SIMPLANT NAVIGATOR PERSONALIZED DENTAL CARE SYSTEM
Applicant
Materialise Dental NV
Product Code
LLZ · Radiology
Decision Date
Feb 22, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
SimPlant Navigator Personalized Dental Care System is intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner such as a CT scanner or a Magnetic Resonance scanner. It is also intended as pre-planning software for dental implant placement and surgical treatment. SurgiGuide® guides and the BIOMET 3i Navigator Surgical Kit, which are used intra-operatively to prepare the osteotomy for placement of BIOMET 3i implants pre-operatively determined in the software.
Device Story
SimPlant Navigator is a software-based system for dental implant pre-operative planning. It imports CT or MRI scan data to create 3D anatomical models of the jaw. Clinicians use the software to perform image segmentation, visualize anatomical structures, and plan dental implant positions and orientations. The system includes a library of dental implants and instruments. Output includes a surgical plan and the design for patient-specific SurgiGuide® templates. These guides, along with the BIOMET 3i Navigator Surgical Kit, are used by surgeons intra-operatively to transfer the virtual plan to the patient by guiding drill locations and osteotomy preparation. The device assists in the placement of implants to support temporary or final prostheses. It is intended for use by qualified surgeons in a clinical setting.
Clinical Evidence
No clinical data provided. Performance was established through software validation and bench-top testing to ensure compatibility of system components, including biocompatibility, sterilization, and dimensional stability testing.
Technological Characteristics
Software-based image processing and planning system. Hardware components include polyamide patient-specific guides and stainless steel surgical instruments. Connectivity via DICOM standards. Software developed in C++ for Windows. Complies with ISO 14971 (risk management), ISO 13485 (quality management), and ISO 10993 (biocompatibility).
Indications for Use
Indicated for use as a software interface and image segmentation system for transferring medical imaging (CT/MRI) and as pre-planning software for dental implant placement and surgical treatment. Used with SurgiGuide® and BIOMET 3i Navigator Surgical Kit for intra-operative osteotomy preparation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 892.2050.
| Submitter: | Materialise Dental NV<br>Technologielaan 15<br>Leuven<br>Belgium |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Reg. Number: | 3006638827 |
| Contact: | Carl Van Lierde<br>QA/RA Manager<br>Materialise Dental NV<br>Technologielaan 15<br>Leuven<br>Belgium<br>Tel. +32 16 39 67 14<br>Fax. +32 16 39 66 22<br>Email: carl.vanlierde@materialise.be |
| Date Prepared: | September 12, 2011 |
| Trade/<br>Proprietary<br>Name: | SimPlant Navigator Personalized Dental Care System |
| Common/Usual<br>Name: | System, Image processing. The product uses images acquired from Computerized<br>Tomography (CT) scanners |
| Classification<br>Name/ FDA<br>Reviewing<br>Branch: | Radiology branch |
| Device<br>Classification/<br>Code: | Class II - 21 CFR §892.2050 LLZ |
{1}------------------------------------------------
K112679
ge 2 of 7.
