K112392 · Hivox Biotek, Inc. · NUH · Mar 19, 2012 · Neurology
Device Facts
Record ID
K112392
Device Name
HIVOX BIOTEK INC.
Applicant
Hivox Biotek, Inc.
Product Code
NUH · Neurology
Decision Date
Mar 19, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HIVOX Electric Stimulator OTC TENS, Pennypad 904 and Pennypad 909 are indicated for temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and work activities. The HIVOX Electric Stimulator OTC TENS, Pennypad 907 is indicated for Arm & Leg Pain Relief is intended for temporary relief of pain associated with sore and aching muscles in the upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
Device Story
HIVOX Electric Stimulator OTC TENS (Pennypad 904, 907, 909) is a transcutaneous electrical nerve stimulator; generates small electrical pulses; delivered via skin electrodes to activate underlying nerves; intended for temporary pain relief. Device is compact (113 x 70 x 9.7 mm); operated by patient in home/non-clinical setting. Output parameters (voltage, current, frequency) are user-adjustable; Pennypad 909 includes a 'Burst Mode'. Device provides symptomatic relief of muscle pain; helps patients manage discomfort from strain without prescription intervention.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by bench testing, including electrical safety, electromagnetic compatibility, and waveform performance testing.
Technological Characteristics
Transcutaneous electrical nerve stimulator; symmetrical biphasic rectangular waveform; dimensions 113 x 70 x 9.7 mm. Complies with IEC/EN 60601-1 (safety), IEC/EN 60601-1-2 (EMC), and IEC/EN 60601-2-10 (nerve/muscle stimulator safety).
Indications for Use
Indicated for temporary relief of pain in sore/aching muscles of the lower back (Pennypad 904/909) or upper/lower extremities (Pennypad 907) caused by strain from exercise or daily activities. For OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Limited Function OTC TENS, WELL-LIFE Healthcare Limited (K040512)
Submission Summary (Full Text)
{0}------------------------------------------------
HIVOX BIOTEK INC.
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan . Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
## SUMMARY OF SAFETY AND EFFECTIVENESS for HIVOX Electric Stimulator OTC TENS
HIVO>
MAR 1 9 2012
## 510(k): K112392
| DATE OF | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMISSION: | July 13, 2011 |
| SUBMITTER: | HIVOX BIOTEK INC. |
| | 5F, No.123, Shinde Road, Sanchong Dist.,<br>New Taipei City, 24158, TAIWAN, R.O.C.<br>TEL: 886-2-85112668 FAX:886-2-85112669 |
| ESTABLISHMENT<br>REGISTRATION NO: | 9611558 |
| OFFICIAL<br>CONTACT: | Dr. JEN, KE-MIN<br>TEL: 886-2-85112668 FAX:886-3-5209783<br>Email: ceirs.jen@msa.hinet.net |
| TRADE NAME: | HIVOX Electric Stimulator OTC TENS<br>Pennypad 904 and Pennypad 909 : OTC TENS - Lower<br>Back pain relieve<br>Pennypad 907: OTC TENS - Arm and Leg pain relief |
| COMMON/USUAL<br>NAME: | Stimulator, Nerve, Transcutaneous,<br>Over-The-Counter |
| CLASSIFICATION<br>NAME: | Transcutaneous Nerve Stimulator |
| REGULATION<br>NUMBER: | NUH, Class II, 882.5890 |
| SECONDARY<br>PRODUCT CODE: | GZJ, Class II, 882.5890 |
| PREDICATED<br>DEVICE: | Therapeutic Massage Companion, Endurance Therapeutics<br>1) K060846, OTC TENS for Lower Back, Arm and Leg<br>Pain Relief, T1040<br>WELL-LIFE Healthcare Limited<br>2) K040512, Limited Function OTC TENS, for Lower<br>Back Pain Relief, WL-2402, WL-2403 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "HIVOX" in bold, sans-serif font. To the right of the word is a symbol that looks like a plus sign made of squares. The symbol is also in bold and has a similar line thickness to the letters.
