ORTHOLOC™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Device Story
ORTHOLOC™ Bone Screws are titanium alloy, single-use, non-locking cortical bone screws (3.5mm and 4.0mm). Used by surgeons in clinical settings for bone fixation, reconstruction, and joint fusion. The device functions as a mechanical fastener to stabilize bone segments. The subject device features a modified neck radius compared to the predicate, resulting in the removal of approximately one distal thread. Performance is verified through worst-case pull-out testing and design comparison to ensure equivalence to predicate systems.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing (worst-case pull-out testing) and design characteristic comparisons.
Technological Characteristics
Titanium alloy, solid core, non-locking cortical bone screws. Available in 3.5mm and 4.0mm diameters. Single-use. Mechanical fixation principle.
Indications for Use
Indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation in patients requiring bone fixation, size-dependent.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
ORTHOLOC™ 3Di Ankle System and ORTHOLOC™ Bone Screws (K102429)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a stylized graphic consisting of three angled shapes. At the top of the image, there is a handwritten number "K112772(1/2)".
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
OCT - 7 2011
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the URTHOLOC™ Bone Screws.
| A.1. Submitted By: | Wright Medical Technology, Inc.<br>5677 Airline Rd<br>Arlington, TN 38002 |
|-----------------------------------|---------------------------------------------------------------------------|
| Date: | August 29, 2011 |
| Contact Person: | Kellen Hills<br>Regulatory Affairs Specialist<br>(901) 290-5816 |
| A.2. Proprietary Name: | ORTHOLOC™ Bone Screws |
| Common Name: | Bone Screw |
| Device Classification Regulation: | 21 CFR 888.3040-Class II |
| Device Product Code & Panel: | HWC: Screw, Fixation Bone<br>87 Orthopedics |
| A.3. Predicate Device: | K102429-ORTHOLOC™ 3Di Ankle<br>System and ORTHOLOC™ Bone Screws |
K082320-WrightTM Compression Screws
#### A.4. Device Description
The ORTHOLOC™ Bone Screws subject to this premarket notification include the nonlocking 3.5mm low profile cortical bone screws and non-locking 3.5 and 4.0 cortical bone screws.
These screws are manufactured from titanium alloy and have a solid core. The implants are single use only devices.
#### A.5. Intended Use
ORTHOLOC™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002 901.867.9971 phone
. ・ ...
www.wmt.com
Plating
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a graphic design of three horizontal lines that are stacked on top of each other. Above the word "WRIGHT." is the text "K 112772 (2/2)".
#### A.6. Technological Characteristics Comparison
The subject ORTHOLOC™ Bone Screws are identical to the predicate ORTHOLOC™ Bone Screws with the exception of the radii at the transition from the screw head to the screw shaft. This change resulted in approximately one thread being removed from the distal portion of the screw.
#### B.1. Substantial Equivalence - Non-Clinical Evidence
Substantial equivalence is shown through worst-case pull out testing, materials information, and comparison of design characteristics. The results show that the subject ORTHQLOC™ Bone Screws can be expected to perform at least as well as the legally marketed predicate ORTHOLOCTM Bone Screws and Wright Compression Screws.
#### B.2. Substantial Equivalence - Clinical Evidence
N/A
#### B.3. Substantial Equivalence - Conclusions
Substantial equivalence is shown through worst-case pull out testing, materials information, and comparison of design characteristics. The subject screws are identical in indication for use, diameter, size range and material to the predicates, and differ only in screw neck radii. No new types of safety and effectiveness questions can be expected. From the evidence given in the Premarket Notification, the subject devices can be expected to perform at least as well as the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the wings of the serpent. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Wright Medical Technology, Inc % Kellen Hill Regulatory Affairs Specialist 5677 Airline Road Arlington, Tennessee 38002
OCT - 7 2011
Re: K112772
Trade/Device Name: ORTHOLOC"™ Bone Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class H Product Code: HWC Dated: September 12, 2011 Received: September 23, 2011
Dear Kellen Hill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and lisuing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
### Page 2 – Kellen Hill
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QS) regulation (SF OF to Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parl 801), please If you desire specific advice for your cevree on var acconsideratives/ucm l 1589.html for go to http://www.fda.gov/Aboutr1787Ccmcrs0fffcs) Office of Connyliance. Also, please
the Center for Devices and Radiological Health's (CDRH's) workst natification" (21CFF) P the Center for Devices and Radionogical Health Steans to premarket notification" (21CFR Part note the regulation entitled, "Misolanding of reference to promised in MDR regulation (21 CFR Part 803), please go to
CFR Part 803), please go to Illip://www.ida.2017/01/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11
You may obtain other general information on your responsibilities under the Act from the You may obtain oner general mornation on your responsion.
Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address (800) 038-2041 of (2017) 720 Devices/Resourcesfor You/Industry/default.h1m.
Sincerely yours,
Per
E. Melkerson
Mark N. Melkers Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# K || 2772 (11)
# Indications for Use
510(k) Number (if known):
Device Name: ORTHOLOC™ Bone Screws
Indications For Use:
ORTHOLOC™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mutheal Threis for mxm
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112772
1 of 1
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.