K112351 · Johari Digital Healthcare , Ltd. · NFO · Mar 30, 2012 · Neurology
Device Facts
Record ID
K112351
Device Name
ULTRA BEAUTY
Applicant
Johari Digital Healthcare , Ltd.
Product Code
NFO · Neurology
Decision Date
Mar 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ultra Beauty device is intended for facial stimulation and is indicated for over-the-counter cosmetic use.
Device Story
Ultra Beauty is a portable, battery-powered microcurrent device for cosmetic facial stimulation. It features a base module with an LCD display and MKB control panel, connected to two hand pieces via conductive cables. The device utilizes two types of probes (T-bar and two-prong) to deliver electrical stimulation to the face. Users select from three pre-programmed modes to customize treatment protocols. A high-speed controller manages parameters, including pulse width, rate, and amplitude, while providing feedback-based safety monitoring; the system shuts down upon component failure. The device is intended for over-the-counter use, allowing consumers to perform facial treatments at home. Output intensity and treatment progress are monitored via the LCD screen and LED indicators on the hand pieces. The device is powered by a 3.7V rechargeable Lithium-ion battery.
Clinical Evidence
Bench testing only. The device underwent electrical safety and electromagnetic compatibility testing per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10. Software verification was performed according to FDA guidance. No clinical data provided.
Technological Characteristics
ABS enclosure; 3.7V/1200mAh rechargeable Lithium-ion battery; microcurrent stimulation; two hand pieces with T-bar and two-prong probes; LCD display; MKB control panel; high-speed controller for parameter management and feedback-based safety shutdown. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971:2007.
Indications for Use
Indicated for over-the-counter cosmetic facial stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
Rejuvenique
NuFACE
FaceMaster
Submission Summary (Full Text)
{0}------------------------------------------------
510(k): k 112351
MAR 3 0 2012
# Summary Traditional 510(k)
: JOHARI DIGITAL HEALTHCARE LTD. Submitter's Name Electronic Hardware Technology Park G-582, 583, E.P.I.P., Boranada, Jodhpur, INDIA-342008 : Pooja Johari · Contact person : pjohari@betterlifedevices.com, 3 E-mail pooja.johari@7ellc.com : 858 -6579226 Phone Fax : 858-220-7523 Date of Summary Submission : August 05, 2011
Resubmitting on.
: N.A.
{1}------------------------------------------------
| NEW DEVICE FOR WHICH SUBMITTING | |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | : Ultra Beauty |
| Common Name | : Transcutaneous Electrical Nerve Stimulator |
| Classification Name | : Stimulator, Nerve, Transcutaneous, Aesthetic Purposes |
| Device's Classification Panel | : Neurology (As Per 21 CFR Section 882.5890) |
| Regulatory Class | : Class II |
| Product Code | : NFO |
| LEGALLY MARKETING DEVICE | |
| Rejuvenique<br>Manufacturer<br>Address | : Transcutaneous Electrical Nerve Stimulator<br>: Salton Inc<br>: 1955 W. Field Court<br>Lake Forest, IL 60045 |
| NuFACE<br>Manufacturer<br>Address | : Transcutaneous Electrical Nerve Stimulator<br>: Carol Cole Company<br>: 147 Basil Street<br>Encinitas, California 92024 |
| FaceMaster<br>Manufacturer<br>Address | : Transcutaneous Electrical Nerve Stimulator<br>: FaceMaster of Beverly Hills Inc<br>: 2603 Main Street 1170<br>Irvine, California 92614 |
#### DESCRIPTION OF NEW DEVICE ULTRA BEAUTY
Ultra Beauty is a microcurrent device that provides facial stimulation for cosmetic purposes. Ultra Beauty provides a suite of treatment options, combining the different modes and applications from various devices available in the market. The device provides three easy to use pre-programmed modes and the device manual illustrates in detail the suggested treatment protocols for the desired results. The unit features a portable battery powered base module that features a large LCD display and various buttons for easy navigation through the different treatment settings. Ultra Beauty utilizes two different types of probes and to enhance the ease of use and the maneuverability, each probe is attached to the two separate hand pieces. Stimulations generated by the base unit are delivered to the hand pieces through conductive cables.
{2}------------------------------------------------
# Features:
- Sleek and Modular Design .
