← Product Code [MAX](/productcode/MAX) · K112316

# 4-WEB ALIF SPINAL TRUSS SYSTEM (STS) (K112316)

_4-Web, Inc. · MAX · Dec 14, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K112316

## Device Facts

- **Applicant:** 4-Web, Inc.
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Dec 14, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The ALIF Spinal Truss Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation and must be used with autograft bone. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).

## Device Story

ALIF Spinal Truss Interbody Fusion Device is a titanium (Ti6Al4V) implant providing mechanical support to the lumbar spine during biologic fusion. Features an 'open architecture' truss design to facilitate bone growth. Implanted by surgeons in a clinical setting to treat DDD. Device provides structural stability while autograft bone promotes fusion; requires supplemental fixation. Available in various footprints, lordotic angles (0, 6, 8, 12 degrees), and heights (8-17mm).

## Clinical Evidence

Bench testing only. Mechanical performance testing included static compression, static compression shear, static torsion, dynamic axial compression, and dynamic compressive shear per ASTM F2077. Additional testing included subsidence per ASTM F2267-04 and industry-standard expulsion testing.

## Technological Characteristics

Material: Ti6Al4V alloy. Design: Open architecture truss structure. Dimensions: 3 footprints, 4 lordotic angles (0-12 degrees), 9 heights (8-17mm). Mechanical testing standards: ASTM F2077 (compression, shear, torsion), ASTM F2267-04 (subsidence).

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- 4-Web Spine STS ([K083894](/device/K083894.md))
- DePuy/AcroMed Brantigan Cage ([P960025](/device/P960025.md))
- Advanced Medical Technologies Distractable Wave Cage ([K083626](/device/K083626.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for 4WEB. The logo consists of a geometric shape on the left and the text "4WEB" on the right. The geometric shape is a three-dimensional object made up of smaller triangles. The text "4WEB" is in a bold, sans-serif font.

K112316 Page 10

DEC 14 2011

# 510(k) Summary according to 807.92(c)

| Date prepared:  | January 17, 2011                                 |
|-----------------|--------------------------------------------------|
| Contact:        | Jessee Hunt, President                           |
|                 | 4-Web Spine, Inc.                                |
|                 | 6629 Whispering Woods Ct.                        |
|                 | Plano, TX 75024                                  |
|                 | 972-841-6126                                     |
| Trade Name:     | ALIF Spinal Truss System ® Interbody Fusion      |
| Product Class:  | Class II                                         |
| Classification: | 21 CFR §888.3080 Orthosis, intervertebral fusion |
| Product Codes:  | MAX                                              |
| Panel Code:     | 87                                               |

#### Indications for Use:

The ALIF Spinal Truss Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation and must be used with autograft bone. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).

## Device Description:

The ALIF STS Interbody Fusion Device is a titanium implant that is designed to provide mechanical support to the lumbar spine while biologic fusion takes place. The device is an "open architecture" design consisting of trusses mathematically designed to provide maximum support with the greatest amount of open space throughout the implant for bone growth and fusion. The implant is made from Ti6Al4V alloy.

The device is available in three basic "footprint" sizes, small, medium and large. These sizes are available in 0, 6, 8 and 12 degree lordosis and each of these in 9 heights ranging from 8mm to 17mm in 1mm increments.

## Predicate Device(s):

The ALIF STS® Interbody Fusion Device was shown to be substantially equivalent to previously cleared devices and has the same indications for use, design, function, and materials used. The predicate devices include the 4-Web

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for 4WEB. The logo consists of a geometric shape on the left and the word "4WEB" on the right. The geometric shape is a three-dimensional object made up of triangles. The word "4WEB" is in a bold, sans-serif font. There is a trademark symbol to the right of the word WEB.

KI12316

ge 1 of 2

Spine STS (K083894), the DePuy/AcroMed Brantigan Cage (P960025) and the Advanced Medical Technologies Distractable Wave Cage (K083626).

#### Performance Testing:

Preclinical testing performed on the 4-Web STS® Interbody Fusion Device included static compression, static compression shear, static torsion, dynamic axial compression, and dynamic compressive shear mechanical testing per ASTM F2077. Other mechanical tests included subsidence per ASTM F2267-04 and expulsion testing per an industry accepted methodology.

#### Design Changes to (K083894):

Design changes include reduced struts diameter, removal of struts, rounding the top and bottom rim, adding domed contact surfaces, rounding the posterior edge, adding thread ports, increasing the number of lordotic angles and heights available as well as adding one additional cross strut under the posterior rims. The modified design underwent the same performance testing as to the previously approved device. The results indicate equal to or better than performance to the original design or other previously approved devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 1 4 2011

4-Web Spine, Inc. % Silver Pine Consulting, Ltd. Rich Jansen, Pharm.D. 13450 Guild Avenue Apple Valley, Minnesota 55124

Re: K112316

Trade/Device Name: ALIF Spinal Truss System® Interbody Fusion Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: . Class II Product Code: MAX Dated: December 08, 2011 Received: December 09, 2011

Dear Dr. Jansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

Page 2 -- Rich Jansen, Pharm.D.

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson
Director

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

4-Web. Inc. 510(k)

# Statement of Indications for Use

510(k) Number (if known): K112316

Indications for Use:

The ALIF Spinal Truss Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation and must be used with autograft bone. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ・ 

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices

KILZ316 510(k) Number_

---

**Source:** [https://fda-staging.innolitics.com/device/K112316](https://fda-staging.innolitics.com/device/K112316)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
