COLLAGEN TENDON SHEET

K112423 · Rotation Medical, Inc. · FTM · Dec 22, 2011 · General, Plastic Surgery

Device Facts

Record IDK112423
Device NameCOLLAGEN TENDON SHEET
ApplicantRotation Medical, Inc.
Product CodeFTM · General, Plastic Surgery
Decision DateDec 22, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Collagen Tendon Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.

Device Story

Collagen Tendon Sheet is a resorbable type I collagen matrix; provides protective layer between injured tendon and surrounding tissue. Device is soft, pliable, nonfriable, porous sheet when hydrated. Provided sterile, non-pyrogenic, for single use. Intended for surgical management of tendon injuries where tendon tissue remains intact. Clinical benefit derived from physical barrier properties protecting healing tendon.

Clinical Evidence

No clinical tests performed. Substantial equivalence supported by in vitro characterization, biocompatibility studies, animal efficacy study, and clinical history of the predicate device.

Technological Characteristics

Resorbable type I collagen matrix; porous, soft, pliable, nonfriable sheet. Sterilized, single-use. Evaluated per FDA Guidance for Surgical Mesh and ISO 10993-1. No software or electronic components.

Indications for Use

Indicated for management and protection of tendon injuries without substantial loss of tendon tissue.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K112423 > | of 2 Image /page/0/Picture/1 description: The image shows the logo for Rotation Medical. The logo consists of a circular arrangement of dots above the word "rotation" in a stylized font. Below "rotation" is the word "MEDICAL" in a smaller, sans-serif font. DEC 2 2 2011 # 510(k) Summary #### Applicant Information Applicant Name: Applicant Address: Telephone: Fax: Contact Person: Rotation Medical, Inc. 15350 25th Avenue North, Suite 100 Plymouth, MN 55447 763-746-7502 763-746-7501 Jeff Sims Vice President, Clinical Programs and Regulatory Affairs August 22, 2011/November 14, 2011/December 13, 2011 Date Prepared/Revised: ### Name of Device Device Common Name: Device Trade Name: Device Classification Name: Tendon Protector Collagen Tendon Sheet Mesh, Surgical 878.3300 Class II FTM ### Legally Marketed Devices to Which Substantial Equivalence is Claimed Predicate Device(s): Collagen Tendon Wrap. K0080452 Collagen Matrix, Inc., Oakland, NJ #### Description of the Device Collagen Tendon Sheet is a resorbable type I collagen matrix that provides a layer of collagen over injured tendons. Collagen Tendon Sheet is designed to provide a layer between the tendon and the surrounding tissue. When hydrated, Collagen Tendon Sheet is an easy-to-use, soft, pliable, nonfriable, porous collagen sheet. Collagen Tendon Sheet is provided sterile, non-pyrogenic, for single use only, in a variety of sizes, in double peel packages. #### Intended Use Collagen Tendon Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. {1}------------------------------------------------ #### Summary/Comparison of Technical Characteristics Collagen Tendon Sheet and its predicate device have the same technological characteristics. In particular, Collagen Tendon Sheet and its predicate are the same with respect to intended use, design, material characterization. The substantial equivalence of the Collagen Tendon Sheet and its predicate was demonstrated primarily based on in vitro characterization studies, biocompatibility studies, an animal efficacy study, and clinical experience of the predicate device. In vitro characterization studies included evaluation of material properties, biological properties, chemical and physical properties. The Collagen Tendon Sheet and its predicate device are manufactured with similar processing, in the same facilities, by the same manufacturer, using the same raw materials. They vary with respect to size and thickness. In addition, the Collagen Tendon sheet is provided in a flat form. Collagen Tendon Sheet and its predicate have been characterized for chemical composition, purity, density, and strength to demonstrate substantial equivalence. Testing was conducted in accordance to FDA's Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh. Collagen Tendon Sheets have been evaluated in a number of in vitro and in vivo tests to assess its safety/biocompatibility. The device passed all applicable FDA Blue Book Memorandum G95-1and ISO 10993-1 testing for the biological evaluation of medical devices. An animal efficacy study was conducted to evaluate the device as compared to its predicate device. No clinical tests were performed on the product; however clinical history of the predicate device was referenced in the submission. Viral inactivation studies were performed to ensure the viral safety of the product. #### Conclusion of Non-clinical Studies The results of the in vitro product characterization studies, in vitro and in vivo biocompatibility studies, as well as the animal efficacy study show that Collagen Tendon Sheet is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing health, services, and people. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the left side of the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 DEC 2 2 2011 Rotation Medical, Incorporated % Mr. Jeff Sims Vice President, Clinical Programs and Regulatory Affairs 15350 25th Avenue North, Suite 100 Plymouth, Minnesota 55447 Re: K112423 Trade/Device Name: Collagen Tendon Sheet Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTM Dated: December 16, 2011 Received: December 16, 2011 Dear Mr. Sim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ # Page 2 - Mr. Jeff Sims CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Clice-related adverse events) (21 CFR 803); good manufacturing practice requirements as set de rice related duverse oversion (21 CFR Part 820); and if applicable, the electronic forth in the quand on control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you dosire specific ad 100 500 your PDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misoranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vor For Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Confidential Rotation Medical, Inc. # Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: __Collagen Tendon Sheet Indications for Use: Collagen Tendon Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Keane for MKM (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K112423 Page 1 of _1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...