K112396 · Kitazato Medical Co., Ltd. · MQF · Nov 2, 2012 · Obstetrics/Gynecology
Device Facts
Record ID
K112396
Device Name
IUI CATHETERS
Applicant
Kitazato Medical Co., Ltd.
Product Code
MQF · Obstetrics/Gynecology
Decision Date
Nov 2, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6110
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Kitazato IUI Catheters consist of the following versions: - Kitazato IUI Catheter with Stainless Steel Core Type, 10 cm, model number Type 1-v1 . - Kitazato IUI Catheter with Stainless Steel Core Type, 7 cm, model number Type 1-v2 . - Kitazato IUI Catheter without Stainless Steel Core Type, 10 cm, model number Type 2-v1 . - Kitazato IUI Catheter without Stainless Steel Core Type, 7 cm, model number Type 2-v2 . Kitazato IUI Catheters are used for the introduction of washed spermatozoa into the uterine cavity through the cervix.
Device Story
Sterile, single-use catheter for intrauterine insemination (IUI). Device consists of a nylon catheter shaft and a connector with a 6% taper for syringe coupling. Washed spermatozoa are aspirated into the catheter via a syringe and injected into the uterine cavity through the cervix. Models include versions with or without an internal stainless steel core to provide rigidity during placement, similar to a stylet. Some models feature a depth mark and stopper to assist in setting insertion depth. Used by clinicians in a clinical setting. Benefits include controlled delivery of sperm to the uterine cavity to facilitate conception.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by non-clinical bench testing: mechanical tensile strength (4.9N), dimensional verification, endotoxin levels (≤20 EU/device), sterility, and Human Sperm Survival Assay (≥70% motility at 24 hours).
Technological Characteristics
Materials: 12 Nylon catheter body, stainless steel core (select models). Dimensions: 1.65 mm (5 Fr) outer diameter, 7 cm or 10 cm length. Features: Closed rounded tip with side hole, 6% taper connector. Sterilization: Radiation. Single-use.
Indications for Use
Indicated for intrauterine artificial insemination procedures in patients requiring the introduction of washed spermatozoa into the uterine cavity through the cervix.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
Predicate Devices
Gynetics Medical Products N.V. – Smooze Model 4225 (K013501)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
For
# NOV
# Kitazato IUI Catheters – K112396
## 1. Submission Sponsor
KITAZATO Medical Co., Ltd. 2-15-12 Hongo, Bunkyou-ku Tokyo 113-0033 JAPAN Phone: +(81) 3-5800-6055 Fax: +(81) 3-5800-6056 Contact: Mari Yazaki, Quality Assurance Manager
### 2. Submission Correspondent
Emergo Group 611 West 5th Street, Third Floor Austin, TX 78701 Cell Phone: (508) 838.9139 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Richard Vincins, Vice President, QA Email: richard@emergogroup.com
## 3. Date Prepared
30 October 2012
#### 4. Device Name
| Trade/Proprietary Name: | Kitazato IUI Catheter with Stainless Steel Center Core Type, 10<br>cm, model number Type 1-v1<br>Kitazato IUI Catheter with Stainless Steel Center Core Type, 7<br>cm, model number Type 1-v2<br>Kitazato IUI Catheter without Stainless Steel Center Core Type,<br>10 cm, model number Type 2-v1<br>Kitazato IUI Catheter without Stainless Steel Center Core Type,<br>7 cm, model number Type 2-v2 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Kitazato Intrauterine Insemination (IUI) Catheter |
| Classification Name: | Assisted Reproduction Catheter |
| Classification Regulation: | 884.6110 |
| Classification Panel: | Obstetrics/Gynecology |
{1}------------------------------------------------
Product Code: MQF Device Class: =
### 5. Predicate Devices
Gynetics Medical Products N.V. – Smooze Model 4225 – K013501
### 6. Indication for Use
Kitazato IUI Catheters consist of the following versions:
- Kitazato IUI Catheter with Stainless Steel Core Type, 10 cm, model number Type 1-v1 .
- Kitazato IUI Catheter with Stainless Steel Core Type, 7 cm, model number Type 1-v2 .
