K112572 · Geistlich Pharma AG · NPM · Dec 28, 2011 · Dental
Device Facts
Record ID
K112572
Device Name
GEISTLICH COMBI-KIT COLLEGEN
Applicant
Geistlich Pharma AG
Product Code
NPM · Dental
Decision Date
Dec 28, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Geistlich Bio-Gide® is intended for the following uses: - Augmentation around implants placed in immediate extraction sockets: - Augmentation around implants placed in delayed extraction sockets; - Localized ridge augmentation for later implantation; - Alveolar ridge reconstruction for prosthetic treatment; - Filling of bone defects after root resection, cystectomy, removal of retained teeth: - Guided bone regeneration in dehiscence defects; - Guided tissue regeneration procedures in periodontal defects. Geistlich Bio-Oss Collagen® is intended for the following uses: - Augmentation or reconstructive treatment of the alveolar ridge: - Filling of periodontal defects; - Filling of defects after root resection, apicoectomy, and cystectomy; - Filling of extraction sockets to enhance preservation of the alveolar ridge; - Elevation of the maxillary sinus floor; - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR). Geistlich Bio-Oss® is intended for the following uses: - Augmentation or reconstructive treatment of the alveolar ridge; - Filling of infrabony periodontal defects; - Filling of defects after root resection, apicoectomy, and cystectomy; - Filling of extraction sockets to enhance preservation of the alveolar ridge; - Elevation of the maxillary sinus floor; - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
Device Story
Geistlich Combi-Kit Collagen is a convenience kit containing one unit of Geistlich Bio-Oss Collagen (purified bovine bone mineral granules with 10% collagen) and one unit of Geistlich Bio-Gide (bilayer collagen membrane). The kit is designed for ease of use by clinicians during regenerative dental procedures. The Bio-Oss component acts as a bone graft matrix, while the Bio-Gide membrane provides a barrier to prevent fibrous tissue ingrowth into bone defects. The kit is sterilized by gamma irradiation. The device is used in clinical settings by dental professionals to facilitate bone and tissue regeneration in the alveolar ridge and peri-implant areas. The kit simplifies the surgical workflow by providing the necessary graft and barrier materials in a single package.
Clinical Evidence
No clinical data provided; substantial equivalence is based on the identical technological characteristics and materials of the individual components to their respective predicate devices.
Technological Characteristics
Convenience kit containing bovine-derived bone mineral matrix (Bio-Oss) and collagen membrane (Bio-Gide). Bio-Oss: purified spongiosa granules/blocks, non-antigenic, porous. Bio-Gide: bilayer collagen membrane, non-cross-linked. Sterilization: gamma irradiation. No software or electronic components.
Indications for Use
Indicated for patients requiring alveolar ridge augmentation/reconstruction, periodontal defect filling, extraction socket preservation, maxillary sinus floor elevation, or guided bone/tissue regeneration in dental/oral surgical procedures.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
# K112572
DEC 2 8 2011
### 2. 510(k) Summary
This 510(k) was submitted for two reasons: (1) to provide a convenience kit Geistlich Combi-Kit Collagen (composed of one unit of Geistlich Bio-Oss Collagen® and one unit of Geistlich Bio-Gide®), and (2) to modify the Instructions for Use of Geistlich Bio-Oss Collagen®, Geistlich Bio-Gide®, and Geistlich Bio-Oss®.
## SPONSOR
Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland
| Contact Person: | Daniel Kracov, Arnold & Porter, LLP |
|-----------------|-------------------------------------|
| e-mail: | Daniel.Kracov@aporter.com |
| Phone: | (202)942-5120 |
| Date Prepared: | December 12, 2011 |
#### DEVICE NAME
| Proprietary Name: | Geistlich Combi-Kit Collagen<br>- Geistlich Bio-Oss Collagen®<br>- Geistlich Bio-Gide®<br>Geistlich Bio-Oss® |
|----------------------|--------------------------------------------------------------------------------------------------------------|
| Common/Usual Names: | Natural Bone Grafting Material Plus Collagen<br>Resorbable Bilayer Membrane for GTR/GBR |
| Classification Name: | Bone grafting material, animal source (NPM)<br>Barrier, animal source, intraoral (NPL) |
{1}------------------------------------------------
## PREDICATE DEVICES
Bio-Oss Collagen® (K092428, K033815, K974399) Bio-Gide® (K050446, K042197, K960724) Bio-Oss® and Bio-Oss® Blocks (K033815, K970321, K952618)
## INTENDED USE
Geistlich Bio-Gide® is intended for the following uses:
- Augmentation around implants placed in immediate extraction sockets:
- Augmentation around implants placed in delayed extraction sockets;
- Localized ridge augmentation for later implantation;
- Alveolar ridge reconstruction for prosthetic treatment;
- Filling of bone defects after root resection, cystectomy, removal of retained teeth:
- Guided bone regeneration in dehiscence defects;
- Guided tissue regeneration procedures in periodontal defects.
