K112774 · Becton, Dickinson and Company · MMK · Mar 28, 2012 · General Hospital
Device Facts
Record ID
K112774
Device Name
BD RECYKLEEN SHARPS COLLECTORS
Applicant
Becton, Dickinson and Company
Product Code
MMK · General Hospital
Decision Date
Mar 28, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
Sharps collectors are intended to be used for the disposal of contaminated medical sharps in health care facilities.
Device Story
BD Recykleen™ Sharps Collectors are single-use containers for safe disposal of contaminated medical sharps in healthcare facilities. Device design is identical to predicate BD Guardian™ Sharps Collectors, featuring injection-molded polypropylene/polyolefin plastic construction. Key modification involves using recycled resin grades recovered from hospital/healthcare facility medical sharps waste-stream via the BD ecoFinity™ Life Cycle Solution. Containers include features for puncture resistance, leakproof sides/bottom, and overfill indication; optional accessories include wall/cart brackets and stabilizers. Device is used by healthcare personnel to safely contain sharps waste; output is a secured, puncture-resistant container that prevents accidental needle sticks and facilitates safe disposal. Benefits include reduced environmental impact through plastic recycling while maintaining equivalent safety and performance standards as virgin-resin collectors.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including puncture resistance (ASTM F2132-01), impact resistance (drop testing), leak resistance (OSHA 29 CFR 1910.1030), and transit damage resistance. Safety testing included biocompatibility (ISO 10993), bioburden analysis (ISO 11737-1), chemical extractables/leachables (LC/DAD/MS, GC/MS), and heavy metal analysis (ICP/MS, XRF).
Technological Characteristics
Materials: Polypropylene and polyolefin plastics (virgin, post-consumer, post-industrial, and post-hospital recycled). Construction: Injection-molded container and lids. Standards: ASTM F2132-01 (puncture resistance), ISO 10993 (biocompatibility), ISO 11737-1 (bioburden). Features: Puncture-resistant, leakproof, closable, overfill indication, optional wall/cart brackets and stabilizers. Single-use. No electronic or software components.
Indications for Use
Indicated for the disposal of contaminated medical sharps in health care facilities.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
# MAR 2 8 2012
### K112774 510k SUMMARY BD Recykleen™ Sharps Collector
Date: March 28, 2012
1. Submitted By
BD Medical Surgical Systems Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, New Jersey, 07417 USA
### 2. Contact Information
Primary Contact Eileen Hiller Director, Regulatory Affairs BD Medical Surgical Systems 1 Becton Drive Franklin Lakes. New Jersey 07417 USA 201-847-4570 Eileen Hiller@BD.com
### Secondary Contact
Sergio Gadaleta Vice President, Regulatory Affairs BD Medical Surgical Systems 1 Becton Drive Franklin Lakes, New Jersey 07417 USA 201-847-4496 Sergio Gadaleta@BD.com
| 3 .- Device identification :- | |
|-----------------------------------------|----------------------------------------------------------------------------------|
| -Trade Name: | BD-Rec |
| Common Name: | Sharps |
| Classification: | MMK - |
| | needles |
| Allian William Harris, College, Comment | And Antiques and the manufacturer of the contribution of the contribution of the |
CFR Reference: Classification Panel:
cykleen TM-Sharps Container-Container Accessory to hypodermic single lumen 21CFR 880.5570 - Class II General Hospital
#### 4. Predicate devices:
Substantial equivalence is being claimed to the following legally marketed devices:
Trade Name: BD Guardian™ Sharps Collector Common Name: Sharps Container Classification: MMK - Accessory to hypodermic single lumen needles Predicate 510k No K943134 Cleared 11/07/1994 21CFR 880.5570 - Class II CFR Reference:
{1}------------------------------------------------
Classification Panel:
General Hospital
### 5. Product Description:
The BD Recykleen™ Sharps Collector is a single use device designed for the safe disposal of regulated medical waste such as contaminated sharps waste. All materials, including sides, bottom, and top are manufactured with polypropylene or polyolefin plastics. the same type of material used in current BD Sharps Collectors and BD Recykleen™ Sharps Collectors. The predicated devices were previously cleared under K943134.