- Two separate hand pieces for maneuverability and ease of use 8
- LED indicators on the hand pieces and a conductivity bar on the display screen that. . shows the micro-current is working
- Two different types of probes l
- Large LCD display .
- Battery life indicator l
#### Accessories:
| S.<br>No. | Particulars | Qty. |
|-----------|---------------------------------------------------------------------------------------|---------|
| 1. | Hand pieces wire cable(Hand piece size 13.9 cm ×<br>2.5 cm, Cable: 4 core 1.25 meter) | 02 Nos. |
| 2. | T Bar( Cylindrical size: Diameter: 0.62 cm, Length:<br>5 cm) | 02 Nos. |
| 3. | Two Prong Applicator( Diameter size: 2× 0.508<br>cm=1.016 cm) | 04 Nos. |
| 4. | 5 Volt Battery Charger | 01 No. |
| 5. | Conductive Gel | 250gram |
## INDICATION FOR USE
- Ultra Beauty device is intended for facial stimulation and is indicated for over-. the-counter cosmetic use
{3}------------------------------------------------
#### 3. DISPLAY
ULTRA BEAUTY comes with LCD display to show parameter of selected step, Phase, Time and Output intensity, all parameter can be modified by using MKB panel on top of the unit.
NuFACE uses LED of power on/off indication and power intensity can be controlled by potentiometer.
Face master uses LCD display to show treatment time and power intensity and can be controlled by using buttons on the unit.
Rejuvenique uses LCD display to show treatment time and power intensity and can be controlled by using button on unit.
#### 4. Power
ULTRA BEAUTY use 3.7V /1200MA rechargeable Lithium ion battery and all predicated devices use +9V battery.
#### 5. KEY PADS
In ULTRA BEAUTY all parameter can be modified by using MKB panel on top of the unit.
In NuFACE power intensity can be controlled by potentiometer. In Face master power intensity can be controlled by using button on unit. In Rejuvenique power intensity can be controlled by potentiometer.
#### 6. CASING
ULTRA BEAUTY is packaged in an attractive ABS body enclosure. It is strong and sturdy. It is ergonomically designed.
FaceMaster Rejuvenique and NuFACE is fitted in ABS enclosures.
#### SAFETY:
- 1. In ULTRA BEAUTY the micro-current parameter is controlled by feedback signals, any component failure system gets off.
- 2. Since ULTRA BEAUTY operates only on battery, it makes it safe.
- 3. A reliable high speed controller used to get interface between all parameters.
- 4, Circuitry efficiency is high and reliable.
ULTRA BEAUTY functions normally after open and short circuited conditions between output jacks, with the device operating for maximum of 15 minutes, in each condition at the maximum available setting of pulse width, pulse rate and pulse amplitude. A concise detailed design control activities, verification and validation activities are described in next section.
{4}------------------------------------------------
### NON-CLINICAL TESTS PERFORMED:
Ultra Beauty complies with international standards for electrical safety and electromagnetic compatibility. Compliance to applicable voluntary standards includes IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and ISO 14971: 2007. Comprehensive risk analysis has been carried out for the device with regards to safety and effectiveness.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
#### SUBSTANTIAL EQUIVALENCE:
Based on the foregoing, the Ultra Beauty device was found to have the same intended use and indication for use as the predicate devices. The device also has similar technological characteristics to its predicate devices. Minor differences in the technological characteristics of the Ultra Beauty device and the predicate devices do not raise any issues of safety or effectiveness. Thus Ultra Beauty device is found to be substantially equivalent to the legally marketed predicate devices labeled for over-thecounter cosmetic use
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is composed of thick, black lines, and the text is in a simple, sans-serif font. The overall design is clean and professional.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR 3 0 2012
Johari Digital Healthcare, Ltd. c/o Ms. Pooja Johari Vice President, Marketing Electronic Hardware Technology Park G-582, 583 E.P.I.P. Borananda Jodhpur, 3420008 Rajasthan, India
Re: K112351
Trade/Device Name: Ultra Beauty Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: NFO Dated: March 6, 2012 Received: March 12, 2012
#### Dear Ms. Johari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 – Ms. Pooja Johari
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# INDICATIONS FOR USE
- Ultra Beauty device is intended for facial stimulation and is indicated for over-the-. counter cosmetic use.
Prescription Use _ AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number k112351
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.