ﻪ ﻣﺮ
## HIVOX BIOTEK INC.
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
| Intended Use: | The HIVOX Electric Stimulator OTC TENS, Pennypad 904 and<br>Pennypad 909 are indicated for temporary relief of pain associated<br>with sore and aching muscles in the lower back due to strain from<br>exercise or normal household and work activities.<br>The HIVOX Electric Stimulator OTC TENS, Pennypad 907 is<br>indicated for Arm & Leg Pain Relief is intended for temporary<br>relief of pain associated with sore and aching muscles in the upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description of Device: | The HIVOX Electric Stimulator OTC TENS, Pennypad 904,<br>Pennypad 907, and Pennypad 909 can generate small pulses of<br>electrical current. Delivered these pulses pass through the skin and activated underlying nerves. |
| Performance Tests<br>Submitted: | The relevant standards including:<br>1. IEC/EN 60601-1 : Medical electrical equipment Part 1.<br>General requirements for safety, 1996.<br>2. IEC/EN 60601-1-2 : Medical electrical equipment, Part 2.<br>Electromagnetic compatibility - Requirements and tests, 2007.<br>3. IEC/EN 60601-2-10 : Medical electrical equipment, Part 2-10:<br>Particular requirements for safety of nerve and muscle<br>stimulators, 2001. |
| Non-Clinical Tests<br>Submitted: | The HIVOX Electric Stimulator OTC TENS, Pennypad 904,<br>Pennypad 907, and Pennypad 909 have been tested in accordance<br>with applicable standards for medical device electrical safety,<br>electromagnetic compatibility, and the particular requirements for Accessories also meet safety requirements: 510(k) electrodes are<br>specified the safety requirements.<br>System level testing including waveform testing was performed in combination the HIVOX Electric Stimulator OTC TENS. |
| Clinical Tests<br>Submitted: | None |
| Conclusion: | As the product description and tests as above, the new devices:<br>HIVOX Electric Stimulator OTC TENS, Pennypad 904, Pennypad<br>907, and Pennypad 909 are as safe and effective as, and the<br>function in a manner equivalent to the predicate devices: K060846,<br>OTC TENS for Lower Back, Arm and Leg Pain Relief, T1040; and<br>K040512, Limited Function OTC TENS, for Lower Back Pain<br>Relief, WL-2402, WL-2403.<br>Thus the new device is substantially equivalent to the predicate devices in this aspect. |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "HIVOX" in a bold, sans-serif font. To the right of the "X" is a small, stylized graphic that resembles a plus sign made up of small squares. The letters are uniformly black, providing a stark contrast against the white background.
#### HIVOX BIOTEK INC.
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 - Fax: +886 2 8511 2669
# Comparison for Predicate Device & Subject Device
## ● COMPARISON TABLE 1
We present the relevant information for the predicate device here for demonstrating the characteristics of the predicate device.