K\\2679 | K\12679 | K\12679
Special 510(k) Premarket Notification | Rape 2 of 2 of 2 of 2 | 2 of 21 | Plant Care System
T
| Predicate Device<br>Manufacturer: | SimPlant® 2011; (K110300)<br>Nobel Biocare Guided Surgery Concept; (K050393) |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Purpose of the<br>SPECIAL 510(k)<br>notice: | The reason for this Special 510k submission is to request clearance for a modificatio<br>to a device that has been cleared under the 510(k) process referred to herein a<br>SimPlant Navigator Personalized Dental Care System (Image processing system<br>referenced under 21 CFR §892.2050 and are considered Class II devices. |
| Device<br>Description: | SimPlant Navigator Personalized Dental Care System provides a method o<br>importing medical imaging information from radiological imaging systems such as<br>Computer Tomography (CT) or Magnetic Resonance Imaging (MRI) to a computer fi<br>that is usable in conjunction with other diagnostic tools and expert clinical judgmen<br>Visual representations of the imaged anatomical structures (e.g. the jaw) are derive<br>allowing for a three-dimensional assessment of the patient without patient contac<br>Dental implant positions including orientations are planned pre-operatively. Compute<br>visualization of the 3D anatomical jaw models, planned implants, planned tooth setu<br>and numerical measurements assist the surgeon in the creation and approval of a pre<br>surgical plan. SurgiGuide® guides and the BIOMET 3i Navigator Surgical Kit are use<br>intra-operatively to prepare the osteotomy for placement of BIOMET 3i implants pre<br>operatively determined in the software. |
| Indications<br>for<br>Use: | SimPlant Navigator Personalized Dental Care System is intended for use as a softwar<br>interface and image segmentation system for the transfer of imaging information from<br>a medical scanner such as a CT scanner or a Magnetic Resonance scanner. It is als<br>intended as pre-planning software for dental implant placement and surgical treatment.<br>SurgiGuide® guides and the BIOMET 3i Navigator Surgical Kit, which are used intra<br>operatively to prepare the osteotomy for placement of BIOMET 3i implants pre<br>operatively determined in the software. |
| Technological<br>Characteristics: | The predicate devices, SimPlant® 2011 and Nobel Biocare Guided Surgery Conce<br>have a number of very similar and equivalent design /technological characteristics:<br>which are very similar and equivalent with the SimPlant Navigator Personalized Dent<br>Care System (see Substantial Equivalence comparison table in Section 14). |
| Performance<br>Data: | Software Validation in addition to bench top performance testing was conducted t<br>ensure the compatibility of all system components. |
| Clinical Data: | N/A |
{2}------------------------------------------------
| Performance<br>Standards: | a. DICOM NEMA PS3.1-3.18: Digital imaging and communication in medicine:<br>2009<br>b. ISO14971: Applications of risk management to medical devices: 2007<br>c. ISO13485: Medical Devices Quality Management System: 2003<br>d. ISO9001: Quality Management System: 2008<br>e. ISO10993: Biological evaluations of medical devices: 1992 | | Intended Use | | | | | | | | | | | | | | | | | | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|---------------------------------------------------------|--|--|
| Substantial<br>Equivalence: | Materialise Dental NV's SimPlant Navigator Personalized Dental Care System<br>included in this submission have the same intended use, fundamental scientific<br>technology, similar indications, and principles of operation as the previously cleare<br>SimPlant® 2011; (K110300).<br>Refer to the following substantial equivalence data table: | | The SimPlant Navigator<br>Navigator Personalized Dental<br>Care System is intended for<br>use as a software interface<br>and image segmentation<br>system for the transfer of<br>imaging information from a<br>medical scanner such as a CT<br>scanner or a Magnetic<br>Resonance scanner. It is also<br>intended as pre-planning<br>software for dental implant<br>placement and surgical<br>treatment.<br><br>The SimPlant Navigator<br>Personalized Dental Care<br>System can be used with the<br>following Biomet 3i<br>instrument kits and their<br>respective components:<br>implant mounts, cortical<br>perforator, tissue punches,<br>drill positioning handles, twist<br>drills, countersink drills,<br>shaping drills, bone taps, bone<br>profilers, drivers, and ratchets. | Materialise Dental's<br>SimPlant® 2011 software is<br>intended for use as a<br>software interface and<br>image segmentation system<br>for the transfer of imaging<br>information from a medical<br>scanner such as a CT scanner<br>or a Magnetic Resonance<br>scanner. It is also intended<br>as pre-planning software for<br>dental implant placement<br>and surgical treatment. | The Guided Surgery<br>Concept and Teeth-in-<br>an-Hour are indicated<br>for the treatment of<br>single, partially and<br>totally edentulous jaws<br>for placement of implant<br>fixtures with immediate<br>function to restore<br>patient esthetics and<br>chewing function. The<br>following prerequisites<br>must be fulfilled:<br>- Adequate amount of<br>jaw bone<br>- The quality of jaw bone<br>must be judged as<br>adequate | | | | | | | | | | | | | | | | |
| | Device for<br>premarket<br>notification K110300 K050393 Trade<br>name SimPlant® Navigator<br>Personalized Dental Care<br>System SimPlant® 2011 Nobel Biocare The<br>Guided Surgery Concept Common<br>name SimPlant® Software Classification Product Code: LLZ<br>21 CFR. § 892.2050<br>Classification Panel: Radiology<br>Device Class: II Product code: LLZ<br>21 CFR. § 892.2050<br>Classification Panel:<br>Radiology<br>Device Class: II Product code: DZE<br>21 CFR. § 872.3640<br>Classification Panel:<br>Dental<br>Device Class: II | | | | | | | | | | | | | | | | | | SurgiGuide® guides are<br>intended for single use only. | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | Software – magnetic media | Software – Magnetic media | Software – Magnetic<br>media | | | | | | | | | | | | | | | | | | |
| Material | Hardware –<br>polyamide guides –<br>biocompatible<br>material stainless steel<br>tubes/sleeves<br>medical grade | | Hardware<br>Polyamide<br>guides<br>biocompatible<br>material | | | | | | | | | | | | | | | | | | |
ﻟﻢ
{3}------------------------------------------------
Special 510(k) Premarket Notification -The SimPlant® Navigator Personalized Dental Care System
K112679 Page র্য
.
