STYKER CONSOLIDATED OPERATING ROOM EQUIPMENT (CORE) SYSTEM
Applicant
Stryker Corporation
Product Code
ERL · Ear, Nose, Throat
Decision Date
May 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker ® Consolidated Operating Room Equipment (CORE) System is intended for use in the cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement and teeth in a variety of surgical procedures, including but not limited to, dental, ENT (ear, nose, throat), neuro, spine, and endoscopic applications. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
Device Story
Stryker CORE System is an electrically powered surgical console used in operating rooms by surgeons. It drives various handpieces—saws, drills, wire drivers, microdebriders, bone mills—for bone/tooth manipulation and fixation device placement. The console features three handpiece connectors (supporting two simultaneous handpieces) and two footswitch connectors (supporting wired or wireless footswitches). It provides 400W power output. Surgeons interact with the system via an LCD screen to set operating parameters, which are also automatically identified from attached accessories using wireless tag technology. The system allows for torque control to assist in screw implantation. By providing precise power and control to surgical instruments, the device facilitates bone and tissue modification, aiding in orthopedic, dental, and neurosurgical procedures.
Clinical Evidence
Bench testing only. No clinical data was required or provided. Safety and effectiveness were established through software code review, hazard analysis, and verification/validation testing of software revisions, confirming the device performs as well as or better than the predicate devices.
Technological Characteristics
Electrically powered console; 400W power output. Features three handpiece connectors, two footswitch connectors (wired/wireless support), and LCD interface for parameter control. Utilizes wireless tag technology for accessory identification. Software-controlled torque profiles. Non-patient contact device; biocompatibility evaluation not required per ISO 10993-1:2003.
Indications for Use
Indicated for cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement, and teeth in dental, ENT, neuro, spine, and endoscopic surgical procedures; also for placement or cutting of screws, metal, wires, pins, and other fixation devices. No specific age or gender contraindications provided.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
Predicate Devices
Stryker® Consolidated Operating Room Equipment (CORE) System (K040300, K040369, K032303)
{0}------------------------------------------------
MAY - 1 2012
stryker
·
**
4100 E. Milham Avenue
Kalamazoo, MI 49001
t: 269 323 7700 f: 269 389 5412
www.stryker.com
#### Instruments ·
# 510(k) Summary
| 510(k) Owner: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI 49001<br>(p) 269-323-7700<br>(f) 269-389-5412 | | | | |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|--------------------|-------|--------------------|
| Contact Person: | Michelle Jump | | | | |
| Registration No.: | 1811755 | | | | |
| Trade Name: | Stryker® Consolidated Operating Room Equipment (CORE) System | | | | |
| Common Name: | Console | | | | |
| Classification Name and Regulation Number | Primary Product Code | | | | |
| | PRODUCT CODE | DEVICE | REGULATION NUMBER | CLASS | REVIEW PANEL |
| | ERL | Drill surgical ent<br>(electric or pneumatic)<br>including handpiece | 21 CFR<br>874.4250 | II | EAR NOSE<br>THROAT |
| | Secondary Product Codes | | | | |
| | PRODUCT CODE | DEVICE | REGULATION NUMBER | CLASS | REVIEW PANEL |
| | DZI | Drill, bone, powered | 21 CFR<br>872.4120 | II | DENTAL |
| | DZJ | Driver, wire, and bone<br>drill, manual | 21 CFR<br>872.4120 | II | DENTAL |
| | HBE | Drills, burs, trephines<br>& accessories (simple<br>powered) | 21 CFR<br>882.4310 | II | NEUROLOGY |
| Predicate Devices: | Stryker® Consolidated Operating Room Equipment (CORE) System<br>(K040300, K040369, K032303); Synthes Electric Pen Drive (EPD) System (K043310);<br>and Stryker® Total Performance System (K032117, K942956, K943589, K943569,<br>K943540) | | | | |
| Device Description: | The Stryker Consolidated Operating Room Equipment (CORE) System is an<br>electrically powered system. This system includes a console, irrigation accessories,<br>footswitches, handswitches, and handpiece cords which drive a variety of devices | | | | |
{1}------------------------------------------------
| including saws, drills, wire drivers, microdebriders, and bone mills. | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | The Stryker ® Consolidated Operating Room Equipment (CORE) System is intended<br>for use in the cutting, drilling, reaming, decorticating, shaping, and smoothing of bone,<br>bone cement and teeth in a variety of surgical procedures, including but not limited to<br>dental, ENT (ear, nose, throat), neuro, spine, and endoscopic applications. It is also<br>usable in the placement or cutting of screws, metal, wires, pins, and other fixation<br>devices. | | | | |
