K112548 · Altomec Endoscopy, Inc. · HRX · Aug 15, 2012 · Orthopedic
Device Facts
Record ID
K112548
Device Name
ALTOMEC ARTHROSCOPE
Applicant
Altomec Endoscopy, Inc.
Product Code
HRX · Orthopedic
Decision Date
Aug 15, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Arthroscopes are rigid endoscopic devices introduced into a patient to provide an internal view or image of the interior of a joint for examination, diagnosis, and/or therapy. Arthroscopes are indicated for use in arthroscopic procedures performed in the hip, knee, shoulder, wrist (carpel tunnel syndrome), temporal mandibular joint, ankle, elbow, and feet (plantar fascia release).
Device Story
Altomec Arthroscope is a non-deflectable rigid endoscope for internal joint visualization during examination, diagnosis, or therapy. Device consists of a long tube with objective lens, relay lenses, and proximal coupling lens. Input: optical image of joint interior. Transformation: optical relay of image to proximal end for connection to external CCD camera and monitor system. Output: magnified visual image on monitor. Used in clinical settings by physicians. Light transmission via internal glass fibers connected to external light source. No working channel; irrigation provided via external sheaths. Magnification (2x-5x) depends on tip-to-object distance. Benefits include minimally invasive visualization of joint structures.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing and comparison of technological characteristics and materials to the predicate device.
Technological Characteristics
Rigid endoscope; 4.0mm outer diameter; 140mm working length; 105° FOV; 30° direction of view. Materials: medical grade stainless steel, high-quality optical glass, glass fiber bundles, Ultem (polyetherimide) eyepiece, sapphire window, EPO-TEK 353ND adhesive. Light transmission via fiber optics. Reusable. No software or electronic components.
Indications for Use
Indicated for patients undergoing arthroscopic procedures in the hip, knee, shoulder, wrist (carpal tunnel syndrome), temporal mandibular joint, ankle, elbow, and feet (plantar fascia release).
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
# K117548
Page 1 of 3
AUG 15 2012
## Section 5: 510(k) Summary
August 31. 2011 1. Preparation Date:
- Altomec Endoscopy Inc 2. Submitted by: 3310 Miller Road Kalamazoo, MI 49001 Owner/Operator #: N/A - Not yet marketing the device Small Business SBD118566
#### Contact Person/Prepared by:
Darren Reeves 510(k) Submitter Phone (866) 393-4954 Fax (866) 393-4954 E-mail: dpdist@bww.com
#### 3. Device Identification:
Altomec Arthroscope Trade Name: Common Name: Arthroscope Classification Name: Arthroscope (21 CFR 888.1100; Product Code HRX)
Stryker Arthroscope (K093677) 4. Predicate Device:
#### 5. Device Description:
Altomec Arthroscopes are Non-deflectable rigid endoscopic devices, introduced into a patient to provide an internal view or image of the interior of a joint for examination, diagnosis, and/or therapy. The Altomec Rigid Arthroscope is a long tube containing a series of lenses. At the distal end, an objective lens captures the image of the object. Lens along the rod relay the image. At the proximal end, a proximal coupling lens relays the image to a CCD (Camera). The Altomec Arthroscope has 140mm working length with an outside diameter of 4.0mm. The inner or optical tube that holds the optical system has a diameter of 3.25mm. The Altomec Arthroscope has a 105 degree field of view and 30 degree direction of view. The direction of view enables viewing of different parts.
Operating site is magnified two to five times of its actual size, depending on the distance between the tip of the endoscope and the object that is inspected. Magnified image is viewed on a monitor by connecting device's eyepiece to a video coupler of a camera monitor system. Image size on the monitor also depends on the type of video coupler used with camera system. Image can be recorded in a variety of formats depending on the camera system used.
Light is transmitted through glass fibers running between inner and outer tubes from distal end to the light post. Light transmission is achieved by connecting one end of a light source and other end to the light post of the device. Device has no working channel. Sheaths with locking mechanism identical to predicate device, are used for irrigation.
{1}------------------------------------------------
K112548 Page 2 of 3
## 6. Intended Use:
Arthroscopes are rigid endoscopic devices introduced into a patient to provide an internal view or image of the interior of a joint for examination, diagnosis, and/or therapy.
Arthroscopes are indicated for use in arthroscopic procedures performed in the hip, knee, shoulder, wrist (carpel tunnel syndrome), temporal mandibular joint, ankle, elbow, and feet (plantar fascia release).
