K112636 · B.Braun Medical, Inc. · FPA · Nov 26, 2011 · General Hospital
Device Facts
Record ID
K112636
Device Name
CYTOGUARD(TM) CLOSED LUER CONNECTOR
Applicant
B.Braun Medical, Inc.
Product Code
FPA · General Hospital
Decision Date
Nov 26, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The CytoGuard™ Closed Luer Connector is a bi-directional flow, needle-free, swabable luer access device used as an accessory to an IV Administration Set. It is intended for the administration of parenteral fluids, medication (including hazardous drugs), blood and blood products. With the CytoGuard attached to the standard male luer of a dosing device or container, the CytoGuard valve is in a closed position. Upon connection of its male luer fitting to the female luer fitting of a patient's administration set or catheter, the valve opens to allow medication delivery to the patient. The fluid path then closes, upon disconnect.
Device Story
CytoGuard is a needle-free, swabable, bi-directional luer connector; used by healthcare professionals during medication preparation, admixture, fluid transfer, administration, or disposal. Device attaches to dosing device male luer; provides closed fluid path via internal valve. When connected to patient administration set/catheter, valve opens for fluid flow; disconnect closes valve, preventing flow. Features spinning locking mechanism on female luer to prevent accidental disconnection; translucent orange male luer nut for visual identification. MRI safe; PVC, DEHP, and latex-free. Reduces inadvertent fluid exposure to clinicians and patients. Used in clinical settings for parenteral therapy.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing, including functional performance testing, evaluation of the locking feature, and a simulated clinical use study. Compatibility testing was conducted with various fluids, including blood, general IV fluids, and hazardous chemicals identified in the NIOSH Alert 2004-165.
Indicated for administration of parenteral fluids, medications (including hazardous drugs), blood, and blood products via IV administration sets in patients requiring fluid or medication therapy.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
NOV 2 6 2011
K112636
B. Braun Medical Inc. 510(k) Premarket Notification CytoGuardTM Closed Luer Connector
#### 5. 510(k) SUMMARY
DATE:
September 7, 2011
SUBMITTER:
B. Braun Medical Inc. 901 Marcon Boulevard Allentown, PA 18109-9341 610-266-0500
Contact: Angela J. Caravella, Regulatory Affairs Specialist Phone: (610) 596-2966 Fax: (610) 266-4962 E-mail: angela.caravella@bbraun.com
DEVICE NAME:
CytoGuard™ Closed Luer Connector
COMMON OR USUAL NAME:
Closed Luer Connector, Needle free
# DEVICE
Class II, per 21 CFR §880.5440, Product Code: FPA CLASSIFICATION:
PREDICATE DEVICES:
Spiros™, ICU Medical Inc., K070532, Class II, FPA, 880.5440. Spinning Spiros™, ICU Medical Inc., K082806, Class II, FPA, 880.5440, CML TM, ICU Medical Inc., K051437, Class II FPA, 880.5440
DESCRIPTION:
CytoGuard is a needle-free, bi-directional flow, luer connector that is swabable, MRI Safe and does not contain PVC, DEHP, or natural rubber latex. The CytoGuard™ Closed Luer Connector is designed to be used by healthcare professionals during medication preparation, admixture, fluid transfer, administration or disposal. Using CytoGuard's female luer end CytoGuard is attached to a dosing device's male luer connector, creating a luer locking connection. With the CytoGuard attached, the dosing device is provided with a closed fluid path due to the presence of the CytoGuard's internal valve. The closed valve helps to reduce inadvertent fluid exposure to patients, clinicians and healthcare personnel.
When the CytoGuard's male luer nut is connected to a standard female luer, the fluid path is opened and allows for the movement of fluid. When the female luer is disconnected from the proposed device, the valve and fluid path closes, preventing the flow of fluid through the CytoGuard.
24
{1}------------------------------------------------
SUBSTANTIAL
EQUIVALENCE:
K112636
ge 2 of 3
The female luer end of the CytoGuard possesses a spinning locking feature which will not allow the device to be easily removed, further aiding the reduction of fluid exposure. This feature helps to reduce the risk that the female luer end may inadvertently become disconnected or be removed from the male luer of the dosing device.
