NON-STERILE, POWER FREE YELLOW LATEX EXAMINATION GLOVES

K112693 · Sgmp Co., Ltd. · LYY · Sep 12, 2012 · General Hospital

Device Facts

Record IDK112693
Device NameNON-STERILE, POWER FREE YELLOW LATEX EXAMINATION GLOVES
ApplicantSgmp Co., Ltd.
Product CodeLYY · General Hospital
Decision DateSep 12, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Non-Sterile, Powder-Free Yellow Latex Examination Glove With Tutti Flavour and With Protein Labeling Claim of <50 ug/dm² per Glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable, non-sterile, powder-free yellow latex examination glove; features Tutti Frutti flavor; protein content labeled <50 ug/dm². Used in clinical settings by healthcare professionals to prevent cross-contamination between patient and examiner during examinations. Device acts as a physical barrier.

Clinical Evidence

Bench testing only.

Technological Characteristics

Material: Latex. Form: Powder-free, yellow, flavored (Tutti Frutti). Protein content: <50 ug/dm². Non-sterile.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SGMP Company, Limited C/O Ms. Janna P. Tucker Tucker & Associates 198 Avenue De La D'Emerald Sparks, Nevada 89434 SEP 12 2012 Re: K112693 Trade/Device Name: Non-Sterile, Powder-Free Yellow Latex Examination Gloves with Tutti Frutti Flavour and with Protein Labeling Claim of <50 ug/dm- Per Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: August 19, 2012 Received: August 21, 2012 Dear Ms. Tucker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## Page 2- Ms. Tucker Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing. practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. R. O'Shell windett Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## INDICATIONS FOR USE ## SGMP COMPANY LIMITED APPLICANT: ## K112693 510(K) NUMBER: Non-Sterile; Powder-Free Yellow Latex Examination Gloves With Tutti DEVICE NAME: Frutti Flavour and with Protein Labeling Claim of <50 ug/dm² Per Glove INDICATIONS FOR USE: The Non-Sterile, Powder-Free Yellow Latex Examination Glove With Tutti Flavour and With Protein Labeling Claim of <50 ug/dm² per Glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801.Subpart D and/or Over-The-Counter 21 CFR 801.Subpart C Concurrence of CDRH, Office of Device Evaluation (ODE) Ralió banguhim Division of Anesthesiology, General Hospitai Infection Control, Dental Devices 510(k) Number: K112643
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...