← Product Code [NRB](/productcode/NRB) · K112782

# MISONIX SONICONE(R) ULTRASONIC WOUND CARE SYSTEM AND ACCESSORIES (K112782)

_Misonix, Inc. · NRB · Dec 1, 2011 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K112782

## Device Facts

- **Applicant:** Misonix, Inc.
- **Product Code:** [NRB](/productcode/NRB.md)
- **Decision Date:** Dec 1, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4410
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Misonix SonicOne Ultrasonic Would Care System and Accessories are indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers, bedsores and vaginal ulcers, soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgment would require the use of an ultrasonic aspirator with sharp debridement.

## Device Story

System comprises generator, handpiece, and irrigation unit. Generator feeds 22.5 kHz electrical signal to piezoelectric crystals in handpiece; crystals vibrate at same frequency. Titanium tip amplifies vibration for tissue ablation. Irrigation unit provides solution to operative site. Used by physicians in clinical settings for wound debridement and surgical site cleansing. Output is mechanical ultrasonic energy for sharp debridement. Benefits include effective tissue removal and site cleansing.

## Clinical Evidence

No clinical data obtained. Substantial equivalence based on design, hardware, and performance testing identical to predicate AUSS-6.

## Technological Characteristics

System uses 22.5 kHz ultrasonic energy. Components: generator, piezoelectric handpiece, titanium tip, irrigation unit. Materials: titanium tip. Standards: IEC 60601-1, IEC 60601-1-2, FCC Part 18, ISO 10993-1, ISO 10993-7, ISO 15223-1. Sterilization: Ethylene oxide.

## Regulatory Identification

A low energy ultrasound wound cleaner is a device that uses ultrasound energy to vaporize a solution and generate a mist that is used for the cleaning and maintenance debridement of wounds. Low levels of ultrasound energy may be carried to the wound by the saline mist.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Low Energy Ultrasound Wound Cleaner.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Arobella Medical LLC AR 1000 Ultrasonic Wound Therapy System ([K062544](/device/K062544.md))
- Misonix Inc. Alliger Ultrasonic Surgical System AUSS-6 ([K050776](/device/K050776.md))
- Microtek Medical Inc. Equipment Drapes ([K050322](/device/K050322.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEC - 1 2011 K112782

### Exhibit E 510(k) SUMMARY - Misonix SonicOne Ultrasonic Wound Care System and Accessories

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92.

#### 1. Submitter's Identification

| Submitter's Name: | MISONIX INCORPORATED                    |
|-------------------|-----------------------------------------|
| Address:          | 1938 New Highway, Farmingdale, NY 11735 |
| Telephone Number: | 516-694-9555                            |
| Contact Person:   | Ronald R. Manna                         |
| Date Prepared:    | August 9, 2010                          |

#### 2. Name of Device

| Proprietary Name:    | Misonix SonicOne Ultrasonic Wound Care System<br>and Accessories |
|----------------------|------------------------------------------------------------------|
| Common/Usual Name:   | Ultrasonic Surgical System<br>Ultrasonic Surgical Aspirator      |
| Classification Name: | Instrument, Ultrasonic Surgical                                  |

#### 3. Predicate Device Information

| Predicate Devices | Arobella Medical LLC AR 1000 Ultrasonic Wound<br>Therapy System K062544<br>Misonix Inc. Alliger Ultrasonic Surgical System<br>AUSS-6 K050776<br>Microtek Medical Inc. Equipment Drapes K050322 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 4.                | Device Description                                                                                                                                                                             | The SonicOne Ultrasonic Wound Care System is<br>comprised of a generator, which feeds a 22.5 kHz<br>electrical signal to a piezoelectric crystals mounted<br>in a hand-held handpiece; the crystals then vibrate<br>at the same frequency. The titanium tip attached to<br>the handpiece amplifies the vibration. An irrigation<br>unit is provided to introduce irrigation solution to<br>the operative site. Accessories include probe tips,<br>wrenches, sterile and non sterile tube sets and sterile<br>Surgical Procedure bags and handpiece sheaths. |

\$\beta. 1 (3\$

{1}------------------------------------------------

# KII2782

- The Misonix SonicOne Ultrasonic Would Care System and 5. Intended Use: Accessories are indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers, bedsores and vaginal ulcers, soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgement would require the use of an ultrasonic aspirator with sharp debridement.
## Comparison to Predicate Device SonicOne Ultrasonic Wound Care System 6. and Accessories are similar in design, material and operating parameters to the Misonix Inc. AUSS-6 Ultrasonic Surgical Aspirator and the Arobella Medical LLC AR 1000 Ultrasonic Wound Therapy System.

#### 7. Safety and Performance Data

The Misonix SonicOne Ultrasonic Wound Care System and Accessories have been designed and tested to pass the following Voluntary Standards:

| IEC 60601-1*             | Medical Electrical Equipment, Part 1: General Requirements for Safety                                                                  |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2:2001       | Medical Electrical Equipment General Requirements for EMC                                                                              |
| FCC Part 18              | EMC Requirements                                                                                                                       |
| ISO 10993-1:2009         | Biological evaluation of medical devices —Part 1: Evaluation and testing                                                               |
| ISO 10993-7:2008         | Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals                                                |
| ISO 15223-1:2007/A1:2008 | Medical devices —Symbols to be used with medical device labels, labeling, and information to be supplied —Part 1: General requirements |

- This device does not contain software. 7. Software Validation
- Validation statements are contained in Exhibit J. 8. Sterilization Validations
- 9. Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

The SonicOne is identical to the AUSS-6 Ultrasonic Surgical Aspirator cleared under 510K #K050776. Therefore, no new Non-Clinical Tests have been performed in anticipation of this submission.

P. 2/3

{2}------------------------------------------------

K112782

#### Discussions of Clinical Tests Performed 10.

The FDA has cleared all indications for use in the predicates. As such, no additional clinic data was obtained in anticipation of this submission.

#### Conclusions 11.

Misonix Inc. can state that the SonicOne is substantially equivalent in Mode of Operation, Hardware Design and Output Parameters to the Arobella Medical AR 1000 and the Misonix Inc. AUSS-6. Based upon the system and hardware validations described herein, the Misonix Inc. believes the SonicOne Ultrasonic Wound Care System and Accessories pose no new issues of safety or efficacy when used for soft and hard tissue ablation during wound treatment.

P.3/3

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol associated with healthcare and medicine. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC - 1 2011

Misonix Inc. % Mr. Ronald R. Manna 1938 New Highway Farmingdale, New York 11735

Re: K112782

Trade/Device Name: Misonix SonicOne® Ultrasonic Wound Care System and Accessories Regulation Number: 21 CFR 878.4410 Regulation Name: Low energy ultrasound wound cleaner Regulatory Class: Class II Product Code: NRB, FQH Dated: November 02, 2011 Received: November 03, 2011

Dear Mr. Manna:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 - Mr. Ronald R. Manna

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Sincerely yours,
Erine Keith.

for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Exhibit C Indications for Use Statement

i12782

Device Name: Misonix SonicOne ® Ultrasonic Wound Care System and Accessories

The Misonix SonicOne Ultrasonic Would Care System and Accessories are indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers, bedsores and vaginal ulcers, soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgment would require the use of an ultrasonic aspirator with sharp debridement.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X. (Per 21 CFR 801.109)

or

Over-The-Counter Use

Neil R.P. Dych for Nixon

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K112782

8

---

**Source:** [https://fda-staging.innolitics.com/device/K112782](https://fda-staging.innolitics.com/device/K112782)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).