AMSCO V-PRO MAX LOW TEMPERATURE STERILIZATION SYSTEM
K112760 · STERIS Corporation · MLR · Nov 22, 2011 · General Hospital
Device Facts
Record ID
K112760
Device Name
AMSCO V-PRO MAX LOW TEMPERATURE STERILIZATION SYSTEM
Applicant
STERIS Corporation
Product Code
MLR · General Hospital
Decision Date
Nov 22, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Amsco® V-PRO™ maX Low Temperature Sterilization System, with VAPROX™ HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The three pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.
Device Story
Amsco V-PRO maX is a vaporized hydrogen peroxide (VHP) sterilizer for heat/moisture-sensitive reusable medical instruments. Device utilizes VAPROX HC Sterilant; operates via three pre-programmed cycles: Lumen, Non Lumen, and Flexible. System performs conditioning, sterilization, and aeration phases at low pressure/temperature. Used in healthcare facilities by trained staff. Output is sterile, ready-to-use instruments. Benefits include rapid processing without cool-down/aeration periods. Software modifications enhance usability and address customer requests.
Clinical Evidence
Bench testing only. Validation studies performed using specific load configurations (e.g., 19.65 lbs for Lumen/Non-Lumen cycles; 24.0 lbs for Flexible cycle). Testing confirmed sterilization efficacy for specified lumen dimensions and load types using biological indicators (Verify V24 SCBI) and process indicators.
Technological Characteristics
Vaporized hydrogen peroxide sterilizer; uses VAPROX HC Sterilant. Operates at low pressure/temperature. Features three cycles: Lumen, Non Lumen, Flexible. Chamber/vaporizer temperature 50/60°C. Software-controlled cycles with automated conditioning, sterilization, and aeration phases.
Indications for Use
Indicated for terminal sterilization of cleaned, rinsed, and dried reusable metal and nonmetal medical devices in healthcare facilities. Includes rigid endoscopes, hinged instruments (forceps/scissors), and flexible endoscopes (ENT, Urology, bronchoscopes) with specific lumen diameter/length constraints. Contraindicated for devices not compatible with vaporized hydrogen peroxide or those not cleaned/dried per instructions.
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
Predicate Devices
Amsco® V-PRO™ maX Low Temperature Sterilization System (K102330)
{0}------------------------------------------------
K112760
STERIS®
NOV 222 2011
510(k) Summary For Amsco® V-PRO™ maX Plus Low Temperature Sterilization Systems
t
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact:
Robert Sullivan Senior Director, FDA Regulatory Affairs Tel: 440-392-7695 Fax: 440-357-9198
Submission Date:
September 21, 2011
STERIS Corporation = 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{1}------------------------------------------------
#### 1. Device Name
| Trade Name: | Amsco V-PRO maX Low Temperature Sterilization System |
|------------------------|------------------------------------------------------|
| Common/usual Name: | Vapor Phase Hydrogen Peroxide Sterilizer |
| Classification Name: | Sterilizer, Ethylene Oxide Gas |
| Classification Number: | 21 CFR 880.6860 |
| Product Code: | MLR |
#### Predicate Device 2.
Amsco® V-PROTM maX Low Temperature Sterilization System (K102330)
#### 3. Description of Device
The Amsco V-PRO maX Low Temperature Sterilization System is a new sterilizer model to be added to the Amsco V-PRO family of sterilizers. The V-PRO product line currently consists of the Amsco V-PRO 1 and Amsco V-PRO 1 Plus Sterilizers.
The V-PRO maX Sterilizer has three pre-programmed cycles: the Lumen Cycle (K062297), the Non Lumen Cycle (K083097) and the Flexible Cycle (K102330). The V-PRO maX Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities.
The V-PRO maX Sterilizer uses VAPROX HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Its three preprogrammed cycles all utilize a conditioning phase, a sterilization phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle; no cool down or aeration period is required.
The Flexible Cycle has been designed to sterilize surgical flexible endoscopes (such as those used in ENT, Urology and Surgical care) and bronchoscopes. The flexible endoscopes can be single or dual channeled and they may be processed in a flexible endoscope-only load or in combination with non-lumened medical instruments. The cycle is approximately 35 minutes long.
The Verify V24 SCBI (K073244 and K09051), Verify VH2O2 Process Indicator (K091174), V-PRO Sterilization Trays (K070769), and Vis-U-All Tyvek pouches (K070765, K071087, and K090371) have been validated for use in the V-PRO maX Sterilizer.
{2}------------------------------------------------
#### 4. Intended Use
The Amsco V-PRO maX Low Temperature Sterilization System, with VAPROXTM HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The three pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.
The Amsco V-PRO maX Low Temperature Sterilizer System's Lumen Cycle, cleared under K062297, can sterilize:4
- . Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
- . Medical devices, including rigid endoscopes, with a single stainless steel lumen with:
- an inside diameter of 1 mm or larger and a length of 125 mm or o shorter
- an insider diameter of 2 mm or larger and a length of 250 mm or o shorter
- an inside diameter of 3 mm or larger and a length of 400 mm or o shorter
- a The validation testing for all lumen sizes was conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.
The Amsco V-PRO maX Low Temperature Sterilization System's Non Lumen Cycle, cleared under K083097, can sterilize: 0
Non-lumened instruments including non-lumened rigid endoscopes and non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.
- b The validation studies were conducted using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.