- Kitazato IUI Catheter without Stainless Steel Core Type, 10 cm, model number Type 2-. v 1
- . Kitazato IUI Catheter without Stainless Steel Core Type, 7 cm, model number Type 2v 2
Kitazato IUI Catheters are used for the introduction of washed spermatozoa into the uterine cavity through the cervix.
### 7. Device Description
Kitazato IUI Catheters are sterile, single-use catheters for use in infusion of washed spermatozoa into the uterine cavity. Catheters are composed of a catheter shaft and a connector. The connector is connected to a syringe (not included with the catheter), and washed spermatozoa are aspirated through the tip of the catheter shaft connected to the syringe. During the insemination procedure, the shaft of catheter is introduced into the uterine cavity through the cervix, and then spermatozoa are injected into the uterine cavity.
The Kitazato IUI Catheter has the following types; IUI Catheter with Stainless Steel Center Core Type and IUI Catheter without Stainless Steel Center Core Type.
IUI Catheter with Stainless Steel Center Core Type: Catheters of this type consist of a catheter body and the connector. The rounded tip of the catheter shaft has a side hole used for aspiration and delivery of sperm. These catheters incorporate a stainless steel core in the center of the catheter shaft to provide rigidity to the catheter during device placement procedures. The connector has a 6% taper that allows it to be coupled with a standard syringe. One model is offered with the shaft of the catheter containing a depth mark and a stopper that aids in setting catheter insertion depth.
| Model | Trade Name | Catheter<br>Body | Catheter<br>Length | Outer<br>Diameter | Depth<br>Mark | Stopper |
|-----------|-------------------------------------------------------|------------------|--------------------|-------------------|---------------|---------|
| Type1 -v1 | IUI Catheter with Stainless<br>Steel Center Core Type | 12 Nylon | 10 cm | 1.65 mm / 5 Fr | Yes @<br>7 cm | Yes |
| Type1 -v2 | IUI Catheter with Stainless<br>Steel Center Core Type | 12 Nylon | 7 cm | 1.65 mm / 5 Fr | No | No |
{2}------------------------------------------------
IUI Catheter without Stainless Steel Center Core Type: Catheters of this type consist of a catheter body and the connector. The rounded tip of the catheter shaft has a side hole used for aspiration and delivery of sperm. These catheters do not include a stainless steel center core in the body of the catheter. The connector has a 6% taper that allows it to be coupled with a standard syringe. One model is offered with the shaft of the catheter containing a depth mark and a stopper that aids in setting catheter insertion depth.
| Model | Trade Name | Catheter<br>Body | Catheter<br>Length | Outer<br>Diameter | Depth<br>Mark | Stopper |
|-----------|-------------------------------------------------------------|------------------|--------------------|-------------------|---------------|---------|
| Type2 -v1 | IUI Catheter without<br>Stainless Steel Center<br>Core Type | 12 Nylon | 10cm | 1.65 mm / 5 Fr | Yes @<br>7 cm | Yes |
| Type2 -v2 | IUI Catheter without<br>Stainless Steel Center<br>Core Type | 12 Nylon | 7cm | 1.65 mm / 5 Fr | No | No |
## Comparison Table - Kitazato IUI Catheter with Stainless Steel Center Core, 10 cm, Type 1-v1
| Manufacturer | KITAZATO Medical Co., Ltd. | Gynetics Medical Products NV | Kitazato IUI Catheter<br>Comparison to Predicate |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Kitazato IUI Catheter with<br>Stainless Steel Center Core | Intra-Uterine Insemination and<br>GIFT Catheters, Smooze #4225 | N/A |
| 510(k) Number | K112396 | K013501 | Same |
| Product Code | MQF | MQF | Same |
| Regulation<br>Number | 884.6110 | 884.6110 | Same |
| Regulation<br>Name | Assisted Reproduction Catheter | Assisted Reproduction Catheter | Same |
| Indications for<br>use: | The Kitazato IUI Catheter with<br>Stainless Steel Center Core is<br>used for the introduction of<br>washed spermatozoa into the<br>uterine cavity through the<br>cervix. | The Intra-Uterine Insemination<br>Cannula is to be used for intra<br>uterine artificial insemination<br>procedures utilizing washed<br>spermatozoa. | Same |
