K112799 · Benlan, Inc. · FPA · Dec 23, 2011 · General Hospital
Device Facts
Record ID
K112799
Device Name
MED-RX EXTENSION SET WITH T-CONNECTOR
Applicant
Benlan, Inc.
Product Code
FPA · General Hospital
Decision Date
Dec 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MED-RX Extension Set with T-Connector is intended to be used for administration or withdrawal of fluids to or from a neonatal patient as directed by a physician.
Device Story
Intravascular administration set; used for fluid administration or withdrawal in neonatal patients. Device consists of tubing (4" or 38" length), proximal T-connector with injection port and rotating male luer lock, and distal female luer lock fitting. Includes slide clamp. Operated by clinicians in hospital settings. Connects to standard luer fittings. Facilitates fluid delivery/access; single-use; sterile.
Clinical Evidence
Bench testing only. Performance verified per ISO 8536-4:2007, including tensile strength, leakage under pressure, liquid leakage, and particulate contamination. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity) and ASTM F-756-00 (hemolysis). Sterilization validated to SAL 10^-6 per ISO 11135-1:2007. No clinical data.
Technological Characteristics
Tubing: Polyvinyl Chloride (PVC). Connectors: Medical grade plastic. Dimensions: 0.040" ID, 0.095" OD; 4" or 38" length. Features: T-connector, injection site, rotating male luer lock, female luer lock, slide clamp. Sterilization: Ethylene Oxide (EO). Standards: ISO 8536-4:2007, ISO 10993, ISO 11135-1:2007.
Indications for Use
Indicated for administration or withdrawal of fluids to or from neonatal patients as directed by a physician.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Baxter Interlink T-Connector Extension Set (K060074)
Submission Summary (Full Text)
{0}------------------------------------------------
Set with T-Connector
DEC 2 3 2011
### SECTION 5 - 510(K) SUMMARY
Date of Summary: September 23, 2011
| Benlan Inc | | |
|----------------------|-------------------------------------------------------------------------|--|
| 2760 Brighton Road | Tel (905) 829-5004 | |
| Oakville, ON L6H 5T4 | Fax (905) 829-5006 | |
| Official Contact: | Cheryl Brown - QA / RA Manager | |
| Proprietary Name: | MED-RX Extension Set with T-Connector | |
| Common Name: | Intravascular Administration Set | |
| Classification Name: | Intravascular Administration Set, 880.5440 Intravascular Administration | |
| | Set. | |
| Class: | Class II | |
| Product Code: | FPA | |
| Predicate Device: | Baxter Interlink T-Connector Extension Set (K060074) | |
### Device Description
The MED-RX Extension Set with T-Connector is offered in two configurations, identical except for the tubing length. The proximal end of the MED-RX Extension Set with T-Connector features a T-Connector with an injection port and an integral rotating male luer lock. The distal end has a female luer lock fitting. The inner and outer diameter of both devices is consistent at 0.040" and 0.095" respectively, while the lengths available are either 4" or 38". Each of the luer fittings (distal female luer lock and proximal rotating male luer lock) are provided with protective caps. The MED-RX Extension Set with T-Connector is provided sterile and is single use.
### Indications for Use
The MED-RX Extension Set with T-Connector is intended to be used for administration or withdrawal of fluids to or from a neonatal patient as directed by a physician.
### Substantial Equivalence
The information provided in the premarket notification demonstrates that the proposed device is substantially equivalent to legally marketed devices. The proposed MED-RX Extension Set with T-Connector is substantially equivalent to the predicate Baxter Interlink T-Connector Extension Set (K060074). Both devices have the same intended use for intravascular administration or withdrawal of fluids to or from a neonatal patient as directed by a physician. Both devices are provided sterile and are single use.
A comparison of features and principles of operation between the proposed device and predicate device is provided in Table 1 below.