| Testing | The clarification of the Stryker® Consolidated Operating Room Equipment (CORE)<br>System indications for use statement is not the result of any design changes that have<br>been made to the console. In addition, the use of this device in orthopedic and spine<br>applications do not present any new risks that would require testing. Therefore, no<br>additional performance testing was performed to evaluate this change. | | | | |
| Each software revision of the Stryker® CORE Console has undergone a code review<br>by a team of internal experts to review source code, evaluate error messaging, and<br>verify adherence to standards. An analysis of new hazards and software requirements<br>emerging as a result of the changes is evaluated for each revision. This is<br>documented in the CORE Console's Software Requirements Specifications (SRS) and<br>risk analysis, which can be found in Section 16. Subsequent verification/ validation<br>testing, including system and Unit box testing, is performed on the console to confirm<br>the continued safety and effectiveness of the Stryker® CORE Console prior to the<br>release of each software revision to the market. | | | | | |
| The results of this testing demonstrate that the device is as safe, as effective, and<br>performs as well as or better than the predicate device. | | | | | |
| Biocompatibility | Per ISO 10993-1:2003 and FDA G95-1:1995 guidelines, a biocompatibility<br>evaluation for the CORE Console is not required due to the non-patient contact<br>intended use of this medical device. | | | | |
| Substantial<br>Equivalence<br>(SE) Rationale: | The Stryker® Consolidated Operating Room Equipment (CORE) System when<br>compared to its predicates has a similar intended use and technology characteristics<br>as the predicate devices. Therefore, this device is substantially equivalent to predicate<br>devices. | | | | |
| Predicate Device Comparison Table | | | | | |
| Element of<br>Comparison | Stryker ® CORE<br>Console (Under<br>Review) | Stryker ®<br>System<br>(K040300,<br>K040369,<br>K032303) | Synthes Electric Pen<br>Drive (EPD) System<br>(K043310) | Stryker ® Total<br>Performance System<br>(TPS) (K942956,<br>K943589, K943569,<br>K943540, K032117) | Substantial<br>Equivalence |
| Indications for Use | The Stryker ®<br>Consolidated Operating<br>Room Equipment<br>(CORE) Console is<br>intended for use in the<br>cutting, drilling,<br>reaming,<br>decorticating, shaping,<br>and smoothing of bone,<br>bone cement and teeth<br>in a variety of surgical<br>procedures, including<br>but not limited to<br>dental, ENT, neuro,<br>spine, and endoscopic<br>applications. It is also<br>usable in the placement<br>or cutting of screws, | The Stryker®<br>Consolidated<br>Operating<br>Room Equipment<br>(CORE) System is<br>intended for use in<br>the cutting,<br>drilling, reaming,<br>decorticating, and<br>smoothing of<br>bone, bone<br>cement and teeth<br>in a variety of<br>surgical<br>procedures,<br>including but not<br>limited to dental,<br>ENT, neuro, and<br>endoscopic | The Synthes Electric<br>Pen Drive (EPD)<br>System is indicated for<br>screw insertion, pin<br>and wire placement,<br>cutting of bone and<br>metal, drilling,<br>reaming, decorticating,<br>shaping, and<br>smoothing of bones<br>and teeth in a wide<br>variety of surgical<br>procedures, including<br>but not limited to<br>general orthopedic<br>trauma, foot, hand,<br>maxifacial,<br>neurosurgical, oral, | The Stryker ® Total<br>Performance System is<br>intended for use in the<br>cutting, drilling,<br>reaming,<br>decorticating, and<br>smoothing of bone,<br>teeth and other bone<br>related tissue in a<br>variety of surgical<br>procedures,<br>including but not<br>limited to dental, ENT,<br>neuro and endoscopic.<br>It is<br>also usable in the<br>placement or cutting of<br>screws, wires, pins, | The indications for<br>use for all three<br>devices are similar. |
| | other fixation devices. | applications. It is<br>also usable in the<br>placement or<br>cutting of screws,<br>metal, wires, pins,<br>and other fixation<br>devices. | reconstructive and<br>spine surgery. | devices. It can also be<br>used to cut metal. | |
| Handpiece<br>Connectors | Three connectors allow<br>handpiece cords to be<br>connected to the<br>console. The console is<br>capable of running two<br>handpieces<br>simultaneously. | Three connectors<br>allow handpiece<br>cords to be<br>connected to the<br>console. The<br>console is capable<br>of running two<br>Handpieces<br>simultaneously. | Two connectors allow<br>for two Electric Pen<br>Drive units to be<br>connected. | Three connectors allow<br>handpiece cords to be<br>connected to the<br>console. | Similar for all three<br>devices.<br>Design feature is<br>identical to<br>previously-cleared<br>Stryker® CORE<br>Console. |
| Footswitch<br>Connectors | Two connectors allow<br>up to two footswitches<br>to be run on the<br>console simultaneously.<br>The console can also<br>be controlled by a<br>wireless footswitch. | Two connectors<br>allow up to two<br>footswitches to be<br>run on the console<br>simultaneously. | One connector allows<br>one footswitch to<br>control one drive unit. | One connector allows<br>one footswitch to<br>control one drive unit. | Similar for all three<br>devices. Stryker®<br>CORE Console<br>under review has<br>the capability of<br>connecting to a<br>wireless<br>footswitch. |
