K112544 · Unimed Medical Supplies, Inc. · DXQ · Nov 22, 2011 · Cardiovascular
Device Facts
Record ID
K112544
Device Name
UNIMED BLOOD PRESSURE CUFF
Applicant
Unimed Medical Supplies, Inc.
Product Code
DXQ · Cardiovascular
Decision Date
Nov 22, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The Unimed Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
Device Story
Device consists of inelastic sleeve with integrated inflatable bladder, tubing, and hook-and-loop closure; wraps around patient limb. Connects to non-invasive blood pressure measurement system to facilitate indirect blood pressure monitoring. Used in clinical settings; operated by healthcare professionals. Cuff inflates to occlude artery and deflates to allow system to detect pressure oscillations. Output is pressure data processed by the connected measurement system to determine systolic/diastolic blood pressure. Benefits include non-invasive, reusable blood pressure monitoring across diverse patient sizes (neonatal to adult).
Clinical Evidence
Bench testing only. Device tested for performance per ANSI/AAMI SP10 (2002+A1:2003+A2:2006+(R)2008) and biocompatibility per ISO 10993-1, ISO 10993-5, and ISO 10993-10. No clinical data presented.
Technological Characteristics
Materials: PU synthetic leather cuff, TPU film bladder, PVC tubing, molded nylon hook/loop. Dimensions: Neonatal to adult thigh sizes (6-66cm). Connectivity: Pneumatic connection to non-invasive blood pressure monitor. Pressure limit: 0-300mmHg. Reusable; non-sterile. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for indirect blood pressure measurement in pediatric and adult patient populations when used with compatible non-invasive blood pressure measurement systems.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
<11 2.5 44
Image /page/0/Picture/1 description: The image shows a logo with the words "Med" on top of "Uni". The logo is contained within a black box with rounded corners. The text is white, and the background of the box is black.
Unimed Medical Supplies Inc. No. 37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26828795 Fax: 86 755 26697984 Email: info@unimed.cn Website: www.unimed.cn
# Section 3
# 510(K) Summary
Summary prepared Date: April 2, 2011
### Submitter Information:
Unimed Medical supplies Inc.
No.37, Yanshan Road, Shekou, Shenzhen, China 518067
#### Contact Person:
Xinmei Tan, QA manager No.37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: (86) 755 26695165 Fax: (86) 755 26697984 Email: info@unimed.cn, tanxinmei1979@163.com
### Device Name
Trade Name: Unimed Blood Pressure Cuff Common Name: Non-invasive Blood pressure cuff Product Code/Classification: DXQ/21 CFR870.1120 Review Panel: Cardiovascular
## Predicate Device
Sensa-Cuff (K022482) GE Medical Systems Information Technologies
### Intended Use
The Unimed Blood Pressure Cuff is an accessory used in conjunction with non-invasive
Confidential
Unimed Blood Pressure Cuff - 510(k) Submission
Page 1 of 4
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image is a logo for MedUni. The logo is in a rectangular shape with rounded corners. The text "Med" is on the top line, and the text "Uni" is on the bottom line. There is a horizontal line separating the two lines of text.
Unimed Medical Supplies Inc. No. 37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26828795 Fax: 86 755 26697984 Website: www.unimed.cn Email: info@unimed.cn
blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
# Device Description
The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system.
| Features | Unimed Medical | GE Medical |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Indirect measurement of<br>blood pressure | Indirect measurement of<br>blood pressure |
| Patient Populations | Adults/Pediatrics | Adults/Pediatrics |
| Material | Cuff: PU Synthetic Leather<br>Bladder: Transparent<br>Polyurethane (TPU Film)<br>Tubing: PVC<br>Hook: Molded Nylon<br>Loop: Nylon | Cuff: Woven nylon<br>Film (bladder): Ethylene vinyl<br>acetate copolymer (EVA)<br>Tubing: PVC<br>Ribbon: Textured polyester<br>Hook: Molded Nylon<br>Loop: Nylon |
| Tube Configuration | One or two tube | One or two tube |
| Size<br>(Range in cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal (6-11cm)<br>Infant (10-19cm) | Conform to AHA bladder sizes<br>recommendations<br>Infant (8-13cm)<br>Child (12-19cm) |
# Comparison to the Predicate Device
Confidential
UnImed Blood Pressure Cuff - 510(k) Submission
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the text "Med Uni" in a bold, sans-serif font. The text is arranged in two lines, with "Med" on the top line and "Uni" on the bottom line. The logo is enclosed in a rounded rectangle with a thick black border. The background of the logo is white, while the text is black.
