REVIVE INTERMEDIATE CATHETER (REVIVE IC)

K112828 · Codman & Shurtleff, Inc. · DQY · Oct 25, 2011 · Cardiovascular

Device Facts

Record IDK112828
Device NameREVIVE INTERMEDIATE CATHETER (REVIVE IC)
ApplicantCodman & Shurtleff, Inc.
Product CodeDQY · Cardiovascular
Decision DateOct 25, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The REVIVE™ Intermediate Catheter is intended for use in the peripheral, coronary, and neuro vasculature for the intravascular introduction of interventional/diagnostic devices.

Device Story

REVIVE™ Intermediate Catheter; variable stiffness, single lumen catheter; aids physician in vascular access and delivery/retrieval of interventional/diagnostic devices. Features: highly flexible tip to semi-rigid proximal section; hydrophilic outer coating for friction reduction; PTFE-lined inner lumen for device/guide wire movement; luer fitting for accessory attachment. Used in peripheral, coronary, and neurovasculature; operated by physicians. Delivered over guide wires or through larger bore guiding catheters. Provides trackability and support for secondary devices; benefits patient by facilitating minimally invasive interventional procedures.

Clinical Evidence

Bench testing only. No clinical data required. Verification and validation testing included shaft kink resistance, lumen patency, trackability, fluoroscopic verification, tensile strength, burst testing, and flow rate. Biocompatibility testing performed per ISO 10993 and G95-1, including cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, thrombogenicity, and genotoxicity.

Technological Characteristics

Single lumen catheter; nylon body with stainless steel reinforcement; PTFE-lined inner lumen; hydrophilic outer coating; radiopaque distal marker. Variable stiffness profile (flexible tip to semi-rigid proximal). EtO sterilized. Dimensions/form factor: variable stiffness segments. Connectivity: none (mechanical device).