To further assist in the appropriate use of the device, the male luer nut has been colored a translucent orange to alert users which end of the device that connections and disconnections can be made. The Instructions for Use references an "orange end" to aid in the easy identification of this part of the device.
The CytoGuard will be available for sale as an individually packaged item and a non luer-activating dust cover cap to prevent touch contamination is provided as a separately sold accessory for the CytoGuard for use during transport.
INTENDED USE: The CytoGuard™ Closed Luer Connector is a bi-directional flow, needle-free, swabable luer access device used as an accessory to an administration device. It is intended for the administration of parenteral fluids, medication (including hazardous drugs), blood and blood products. With the CytoGuard attached to the standard male luer of a dosing device or container, the CytoGuard valve is in a closed position. Upon connection of its male luer fitting to the female luer fitting of a patient's administration set or catheter, the valve opens to allow medication delivery to the patient. The fluid path then closes, upon disconnect.
> Three predicate devices are utilized for substantial equivalence. ICU Medical Inc.'s Spiros™ (K070532), Spinning Spiros™ (K082806) and CML™ (K051437).
> The Cytoguard Closed Luer Connector has similar intended uses to ICU Medical's Spiros, Spinning Spiros, and CML in that they each act as a closed male luer that provides access for the administration of fluids from a container (or dosing device) to the patient. Each of these devices has a normally closed design that is intended to prevent the leakage of fluid or ingress of air when in an inactivated state. When activated, they all provide a 2-way fluid flow. The CytoGuard and Spiros devices contain polycarbonate female luers.
{2}------------------------------------------------
Similar to the Spinning Spiros, the CytoGuard has a locking feature on the female luer end to prevent the device's disconnection during use.
1-112636
Each of these devices has an open and closed state; however, the mechanism used to achieve it differs between the proposed and predicate devices. Therefore, performance testing has been conducted. In addition, testing was performed to support an expanded variety of usage conditions of the CytoGuard. Several different fluids were evaluated for compatibility, including blood, general IV fluids, and chemicals identified on the "NIOSH Alert" drugs list, publication number 2004-165 dated September 2004 which is provided within this submission.
CONCLUSION: Testing was conducted to demonstrate the performance of the CytoGuard Closed Luer Connector and substantial equivalence with the predicate devices: ICU Medical's Spinning Spiros™, Spiros™ and CMLTM. Functional performance testing, a simulated clinical use study, and testing of the locking feature and colored design features of the proposed device, have each demonstrated that the CytoGuard Closed Luer Connector does not raise new issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white and has a simple, clean design. The eagle is facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002
Ms. Angela J. Caravella Regulatory Affairs Specialist B. Braun Medical, Incorporated 901 Marcon Boulevard Allentown, Pennsylvania 18109-9341
NOV 2 6 2011
Re: K112636
Trade/Device Name: CytoGuard™ Closed Luer Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: September 7, 2011 Received: September 9, 2011
Dear Ms. Caravella
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2- Ms. Caravella
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reportung (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Anatra
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{5}------------------------------------------------
## 4. INDICATIONS FOR USE STATEMENT
Page ________________of
K112636 510(k) Number (if known):
Device Name:
CytoGuard™ Closed Luer Connector
Indications For Use:
The CytoGuard™ Closed Luer Connector is a bi-directional flow, needle-free, swabable luer access device used as an accessory to an IV Administration Set. It is intended for the administration of parenteral fluids, medication (including hazardous drugs), blood and blood products. With the CytoGuard attached to the standard male luer of a dosing device or container, the CytoGuard valve is in a closed position. Upon connection of its male luer fitting to the female luer fitting of a patient's administration set or catheter, the valve opens to allow medication delivery to the patient. The fluid path then closes, upon disconnect.
| Prescription Use | X |
|----------------------|---|
| (Per 21 CFR 801.109) | |
| OR | |
| Over-The-Counter Use | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard C. Chapman 11/28/4
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K 112636
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.