The basis of this new Traditional 510(k) submission is for a material change of the sharps collector. The material change, which will incorporate the use of a new resin grade of recycled plastic recovered from the hospital and healthcare facility medical sharps waste-stream. The use of post-hospital recycled resins in the manufacture of the BD Recykleen Sharps Collectors has been demonstrated to be safe and effective. BD has developed the BD ecoFinity™ Life Cycle Solution, which is a program for recycling clinically used hospital-generated medical sharps waste. The basics of this new recycling program are that plastics, recovered from recycling post-hopital sharps waste, are utilized to manufacture new sharps collectors rather than being treated and disposed in landfills or incinerated. After sanitization, the treated sharps waste is separated into metal, light plastics, and other materials. The light plastics are recovered, extruded and pellitized into a final resin pellet form. The recycled resin pellets are then injection molded into components that make up the finished BD Recykleen™ Sharps Collectors.
The technological and performance characterisitics of the BD Recykleen™ Sharps-Collectors (new-device) is similar-to-BD-Guardian Sharps-Collectors. (predicate device). There is no change to the intended use of the new or predicate Sharps Collectors, the collector design and technological and performance characteristics remain unchanged.
BD Sharps Collectors may be used with optional accessories, stabilizer and brackets, that provide the customer choices for further security and enhanced stability during usage. The 8Qt and 6 Gallon Product Instructions for Use list the compatible brackets and stabilizer accessories available.
Brackets are designed to secure the collector to a wall or cart. Some bracket accessories include additional security via a locking mechanism.
Stabilizers: Stabilizers are designed to stabilize collectors on a countertop or other flat surface.
{2}------------------------------------------------
# 6. Intended Use:
The BD Recykleen™ Sharps Collectors intended use:
Sharps Collectors are intended to be used for disposal of contaminated medical sharps in health care facilities
### 7. Comparison to Predicate Devices:
The sharps container is similar to the predicated in intended use, materials, measuring principle and performance.
| FEATURE | MODIFIED DEVICE<br>BD RECYKLEENTM<br>SHARPS COLLECTOR | PREDICATE DEVICE<br>BD GUARDIANTM SHARPS<br>COLLECTOR |
|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Closable | Same | Yes, Snap fit cap/closure |
| Puncture Resistant | Same<br>ASTM F2132-01 (2008),<br>"Standard Specification for<br>Puncture Resistance of<br>Materials Used in Containers<br>for Discarded Medical<br>Needles and Other Sharps". | Yes Passes proposed ASTM<br>F04.65.01 procedure<br>(Current draft in 1994) |
| Leakproof on sides<br>and<br>Bottom | Same | Yes |
| Labeled or Color-<br>Coded: | Same | Sharps/Infectious Waste<br>Yes |
| Biohazard Warning<br>Label | | |
| Fluorescent orange<br>or orange-red with<br>lettering in<br>contrasting color | Same | Yes |
| Affixed to container | Same | Yes |
| Red Container or<br>label | Same | Yes |
| Recycled Content<br>Label | Patient Room Collectors<br>Recycled Content- Minimum<br>20%<br><br>Nestable Collectors<br>Minimum 50% Recycled<br>Content | NA |
| FEATURE | MODIFIED DEVICE<br>BD RECYKLEENTM<br>SHARPS COLLECTOR | PREDICATE DEVICE<br>BD GUARDIANTM SHARPS<br>COLLECTOR |
| Capable of<br>maintaining stable,<br>upright position. | Same | Yes |
| No feature to bend,<br>break, or shear<br>needle. | Same | No feature present. |
| Unwinder | Same | Unwinder is designed for one-<br>handed operation on containers.<br>Labeling is "Single Use Only" |
| Reusable Sharp<br>Container | Same | |
| Overfill Indication | Same | "Do Not Overfill" or "Fill to this<br>Level Only" is labeled or<br>embossed on container at the<br>location of overfill. Labeling<br>includes a "Fill Line." |
| Locking Enclosure | Same | All models have the option for<br>use of security locks and keys |
| Holder to secure to<br>walls | Same | Product is capable of attaching to<br>wall-mount or cart-mount. |
| Materials | Polyolefinic resins:<br>Virgin polypropylene<br>Post consumer<br>Post industrial<br>Post hospital | Polyolefinic resins- Virgin<br>Polypropylene |
| Construction | Same | Injection molded Container,<br>Injection molded Lids/Closures |
| Clarity | Same | Each Collector has a minimum<br>of one translucent component,<br>either base or top. Some models<br>with clear "see thru" tops. |
| Recycled Content | Nestable Sharps Collectors<br>minimum 50% | N/A |
{3}------------------------------------------------
The Device Models Subject to this submission are listed below. The BD Recykleen™ Sharps collector's dimensions, access openings, features and capacity are essentially equivalent to the predicate devices.