| Overall Design | The device consists of Catheter<br>with Stainless Steel Center Core.<br>The catheters are packaged in a<br>barrier sterilization pouch<br>/wrapping. A syringe is not<br>included in the products. | The device consists of Catheter<br>without Stainless Steel Center<br>Core. The catheters are<br>packaged in a barrier<br>sterilization pouch /wrapping. A<br>syringe is not included in the<br>products. | There are slight differences<br>that are discussed; the<br>stainless steel core is to add<br>rigidity for the catheter to<br>assist if the uterine cervix is<br>curved. This does not impact<br>the use of the device and<br>adds no safety or efficacy<br>concerns as the Stainless<br>Center Core is inside the<br>catheter similar to the<br>function of a stylet that is<br>commonly used with invasive<br>catheters. |
| Sterile | Radiation | Radiation | Same |
| Single-Use | Yes | Yes | Same |
| French Size | 1.65 mm (5 Fr) | 2 mm (6 Fr) | These have similar size outer<br>diameter of catheter; the<br>size difference of 0.35 mm or<br>one French size does not |
| Manufacturer | KITAZATO Medical Co., Ltd. | Gynetics Medical Products NV | Kitazato IUI Catheter<br>Comparison to Predicate |
| Trade Name | Kitazato IUI Catheter with<br>Stainless Steel Center Core | Intra-Uterine Insemination and<br>GIFT Catheters, Smooze #4225 | |
| | | | impact the use of the device<br>and adds no safety or<br>efficacy concerns. |
| Length | 10 cm | 7 cm, 14 cm, 20 cm | The length of the Kitazato IUI<br>catheter is within the range<br>of the predicate device. |
| Depth Marks | Reference mark is located at 7<br>cm from tip | No Depth Mark | The mark at 7 cm from the<br>tip is used as a reference<br>point; the physician<br>measures the length of<br>uterus prior to insertion of<br>the catheter and the mark is<br>used as a guide. This does<br>not impact the use of the<br>device and adds no safety or<br>efficacy concerns. |
| Tip | Closed and smoothly rounded;<br>one side hole end type | Closed and smoothly rounded;<br>one side hole end type | Same |
| Stylet | No; inner stainless steel center<br>core | No | The stainless steel core is to<br>add rigidity for the catheter<br>to assist if the uterine cervix<br>is curved. This does not<br>impact the use of the device<br>and adds no safety or<br>efficacy concerns as the<br>Stainless Steel Center Core is<br>inside the catheter similar to<br>the function of a stylet that is<br>commonly used with invasive<br>catheters. |
| Stopper | Yes | No | The stopper is used as a<br>reference point; the<br>physician measures the<br>length of uterus prior to<br>insertion of the catheter and<br>utilizes the stopper as a<br>guide. This does not impact<br>the use of the device and<br>adds no safety or efficacy<br>concerns |
| Manufacturer | KITAZATO Medical Co., Ltd. | Gynetics Medical Products NV | Kitazato IUI Catheter Comparison to Predicate |
| Trade Name | Kitazato IUI Catheter with<br>Stainless Steel Center Core | Intra-Uterine Insemination and<br>GIFT Catheters, Smooze #4225 | N/A |
| 510(k) Number | K112396 | K013501 | N/A |
| Product Code | MQF | MQF | Same |
| Regulation<br>Number | 884.6110 | 884.6110 | Same |
| Regulation<br>Name | Assisted Reproduction Catheter | Assisted Reproduction Catheter | Same |
| Indications for<br>use: | The Kitazato IUI Catheter with<br>Stainless Steel Center Core is<br>used for the introduction of<br>washed spermatozoa into the<br>uterine cavity through the<br>cervix. | The Intra-Uterine Insemination<br>Cannula is to be used for intra<br>uterine artificial insemination<br>procedures utilizing washed<br>spermatozoa. | Same |
| Overall Design | The device consists of Catheter<br>with Stainless Steel Center Core.<br>The catheters are packaged in a<br>barrier sterilization pouch<br>/wrapping. A syringe is not<br>included in the products. | The device consists of Catheter<br>without Stainless Steel Center<br>Core. The catheters are<br>packaged in a barrier<br>sterilization pouch /wrapping. A<br>syringe is not included in the<br>products. | There are slight differences<br>that are discussed; the<br>stainless steel core is to add<br>rigidity for the catheter to<br>assist if the uterine cervix is<br>curved. This does not impact<br>the use of the device and<br>adds no safety or efficacy<br>concerns as the Stainless<br>Steel Center Core is inside<br>the catheter similar to the<br>function of a stylet that is<br>commonly used with invasive<br>catheters. |