{1}------------------------------------------------
K112799
page 2 of 5
n Set with T-Connector
Table 1: Comparison between MED-RX Extension Set with T-Connector and Baxter Interlink T-Connector Extension Set (K060074)
| ATTRIBUTE | PROPOSED DEVICE –<br>MED-RX Extension Set with T-<br>Connector | PREDICATE DEVICE –<br>Baxter Interlink T-Connector Extension<br>Set (K060074) |
|------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| General Indications | | |
| Indications for Use | For administration or withdrawal of<br>fluids to or from a neonatal patient as<br>directed by a physician | Same |
| Type of Placement | Intravascular | Same |
| Intended for single<br>patient use | Yes | Yes |
| Prescription | Yes | Yes |
| Intended Population | Neonatal | Same |
| Intended Environment of<br>Use | Hospital | Same |
| Non Pyrogenic | Yes | Yes |
| Physical Specifications | | |
| Tubing outer diameter | 0.095" | 0.090" |
| Tubing inner diameter | 0.040" | 0.040" |
| Approximate length | 4" or 38" | 6" |
| Design Features | | |
| Distal Configuration | Female luer lock fitting | Same |
| Proximal Configuration | T-Connector with injection site and<br>rotating male luer lock | T-connector with injection site and male<br>luer slip adaptor or rotating male luer<br>lock adaptor |
| Clamp | Slide clamp | Same |
| Caps | Protective luer caps as necessary | Same |
| Compatibility | Compatible with standard luer fittings | Same |
| Material Composition | | |
| Tubing | Polyvinyl Chloride | Same |
| Connectors | Medical grade plastic | Medical grade plastic |
| Packaging and Sterilization | | |
| Sterile | Yes | Yes |
| Sterilization Method | Ethylene Oxide (EO) | Gamma Radiation |
| Packaging Configuration | Medical grade peelable paper/ poly<br>pouch | Polypropylene pouch |
### Summary of Differences
There are no significant differences between the proposed MED-RX Extension Set with T-Connector and the predicate device Baxter Interlink T-Connector Extension Set (K060074). Similarities between the proposed device and the predicate devices include identical indications for use and duration of use. The MED-RX Extension Set with T-Connector and the Baxter Interlink T-Connector Extension Set are sterile, single use devices, packaged in peelable pouches.
{2}------------------------------------------------
# sion Set with T-Connector
page 3 of 5
The proposed MED-RX Extension Set with T-Connector is to be offered in two different lengths: 4″ and 38″, whereas the predicate device is only available in a 6″ length. The additional lengths of extension sets to be offered for the proposed device can be considered to have no impact on safety or effectiveness as the additional length is merely intended to increase the amount of accessible tubing for use. The proposed device has a slightly larger tubing outer diameter than the predicate device but the dimensions are effectively equivalent. Both devices feature luer connectors with protective caps as required and slide clamps are placed on each set. The design features of both the proposed device and the predicate device can be also considered equivalent as the configuration of the devices are identical.
Therefore, any minor differences between the proposed device and the predicate have been evaluated to have no impact on safety or effectiveness of the MED-RX Extension Set with T-Connector. Therefore the proposed device can be considered substantially equivalent to legally market devices.
### Non-Clinical Test Summary
Verification of functional performance of the MED-RX Extension Set with T-Connector has been performed as per ISO 8536-4: 2007. The MED-RX Extension Set with T-Connector has successfully completed all required performance testing following the applicable guidelines of ISO 8536-4: 2007 including tensile strength, resistance to leakage under pressure, resistance to liquid leakage, particulate contamination, and chemical requirements. The MED-RX Extension Set with T-Connector was also tested for natural rubber latex content. Please refer to Table 2.
