presented.
Technological Characteristics
Medical charged-particle radiation therapy system. Features FFF (Flattening Filter Free) high-intensity photon mode. Connectivity includes control software (C-Series 9.0). Safety features include hardware and secondary software interlocks for dose delivery and collision protection. Supports extended couch travel range. Operates via programmable dose delivery (up to 9999 MU) with maximum rates up to 2400 MU/min for 10x FFF mode.
Indications for Use
Indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body requiring radiation treatment.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
K112839
#### 510(k) Summary
# NOV 1 4 2011
The information below is provided for the Varian High Energy Linear Accelerator, following the format of 21 CFR 807.92.
| 1. | Submitter: | Varian Medical Systems<br>3100 Hansen Way, M/S e110<br>Palo Alto, CA 94304<br>Contact Name: Vy Tran<br>Phone: 650/424.5731<br>Fax: 650/842.5040<br>E-mail: vy.tran@varian.com |
|----|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Name of the Device:<br>Trade / Proprietary Names: | Varian High Energy Linear Accelerator<br>Novalis Tx, Trilogy, Trilogy Tx<br>Clinac iX, Clinac Cx<br>Clinac 2100C, 2100 C/D, 2300 C/D<br>Clinac 21 EX, 23 EX<br>Clinac DHX, DMX |
|----|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Common or Usual Names: | Novalis Tx, Trilogy, Trilogy Tx<br>Clinac iX, Clinac Cx<br>Clinac 2100C, 2100 C/D, 2300 C/D<br>Clinac 21 EX, 23 EX<br>Clinac DHX, DMX |
| | Classification Name: | Medical Charged Particle Radiation Therapy System<br>21 CFR §892.5050<br>Class II |
| | Product Code: | 90 IYE |
- 3. Predicate Device: Varian High Energy Linear Accelerator K100890
- 4. Description of the Device:
The Varian High Energy Linear Accelerator models provide various selections among the features, specifications, and accessories that have been most recently cleared as the Varian High Energy Linear Accelerator, K100890.
The changes to the Varian High Energy Linear Accelerator provide support for treatments using the high intensity photon treatment mode, also known as FFF (Flattening Filter Free). Additionally, couch motion rules are modified to enable the use of patient support systems where the targeted anatomy and the treatment isocenter area below the surface of the couch top. Modifications that augment existing safety controls are also included.
{1}------------------------------------------------
All other features of the Varian High Energy Linear Accelerator models remain as cleared by K100890.
- 5. Intended Use Statement
The Varian High Energy Linear Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Indications for Use Statement 6.
The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
#### 7. Substantial Equivalence
The modified device, the Varian High Energy Linear Accelerator, is substantially equivalent to the predicate device, the Varian High Energy Linear Accelerator (K100890).
The intended use and indications for use for the device are unchanged - see items 5 and 6 above.
The functionality of the Varian High Energy Linear Accelerator is equivalent to the functionality of the Varian predicate device in safety and effectiveness. Compared with the predicate device, the Varian High Energy Linear Accelerator (K100890), the basic operation is the same.
The comparison table below illustrates the substantial equivalence of the cleared device, Varian High Energy Linear Accelerator (K100890) and the modified device, Varian High Energy Linear Accelerator.
| Feature and/or Specification | Cleared Device (High<br>Energy Linear<br>Accelerator K100890) | Device with Change<br>(High Energy Linear<br>Accelerator) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Release version of control software | C-Series 8.0 | C-Series 9.0 |
| For gantry rotation from outside the treatment<br>room, collision protection between the gantry<br>& couch, when the couch is outside the<br>boundary conditions set by the user | motion rules are<br>enforced when the<br>imaging arms are<br>extended | motion rules are enforced<br>whether the imaging arms<br>are extended or retracted |
| Extended Travel Range Zone includes the 10<br>cm additional height needed to support the<br>Third-party Prone Breast Couch Insert | No | Yes |
| Interlock preventing additional dose from<br>being delivered after beam hold has been set | Hardware control | Hardware control plus<br>additional secondary<br>software check |
| Software and hardware support for FFF / HIM<br>(Flattening Filter Free/High Intensity Mode) | No | Yes |
| Maximum rate at which dose delivery occurs<br>for 10x photon energy | 600 MU per minute | 600 MU per minute |
510(k) Summary - Varian High Energy Linear Accelerator
{2}------------------------------------------------
| Feature and/or Specification | Cleared Device (High Energy Linear Accelerator K100890) | Device with Change (High Energy Linear Accelerator) |
|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Maximum rate at which dose delivery occurs for 10x FFF photon energy | None | 2400 MU per minute |
| Maximum rate at which dose delivery occurs for 6x SRS photon energy | 1000 MU per minute | 1000 MU per minute |
| Maximum rate at which dose delivery occurs for 6x FFF photon energy | None | 1400 MU per minute |
| Maximum field size | 3D Conformal Radiation Therapy: 40cm x 40cm.<br>IMRT: 34cm x 40cm.<br>SRS: 15cm x 15cm | 3D Conformal Radiation Therapy: 40cm x 40cm.<br>IMRT: 34cm x 40cm.<br>SRS: 15cm x 15cm |
| Maximum allowable dose limit for fixed X treatment type (for non-SRS and non-FFF treatment types) | 1999 MU | 1999 MU |
| Maximum programmable dose | 9999 MU | 9999 MU |
#### 8. Summary of Performance Testing
The design control procedures applied to the development of the Varian High Energy Linear Accelerator include requirements reviews, risk analysis, and verification and validation testing,
The verification and validation testing results from bench testing support the substantial equivalence of the modified device with its predicate, the unmodified device. Testing included testing against functional requirements, validation of use cases, hazard mitigation and control testing for compliance with applicable international medical device standards including the general safety standard IEC 60601-1, the electromagnetic compatibility standard IEC 60601-1-2, and the medical accelerator safety standard IEC 60601-2-1.
The objectives of the testing were to ensure that the pre-defined acceptance criteria and pass/fail criteria were met, including all specifications, functional requirements, use cases, hazard mitigations, and compliance with applicable international standards.
The conclusion from the results of the performance testing summarized above is that the defined criteria have been met, and that the specifications have been substantiated by the testing. The table in item 7 presents the key specifications relevant to the modifications for the Varian High Energy Linear Accelerator.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Vy Tran Vice President of Regulatory Affairs and Ouality Systems Varian Medical Systems 3100 Hansen Way, M/S E-110 PALO ALTO CA 94304-1129
NOV 1 4 2011
Re: K112839
Trade/Device Name: Varian High Energy Linear Accelerator Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 26, 2011 Received: September 28, 2011
Dear Ms. Tran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 5.10(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
## Indications for Use
·············································································································································································· 510(k) Number (if known):
Device Name: Varian High Energy Linear Accelerator
Indications for Use:
The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Office of In Vitro Die
K112839
510K
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.