{4}------------------------------------------------
K\267-9.
Page 5 of 7
Special S10(k) Premarket Notification – Page 5 – F
Special S10(k) Premarket Notification – The SimPlant® Navigator Personalized Dental Care System
| | | Software for use in pre-<br>operative planning. | Software for use in pre-<br>operative planning. | Software for use in pre-<br>operative planning. |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Design | The SimPlant Navigator<br>Personalized Dental Care<br>System includes SimPlant®<br>software, which provides a<br>means for the clinician for<br>image segmentation and<br>advanced pre-operative<br>planning. This enables the<br>clinician to view three-<br>dimensional CT-scan data as<br>well as to plan the case in a<br>virtual three-dimensional<br>environment. | SimPlant® software provides<br>a means for image<br>segmentation and advanced<br>pre-operative planning.<br>Surgical templates may be<br>fabricated based on the<br>output of the pre-operative<br>planning. | The Guided Surgery<br>Concept includes a 3D-<br>Planning Software that<br>enables the clinician to<br>view three-dimensional<br>CT-scan data as well as<br>to plan the case in a<br>virtual three-<br>dimensional<br>environment. |
| | | The case planning can be<br>used to produce a Surgical<br>Template, thus transferring<br>the virtual case planning into<br>physical tools enabling the<br>surgical installation according<br>to the virtual case planning. | | This case planning can<br>be used to produce a<br>Surgical Template, thus<br>transferring the virtual<br>case planning into<br>physical tools enabling<br>the surgical installation<br>according to the virtual<br>case planning. |
| | | The SimPlant Navigator<br>Personalized Dental Care<br>System is based upon<br>knowledge of the location<br>and orientation of the<br>implant(s) prior to the<br>surgery. This knowledge<br>enables the production of a<br>SurgiGuide. | | The Guided Surgery<br>Concept is based upon<br>knowledge of the<br>location and orientation<br>of the implant(s) prior<br>to the surgery. This<br>knowledge enables the<br>production of a Surgical<br>Template. |
| | | Aided by the SurgiGuide, the<br>sites can be prepared and the<br>implants placed in the<br>predetermined locations<br>enabling the immediate<br>attachment of the<br>prefabricated temporary or<br>final prosthesis. | | Aided by the Surgical<br>Template, the sites can<br>be prepared and the<br>implants placed in the<br>predetermined locations<br>enabling the immediate<br>attachment of the<br>prefabricated temporary<br>or final prosthesis. |
| | | | | |
| | The SimPlant® software<br>component is used to<br>incorporate the images from<br>either an MRI or CT scan of<br>the affected joint into the<br>specialized planning<br>software. | SimPlant® software is used to<br>incorporate the images from<br>either an MRI or CT scan of<br>the affected joint into the<br>specialized planning software. | | |
| | The SimPlant® software is<br>used by a qualified surgeon<br>to plan, inspect, fine-tune<br>and approve the pre-surgical<br>plan. The software is used<br>pre-operatively. | The SimPlant® software is<br>used by a qualified surgeon to<br>plan, inspect, fine-tune and<br>approve the pre-surgical plan.<br>The software is used pre-<br>operatively. | | |
| Function | SimPlant® software contains<br>a library of dental implants,<br>and additional instruments<br>for the placement of<br>implants. | SimPlant® software contains<br>a library of dental implants,<br>and additional instruments for<br>the placement of implants. | | |
| | A SurgiGuide® guide may be<br>designed and fabricated<br>based on the output of the<br>pre-operative planning. | A SurgiGuide® guide may be<br>designed and fabricated<br>based on the output of the<br>pre-operative planning. | | |
| | SurgiGuide® guides are<br>patient specific templates<br>that are intended to transfer<br>the pre-operatively<br>determined positioning of<br>the dental implants to the<br>patient intra-operatively,<br>assisting the surgeon in<br>placing dental implants by<br>guiding and marking drill<br>locations. | | | |