Unimed Medical Supplies Inc. No. 37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26828795 Fax: 86 755 26697984 Website: www.unimed.cn Email: info@unimed.cn
| | Pediatric (18-26cm) | Small Adult (17-25) |
|--------------------|------------------------------|------------------------------|
| | Small Adult (20-28cm) | Adult (23-33cm) |
| | Adult (25-35cm) | Large Adult (31-40cm) |
| | Adult Long (25-35cm) | Thigh (38-50cm) |
| | Large Adult (33-47cm) | |
| | Large Adult Long (33-47cm) | |
| | Adult Thigh (46-66cm) | |
| Repeated inflation | 10,000 inflations | 10,000 inflations |
| | 3,000 hook and loop closures | 3,000 hook and loop closures |
| Pressure limits | 0-300mmHg | 0-300mmHg |
| Sterility | Non-sterile | Non-sterile |
| Biocompatibility | Comply with ISO 10993 | Comply with ISO 10993 |
| | biocompatibility evaluation | biocompatibility evaluation |
The Unimed Blood Pressure Cuff has the same Intended Use, basic construction, and technology specification as the predicated devices are wrapped the patient's arm or leg and secured by a hook and loop fastener commonly called Velcro. Both devices are available in the same size and range and are intended for the same patient populations.
(1) Material: Although the patient materials for subject device and predicate device are not all the same, but they are all conformed to ISO 10993.
(2) Size: We extend the length of the cuffs in order to accommodate special groups, such as neonates and overweight subjects. Based on the performance testing in this submission, the slight difference on the range of these blood pressure cuffs does not raise any safety or effectiveness issue.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the words "Med Uni" stacked on top of each other. The logo is enclosed in a rounded rectangle with a thick black border. There is a horizontal line separating the two words. The text is in a bold, sans-serif font.
Medical Supplies Inc. Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26828795 Fax: 86 755 26697984 Website: www.unimed.cn Email: info@unimed.cn
# Performance Summary
The Unimed Blood Pressure Cuff has been tested according to the following standards:
- � ANSI/AAMI SP10, Manual, electronic or automated sphygmomanometers, 2002+A1:2003+A2:2006+(R)2008
- � ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and testing, 2003
- � ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, 2009
- ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for � irritation and delayed-type hypersensitivity, 2002+A1:2006
# Conclusion
The subject device Unimed Blood Pressure Cuff has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject devices.
Thus, the subject device is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in black and is in all capital letters. There is a logo to the left of the text.
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the circle's perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Unimed Medical Supplies, Inc. c/o Mr. Clare He Underwriters Laboratories Inc. Electronic Building, Parage Electronic Industrial Park No. 8 Nanyun Er Road, Gangzhou 510663 CHINA
NOV 22 2011
Re: K112544
Trade/Device Name: Unimed Blood Pressure Cuffs: U1883S, U1883D, U1882S, U1882D, U1881S, U1881D, U1885S, U1885D, U1880S, U1880D, U1886S, U1886D, U1869S, U1869D, U1889S, U1889D, U1884S, and U1884D (18 models) Regulatory Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II (Two) Product Code: DXO Dated: November 4, 2011 Received: November 8, 2011
Dear Mr. He:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Clare He
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Med Uni. The logo is in a rounded rectangle shape. The words "Med" and "Uni" are stacked on top of each other in the center of the logo.
Unimed Medical Supplies Inc. No. 37, Yanshan Road, Shekou, Shenzhen, China 518067 Tel: 86 755 26828795 Fax: 86 755 26697984 Website: www.unimed.cn Email: info@unimed.cn
#### Section 4 Indications for Use Statement
510(k) Number:
KII2544
Device Name: Unimed Blood Pressure Cuff
# Indications for Use
The Unimed Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR
/OR
Over-The-Counter Use _
(Part 21 CFR 801 Sub part D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
Page 1 of 1
**510(k) Number** K112544
Confidential
Unimed Blood Pressure Cuff - 510(k) Submission
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.