Indications for Use

Indicated for use in peripheral, coronary, and neurovasculature for intravascular introduction of interventional or diagnostic devices.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 112828 OCT 2 5 2011 # SECTION 5. 510(k) Summary (As Required By 21 CFR 807.92(a)) # A. Submitter Information | Submitter's name: | Codman & Shurtleff, Inc. | |---------------------|------------------------------------------| | Address: | 325 Paramount Drive<br>Raynham, MA 02767 | | Telephone: | 508-828-3312 | | Fax: | 508-977-6428 | | Contact Person: | Catherine Kilshaw | | Date of Submission: | September 23, 2011 | | B. Trade/Device Name: | REVIVE™ Intermediate Catheter | |-----------------------|-------------------------------| | Common Name: | REVIVE™ Intermediate Catheter | | Classification Name: | Percutaneous Catheter | | Regulation Number: | Class II per 21 CFR 870.1250 | ### C. Predicate Devices: | Device | Company | 510(k) Number/<br>Concurrence Date | Product<br>Code | Predicate For: | |-------------------------------|-------------------------------------------------------------------------|------------------------------------|-----------------|------------------------------------------------------| | ENVOY®<br>Guiding<br>Catheter | Codman & Shurtleff,<br>Inc. | K982770 | DQY | Intended Use<br>Design<br>Materials<br>Sterilization | | Courier ®<br>Microcatheter | Codman & Shurtleff,<br>Inc (Previously<br>Micrus<br>Endovascular, Corp) | K060116 | DQY | Materials | -1.2 -た。 189 #### D. Device Description: The REVIVE™ Intermediate Catheter is a variable stiffness, single lumen catheter designed to aid the physician in access to vasculature and in the delivery and retrieval of devices. Multiple levels of stiffness ranging from a highly flexible tip to a semi-rigid proximal section along the length of the catheter are designed to aid the physician in tracking over guide wires. The Revive IC may also be delivered through the lumen of a larger bore guiding catheter. The catheter has an outer hydrophilic coating that reduces friction during manipulation in the vessel. The lubricious PTFE lined inner lumen is designed to facilitate movement of the guide wires and other devices. A luer fitting located on the end of the catheter hub can be used to attach accessories. {1}------------------------------------------------ #### E. Intended Use: The REVIVE™ Intermediate Catheter is intended for use in the peripheral, coronary, and neuro vasculature for the intravascular introduction of interventional/diagnostic devices. # F. Summary of technological characteristics of the proposed to the predicate device: The REVIVE Intermediate Catheter is substantially equivalent to the Codman ENVOY Guiding Catheter and the Courier Microcatheter. No new technological characteristics are being introduced with the proposed device. The REVIVE IC has the same intended use as the ENVOY Guiding Catheter and it is similar with regard to design, materials, manufacturing and sterilization process. Both the REVIVE IC and the ENVOY feature a nylon body which is reinforced with stainless steel. REVIVE's multiple levels of stiffness from the highly flexible tip to a semi-rigid proximal section along the length of the catheter are designed to aid the physician in tracking over guide wires. This is similar to ENVOY which has three transition segments comprised of various durometers, along with a radiopaque filler. Both devices contain a soft, atraumatic distal tip. In addition, the REVIVE IC is similar to the Courier Microcatheter with regard to materials. The base materials of the four major components of the REVIVE IC (hub & shaft, PTFE liner, stainless steel reinforcement, hydrophilic coating) are identical to the Courier. All three devices are PTFE lined, have either a metallic braid or coil reinforcement, a polymer jacket of decreasing durometer from proximal to distal end, and a radiopaque marker at the distal tip. They are all EtO sterilized, packaged in a tyvek pouch and mounted either in a mounting card or plastic hoop. # G. Testing Summary: Preclinical testing data to demonstrate that the device performs according to its description and intended use were used to establish the performance characteristics : {2}------------------------------------------------ of the modifications to this device. Clinical testing was not required to establish substantial equivalence. Results of verification and validation conducted on the REVIVE IC demonstrate that it performs as designed, is suitable for its intended use, is substantially equivalent to the predicate device and therefore, does not raise any new issues of safety and effectiveness. The following tests were conducted: - Shaft Kink Resistance . - Kink Resistance at Proximal End . - Catheter Lumen Patency . - Ability to Track to Target Sites . - Fluoroscopic Verification . - Trackability . - Catheter Length Verification . - Catheter Diameter Verification . - Compatibility with Guiding Catheters � - . Compatibility with Microcatheters - Luer Tapering . - Hub Transitioning ● - Hub Attachment Tensile Strength® > � - Shaft Tensile Strength . - . System Liquid Air Leakage - System Air Leakage . - Tip Atraumaticity . - . Coating Integrity - . Burst Test - Flow Rate Test . The REVIVE IC met all the biocompatibility requirements as specified by the ISO 10993 Part I, and the General Program Memorandum # G95-1 on Biological Evaluation of Medical Devices. The following tests were conducted: - Cytotoxicity . - Sensitization . - Irritation . - . Systemic Toxicity - Hemocompatibility . {3}------------------------------------------------ - Thrombogenicity . - Genotoxicity . Based upon the design, materials, function, intended use, comparison with currently marketed devices and the non-clinical testing performed by Codman & Shurtleff, Inc., it is concluded that the REVIVE Intermediate Catheter is substantially equivalent to the ENVOY Guiding Catheter and the Courier Microcatheter and therefore, does not raise any new issues of safety and effectiveness. : {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 2 5 2011 Codman & Shurtleff, Inc. c/o Ms. Catherine Kilshaw Regulatory Affairs Specialist II 325 Paramount Dr. Raynham, MA 02767-0350 Re: K112828 Trade/Device Name: REVIVE™ Intermediate Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: September 26, 2011 Received: September 28, 2011 Dear Ms. Kilshaw: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. # # Melvin R. Feldman, M.D. Malvina B. Evdelman. M Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): K112828 Device Name: REVIVE™ Intermediate Catheter Indications For Use: The REVIVE™ Intermediate Catheter is intended for use in the peripheral, coronary, and neurovasculature for the intravascular introduction of interventional/diagnostic devices. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) JOE HUTTER (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices -Page 1 of KII2828 510(k) Number .
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...