{4}------------------------------------------------
| Family | BD<br>Recykleen™Sharps<br>Collector | Access Opening<br>and Closure | Access<br>Opening<br>Size | Length (in) | Width<br>(in) | Height<br>(in) | Weight<br>(grams) | Capacity |
|----------------------------------|-------------------------------------|--------------------------------------------------------------|------------------------------------------------|-------------|---------------|----------------|-------------------|-----------|
| Nestable<br>Sharps<br>Collectors | 8 Quart<br>Model No. 305059 | Funnel Top, with<br>restricted access<br>petals<br>Hinge Cap | 1.5 "x<br>1.125"<br>with<br>flexible<br>petals | 9.75 | 10.00 | 6.75 | 446 | 8 Quarts |
| | 6 Gallon<br>Model No. 305160 | Open top with<br>restricted access<br>opening<br>Hinge Cap | 2.5"<br>diameter | 17.50 | 12.50 | 8.50 | 935 | 6 Gallons |
### 8. Discussion of Similarities and Differences in New and Predicate Product
### Intended Use Comparison
The indications for use of the new BD Recykleen™ Sharps Collector are not new indications in that they are the same as those for the predicated devices. Sharps collectors-and-other-predicated-devices-are-containers-intended-for-the-disposal-ofcontaminated medical waste within healthcare facilities.
### \ Design and Materials Comparison
The design and functionality of the sharps collectors and predicated devices are identical. They are constructed from polypropylene and contain a component(s) made from recycled polyolefin plastics and are intended for single use only. All of the devices conform to recognized standard, ASTM F-2132-01 for needle penetration resistance. They have features to prevent contact between user and the contents, and are designed for easy and safe determination of fullness. None of the devices have features that bend, break, or shear needles.
### 9. Summary of Performance Bench Testing of Device Modifications
#### 9.1 Performance Standards:
{5}------------------------------------------------
No performance standards have been established under Section 514 for this product code. All recognized standards and other regulations and guidance documents that were used in this 510(k) have been listed in Section 20.
The performance testing demonstrates compliance with the recognized consensus standard, ASTM F 2132-01 (Reapproved 2008) e1, "Standard Specification for Puncture Resistance of Materials Used in Collectors for Discarded Medical Needles and Other Sharps." In addition the relevant FDA guidance document, "Guidance on the Content and Format of Premarket Notification [510(K)] Submission for Sharps Collectors" dated October 1993, was used to identify applicable physical and mechanical features of the modified and predicate devices.
All applicable standards have been used to show that the BD Recykleen Container family is substantially equivalent to the appropriately listed predicate devices.
The performance testing summary demonstrates substantial equivalence between the modified device and the predicate devices. The new sharps collectors have been tested by appropriate methods with respect to relevant FDA guidance documents, FDA recognized ASTM standards F 2132-01 and OSHA regulations 29 CFR Part 1910:1030. No new issues of safety and effectiveness were raised with the testing performed, and the BD Recykleen Containers are considered substantially equivalent to its predicate device.
Performance Testing (Bench) -Product Testing 9.2
The BD Recykleen™ Sharps Containers incorporate the identical collector design features, performance characteristics, manufacturing technology. The results of product_performance_testing_demonstrated_equivalent_performance_to_predicate_device performance and no new issues of concern-were raised.
#### Impact Resistance
ﺔ ﻣﻤﺜﻠ
Impact resistance was assessed such that no open fractures or disassembly results when a filled collector is dropped. The container is filled with assorted sharps or syringes with or without needles, or equivalent weight using resin beads and sealed as if ready for transport. The collector is dropped 39" from the base onto a hard surface floor.
Pass/Fail Criteria: No gross fractures (external wall of container is open enough to allow escape of solid contents) or loss of contents are permitted. Stress fractures (crack in external wall where wall does not separate to release solid contents) are permitted.
#### Needle Penetration Resistance
{6}------------------------------------------------
Needle Penetration Resistance is based on ASTM F2132-01, "Standard Specification for Puncture Resistance of Materials Used in Containers for Discarded Medical Needles and Other Sharps". for minimum and average needle penetration force. The test method involves cutting 12 samples from needle containment areas of the sharps collector. Each sample is tested using a motorized mechanical tester (Instron, ATS or equivalent) with a new 21 gauge x 1" needle, each penetration is tested at 4"/minute.
Pass/Fail Criteria: No one needle puncture force can be less than 2.8 lbf and average puncture force of 3.4 lbf must be met to be acceptable.