| Test | Standard | Results |
|--------------------------------------------|-----------------|----------------------------------------|
| Tensile Strength -Tubing/T-Connector | ISO 8536-4:2007 | Pass |
| Tensile Strength - Tubing/Female Luer Lock | ISO 8536-4:2007 | Pass |
| Leakage under Pressure | ISO 8536-4:2007 | Pass |
| Liquid Leakage | ISO 8536-4:2007 | Pass |
| Particulate Contamination | ISO 8536-4:2007 | Samples met contamination index limit. |
Table 2: Non-Clinical Test Summary
{3}------------------------------------------------
K112799
# n Set with T-Connector
| Test | Standard | Results |
|------------------------------|-------------------------------------|-------------------------------------------------|
| Chemical Requirements | ISO 8536-4:2007 per<br>Clause 5 & 7 | Pass |
| Natural Rubber Latex Content | Modified Lowry Method | Device does not contain<br>natural rubber latex |
### Summary of Sterilization
Each MED-RX Extension Set with T-Connector is individually packaged using a medical grade peelable synthetic polymer reinforced paper with a film backing, and sterilized using ethylene oxide. Please see Table 3 for a summary.
Table 3: Sterilization Summary
| Test Description | Standard | Results |
|----------------------------|-----------------------------|---------------------------------------------------------|
| Method of Validation | ANSI/AMMI/ISO 11135-1: 2007 | Validated to a Sterility Assurance<br>Level of 1 x 10-6 |
| EO Sterilization Residuals | ISO 10993-7: 2008 | Pass |
| Bacterial Endotoxins | ANSI/AAMI ST72: 2002 | Pass |
### Summary of Biocompatibility Tests
Biocompatibility testing was successfully completed on sterile finished devices. The MED-RX Extension Set with T-Connector is classified as a prolonged duration, indirect blood path contacting device. A summary of the testing completed and the relevant standards are listed in Table 4.
3 of 5
{4}------------------------------------------------
Benlan™
MED-RX® Extension Set with T-Connector
K1127991
page ్ర 50
Table 4: Biocompatibility Test Summary
| Test Description | Standard | Results |
|---------------------------------------------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------|
| ISO MEM Elution with L-929<br>Mouse Fibroblast Cells<br>(Cytotoxicity) | ISO 10993-5: 2009 | Product code 10-1053XLU is considered<br>non-toxic. |
| ISO Guinea Pig Maximization<br>Sensitization Test (Method<br>of Biomaterial Extracts) | ISO 10993-10:2002 | Product code 10-1053XLU did not elicit a<br>sensitization response. |
| ISO Intracutaneous<br>Reactivity Test | ISO 10993-10:2002 | Product code 10-1053XLU would be<br>considered a non-irritant. |
| ISO Acute Systemic Injection<br>Test | ISO 10993-11: 2006 | The findings indicate that the<br>requirements of the ISO Acute Systemic<br>Injection Test have been met. |
| ASTM Hemolysis Assay -<br>Extract Method | ASTM F-756-00 | Product code 10-1053XLU is considered<br>non-hemolytic and passes the test. |
| Materials Mediated Rabbit<br>Pyrogen Test | USP 32: 2009 <151> | Product code 10-1053XLU is determined<br>to be non-pyrogenic. |
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 2 3 2011
Ms. Cheryl Brown Quality Assurance / Regulatory Affairs Manager Benlan Incorporated 2760 Brighton Road Oakville, Ontario CANADA L6H 5T4
Re: K112799
Trade/Device Name: MED-RX Extension Set with T-Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: September 23, 2011 Received: September 27, 2011
Dear Ms. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2 - Ms. Brown
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety /ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov /MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D.m
Anthony D. Watson, B.S., M.S., M.P.
nthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{7}------------------------------------------------
KII2799
## SECTION 4 - INDICATIONS FOR USE
510(K) Number (If Known): Not Known.
Device Name: MED-RX Extension Set with T-Connector
Indications For Use:
The MED-RX Extension Set with T-Connector is intended to be used for administration or withdrawal of fluids to or from a neonatal patient as directed by a physician.
| Prescription Use: | <span style="text-align: center;">✓</span> |
|-----------------------------|--------------------------------------------|
| (Part 21 CFR 801 Subpart D) | |
| AND/OR | |
| Over-the-Counter Use | |
| (21 CFR 801 Subpart C) | |
| N/A | |
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED.
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Richard Chapman Dec 21, 2016
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
510(k) Number: K112799
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.