| Adjustable<br>Operating<br>Parameters | The LCD screen allows<br>the user to set the<br>desired operating<br>parameters. Operating<br>parameters are also<br>obtained from attached<br>accessories. | The LCD screen<br>allows<br>the user to set the<br>desired operating<br>parameters.<br>Operating<br>parameters are<br>also obtained from<br>attached<br>accessories. | Max speed may be<br>controlled using slide<br>controls provided in the<br>console for each drive<br>unit. Torque limiting<br>feature can be<br>controlled using an On/<br>Off/Calibrate knob.<br>Irrigation can be used<br>with either Drive Unit 1<br>or Drive Unit 2 and<br>knob control is<br>available to select<br>between continuous<br>and variable irrigation<br>for each drive unit.<br>Irrigation flow can be<br>controlled using a flow<br>rate adjustment knob.<br>OR<br>User sets the operating<br>parameters using the<br>controls available in<br>the console or using<br>the handswitch or<br>footswitch interface. | The LCD screen allows<br>the user to set the<br>desired operating<br>parameters. | TPS and CORE<br>Console have<br>similar software<br>feature.<br>Design feature is<br>identical to<br>previously-cleared<br>Stryker® CORE<br>Console. |
| Power Output | 400 Watts | 400 Watts | Information not<br>available. | 400 Watts | TPS and CORE<br>Console power<br>output is identical.<br>Design feature Is<br>identical to .<br>previously-cleared<br>Stryker® CORE<br>Console. |
| Wireless Tag<br>Technology | Accessories are<br>recognized and<br>identified on the<br>console screen. | Accessories are<br>recognized and<br>identified on the<br>console screen. | Information not<br>available. | Accessories are<br>recognized and<br>identified on the<br>console screen. | TPS and CORE<br>Console have<br>similar software<br>feature. Design<br>feature is identical<br>to previously-<br>cleared Stryker®<br>CORE Console. |
| Torque Control | Mechanism to allow the<br>user to specify the<br>torque profile to be<br>used for handpieces. | N/A | Mechanism to allow<br>the surgeon to implant<br>and tighten screws<br>under power. The user<br>calibrates and sets the<br>desired torque value<br>using the control<br>available in the<br>console. | N/A | The Stryker®<br>CORE Console<br>and Synthes<br>Electric Pen Drive<br>System torque<br>control features<br>are similar. Both<br>devices allow the<br>user to vary torque<br>settings. |
:
. . .
. |
{2}------------------------------------------------
#
·
:
:
・
・
:
.
、 .
:
.
:
.
・
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
:
:
・
{3}------------------------------------------------
| Console Accessories | Irrigation accessories Handpiece cord Handswitch Footswitch | Irrigation accessories Handpiece cord Handswitch Footswitch | Irrigation accessories Handpiece cord Handswitch Footswitch | Irrigation accessories Handpiece cord Handswitch Footswitch | Similar for all three devices. |
|---------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|--------------------------------|
|---------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|--------------------------------|
| Safety and Effectiveness: | Based upon the comparison to the predicate devices, the Stryker® Consolidated Operating Room Equipment (CORE) System does not raise any new issues of safety and effectiveness and is substantially equivalent to legally marketed devices. |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Michelle Jump<br>Senior Regulatory Affairs Representative |
Signature
| Date Submitted: | May 1, 2012 |
|-----------------|-------------|
|-----------------|-------------|
and the comments of the comments of
and the country of the country of the country
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and is evenly spaced around the circle.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY - 1 2012
Stryker Corporation % Ms. Michelle Jump 4100 E. Milham Avenue Kalamazoo, MI 49001
Re: K112593
Trade/Device Name: Stryker® Consolidated Operating Room Equipment (CORE) System Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, nose, and throat electric or pneumatic surgical drill Regulatory Class: Class II Product Code: ERL Dated: April 24, 2012 Received: May 1, 2012
Dear Ms. Jump:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 read on as made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Edith m.b
Malyina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
## 510(K) Number: K 112593
Device Name: Stryker® Consolidated Operating Room Equipment (CORE) System
Indications for Use:
The Stryker ® Consolidated Operating Room Equipment (CORE) System is intended for use in the cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement and teeth in a variety of surgical procedures, including but not limited to, dental, ENT (ear, nose, throat), neuro, spine, and endoscopic applications. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
Prescription Use: X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
D.M.ls
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
510(k) Number K1125
Section 4 -1
.. 21 CFR 801.109)
otion Use _
Prescription Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.