The Amsco V-PRO maX Low Temperature Sterilization System's Flexible Cycle, the subject of this submission, can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical care) and bronchoscopes in either of two load configurations:
- 1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load.6
The flexible endoscopes may contain either:
{3}------------------------------------------------
### STERIS SPECIAL 510(k) PREMARKET NOTIFICATION Modification of K102330 V-PRO maX Low Temperature Sterilization System
- · a single lumen with an inside diameter of 1 mm or larger and a length of 1050 mm or shorter
- · or two lumens with:
- · one lumen having an inside diameter of 1 mm or larger and a length of 998 mm or shorter
- and the other lumen having an inside diameter of 1 mm or . larger and a length of 850 mm or shorter
- c The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
- 2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments including instruments with diffusion-restricted areas such as the hinged portion of forceps or scissors. "
The flexible endoscope can contain either:
- · a single lumen with an inside diameter of 1 mm or larger and a length of 1050 mm or shorter
- or two lumens with:
- · one lumen having an inside diameter of 1 mm or larger and a length of 998 mm or shorter
- . and the other lumen having an inside diameter of 1 mm or larger and a length of 850 mm or shorter
- ರ್ The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total load weight of 24.0 lbs.
#### Description of Safety and Substantial Equivalence ક.
The Amsco V-PRO maX Low Temperature Sterilization System is the same as the predicate devices (K102330). Minor modifications are proposed to the device software that addresses customer requests and enhance device usability. The following performance testing has been completed to ensure substantial equivalence.
| Device<br>Modification | Testing | Acceptance<br>Criteria | Results |
|---------------------------|---------------------|------------------------------------------------------------|---------|
| Software<br>Modifications | Software Validation | Software shall be appropriately verified and<br>validated. | PASS |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is written in a bold, sans-serif font. The words are arranged on a single line and centered horizontally. The text is black against a white background.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 222 2011
Mr. Robert Sullivan Senior Director, FDA Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060
Re: K112760
Trade/Device Name: Amsco® V-PRO™ maX Low Temperature Sterilization System . Regulation Number: 21 CFR 880.6860 Regulation Name: Etylene Oxide Gas Sterilzer Regulatory Class: II Product Code: MLR Dated: October 24, 2011 Received: October 25, 2011
Dear Mr. Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Sullivan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Cintron O. waton
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name:
## Amsco® V-PROTM maX Low Temperature Sterilization System
Indications For Use:
The Amsco® V-PRO™ maX Low Temperature Sterilization System, with VAPROXTM HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The three pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.
The Amsco V-PRO maX Low Temperature Sterilization System's Lumen Cycle, cleared under K062297, can sterilize:4
- . Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
- Medical devices, including rigid endoscopes, with a single stainless . steel lumen with:
- an inside diameter of 1 mm or larger and a length of 125 mm or o shorter
- an insider diameter of 2 mm or larger and a length of 250 mm or 0 shorter
- an inside diameter of 3 mm or larger and a length of 400 mm or o shorter
- a The validation testing for all lumen sizes was conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.
The Amsco V-PRO maX Low Temperature Sterilization System's Non Lumen Cycle, cleared under K083097, can sterilize: 9
Non-lumened instruments including non-lumened rigid endoscopes and non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.
Page 1 of 3
{7}------------------------------------------------
- b The validation studies were conducted using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.
The Amsco V-PRO maX Low Temperature Sterilization System's Flexible Cycle, the subject of this submission, can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of two load configurations:
- Two flexible endoscopes with a light cord (if not integral to 1. endoscope) and mat with no additional load. 6.
The flexible endoscopes may contain either:
- a single lumen with an inside diameter of 1 mm or larger and a . length of 1050 mm or shorter
- or two lumens with: .
- one lumen having an inside diameter of 1 mm or larger and a . length of 998 mm or shorter
- and the other lumen having an inside diameter of 1 mm or . larger and a length of 850 mm or shorter
- C The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
- One flexible endoscope with a light cord (if not integral to endoscope) 2. and mat and additional non-lumened instruments including instruments with diffusion-restricted areas such as the hinged portion of forceps or scissors."
The flexible endoscope can contain either:
- a single lumen with an inside diameter of 1 mm or larger and a . length of 1050 mm or shorter
- or two lumens with: . •
- one lumen having an inside diameter of 1 mm or larger and a . length of 998 mm or shorter
- and the other lumen having an inside diameter of 1 mm or . larger and a length of 850 mm or shorter
- d The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total load weight of 24.0 lbs.
Page 2 of 3
{8}------------------------------------------------
### STERIS SPECIAL 510(k) PREMARKET NOTIFICATION Modification of K102330 V-PRO maX Low Temperature Sterilization System
| Sterilization<br>Cycle | Sterilant<br>injection<br>(g) | # of<br>Injections | Sterilant<br>Exposure<br>Time (min) | Chamber<br>Pressure<br>Prior to<br>Injection<br>(Torr) | Chamber/<br>Vaporizer<br>Temperature<br>(°C) |
|------------------------|-------------------------------|--------------------|-------------------------------------|--------------------------------------------------------|----------------------------------------------|
| Lumen | 2.1 | 4 | 32 | 0.4 | 50/60 |
| Non Lumen | 2.1 | 4 | 12 | 1 | 50/60 |
| Flexible | 2.1 | 4 | 12 | 0.4 | 50/60 |
The parameters for the three V-PRO Cycles are as follows:
Prescription Use _ (Part 21 CFR 801 Subpart D)
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
. . . AND/OR
Over-The-Counter Use _ X ____________________________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 3 of 3
Elizabeth F. Clavie-Williams
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112760
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.