| Sterile | Radiation | Radiation | Same |
| Single-Use | Yes | Yes | Same |
| French Size | 1.65 mm (5 Fr) | 2 mm (6 Fr) | These have similar size outer<br>diameter of catheter; the<br>size difference of 0.35 mm or<br>one French size does not<br>impact the use of the device<br>and adds no safety or<br>efficacy concerns. |
| Length | 7 cm | 7 cm, 14 cm, 20 cm | The length of the Kitazato IUI<br>catheter is within the range<br>of the predicate device. |
| Depth Marks | No Depth Mark | No Depth Mark | Same |
| Tip | Closed and smoothly rounded;<br>one side hole end type | Closed and smoothly rounded;<br>one side hole end type | Same |
| Stylet | No; inner stainless steel center<br>core | No | The stainless steel core is to<br>add rigidity for the catheter<br>to assist if the uterine cervix<br>is curved. This does not<br>impact the use of the device<br>and adds no safety or<br>efficacy concerns as the |
{3}------------------------------------------------
and the control control of the control of the control of
1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -
{4}------------------------------------------------
:
.
| | Comparison Table – Kitazato IUI Catheter with Stainless Steel Center Core, 7 cm, Type 1-v2 | | | |
|--|--------------------------------------------------------------------------------------------|--|--|--|
|--|--------------------------------------------------------------------------------------------|--|--|--|
.
·
{5}------------------------------------------------
| Manufacturer | KITAZATO Medical Co., Ltd. | Gynetics Medical Products NV | Kitazato IUI Catheter<br>Comparison to Predicate |
|--------------|-----------------------------------------------------------|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Kitazato IUI Catheter with<br>Stainless Steel Center Core | Intra-Uterine Insemination and<br>GIFT Catheters, Smooze #4225 | |
| | | | Stainless Steel Center Core is<br>inside the catheter similar to<br>the function of a stylet that is<br>commonly used with invasive<br>catheters. |
| Stopper | No | No | Same |
## Comparison Table – Kitazato IUI Catheter without Stainless Steel Center Core Type, 10 cm, Type 2-v1
| Manufacturer | KITAZATO Medical Co., Ltd. | Gynetics Medical Products NV | Kitazato IUI Catheter Comparison to Predicate |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Kitazato IUI Catheter without<br>Stainless Steel Center Core | Intra-Uterine Insemination and<br>GIFT Catheters, Smooze #4225 | N/A |
| 510(k) Number | K112396 | K013501 | Same |
| Product Code | MQF | MQF | Same |
| Regulation<br>Number | 884.6110 | 884.6110 | Same |
| Regulation<br>Name | Assisted Reproduction Catheter | Assisted Reproduction Catheter | Same |
| Indications for<br>use: | The Kitazato IUI Catheter is used<br>for the introduction of washed<br>spermatozoa into the uterine<br>cavity through the cervix. | The Intra-Uterine Insemination<br>Cannula is to be used for intra<br>uterine artificial insemination<br>procedures utilizing washed<br>spermatozoa | Same |
| Overall Design | The device consists of Catheter<br>without Stainless Steel Center<br>Core. The catheters are<br>packaged in a barrier<br>sterilization pouch /wrapping. A<br>syringe is not included in the<br>products. | The device consists of Catheter<br>without Stainless Center Core.<br>The catheters are packaged in a<br>barrier sterilization pouch<br>/wrapping. A syringe is not<br>included in the products. | Same |
| Sterile | Radiation | Radiation | Same |
| Single-Use | Yes | Yes | Same |
| French Size | 1.65 mm (5 Fr) | 2 mm (6 Fr) | These have similar size outer<br>diameter of catheter; the<br>size difference of 0.35 mm or<br>one French size does not<br>impact the use of the device<br>and adds no safety or<br>efficacy concerns. |
.