| Parameter | HIVOX | Well Life<br>(K)040512. | Endurance<br>Therapeutics<br>(K)060846. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|----------------------------------------------------|----------------------|-------------------------|-----------------------------------------|-----------------------|------------------|------|---------------------------------------------|-------|-------|-------|-------|-------------------------|------------------------------------------|-------|-------|-------|-----------------------|------------------|------------|--------------------|----|--------------|-----|----|---------------------|----|---------------|----|----|-------------------------|----|---------------------|----|----|---------------|----|----|----|----|--------------|-----|--|----|----|---------------|---|--|----|----|---------------------|----|--|
| Mode or Program Name | PP909 | PP904 | 2402/2403 | T1040 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Waveform | Symmetrical Biphasic | | Biphasic | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Shape | Rectangular | | Rectangular | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Maximum Output Voltage<br>(Volts) (±20%) | @500Ω | 40.0Vpp | 57.6Vpp | 38.2V rms | 40.7V | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | @2KΩ | 84.0Vpp | 89.6Vpp | 52.8V rms | 105.1V | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | @10KΩ | 92.0Vpp | 96.0Vpp | 84.8V rms | 154.1V | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Maximum Output<br>Current(±20%) | @500Ω | 80.0mApp | 115.2mApp | 76.2mA rms | 81.4mA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | @2KΩ | 42.0mApp | 44.8mApp | 52.8mA rms | 47.8mA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | @10KΩ | 9.2mApp | 9.6mApp | 8.48mA rms | 15.4mA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Duration of primary (depolarizing)<br>phase (µSec) | NA | | NA | NA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Pulse Duration (µSec) | 200µSec (fixed) | | 250 µSec | 4.1~500mS | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Frequency (Hz) | 35 | | 2, 5, and 40 | 25Hz | 245Hz | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Net Charge (µC)<br>per pulse | @500Ω | 0.3200 | 0.2304 | 16.0/8.76 | 4.07 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Maximum Charge (µC) | @500Ω | 16.0 | 23.04 | 12.5/9.2 | 16.9 | Maximum Current Density<br>(mA/cm², r.m.s.) | @500Ω | 1.964 | 2.828 | 0.006 | 2.71 | Maximum Average Power<br>Density (W/cm²) | @500Ω | 0.078 | 0.163 | 0.000848/<br>0.002292 | 5.35<br>(mW/cm²) | Burst Mode | a. Pulse per burst | 1 | NA | NA | NA | b. Burst per second | 4 | NA | NA | NA | c. Burst duration (sec) | 2 | NA | NA | NA | d. Duty Cycle | 8 | NA | NA | NA | ON Time(sec) | 120 | | NA | NA | Off Time(sec) | 0 | | NA | NA | Additional Features | NA | |
| Maximum Charge (µC) | @500Ω | 16.0 | 23.04 | 12.5/9.2 | 16.9 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Maximum Current Density<br>(mA/cm², r.m.s.) | @500Ω | 1.964 | 2.828 | 0.006 | 2.71 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Maximum Average Power<br>Density (W/cm²) | @500Ω | 0.078 | 0.163 | 0.000848/<br>0.002292 | 5.35<br>(mW/cm²) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Burst Mode | a. Pulse per burst | 1 | NA | NA | NA | b. Burst per second | 4 | NA | NA | NA | c. Burst duration (sec) | 2 | NA | NA | NA | d. Duty Cycle | 8 | NA | NA | NA | ON Time(sec) | 120 | | NA | NA | Off Time(sec) | 0 | | NA | NA | Additional Features | NA | | NA | NA | | | | | | | | | | | | | | | | |
| Burst Mode | | a. Pulse per burst | 1 | NA | NA | NA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | b. Burst per second | 4 | NA | NA | NA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | c. Burst duration (sec) | 2 | NA | NA | NA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | d. Duty Cycle | 8 | NA | NA | NA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| ON Time(sec) | 120 | | NA | NA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Off Time(sec) | 0 | | NA | NA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Additional Features | NA | | NA | NA | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Note: The subject device only Pennypad 909 has the "Burst Mode".
Formula
inula
Current Density – ((Max Voltage@500ohm)x1000)/electrode area (40.732cm²)
Power Density – (Amp x V)/ Electrode Area(40.732cm²)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a blurry, black and white close-up of what appears to be a small object or mark. The object is positioned off-center, with some indistinct features. Due to the lack of clarity, it is difficult to determine the exact nature or context of the image.