### Leak Resistance
The Leak Resistance testing is based on OSHA Specification (29 CFR 1910.1030) -The container is filled with water to the labeled fill line and left to stand on its base 1 hour.
Pass/Fail Criteria: No leaks are observed with product standing in upright position.
#### Transit Damage Resistance:
Transit Damage testing demonstrates the product is free from any visible damage that may affect customer usage, safety, or satisfaction when products are packaged in shipping cases and dropped. The appropriate number of products are packaged in the appropriate shipping case and sealed as if ready for shipment. The shipping case is dropped from 24" onto a hard surface floor in 10 different orientations.
Pass/Fail Criteria: The product is free from any visible damage that may affect customer usage, safety, or satisfaction when products are packaged in the shipping case and dropped with a 90% pass rate.
### Performance Testing- Needle Penetration Resistance- Comparison of all Recycled Resin Grades and blend of resins in use
The critical performance characteristic for Sharps Collectors is Needle Penetration Resistance (NPR). Individual NPR regression analyses of wall thickness versus needle puncture force were completed for each of the following: 100% recycled resin grades (PH, PI, and PC); virgin and recycled resin blends. The NPR regression curves were completed for each virgin material, recycled resin blend formulation. Each resin material is characterized to have a unique minimum wall thickness to meet ASTM F 2132-01.
#### ਰੇ ਤੇ Safety Tests
To demonstrate that post hospital recycled resins are substantially equivalent to the PC and PI recycled resins and virgin resins used in the predicate Sharps Collectors. additional safety tests were conducted. These tests were selected to eliminate
BD Recykleen Sharps Collector 510k Summary
{7}------------------------------------------------
potential concerns of residual resin contamination after completion of the sanitization, separation, recycling and extrusion processes.
Pellet samples or test plaques representing 100% PH, 100% PI, and 100% PC recycled and virgin resins were tested according to those listed in Table 6.
All Safety Tests demonstrated equivalent performance between the principle and predicate Sharps Collectors and individual resins tests.
| Safety Test | Standards/Method | Principle Device | Predicate Device |
|---------------------------------------------------------------------------------------------------|-------------------------------------|--------------------------------------------------|--------------------------------------------------|
| Biocompatibility Testing<br>Cell Toxicity Murine Local Lymph Node Assay Primary Dermal Irritation | ISO 10993<br>ISO 10993<br>ISO 10993 | passed<br>passed<br>passed | passed<br>passed<br>passed |
| Bioburden-<br>All test samples- non sterile Samples<br>Resin<br>Pellets | ISO 11737-1 | (PI,PC,PH) | Virgin PP |
| Sharps Collectors<br>Interior exterior | | <46 CFU/10g<br>1.5 CFU/2 swabs<br>10.8 CFU/100mL | <4.8 CFU/10g<br>3.7 CFU/2 swabs<br>3.1 CFU/100mL |
| Chemical Analysis<br>Extractible: liquid extracts (water/alcohol) | LC/DAD/MS | Same | Same |
| Leachable:<br>volatiles/headspace | GC/MS | No significant differences | No significant differences |
| Heavy Metals | ICP/MS<br>XRF | Passed<br>TPCH<br><100ppm | Passed<br>TCPH<br><100ppm |
## Table 6 Summary of Safety Tests
### Conclusion:
BD Recykleen™ Sharps Collectors proves to be substantially equivalent to the respective predicate sharps collectors.
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is simple, using only black and white, which gives it a clean and official appearance.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Eileen Hiller Manager of Regulatory Affairs Becton, Dickinson and Company BD Medical Surgical 1 Becton Drive MC 237 Franklin Lakes, New Jersey 07417
MAR 2 8 2012
Re: K112774
Trade/Device Name: BD Recykleen" Nestable Sharps Collectors Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: March 20, 2012 Received: March 22, 2012
Dear Ms. Hiller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
### Page 2 - Ms. Hiller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
fa
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation · Center for Devices and Radiological Health
#### Enclosure
{10}------------------------------------------------
# Indications for Use Statement
### 510(k) Number (if known):
K112774
Device Name:
BD Recykleen™ Nestable Sharps Collectors
Model Number Size
305059 8 Quart Nestable Sharps Collector
Model Number Size
305160 6 Gallon Nestable Sharps Collector
Indications for Use:
Sharps collectors are intended to be used for the disposal of contaminated medical sharps in health care facilities.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth B. (Lamie-Wells
(Division Sign-Uff) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Confidential & Proprietary
0000006
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.