•
{6}------------------------------------------------
| Manufacturer<br>Trade Name | KITAZATO Medical Co., Ltd.<br>Kitazato IUI Catheter without<br>Stainless Steel Center Core | Gynetics Medical Products NV<br>Intra-Uterine Insemination and<br>GIFT Catheters, Smooze #4225. | Kitazato IUI Catheter<br>Comparison to Predicate |
|----------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Length | 10 cm | 7 cm, 14 cm, 20 cm | The length of the Kitazato IUI<br>catheter is within the range<br>of the predicate device. |
| Depth Marks | Reference mark is located at 7<br>cm from tip | No Depth Mark | The mark at 7 cm from the<br>tip is used as a reference<br>point; the physician<br>measures the length of<br>uterus prior to insertion of<br>the catheter and the mark is<br>used as a guide. This does<br>not impact the use of the<br>device and adds no safety or<br>efficacy concerns. |
| Tip | Closed and smoothly rounded;<br>one side hole end type | Closed and smoothly rounded;<br>one side hole end type | Same |
| Stylet | No | No | Same |
| Stopper | Yes | No | The stopper is used as a<br>reference point; the<br>physician measures the<br>length of uterus prior to<br>insertion of the catheter and<br>utilizes the stopper as a<br>guide. This does not impact<br>the use of the device and<br>adds no safety or efficacy<br>concerns. |
## Comparison Table – Kitazato IUI Catheter without Stainless Steel Center Core Type, 7 cm, Type 2-v2
| Manufacturer | KITAZATO Medical Co., Ltd. | Gynetics Medical Products NV | Kitazato IUI Catheter<br>Comparison to Predicate |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Trade Name | Kitazato IUI Catheter without<br>Stainless Steel Center Core | Intra-Uterine Insemination and<br>GIFT Catheters, Smooze #4225 | |
| 510(k) Number | K112396 | K013501 | N/A |
| Product Code | MQF | MQF | Same |
| Regulation<br>Number | 884.6110 | 884.6110 | Same |
| Regulation<br>Name | Assisted Reproduction Catheter | Assisted Reproduction Catheter | Same |
| Indications for<br>use: | The Kitazato IUI Catheter is used<br>for the introduction of washed<br>spermatozoa into the uterine<br>cavity through the cervix. | The Intra-Uterine Insemination<br>Cannula is to be used for intra<br>uterine artificial insemination<br>procedures utilizing washed<br>spermatozoa | Same |
| Overall Design | The device consists of Catheter<br>without Stainless Steel Center<br>Core. The catheters are<br>packaged in a barrier<br>sterilization pouch /wrapping. A<br>syringe is not included in the | The device consists of Catheter<br>without Stainless Center Core.<br>The catheters are packaged in a<br>barrier sterilization pouch<br>/wrapping. A syringe is not<br>included in the products. | Same |
·
{7}------------------------------------------------
| Manufacturer | KITAZATO Medical Co., Ltd. | Gynetics Medical Products NV | Kitazato IUI Catheter Comparison to Predicate |
|--------------|---------------------------------------------------------------------------|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Kitazato IUI Catheter without<br>Stainless Steel Center Core<br>products. | Intra-Uterine Insemination and<br>GIFT Catheters, Smooze #4225 | |
| Sterile | Radiation | Radiation | Same |
| Single-Use | Yes | Yes | Same |
| French Size | 1.65 mm (5 Fr) | 2 mm (6 Fr) | These have similar size outer<br>diameter of catheter; the<br>size difference of 0.35 mm or<br>one French size does not<br>impact the use of the device<br>and adds no safety or<br>efficacy concerns. |
| Length | 7 cm | 7 cm, 14 cm, 20 cm | The length of the Kitazato IUI<br>catheter is within the range<br>of the predicate device. |
| Depth Marks | No Depth Mark | No Depth Mark | Same |
| Tip | Closed and smoothly rounded;<br>one side hole end type | Closed and smoothly rounded;<br>one side hole end type | Same |
| Stylet | No | No | Same |
| Stopper | No | No | Same |
## 8. Technological Characteristics
The indication for use and technology of the Kitazato IUI Catheters is substantially equivalent to the identified predicate devices.
