MEDICAL FACETS CORICAL SCREWS, MEDICAL FACETS CANCELLOUS SCREWS, MEDICAL FACETS HEADLESS COMPRESSION SCREWS, MEDICAL
K112727 · Medical Facets · HWC · Jan 13, 2012 · Orthopedic
Device Facts
Record ID
K112727
Device Name
MEDICAL FACETS CORICAL SCREWS, MEDICAL FACETS CANCELLOUS SCREWS, MEDICAL FACETS HEADLESS COMPRESSION SCREWS, MEDICAL
Applicant
Medical Facets
Product Code
HWC · Orthopedic
Decision Date
Jan 13, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medical Facet Bone Fixation Screws and Pins are intended to be used as fixation implants for bone fractures, joint fusion, bone reconstruction, or as guide pins for insertion of other implantable devices.
Device Story
Medical Facets Bone Fixation Screws and Pins are orthopedic implants used for bone fracture fixation, joint fusion, bone reconstruction, or as guide pins for other devices. The product line includes cortical screws, cancellous screws, headless compression screws, external fixation screws, Kirschner wires, guide wires, and Steinmann pins. Devices are manufactured from stainless steel, titanium, or cobalt. They are available in various diameters (1.0mm to 7.3mm) and lengths (6mm to 450mm), with solid or cannulated configurations. These devices are used by surgeons in clinical or surgical settings to stabilize bone segments. The implants provide mechanical fixation to facilitate healing or support reconstruction. Performance is verified through mechanical testing against established ASTM standards.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing performed on worst-case configurations according to ASTM F543, ASTM F1264, ASTM F2193, and ASTM F1541. Results demonstrated substantial performance equivalence to predicate devices.
Indicated for patients requiring bone fracture fixation, joint fusion, bone reconstruction, or as guide pins for the insertion of other implantable devices.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Treu Instrumente GmBH Bone Fixation Screws and Pins (K083912)
{0}------------------------------------------------
长川2727 (113)
JAN 1 3 2012
## 6. 510(k) Summary
| Manufacturer: | Medical Facets NC LLC<br>1669 Federal Ave.<br>Gastonia, NC 28052 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | September 15, 2011 |
| Submitted by: | Medical Facets NC LLC |
| Company Contact | Glenn A. Rupp, President<br>1669 Federal Ave.<br>Gastonia, NC 28052<br>(704) 562-7494<br>(704) 853-2105 (FAX) |
| US Agent Information | Orgenix LLC<br>Mr. Donald W. Guthner<br>111 Hill Road<br>Douglassville, PA 19518<br>+1-646-460-2984<br>+1-484-363-5879 (FAX) |
| Classification Name: | Screw, Fixation Bone; Pin, Fixation Smooth or Threaded |
| Common/Usual Name: | Bone Fixation Screws and Pins |
| Proprietary Name: | Medical Facets Cortical Screws<br>Medical Facets Cancellous Screws<br>Medical Facets Headless Compression Screws<br>Medical Facets External Fixation Screws<br>Medical Facets Pins and Guide Wires |
| Performance standards: | The Medical Facets Cortical Screws, Medical Facets Cancellous<br>Screws. Medical Facets Headless Compression Screws, Medical<br>Facets External Fixation Screws, and Medical Facets Pins and<br>Guide Wires were mechanically tested to the following standards -<br>ASTM F543, ASTM 1264, ASTM F2193 and ASTM F1541<br>(current versions). |
| Classification no .: | 21CFR 888.3040<br>HWC, JDW, HTY<br>Class II<br>87/Orthopedic Panel |
{1}------------------------------------------------
Substantial equivalence for Medical Facets Cortical Screws, Substantial Equivalence: Medical Facets Cancellous Screws, Medical Facets Headless Compression Screws, Medical Facets External Fixation Screws, and Medical Facets Pins and Guide Wires is based on the similarities in indications for use, design features, operational principles and material composition when compared to the predicate devices cleared under the following submissions:
- K083912 Treu Instrumente GmBH Bone Fixation Screws and . Pins
- . K090047 Synthes 1.5 Mini Fragment LCP System
- K911505 Simpex, Inc. Fixation Wire .
- K051605 SBI K-wires .
cleared devices.
Predicate Devices:
Device Description:
Medical Facets Bone Screws are available in thread diameters ranging from 1.5mm to 7.3mm, lengths ranging from 6mm to 180mm, and either solid or cannulated, and made of stainless steel or titanium.
The subject device is substantially equivalent to similar previously
Medical Facets External Fixation Screws are available in thread diameters ranging from 4mm to 6mm, lengths ranging from 60mm to 250mm, and thread lengths ranging from 18mm to 80mm, and made of stainless steel or titanium.
Medical Facets Kirschner Wires are available in wire diameters ranging from 1.0mm to 2.5mm, in lengths ranging from 100mm to 150mm. 0.028 inches to 0.062 inches, in lengths ranging from 4.0 inches to 12.0 inches and are partially or completely threaded. or smooth. Made of stainless steel or titanium.
Medical Facets Guide Wires are available in wire diameters ranging from 0.8mm to 2.8mm, in lengths ranging from 100mm to 450mm, are partially threaded or smooth. Made of stainless steel, titanium, or cobalt.
Medical Facets Steinmann Pins are available in wire diameters ranging from 0.078 inches to 0.187 inches, in lengths ranging from 9.0 inches to 12.0 inches, are completely threaded, or smooth. Made of stainless steel or titanium.
Intended Use:
The Medical Facet Bone Fixation Screws and Pins are intended to be used as fixation implants for bone fractures, joint fusion, bone reconstruction, or as guide pins for insertion of other implantable devices.
{2}------------------------------------------------
Summary of The Medical Facet Bone Fixation Screws and Pins are Technological manufactured in Stainless Steel, Titanium and Cobalt (guide wires Characteristics only). The Medical Facet Bone Fixation Screws and Pins were tested Non-Clinical Testing (worse case) according to the following standards: ASTM F543 - Standard Specification and Test Methods for Metallic Medical Bone Screws ASTM F2193 - Standard Specifications and Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal System ASTM F1264 - Standard Specification and Test Methods for Intramedullary Fixation Devices ASTM F1541 - Standard Specification and Test Methods for External Skeletal Fixation Devices Test results indicated substantial performance of the worse case in side-by-side testing with the predicate device. Conclusion The information discussed above demonstrates that the Medical Facet Bone Fixation Screws and Pins are effective and perform as well as or better than the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three overlapping, curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medical Facets NC LLC % Mr. Glenn A. Rupp President 1669 Federal Avenue Gastonia, North Carolina 28052
JAN 1 3 2012
Re: K112727
Trade/Device Name: Medical Facets Cortical Screws Medical Facets Cancellous Screws Medical Facets Headless Compression Screws Medical Facets External Fixation Screws Medical Facets Pins and Guide Wires
Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastner Regulatory Class: II Product Code: HWC, JDW, HTY Dated: December 15, 2011 Received: December 16, 2011
Dear Mr. Rupp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
## Page 2 - Mr. Glenn A. Rupp
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Director
Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 5. Indications for Use
510(k) Number (if known):
Medical Facets Cortical Screws Device Name: Medical Facets Cancellous Screws Medical Facets Headless Compression Screws Medical Facets External Fixation Screws Medical Facets Pins and Guide Wires
The Medical Facet Bone Fixation Screws and Pins are intended to be used as fixation implants for bone fractures, joint fusion, bone reconstruction, or as guide pins for insertion of other implantable devices
2 Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrenee of CDRH, Office of Device Evaluation (ODE)
for (Division Sign-Off)
sion Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112727
ATT. 1-21
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.