| Programming<br>language | C++ | C++ | | |
| Operating<br>system | Windows | Windows | | |
| | | | | |
{5}------------------------------------------------
Special 510(k) Premarket Notification --The SimPlant® Navigator Personalized Dental Care System
Page
X112679
87
{6}------------------------------------------------
| Testing | Hardware testing:<br>• Biocompatibility<br>testing of patient<br>contacting<br>components<br>• Sterilization testing<br>• Sterilization<br>dimensional<br>stability test | |
|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Software testing:<br>• Unit testing<br>• Integration testing<br>• IR testing<br>• Smoke testing<br>• Formal testing<br>• Acceptance testing<br>• Alpha testing<br>• Beta testing | Software testing:<br>• Unit testing<br>• Integration testing<br>• IR testing<br>• Smoke testing<br>• Formal testing<br>• Acceptance testing<br>• Alpha testing<br>• Beta testing |
| Conclusion: | SimPlant Navigator Personalized Dental Care System and its predicate device:<br>SimPlant® 2011 (K110300) and the Nobel Biocare Guided Surgery Concep<br>(K050393), have the same intended use, indications for use, similar technologica<br>characteristics, and principles of operation.<br>SurgiGuide® guides and the BIOMET 3i Navigator® Surgical Kit, which are used intra<br>operatively to prepare the osteotomy for placement of BIOMET 3i implants pre<br>operatively determined in the software.<br>By modifying the SimPlant® 2011 to include the SurgiGuide® and Navigation Surgica<br>Kit does not raise any new questions of safety or effectiveness. | |
| | The differences noted above do not present new issues of safety or effectiveness for th<br>SimPlant® Navigator Personalized Dental Care System. | |
K112679
, 7 of 7
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its body and wings.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Carl Van Lierde QA/RA Manager Materialise Dental NV Technologielaan 15 3001 LEUVEN BELGIUM
FEB 2 2 2012
Re: K112679
Trade/Device Name: SimPlant® Personalized Dental Care System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ, NOF, NDP, and DZA Dated: January 10, 2012 Received: February 16, 2012
### Dear Mr. Van Lierde:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, merelove, manel of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such 11 your device is elassified to regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reads of ac now has a determination that your device complies with other requirements of the Act that I Dr has intact a and regulations administered by other Federal agencies. You must or uny 1 eather the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{8}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice medical do rece forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as set form in the quarty, adevice as described in your Section 510(k) premarket win anow you to begin marial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 You don't openite and the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Special 510(k) Premarket Notification --The SimPlant® Navigator Personalized Dental Care System
# Indications for Use
510(k) Number (if known): K // Ze 79
Device Name: SimPlant® Personalized Dental Care System
Indications for Use:
SimPlant Navigator Personalized Dental Care System is intended for use as a software Simirralit Navigator Tersonation system for the transfer of imaging information from a medical meriate and mage segmentation by Magnetic Resonance scanner. It is also intended as prebearing software for dental implant placement and surgical treatment.
SurgiGuide® guides and the BIOMET 3i Navigator Surgical Kit, which are used intra-Surgioure guides and the Droming of BIOMET 31 implants pre-operatively determined in the software.
Prescription Use
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
AND/OR
(Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510K. | K112679 |
|-------|---------|
|-------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.