K112362 · Silver Bay, LLC · OHS · Aug 14, 2012 · General, Plastic Surgery
Device Facts
Record ID
K112362
Device Name
C100 WRINKLE REDUCTION DEVICE
Applicant
Silver Bay, LLC
Product Code
OHS · General, Plastic Surgery
Decision Date
Aug 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Quasar C100 Wrinkle Reduction Device is intended to emit energy in the red and IR region of the spectrum, specifically indicated for the treatment of periorbital wrinkles.
Device Story
Quasar Calypso C100 is a handheld light therapy device for home use. It utilizes an array of 20 red and 20 infrared LEDs (610nm to 850nm) to deliver light energy to the skin. Powered by a 12V wall-mount supply, the device emits a constant waveform with a total output of 65 mW/cm². The user operates the device by placing the light head directly on or slightly above the skin in the periorbital area for 3 minutes daily, at least 5 days per week. The device is intended to reduce periorbital wrinkles. It is constructed of ABS plastic with polycarbonate lenses. No complex software or AI algorithms are involved; operation is limited to an on/off switch.
Clinical Evidence
Bench testing only. The device complies with IEC/EN 60601-1 (electrical safety), IEC/EN 60601-1-2 (EMC), and IEC 62471 (photobiological safety). No clinical data provided.
Technological Characteristics
Handheld LED light therapy system. 20 red and 20 infrared LEDs (610-850nm). Constant waveform. 65 mW/cm² output. Materials: ABS plastic body, polycarbonate lenses. Power: 12V wall-mount supply. Standards: IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 62471. No software/algorithm.
Indications for Use
Indicated for the treatment of periorbital wrinkles in individuals. Not intended for ocular applications or direct eye exposure.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K112362 P. 1 of 9
1 of 5
# AUG 1 4 2012
# 510(k) Summary
K
This summary of 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 878.4810.
Submission Date: August 1, 2011
1. Submitter Information:
AEGIS Regulatory, Inc. - Robert T. Wagner 1131 Anthem View Lane Knoxville, TN 37922 Tel.: 865-982-5552 Email: bob@fdalistingconsultants.com Silver Bay, LLC d/b/a Quasar Bio-Technologies Attn: Peter Nesbitt 1431 Tallevast Rd. Sarasota, FL 34231 Tel.: 941-306-5812 Email: peter@quasarbiotech.com
# For Manufacturer:
2. General Information
2.1 Classification Name: Wrinkle Reduction Device
2.2 Common/Usual Name: Light Therapy System, Quasar Calypso C100
2.3 Proprietary Names: Quasar Calypso C100 Wrinkle Reduction Device
2.4 Classification: Class II
2.5 Classification Number: 878.4810
2.6 Product Codes: OHS
#### 3. Device Description:
The Silver Bay, Quasar Calypso, C100 Anti-Wrinkle device consists of a collection of red and near infrared diodes (LEDs), packaged in a compact hand held device. The device has a head containing the LED array and an on/off switch.
The device is made of ABS plastic with clear polycarbonate lenses covering the LED light source. The Quasar Calypso C100 uses a 12 volt wall mount power supply.
{1}------------------------------------------------
K112362. p. 2.0+5
The C100 light delivery system used for applying therapy for the use in the treatment of periorbital wrinkles, by emitting at least 65 mW/cm² of red and near infrared (610nm to 850nm) light via an electic light emitting diodes [LEDs] energy source. There are 20 red LEDs and 20 infrared LEDs in the head. The device is not intended for ocular applications or direct eye exposure.
### 4. Indications / Intended Use:
The Quasar Calypso C100 is an Over-The-Counter handheld device intended to emit energy in the red/IR spectrum, specifically indicated for use in the treatment of periorbital wrinkles.
Rx or OTC:
The C100 is an Over the Counter (OTC) device. The labeling, instructions, and User Operations (21 CFR § 801.60 and 61), are designed for layman understanding and use. The predicate device is OTC.