### 7. Comparison to Predicate:
## a) Technical
| | Proposed Device | Predicate Device | |
|--------|----------------------|---------------------|--|
| Device | *Altomec Arthroscope | Stryker Arthroscope | |
| | Technological | Characteristics<br>(Design) | Equivalence | Impact on Safety<br>and<br>Effectiveness |
|--------------------------------------|-------------------|-----------------------------|-------------|------------------------------------------|
| Field of View<br>(FOV) | 105 | 105 | Same | N/A |
| Degrees Direction<br>of View Degrees | 30 | 30 | Same | N/A |
| Outer Diameter | 4mm | 4mm | Same | N/A |
| Working Length | 140mm | 140mm | Same | N/A |
| Single use Or<br>Reusable | Reusable | Reusable | Same | Equivalent |
| Light Guide End<br>Adapter | Storz and Olympus | Storz and Olympus | Same | N/A |
Technological Characteristics: The Altomec Arthroscopes are substantially equivalent in construction and materials to the predicate Stryker Arthroscopes.
## b) Materials
| | Proposed Device | Predicate |
|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Parts/materials | *Altomec<br>Arthroscope | Stryker<br>Arthroscope |
| Optical system | High quality optical glass | High quality optical glass |
| Outer tube | Medical grade Stainless Steel | Medical grade Stainless Steel |
| Inner (optical) tube | Medical grade Stainless Steel | Medical grade Stainless Steel |
| Spacers | Medical grade Stainless Steel | Medical grade Stainless Steel |
| Fiber Optic bundle(Light fibers) | Glass | Glass |
| Light cone | Glass | Glass |
| Recessed cover glass (window) | Soldered glass made from sapphires | N/A |
| Eyepiece | Ultem® (Polyetherimide) | N/A |
| Mechanical parts:<br>Body, Light post, fiber guide and<br>Ocular holder. | Machined on CNC/lathes using<br>medical grade stainless steel | Machined on CNC/lathes using<br>medical grade stainless steel |
| Adhesive | EPO-TEK®353ND (technical data<br>sheet attached): Used for potting<br>fiber optic bundle, light cone and<br>fiber guide | N/A |
| Finished Silver Solder: Part Number<br>SBSK. Composition 3.4-3.8%<br>Silver. Remainder Tin. | Used for soldering Soldered Glass<br>made from Sapphires. | N/A |
{2}------------------------------------------------
K112548 Page 3 of 3
### 8. Similarities/ Differences of the proposed device when compared to the predicate:
Performance Testing: The patient contacting materials are identical to the materials used in the predicated device (Stryker Arthroscope). The Altomec Arthroscopes met all specified design and performance requirements.
Predicate Devices: The Altomec Arthroscopes are substantially equivalent in terms of safety and effectiveness to the currently marketed device, Stryker Arthroscopes.
Substantial Equivalence: The technological differences between the Altomec Arthroscope and Stryker Arthroscopes do not raise new questions of safety or effectiveness. Therefore the Altomec Arthroscopes are substantially equivalent to the previously cleared Stryker Arthroscope. Refer to Section 7.0 for a detailed comparison.
The data within this submission demonstrates that there are no significant differences between the application device and the predicate, indicating that the application device is safe, effective and substantially equivalent for marketing in the U.S.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Altomec Endoscopy, Incorporated % DP Distribution & Consulting, LLC Mr. Darren Reeves Consultant 7305 Hancock Village Drive, Suite 109 Chesterfield, VA 23832
Re: K112548
Trade/Device Name: Altomec Arthroscope Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: August 9, 2012 Received: August 9, 2012
Dear Mr. Reeves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
AUG 15 2012
{4}------------------------------------------------
#### Page 2 - Mr. Darren Reeves
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
y yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
S. Food and Drug Administration
ENTER FOR DEVICES AND RADIOLOGICAL HEALTH
Image /page/5/Picture/1 description: The image shows the logo of the Department of Health and Human Services. The logo consists of a stylized caduceus symbol on the left, with the text "Department of Health and Human Services" to the right of the symbol. The text is stacked in three lines.
## Indications for Use Statement
510(k) Number (if known): __ K112548
Device Name: Arthroscope
Indications for Use:
Intended Use: Arthroscopes are rigid endoscopic devices introduced into a patient to provide an internal view or image of the interior of a joint for examination, diagnosis, and/or therapy. Arthroscopes are indicated for use in arthroscopic procedures performed in the hip, knee, shoulder, wrist (carpel tunnel syndrome), temporal mandibular joint, ankle, elbow, and feet (plantar fascia release).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K112548
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.