## 9. Non-Clinical Testing
The catheter mechanical tensile testing, dimension testing, endotoxin testing, sterility testing and Human Sperm Survival Assay results support that all the specifications have met the acceptance criteria for the device.
- . Mechanical Tensile Testing: Tensile strength to withstand 4.9N
- . Dimensional Testing: Passes outer diameter and length according to specifications
- Endotoxin Testing: Endotoxin values conform to the value ≤20 EU/device
- . Sterility Testing: No microbial growth from sterility testing
- . Human Sperm Survival Assay: ≥70% motility at 24 hours
The Kitazato IUI Catheters passed all testing and supports the claims of substantial equivalence and safe operation.
The Kitazato IUI Catheters complies with the applicable voluntary standards for biocompatibility and sterilization. The device passed all the testing in accordance with national and international standards.
{8}------------------------------------------------
#### 10. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The substantial equivalence of the device is supported by the non-clinical testing. The validation testing of the device biocompatibility and HSSA testing was found to be acceptable and supports the claims of substantial equivalence.
## 11. Conclusion
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that any differences between the Kitazato IUI Catheters and the predicate device do not raise any questions regarding its safety and effectiveness. The Kitazato IUI Catheters, as designed and manufactured, are substantially equivalent to the referenced predicate devices.
{9}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three wavy lines below the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Letter Date: November 2, 2012
KITAZATO Medical Co., Ltd. % Mr. Richard Vincins, CQA, CBA, RAC (US, EU) Vice President, QA Emergo Group, Inc. 611 West 5th Street, Third Floor AUSTIN TX 78701
Re: K112396
Trade/Device Name: Kitazato IUI Catheter with Stainless Steel Core Type, 10 cm, model number Type 1-v1 and 7 cm, model number Type 1-v2 Kitazato IUI Catheter without Stainless Steel Core Type, 10 cm, model number Type 2-v1 and 7 cm, model number Type 2-v2 Regulation Number: 21 CFR§ 884.6110 Regulation Name: Assisted reproduction catheters Regulatory Class: II
Product Code: MQF
Dated: October 22, 2012
Received: October 22, 2012
Dear Mr. Vincins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for answer of to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may rublish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
# Page 2 - Mr. Richard Vincins, CQA, CBA, RAC (US, EU)
Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 lease be advised that I Drisean.teat your device complies with other requirements of the Act that I Dr has Intact a and regulations administered by other Federal agencies. You must or any 1 outlated and the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of it Fart 60 7), accember (21 CFR 803); good manufacturing practice requirements as set de recording systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you desire specific advisiboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regulation untiled, "Mischange of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou may other baller generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Benjamin** R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K112396
Device Name: Kitazato IUI Catheter
Indications for Use:
Kitazato IUI Catheters consist of the following versions:
- Kitazato IUI Catheter with Stainless Steel Core Type, 10 cm, model number Type 1-v1 .
- Kitazato IUI Catheter with Stainless Steel Core Type, 7 cm, model number Type 1-v2 .
- Kitazato IUI Catheter without Stainless Steel Core Type, 10 cm, model number Type 2-v1 .
- Kitazato IUI Catheter without Stainless Steel Core Type, 7 cm, model number Type 2-v2 .
Kitazato IUI Catheters are used for the introduction of washed spermatozoa into the uterine cavity through the cervix.
(Part 21 CFR 801 Subpart D) Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use ______________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| | Benjamin R. Fisher -S |
|--|----------------------------|
| | 2012.11.02 16:12:25-04'00' |
Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K112396
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.