#### 5. Predicate Device:
This device is substantially equivalent to the following predicate devices, which are currently in safe and effective commerce:
1. K101190 - Light for Wrinkles (Led Intellectual Properties, LLC)
#### Predicate Chart
| Device | Light for Wrinkles | C100 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | AAL OCT (K101190)<br>LED Intellectual<br>Properties LLC | Quasar Bio-Tech Inc |
| | Properties LLC | K112362 |
| | A Predicate Device | This Submission |
| Indications | The Light for Wrinkles<br>is an Over-The-Counter<br>handheld device<br>intended for use in the<br>treatment of<br>periorbital wrinkles.<br>The target patient<br>population for the AAL<br>OCT is persons with<br>periorbital wrinkles. It<br>is designed for home<br>use. | The C100 is intended<br>to emit energy in the<br>red and IR region of<br>the spectrum,<br>specifically indicated<br>for use in the<br>treatment of<br>periorbital wrinkles.<br>The target patient<br>population for the<br>C100 is the same as the<br>predicate devices. Like<br>the predicate devices,<br>the C100 is designed |
| Device | Light for Wrinkles | C100 |
| | AAL OCT (K101190)<br>LED Intellectual<br>Properties LLC | Quasar Bio-Tech Inc<br>K112362 |
| | A Predicate Device | This Submission |
| | | for home use. |
| Handheld | Yes | Yes |
| Wavelength<br>s | 605nm, 630nm,<br>660nm, 855nm | 610, 630 and 660,<br>850nm |
| Modes | On/Off | On/Off |
| IR power<br>source | LEDs | LEDs |
| Visible light<br>LEDs | Yes | Yes |
| Waveform | Constant | Constant |
| Energy<br>Source | 70 LED's over 27 sq.<br>cm. | 40 LEDs. Over 20 sq.<br>cm |
| Energy<br>Level | 65 mW total | 65 mW total |
| Power<br>Supply | 115VAC Electric Outlet<br>Power Supply | 115VAC Electric Outlet<br>Power Supply |
| Treatment<br>Time | 3 minutes daily,<br>minimum 5 days per<br>week | 3 minutes daily,<br>minimum 5 days per<br>week |
| Target<br>Population | Individuals with<br>periorbital lines and<br>wrinkles. | Individuals with<br>periorbital lines and<br>wrinkles. |
| Location for<br>Use | OTC | OTC |
{2}------------------------------------------------
Summary of the technological characteristics of the device compared to predicate device:
·
.
{3}------------------------------------------------
U of 5
The C100 and the above referenced predicate. Light for Wrinkles device, are Over the Counter Devices used to treat wrinkles as defined in 21 CFR § 878.4810. These devices utilize red and infrared diodes from 610 to 850 nm to provide narrow bands of light energy to treat wrinkles. The performance achieved by these devices is similar with equal power output. The devices are handheld, and intended to be placed directly on the skin or held just over the skin. They are manufactured out of similar materials. Based upon comparison to the predicate device, the C100 has the same intended uses, with similar technological characteristics as the predicate device. The system performs as intended and does not raise any new safety or effectiveness issues.
#### 6. Biocompatibility:
The only patient contact material in the C100 is the light head and body of the device.
The light head in contact with the face is Polycarbonate and the body is constructed of ABS plastic. the same materials used in the predicate device. The biocompatibility of these materials are well known and considered safe when in contact with heatthy skin. A review of the Biocompatibility decision is shown on the "General Program Memorandum- #G95-1. Attachment C, Biocompatibility Flow Chart for the Selection of Toxicity Tests for 510(k)s."
The conclusion is that the C100 m does not raise any new safety issues.
#### 7. Performance Testing and Standards:
These devices have been tested under and are in compliance with performance standards that have been established for such devices under Section 878 of the Federal Food, Drug, and Cosmetics Act. All electrical and radiological products made by the applicant have been OSHA/NRTL listed, and have received constituent marks.
The device has been tested and is in conformity with IEC/EN 60601-1, IEC/ EN 60601-1-2, IEC 62471 Standards.
#### 8. Statement of Safety and Effectiveness:
The information in this 510(k) submission was used to support the safety and effectiveness of this device with respect to its cited predicates.
#### 9. Substantial Equivalence Discussion
After an analysis of the safety, indications, intended uses,
performance, features, design materials, power output, technological properties, treatment areas, treatment regimes and methods of operation, the manufacturer believes that no
{4}------------------------------------------------
significant differences exist between the device and the predicates listed in Section 5 Therefore substantial equivalency is requested.
:
·
K112362 2.5
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's symbol, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Silver Bay, LLC % Aegis Regulatory, Incorporated Mr. Robert T. Wagner CEO 1131 Anthem View Lane Knoxville, Tennessee 37922
AUG 1 4 2012
Re: K112362
Trade/Device Name: Quasar C100 Wrinkle Reduction Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: August 13, 2012 Received: August 13, 2012
Dear Mr. Wagner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
#### Page 2 - Mr. Robert T. Wagner
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Millman
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Ki12362
# INDICATIONS FOR USE STATEMENT
510(k) Number: K112362
Device Name: Quasar C100 Wrinkle Reduction Device
Indications for Use:
The Quasar C100 Wrinkle Reduction Device is intended to emit energy in the red and IR region of the spectrum, specifically indicated for the treatment of periorbital wrinkles.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96)
Page __ of __
(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic,
Division of Surgices Division of Surgh - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
K112362
510(k) Number
Silver Bay LLC Quasar C100
Premarket Notification
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.