# HIV
## HIVOX BIOTEK INC.
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
### ● COMPARISON TABLE 2
We present the relevant information for the predicate device here for demonstrating the characteristics of the predicate device.
| Parameter | | HIVOX | Endurance<br>Therapeutics<br>(K)060846 |
|-------------------------------------------------------|-------------------------|-------------------------|----------------------------------------|
| Mode or Program Name | | PP907 | T1040 |
| Waveform | | Symmetrical<br>Biphasic | Biphasic |
| Shape | | Rectangular | Rectangular |
| Maximum Output Voltage (Volts) (±20%)<br>Peak to peak | @500Ω | 68.8V | 40.7V rms |
| | @2KΩ | 88.0V | 105.1 V rms |
| | @10KΩ | 95.2V | 154.1V rms |
| Maximum Output Current(±20%)<br>Peak to peak | @500Ω | 137.6mA | 81.4mA rms |
| | @2KΩ | 44.0mA | 47.8mA rms |
| | @10KΩ | 9.52mA | 15.4mArms |
| Duration of primary (depolarizing) phase (µSec) | | NA | NA |
| Pulse Duration (µSec) | | 200µSec (fixed) | 4.1~500mS |
| Frequency (Hz) | | 2 and 40 | 245Hz |
| Net Charge (µC) per pulse | @500Ω | 1.1008 | 4.07 |
| Maximum Charge (µC) | @500Ω | 27.52 | 16.9 |
| Maximum Current Density<br>(mA/cm², r.m.s.) | @500Ω | 3.378 | 2.71 |
| Maximum Average Power Density<br>(W/cm²) | @500Ω | 0.232 | 5.35<br>(mW/cm²) |
| Burst Mode | a. Pulse per burst | NA | NA |
| | b. Burst per second | NA | NA |
| | c. Burst duration (sec) | NA | NA |
| | d. Duty Cycle | NA | NA |
| ON Time(sec) | | 120 | NA |
| Off Time(sec) | | 0 | NA |
| Additional Features | | NA | NA |
#### Formula:
r critical
Current Density – ((Max Voltage(@500ohm)x1000)/electrode area (40.732cm²) Power Density - (Amp x V)/ Electrode Area(40.732cm2)
{4}------------------------------------------------
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
#### DISCUSSION:
・・
The HIVOX Electric Stimulator OTC TENS, Pennypad 904, Pennypad 907, and Pennypad 909 have the same visional appearance, software, and dimensions ( 113 * 70 * 9.7 mm ) the main difference is:
- Pennypad 904 and Pennypad 909 : OTC TENS Lower Back pain relieve ●
- Pennypad 907: OTC TENS Arm and Leg pain relief ●
- Only Pennypad 909 has the "Burst Mode".
As the product description and tests as above, the new devices: HIVOX Electric Stimulator OTC TENS, Pennypad 904, Pennypad 907, and Pennypad 909 are as safe and effective as, and the function in a manner equivalent to the predicate devices: K060846, OTC TENS for Lower Back, Arm and Leg Pain Relief, T1040; and K040512, Limited Function OTC TENS, for Lower Back Pain Relief, WL-2402, WL-2403.. Thus the new devices are substantially equivalent to the predicate devices in this aspect.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
HIVOX BIOTEK, Inc. c/o Dr. Ke-Min Jen Official Correspondent 5F., No. 123, Shinde Road Sanchong District New Taipei City China, Taiwan 24158
MAR 1 9 2012
Re: K112392
Trade/Device Name: HIVOX Electric Stimulator OTC TENS Models: Pennypad 904, 909 -- Lower back pain relief Pennypad 907 - Arm and Leg pain relief Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH Dated: February 24, 2012 Received: March 8, 2012
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be fray of our of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Dr. Ke-Min Jen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Kesia Alexander
\$\fn\$
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number:
KII2392
Device Name:
HIVOX Electric Stimulator OTC TENS Pennypad 904, 909 : OTC TENS – Lower back pain relieve Pennypad 907: OTC TENS – Arm and Leg pain relief
#### Indications for Use :
The Pennypad 904 and Pennypad 909 are indicated for temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and work activities.
The Pennypad 907 is indicated for Arm & Leg Pain Relief is intended for temporary relief of pain associated with sore and aching muscles in the upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
Prescription Use _____________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ V (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---------------------|--|
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
Page 1 of 1K112392
< 1 ( V ) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.