Geistlich Bio-Oss Collagen® is intended for the following uses:
- Augmentation or reconstructive treatment of the alveolar ridge:
- Filling of periodontal defects;
- Filling of defects after root resection, apicoectomy, and cystectomy;
- Filling of extraction sockets to enhance preservation of the alveolar ridge;
- Elevation of the maxillary sinus floor;
- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR);
- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
Geistlich Bio-Oss® is intended for the following uses:
- Augmentation or reconstructive treatment of the alveolar ridge;
- Filling of infrabony periodontal defects;
- Filling of defects after root resection, apicoectomy, and cystectomy;
- Filling of extraction sockets to enhance preservation of the alveolar ridge;
- Elevation of the maxillary sinus floor;
- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR);
{2}------------------------------------------------
- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
The two products are being packaged together as a convenience kit for the ease of clinician use in regenerative procedures.
#### DEVICE DESCRIPTION
Geistlich Combi-Kit Collagen is a convenience kit containing one unit of Geistlich Bio-Oss Collagen® and one unit of Geistlich Bio-Gide®. The two devices are packaged in double blisters in one package and then sterilized by gamma irradiation.
Geistlich Bio-Oss Collagen® is a combination of purified spongiosa (cancellous) natural bone mineral granules and 10% collagen fibers in a block form and is sterilized by gamma irradiation. Geistlich Bio-Oss Collagen® is the same as its predicate device, Bio-Oss Collagen®.
Geistlich Bio-Gide® is a pure collagen membrane with a bilayer structure. The porous surface (facing the bone) allows the ingrowth of bone forming cells, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the bone defect. The membrane is made of collagen without further cross-linking, and is sterilized by gamma irradiation. The size of the Geistlich Bio-Gide® bilayer membrane to be provided in the Geistlich Combi-Kit Collagen convenience kit is 16mm x 22 mm. With the exception of the size of the membrane packaged in the convenience kit, Geistlich Bio-Gide® is the same as its predicate device, Bio-Gide®.
Geistlich Bio-Oss® spongiosa (cancellous) granules and Geistlich Bio-Oss® Block are natural non-antigenic porous bone mineral matrices. They are produced by removal of all organic components from bovine bone. Due to their natural structure, Geistlich Bio-Oss® and Geistlich Bio-Oss® Block, are physically and chemically comparable to the mineralized matrix of human bone. Geistlich Bio-Oss® is the same as its predicate device, Bio-Oss®.
## TECHNOLOGICAL CHARACTERISTICS
{3}------------------------------------------------
Geistlich Bio-Oss Collagen® and Geistlich Bio-Oss® have the same design, same chemical composition, and same materials as their respective predicate devices, and are substantially equivalent to their predicate devices, Bio-Oss Collagen® and Bio-Oss®.
Geistlich Bio-Gide® has the same design, same chemical composition, and same materials as its predicate device, Bio-Gide®. Although the dimensions of the membrane to be included in the Geistlich Combi-Kit Collagen convenience kit have been modified slightly, this difference does not change the intended use of the device. Thus, Geistlich Bio-Gide® is substantially equivalent to its predicate device, Bio-Gide®.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Geistlich Pharma AG C/O Mr. Daniel A. Kravoc Arnold & Porter LLP 555 Twelfth Street, N.W. Washington, Distruct of Columbia 20004
DEC 2 8 2011
Re: K112572
Trade/Device Name: Geistlich Combi-Kit Collagen Geistlich Bio-Oss® Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPM Dated: December 20, 2011 Received: December 21, 2011
Dear Mr. Kracov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Kracov
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety /ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K 112572
# Indications for Use
510(k) Number (if known):
((2572
Device Name: Geistlich Combi-Kit Collagen
Indications For Use:
Indications in of OSE.
Gelstlich Combi-Kit Collagen is a convenience kit that provides one (1) unit Gelstich and seatings Collagen membrane and one (1) unit Geistlich Bio-Oss Collagen® block in one (1) package.
Geistlich Bio-Gide® is intended for the following uses:
- Augmentation around implants placed in immediate extraction sockets;
- Augmentation around implants placed in delayed extraction sockets;
- Localized ridge augmentation for later implantation;
- Alveolar ridge reconstruction for prosthetic treatment;
- Filling of bone defects after root resection, cystectorny, removal of retained teeth;
- Guided bone regeneration in dehiscence defects; and
- Guided tissue regeneration procedures in periodontal defects.
Geistlich Bio-Oss Collagen® is intended for the following uses:
- Augmentation or reconstructive treatment of the alveolar ridge;
- Filling of periodontal defects;
- Filling of defects after root resection, apicoectorny, and cystectomy;
- Filling of extraction sockets to enhance preservation of the alveolar ridge;
- Elevation of the maxillary sinus floor;
- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and
- Filling of per-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
| Prescription Use | _____________________________________________________________________________________________________________________________________________________________ |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AND/OR | |
| Over-The-Counter Use | _________________________________________________________________________________________________________________________________________________________ |
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k112577
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Geistlich Bio-Oss®
Indications For Use:
Geistlich Bio-Oss® is intended for the following uses:
- Augmentation or reconstructive treatment of the alveolar ridge;
- Filling of infrabony periodontal defects;
- Filling of defects after root resection, apicoectorny, and cystectorny;
- Filling of extraction sockets to enhance preservation of the alveolar ridge;
- Elevation of the maxillary sinus floor;
- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and
- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
AND/OR Prescription Use D) (21 CFR 801 Subpart C)
Over-The-Counter Use (Part 21 CFR 801 Subpart
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Parmer
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
...
510(k) Number: K112572
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.